Effectiveness of Sensor-Augmented Insulin-Pump Therapy in Type 1 Diabetes

Richard M. Bergenstal, M.D., William V. Tamborlane, M.D., Andrew Ahmann, M.D., John B. Buse, M.D., Ph.D., George Dailey, M.D., Stephen N. Davis, M.D., Carol Joyce, M.D., Tim Peoples, M.A., Bruce A. Perkins, M.D., M.P.H., John B. Welsh, M.D., Ph.D., Steven M. Willi, M.D., and Michael A. Wood, M.D. for the STAR 3 Study Group

N Engl J Med 2010; 363:311-320July 22, 2010

Abstract

Background

Recently developed technologies for the treatment of type 1 diabetes mellitus include a variety of pumps and pumps with glucose sensors.

Methods

In this 1-year, multicenter, randomized, controlled trial, we compared the efficacy of sensor-augmented pump therapy (pump therapy) with that of a regimen of multiple daily insulin injections (injection therapy) in 485 patients (329 adults and 156 children) with inadequately controlled type 1 diabetes. Patients received recombinant insulin analogues and were supervised by expert clinical teams. The primary end point was the change from the baseline glycated hemoglobin level.

Results

At 1 year, the baseline mean glycated hemoglobin level (8.3% in the two study groups) had decreased to 7.5% in the pump-therapy group, as compared with 8.1% in the injection-therapy group (P<0.001). The proportion of patients who reached the glycated hemoglobin target (<7%) was greater in the pump-therapy group than in the injection-therapy group. The rate of severe hypoglycemia in the pump-therapy group (13.31 cases per 100 person-years) did not differ significantly from that in the injection-therapy group (13.48 per 100 person-years, P=0.58). There was no significant weight gain in either group.

Conclusions

In both adults and children with inadequately controlled type 1 diabetes, sensor-augmented pump therapy resulted in significant improvement in glycated hemoglobin levels, as compared with injection therapy. A significantly greater proportion of both adults and children in the pump-therapy group than in the injection-therapy group reached the target glycated hemoglobin level. (Funded by Medtronic and others; ClinicalTrials.gov number, NCT00417989.)

Media in This Article

Figure 1Glycated Hemoglobin Levels at 3, 6, 9, and 12 Months in All Patients and in Subgroups According to Age.
Figure 2Sensor Use and Change in Glycated Hemoglobin Levels in 244 Patients Using a Sensor-Augmented Insulin Pump with Continuous Glucose Monitoring.
Article

Improved glycemic control can reduce the microvascular and macrovascular complications associated with type 1 diabetes mellitus,1-3 and diabetes practitioners are continuously challenged to optimize glucose control while minimizing severe hypoglycemia and weight gain. Insulin pumps and systems for continuous glucose monitoring represent technologies designed to assist patients with type 1 diabetes in safely reaching glycemic goals. Among adults, the use of an insulin pump has been shown to reduce glycated hemoglobin levels without an increased risk of hypoglycemia, as compared with a regimen of multiple daily insulin injections, but results in children have been inconsistent.4 Recent studies have suggested that patients who used sensor-augmented pump therapy with adherence to continuous glucose monitoring had improved glycated hemoglobin levels without an increased rate of hypoglycemia.5-7 Similarly, in a multicenter trial of continuous glucose monitoring in patients with type 1 diabetes, sponsored by the Juvenile Diabetes Research Foundation (JDRF) (ClinicalTrials.gov number, NCT00406133), the use of a continuous glucose-monitoring device was effective in reducing glycated hemoglobin levels among patients who were 25 years of age or older but not among patients under the age of 25 years.8

Sensor-augmented pump therapy integrates these two technologies into one system and allows patients and clinicians to monitor treatment and response through Internet-based software. Whether, and to what extent, switching directly to sensor-augmented pump therapy might improve metabolic control in patients with type 1 diabetes who were previously unable to reach glycemic targets with a regimen of multiple daily injections and conventional blood-glucose monitoring is unknown. In this unmasked, randomized, controlled trial, called Sensor-Augmented Pump Therapy for A1C Reduction (STAR) 3, we evaluated the use of sensor-augmented pump therapy and injection therapy at 30 diabetes centers in the United States and Canada for 1 year.9

Methods

Patients

Patients with type 1 diabetes were eligible if they were between the ages of 7 and 70 years, had received multiple daily injections that included a long-acting analogue insulin during the previous 3 months, had a glycated hemoglobin level between 7.4% and 9.5%, and had been under the care of the principal investigator or a referring physician for at least 6 months. Patients were required to have access to a computer and to have a history of testing blood glucose an average of four or more times per day for the previous 30 days. Exclusion criteria were the use of insulin-pump therapy within the previous 3 years, a history of at least two severe hypoglycemic events in the year before enrollment, the use of a pharmacologic noninsulin treatment for diabetes during the previous 3 months, and pregnancy or the intention to become pregnant. All patients provided written informed consent.

Treatments

Patients were randomly assigned to receive either sensor-augmented pump therapy (pump therapy) or a regimen of multiple daily injections (injection therapy) with the use of a block design, stratified according to age group: adults (19 to 70 years of age) or children (7 to 18 years of age). Levels of glycated hemoglobin and blood glucose in the two study groups and sensor glucose values in the pump-therapy group were disclosed to investigators, caregivers, and patients in order to optimize glycated hemoglobin levels and to minimize the risk of severe hypoglycemia.

The pump-therapy group used a device that integrates an insulin pump with continuous glucose monitoring (MiniMed Paradigm REAL-Time System, Medtronic) (see Fig. 1 in the Supplementary Appendix, available with the full text of this article at NEJM.org). Before randomization, all patients received training in intensive diabetes management, including carbohydrate counting and the administration of correction doses of insulin. Patients were first placed on insulin-pump therapy for 2 weeks, and then glucose sensors were introduced. During the 5 weeks after randomization, patients in the pump-therapy group completed online insulin-pump training and attended additional visits for insulin-pump and sensor training. This group used insulin aspart (NovoLog or NovoRapid, Novo Nordisk). The injection-therapy group used both insulin glargine (Lantus, Sanofi-Aventis) and insulin aspart under the guidance of the treating clinician. Sensor glucose values were collected for 1-week periods at baseline, 6 months, and 1 year in the two study groups. In the injection-therapy group, a device for continuous glucose monitoring that collected but did not display data (Guardian REAL-Time Clinical, Medtronic) was used.

All patients were seen at 3, 6, 9, and 12 months after randomization and used diabetes-management software (CareLink Therapy Management System for Diabetes–Clinical, Medtronic). Between visits, communication with clinicians was initiated at the discretion of the patient.

At follow-up clinic visits, glucose data were reviewed, therapy was adjusted, glycated hemoglobin was measured, and data on adverse events were collected. Severe hypoglycemia was defined as an episode requiring assistance and was confirmed by documentation of a blood glucose value of less than 50 mg per deciliter (2.8 mmol per liter) or recovery with restoration of plasma glucose. Quest Clinical Trials Laboratory measured glycated hemoglobin by means of immunoturbidimetry.

Study Oversight

The institutional review board at each study site approved the protocol, and the conduct of the study was consistent with the Good Clinical Practice provisions of the Declaration of Helsinki and local regulatory requirements. Data management and statistical analyses were conducted by Parexel International, an independent clinical research organization, which transferred all data to the sponsor, Medtronic. Novo Nordisk supplied all insulin aspart used in the study, and LifeScan, Bayer Healthcare, and Becton Dickinson supplied blood glucose meters.

All authors had access to the data, wrote the first draft of the manuscript with editorial assistance from representatives of the sponsor, subsequently revised the manuscript, and made the decision to submit the manuscript for publication. All authors vouch for the accuracy and completeness of the data and analyses. The STAR 3 steering committee was responsible for the study design and methods.9

The study was conducted in accordance with the original trial protocol, with the following exceptions: the eligibility cutoff level for glycated hemoglobin was lowered from 7.5% to 7.4%, the exclusion criteria were changed from no previous use of insulin-pump therapy to no such use within the previous 3 years, the sample size was increased from 336 patients at 25 centers to 552 patients at 30 centers, results on the Hypoglycemia Fear Survey were moved from a secondary end point to a tertiary end point, the Telemetered Glucose-Monitoring System (Medtronic) was replaced with the MiniLink transmitter (Medtronic), and three visits during the 5 weeks after randomization were removed from the schedule for the injection-therapy group. (The trial protocol is available at NEJM.org.)

Statistical Analysis

The primary outcome was the change from baseline in the glycated hemoglobin level at 1 year. Severe rates of hypoglycemia were analyzed as a secondary outcome. We calculated that the enrollment of 495 patients would provide a power of 90% to detect an absolute difference of 0.35 percentage points in the primary outcome. Analyses were performed in the intention-to-treat population, defined as patients who underwent at least one measurement of glycated hemoglobin after randomization, with the last observation carried forward for the imputation of missing data.

Differences in the change in glycated hemoglobin levels were analyzed with the use of analysis of covariance (ANCOVA) with three categorical variables (study group, pooled investigative site, and sex) and four continuous variables (age, duration of diabetes, body-mass index, and baseline glycated hemoglobin level) as fixed effects. The proportion of patients reaching glycated hemoglobin targets was analyzed with the use of a logistic-regression model with two categorical variables (study group and sex) and the above-mentioned continuous variables as fixed effects. The effect of sensor use on glycated hemoglobin levels was analyzed with the use of an ANCOVA model with three categorical variables (sensor-use categories, pooled investigative site, and sex) and the above-mentioned continuous variables as fixed effects. Changes in weight were analyzed with the use of a general linear model with adjustment for baseline weight. Incidences of adverse events were compared with the use of a logistic-regression model, and rates of adverse events per 100 person-years were compared with the use of an ANCOVA model; both models had two categorical variables (study group and sex) and the above-mentioned continuous variables as fixed effects. The area under the curve is the product of the magnitude and duration of the sensor-measured glucose level above or below a specified cutoff level. Higher values for this calculation indicate more numerous, severe, or protracted glycemic events. The areas under the curve that were calculated from continuous glucose monitoring at 1 year were compared with the use of an ANCOVA model with three categorical variables (study group, pooled investigative site, and sex) and four continuous variables (age, duration of diabetes, body-mass index, and baseline area under the curve) as fixed effects. Baseline characteristics were compared with the use of a two-sample t-test for continuous variables and either the chi-square test or Fisher's exact test for categorical variables.

Analyses were conducted with SAS software, version 9.2 (SAS Institute). All reported P values are two-sided; a P value of less than 0.05 was considered to indicate statistical significance for comparisons of the primary outcome, baseline characteristics, and safety.

Results

Study Recruitment and Baseline Characteristics

From January 2007 through December 2008, a total of 495 patients underwent randomization; follow-up data on glycated hemoglobin levels were missing for 10 patients, who were not included in the primary analysis. Of the remaining 485 patients, 4 (1%) were lost to follow-up, 32 (7%) discontinued the study or were withdrawn, and 6 (1%) did not provide 1-year results for glycated hemoglobin, leaving 443 patients in the primary analysis (Fig. 2 in the Supplementary Appendix). Baseline characteristics were similar in the two study groups, except for weight (P=0.02) and student status (P=0.02) among adult patients (Table 1Table 1Baseline Characteristics of the Patients.).

Primary and Secondary Outcomes

At 1 year, the baseline mean glycated hemoglobin level (8.3% in the two study groups) had decreased to 7.5% in the pump-therapy group (absolute reduction, 0.8±0.8 percentage points), as compared with 8.1% in the injection-therapy group (absolute reduction, 0.2±0.9 percentage points), for a between-group difference in the pump-therapy group of –0.6 percentage points (95% confidence interval [CI], –0.7 to –0.4; P<0.001) (Figure 1Figure 1Glycated Hemoglobin Levels at 3, 6, 9, and 12 Months in All Patients and in Subgroups According to Age.). Among adults, the absolute reduction in the mean glycated hemoglobin level was 1.0±0.7 percentage points in the pump-therapy group and 0.4±0.8 percentage points in the injection-therapy group, for a between-group difference in the pump-therapy group of –0.6 percentage points (95% CI, –0.8 to –0.4; P<0.001). Among children, there was an absolute reduction in glycated hemoglobin of 0.4±0.9 percentage points in the pump-therapy group and an increase of 0.2±1.0 percentage points in the injection-therapy group, for a between-group difference favoring the pump-therapy group of –0.5 percentage points (95% CI, –0.8 to –0.2; P<0.001), with adjustment for the statistical model.

Post hoc analyses that included only data from observed patients and used multiple imputation of missing values yielded similar results (Table 1 in the Supplementary Appendix). In both adults and children in the pump-therapy group, glycated hemoglobin levels fell rapidly from baseline to 3 months and remained lower than levels in the injection-therapy group for the remainder of the study (Figure 1). An increased frequency of sensor use was associated with a greater reduction in glycated hemoglobin levels at 1 year (P=0.003 with adjustment for the baseline glycated hemoglobin level) (Figure 2Figure 2Sensor Use and Change in Glycated Hemoglobin Levels in 244 Patients Using a Sensor-Augmented Insulin Pump with Continuous Glucose Monitoring.).

The numbers of patients who reached a glycated hemoglobin value of 7% or less were 67 of 244 patients (27%) in the pump-therapy group and 23 of 241 patients (10%) in the injection-therapy group (P<0.001); these numbers included 57 of 166 adults (34%) in the pump-therapy group and 19 of 163 adults (12%) in the injection-therapy group (P<0.001) and 10 of 78 children (13%) in the pump-therapy group and 4 of 78 children (5%) in the injection-therapy group (P=0.15) (Figure 3Figure 3Cumulative Distribution of Glycated Hemoglobin Levels at 1 Year among All Patients and in Subgroups According to Age.). In a post hoc analysis that used glycated hemoglobin targets recommended by the American Diabetes Association for children between the ages of 6 and 12 years (<8%) and adolescents between the ages of 13 and 19 years (<7.5%),10 a total of 35 of the 80 children and adolescents (44%) in the pump-therapy group and 16 of the 80 (20%) in the injection-therapy group reached these targets at 1 year (P=0.005). Among adults, weight increased by 2.4 kg in the pump-therapy group and by 1.8 kg in the injection-therapy group (P=0.19).

Adverse Events

Rates of severe hypoglycemia and diabetic ketoacidosis were similar in the two study groups and in the two age groups. The area under the curve that was calculated from continuous glucose monitoring was similar in the two groups at 1 year for patients with hypoglycemia (defined either as <70 mg per deciliter [<3.9 mmol per liter] or as <50 mg per deciliter [<2.8 mmol per liter]) and was significantly lower in the pump-therapy group for patients with hyperglycemia (defined either as >180 mg per deciliter [>10.0 mmol per liter] or as >250 mg per deciliter [>13.9 mmol per liter]) (Table 2Table 2Severe Hypoglycemia, Diabetic Ketoacidosis, and Area under the Curve Calculated from Continuous Glucose Monitoring.). At 1 year, 5 of 100 patients (5%) with a glycated hemoglobin level of 7% or less had severe hypoglycemia, as compared with 33 of 395 patients (8%) with a glycated hemoglobin level of more than 7% (P=0.12). There were no severe hypoglycemic events in either study group among children who had a glycated hemoglobin level of 7% or less at 1 year.

There were two hospital admissions in the pump-therapy group for cellulitis related to insertion-site infections and one death from sudden cardiac arrest in a patient in the injection-therapy group who had a history of cardiovascular disease.

Discussion

In this study comparing sensor-augmented pump therapy with multiple-injection therapy, the between-group difference in glycated hemoglobin levels favored pump therapy and was statistically and clinically significant among both adults and children. More patients in the pump-therapy group reached the prespecified target glycated hemoglobin value of 7% or less, and lower glycated hemoglobin levels were achieved in this group by reducing biochemical hyperglycemia without increasing biochemical hypoglycemia or the rate of severe hypoglycemic events. The incidence of diabetic ketoacidosis was negligible, and there was no significant between-group difference in weight gain among adults.

Among adults, who had a mean baseline glycated hemoglobin level of 8.3±0.5%, the reduction of 1.0 percentage point in the pump-therapy group was significantly greater than the reduction of 0.4 percentage points in the injection-therapy group. In comparison, in the JDRF study of continuous blood glucose monitoring, adults in both the pump-therapy group and the injection-therapy group had a reduction of 0.5 percentage points (mean baseline value, 7.6±0.5%), whereas glycated hemoglobin levels increased slightly in the group undergoing standard blood glucose monitoring.8 Although we cannot directly compare our findings with those of the JDRF study, in our study, the reduction of 0.5 percentage points in glycated hemoglobin levels among children in the pump-therapy group, as compared with the injection-therapy group, differed from the results of the JDRF study, which showed no between-group difference at 6 months among patients who were 8 to 14 years of age or among those who were 15 to 24 years of age. In the JDRF study, the only variables that predicted successful use of continuous glucose monitoring were an older age and an increased frequency of daily blood glucose measurement.11 Also of interest is our finding that nearly half the children in the pump-therapy group reached the American Diabetes Association's age-specific targets for glycated hemoglobin by the end of the study.10

Other studies have compared pump therapy with injection therapy, as well as continuous glucose monitoring with conventional blood glucose monitoring with a meter. In one randomized, controlled trial,5 investigators examined the efficacy of introducing these technologies together in patients who had not used either pumps or sensors. In our study, the decision to introduce these technologies one at a time during the first 5 weeks proved to be an effective strategy. Clinicians may wish to consider this approach when implementing sensor-augmented pump systems in patients who have not undergone such therapy.

Since previous studies have examined the effect of individual components of the sensor-augmented pump system, our study was designed to examine how the combined system compares with optimal injection therapy. In contrast to the findings in previous studies, our results suggest that the effects of the combined system were greater than what would be expected from the individual components alone. In our study, a frequency of sensor use of 41 to 60% was associated with a reduction of 0.64 percentage points in glycated hemoglobin levels, and increasing sensor use to more than 80% doubled the effect. In contrast, in both the JDRF study (in which patients maintained their prestudy insulin-pump or injection regimen) and the STAR 1 study (in which patients maintained their prestudy insulin-pump regimen or switched to sensor-augmented pump therapy), lower glycated hemoglobin levels were observed only in patients who used the sensor 60% or more of the time.6,8 The improvements in metabolic control that we observed were also much greater than expected with insulin-pump therapy alone, since two recent meta-analyses of pump therapy versus injection therapy4,12 have shown reductions of 0.3 and 0.2 percentage points, respectively, in glycated hemoglobin levels with insulin-pump therapy. Additional studies designed to identify the independent benefits of sensors and insulin pumps are warranted. It is also likely that in our study, the use of therapy-management software benefited patients using sensor-augmented pump therapy. However, we limited the effect of this factor by making the data-management program available to all patients.

Patients in our two study groups had much lower rates of severe hypoglycemia than did patients with corresponding glycated hemoglobin levels who underwent intensive insulin treatment in the Diabetes Control and Complications Trial (DCCT) (NCT00360815).1,2 This difference may be due to the multiple advances in diabetes therapy — such as the use of insulin analogues13 — that have occurred since the initiation of the DCCT or to our exclusion of patients who had had two or more episodes of severe hypoglycemia during the previous year. The most striking difference between the two studies was in the number of severe hypoglycemic events among children. In our study, children in the pump-therapy group (mean glycated hemoglobin level, 7.9%) had 9.0 severe hypoglycemic events per 100 patient-years, as compared with adolescents in the DCCT (mean glycated hemoglobin level, 8.1%) undergoing intensive insulin treatment, who had 85.7 such events per 100 person-years.1

Our study has several limitations. First, because of the nature of the medical devices that were used, the interventions were known to patients, investigators, and caregivers. Second, we did not study the effect of insulin-pump therapy alone versus sensor-augmented pump therapy to determine the contribution of each component of the system. Third, the generalizability of the study's results may in part be limited by the use of a mandated range for glycated hemoglobin (7.4 to 9.5%) as an inclusion criterion. Fourth, our patients may have been particularly motivated because they were participating in a study. However, they were generally representative of patients with type 1 diabetes who are considered to be candidates for further intensification of insulin therapy and diabetes care (i.e., those who are not able to reach desired glycemic targets with a regimen of multiple daily insulin injections and appropriate medical support), and the results indicate that sensor-augmented pump therapy is a consideration for patients. Finally, for reasons of technical device training, patients in the pump-therapy group received more contact with clinical staff members than did patients in the injection-therapy group during the first 5 weeks of the study; thereafter, clinical contacts were designed to be identical in the two study groups.

In conclusion, in patients with type 1 diabetes with suboptimal glycemic control, the use of a sensor-augmented insulin pump was associated with significant improvement in glycated hemoglobin levels, as compared with a regimen of multiple daily injections of recombinant insulin analogues.

Supported by Medtronic; Novo Nordisk, which supplied all insulin aspart used in the study; and LifeScan, Bayer Healthcare, and Becton Dickinson, which supplied blood glucose meters used in the study.

Dr. Bergenstal reports that the International Diabetes Center has received consulting fees from Abbott Diabetes Care, Amylin, Bayer, Calibra, Eli Lilly, Intarcia, MannKind, Medtronic, Novo Nordisk, Pfizer, ResMed, Roche, Sanofi-Aventis, and Takeda for his services; grant support from Abbott Diabetes Care, Biodel, Eli Lilly, Hygieia, Intuity, LifeScan, Medtronic, Novo Nordisk, Pfizer, ResMed, Roche, Sanofi-Aventis, Takeda, UnitedHealth Group, and Valeritas; honoraria from Abbott Diabetes Care, Amylin, Bayer, Calibra, Eli Lilly, Intarcia, MannKind, Medtronic, Novo Nordisk, Pfizer, ResMed, Roche, Sanofi-Aventis, and Takeda; royalties on his behalf for the Betty Crocker Diabetes Cookbook; and travel and accommodation reimbursements from Abbott Diabetes Care, Amylin, Bayer, Biodel, Calibra, Eli Lilly, Hygieia, Intarcia, Intuity, LifeScan, MannKind, Medtronic, Novo Nordisk, Pfizer, ResMed, Roche, Sanofi-Aventis, Takeda, UnitedHealth Group, and Valeritas; and having an equity interest in Merck. Dr. Tamborlane reports receiving consulting fees from Medtronic Diabetes as a member of the Diabetes Advisory Board. Dr. Ahmann reports that Oregon Health and Science University has received consulting fees from Biodel, Novo Nordisk, and GlaxoSmithKline for his services; grant support from Amylin, Medtronic, and Novo Nordisk; and honoraria from Amylin, Lilly, Animas, NDEI, and Medical Education Resources. Dr. Buse reports that for his services the University of North Carolina School of Medicine has received consulting fees from Novo Nordisk, Amylin, Becton Dickinson, Eli Lilly, Hoffmann–La Roche, GlycoMark, Wyeth, Daiichi Sankyo, Bristol-Myers Squibb, Bayhill Therapeutics, LipoScience, MannKind, Valeritas, MicroIslet, GlaxoSmithKline, Abbott, Exsulin, and GI Dynamics; funding from Novo Nordisk for expert testimony; grant support from Amylin, Novo Nordisk, Medtronic, Lilly, Novartis, Tolerex, Osiris, Halozyme, Pfizer, Hoffmann–La Roche, Interkrin, Merck, Sanofi-Aventis, Dexcom, Johnson & Johnson, Bristol-Myers Squibb, Fujisawa, and Novartis; and travel and accommodation reimbursement from Novo Nordisk, Amylin, Medtronic, Novartis, Tolerex, Becton Dickinson, Eli Lilly, Hoffmann–La Roche, Fujisawa, GlycoMark, Wyeth, Daiichi Sankyo, Bristol-Myers Squibb, Bayhill Therapeutics, Osiris, Interkrin, Merck, Sanofi-Aventis, Dexcom, Johnson & Johnson, Halozyme, Pfizer, LipoScience, MannKind, Valeritas, MicroIslet, GlaxoSmithKline, Abbott, Exsulin, and GI Dynamics; he also reports having an equity interest in Insulet. Dr. Joyce reports receiving consulting fees from Eli Lilly, Sanofi-Aventis, and AstraZeneca, travel reimbursement from AstraZeneca, and grant support (paid to the Memorial University of Newfoundland) from Eli Lilly, Sanofi-Aventis, AstraZeneca, and Boehringer Ingelheim. Mr. Peoples and Dr. Welsh report being employees of and having an equity interest in Medtronic and receiving compensation for manuscript preparation. Dr. Willi reports that the Children's Hospital of Philadelphia has received grant support from Sanofi-Aventis for his services. Dr. Wood reports receiving consulting fees from Medtronic MiniMed. No other potential conflict of interest relevant to this article was reported. Disclosure forms provided by the authors are available with the full text of this article at NEJM.org.

This article (10.1056/NEJMoa1002853) was published on June 29, 2010, and updated on September 8, 2010, at NEJM.org.

We thank Lianne Johnson, M.S., R.N., Scott W. Lee, M.D., and Brenda Perry, M.S., R.N., employees of Medtronic, who provided editorial assistance with an earlier draft of the manuscript, coordinated data analyses at the authors' request, and coordinated the required forms for manuscript submission.

Source Information

From the International Diabetes Center at Park Nicollet, Minneapolis (R.M.B.); Yale University, New Haven, CT (W.V.T.); Oregon Health and Science University, Portland (A.A.); University of North Carolina School of Medicine, Chapel Hill (J.B.B.); Scripps Institute, La Jolla (G.D.), and Medtronic, Northridge (T.P., J.B.W.) — both in California; University of Maryland School of Medicine, Baltimore (S.N.D.); Memorial University of Newfoundland, Health Science Centre, St. John's, NL, Canada (C.J.); Toronto General Hospital, Toronto (B.A.P.); Children's Hospital of Philadelphia, Philadelphia (S.M.W.); and Helen DeVos Children's Hospital, Grand Rapids, MI (M.A.W.).

Address reprint requests to Dr. Bergenstal at the International Diabetes Center at Park Nicollet, Minneapolis, MN 55416, or at .

Members of the Sensor-Augmented Pump Therapy for A1C Reduction (STAR) 3 Study Group are listed in the Appendix.

Appendix

Members of the STAR 3 Study Group are as follows (asterisks indicate principal investigators): Mountain Diabetes and Endocrine Center, Asheville, NC: W.S. Lane,* K. Arey, T. Przestrzelski; Joslin Diabetes Center, Boston: S. Mehta,* L. Laffel,* J. Aggarwal, K. Pratt; University of North Carolina School of Medicine, Chapel Hill: J.B. Buse,* M. Duclos, J. Largay; Ohio State University College of Medicine, Columbus: K. Osei,* C. Casey-Boyer, H. Breedlove; University of Colorado Denver, Denver: R. Slover, II,* S. Kassels, S. Sullivan; Duke University Medical Center, Durham, NC: J.B. Green,* J. English-Jones; Helen DeVos Children's Hospital, Grand Rapids, MI: M.A. Wood,* E. Gleason, L. Wagner; East Carolina University, Diabetes and Obesity Center, Greenville, NC: R.J. Tanenberg,* C. Knuckey; Rocky Mountain Diabetes and Osteoporosis Center, Idaho Falls, ID: D.R. Liljenquist,* J.E. Liljenquist,* B. Sulik; Kingston General Hospital, Kingston, ON, Canada: R.L. Houlden,* T. LaVallee, A. Breen, R. Barrett; Scripps Institute, La Jolla, CA: G. Dailey,* R. Rosal, J. Shartel; Kentucky Diabetes Endocrinology Center, Lexington: L. Myers,* D. Ballard; Endocrinology Diabetes Clinic, Madison, WI: M. Meredith,* C. Trantow; Diabetes Research Institute, Miami: L.F. Meneghini,* J. Sparrow-Bodenmiller, R. Agramonte; International Diabetes Center at Park Nicollet, Minneapolis: R.M. Bergenstal,* A.B. Criego, S. Borgman; Vanderbilt University, Nashville: M.E. May,* S.N. Davis,* C. Root; Yale University, New Haven, CT: S.A. Weinzimer,* L. Carria, J. Sherr; Children's Hospital of Orange County, Orange, CA: M. Daniels,* J.S. Krantz,* H. Speer, J. Less; Children's Hospital of Philadelphia, Philadelphia: S.M. Willi,* T. Calvano, E. Garth; Oregon Health and Science University, Portland: A. Ahmann,* C. Bogan, V. Chambers, B. Wollam; Mayo Clinic, Rochester, MN: Y.C. Kudva,* B. Wirt; University of Rochester School of Medicine and Dentistry, Rochester, NY: C. Orlowski,* S. Bates, B. Johnson; Endocrine Research Solutions, Roswell, GA: J.C. Reed, III,* J. Tapia, K. Wardell, S. Newsome; Memorial University of Newfoundland, St. John's, NL, Canada: C. Joyce,* D. Gibbons, J. O'Leary; Washington University in St. Louis School of Medicine, St. Louis: N.H. White,* M. Coleman; Children's Hospital of St. Paul, St. Paul, MN: R.C. McEvoy,* C. Girard; Utah Diabetes Center, Salt Lake City: C.M. Foster,* T. Brown, E. Nuttall; Toronto General Hospital, Toronto: B.A. Perkins,* A. Orszag; Endocrine Research, Vancouver, BC, Canada: H. Tildesley,* B. Pottinger; Mid-America Diabetes Associates, Wichita, KS: R.A. Guthrie,* J. Dvorak, B. Childs. Members of the STAR 3 Steering Committee are as follows: T. Battelino, University Children's Hospital, Ljubljana, Slovenia; S.N. Davis, University of Maryland School of Medicine, Baltimore; E.S. Horton, Joslin Diabetes Center, Boston; S.W. Lee, Medtronic, Northridge, CA; R.R. Rubin, Johns Hopkins University, Baltimore; K.A. Schulman, Duke University, Durham, NC; W.V. Tamborlane, Yale University, New Haven, CT.

References

References

  1. 1

    Diabetes Control and Complications Trial Research Group. Effect of intensive diabetes treatment on the development and progression of long-term complications in adolescents with insulin-dependent diabetes mellitus: Diabetes Control and Complications Trial. J Pediatr 1994;125:177-188
    CrossRef | Web of Science | Medline

  2. 2

    Diabetes Control and Complications Trial/Epidemiology of Diabetes Interventions and Complications (DCCT/EDIC) Research Group. Modern-day clinical course of type 1 diabetes mellitus after 30 years' duration: the Diabetes Control and Complications Trial/Epidemiology of Diabetes Interventions and Complications and Pittsburgh Epidemiology of Diabetes Complications experience (1983-2005). Arch Intern Med 2009;169:1307-1316
    CrossRef | Web of Science | Medline

  3. 3

    Diabetes Control and Complications Trial/Epidemiology of Diabetes Interventions and Complications (DCCT/EDIC) Research Group. Intensive diabetes treatment and cardiovascular disease in patients with type 1 diabetes. N Engl J Med 2005;353:2643-2653
    Full Text | Web of Science | Medline

  4. 4

    Misso ML, Egberts KJ, Page M, O'Connor D, Shaw J. Continuous subcutaneous insulin infusion (CSII) versus multiple insulin injections for type 1 diabetes mellitus. Cochrane Database Syst Rev 2010;1:CD005103-CD005103
    Medline

  5. 5

    Hermanides J, Norgaard K, Bruttomesso D, et al. Sensor augmented pump therapy substantially lowers HbA1c: a randomized controlled trial. Diabetologia 2009;52:Suppl 1:S43-S43
    Web of Science

  6. 6

    Hirsch IB, Abelseth J, Bode BW, et al. Sensor-augmented insulin pump therapy: results of the first randomized treat-to-target study. Diabetes Technol Ther 2008;10:377-383
    CrossRef | Web of Science | Medline

  7. 7

    Raccah D, Sulmont V, Reznik Y, et al. Incremental value of continuous glucose monitoring when starting pump therapy in patients with poorly controlled type 1 diabetes: the RealTrend study. Diabetes Care 2009;32:2245-2250
    CrossRef | Web of Science | Medline

  8. 8

    The Juvenile Diabetes Research Foundation Continuous Glucose Monitoring Study Group. Continuous glucose monitoring and intensive treatment of type 1 diabetes. N Engl J Med 2008;359:1464-1476
    Full Text | Web of Science | Medline

  9. 9

    Davis SN, Horton ES, Battelino T, Rubin RR, Schulman KA, Tamborlane WV. STAR 3 randomized controlled trial to compare sensor-augmented insulin pump therapy with multiple daily injections in the treatment of type 1 diabetes: research design, methods, and baseline characteristics of enrolled subjects. Diabetes Technol Ther 2010;12:249-255
    CrossRef | Web of Science | Medline

  10. 10

    American Diabetes Association. Standards of medical care in diabetes -- 2010. Diabetes Care 2010;33:Suppl 1:S11-S61
    Web of Science | Medline

  11. 11

    Juvenile Diabetes Research Foundation Continuous Glucose Monitoring Study Group. Factors predictive of use and of benefit from continuous glucose monitoring in type 1 diabetes. Diabetes Care 2009;32:1947-1953
    CrossRef | Web of Science | Medline

  12. 12

    Pickup JC, Sutton AJ. Severe hypoglycaemia and glycaemic control in Type 1 diabetes: meta-analysis of multiple daily insulin injections compared with continuous subcutaneous insulin infusion. Diabet Med 2008;25:765-774
    CrossRef | Web of Science | Medline

  13. 13

    Hirsch IB. Insulin analogues. N Engl J Med 2005;352:174-183
    Full Text | Web of Science | Medline

Citing Articles (49)

Citing Articles

  1. 1

    Richard R. Rubin, Mark Peyrot. (2012) Health-Related Quality of Life and Treatment Satisfaction in the Sensor-Augmented Pump Therapy for A1C Reduction 3 (STAR 3) Trial. Diabetes Technology & Therapeutics 14:2, 143-151
    CrossRef

  2. 2

    Tadej Battelino, Bruce W. Bode. (2012) Continuous glucose monitoring in 2011. International Journal of Clinical Practice 66, 8-14
    CrossRef

  3. 3

    Shlomit Shalitin, H. Peter Chase. (2012) Diabetes technology and treatment in the paediatric age group. International Journal of Clinical Practice 66, 72-78
    CrossRef

  4. 4

    Jongoh Kim, Se Min Kim, Ha Cam Thuy Nguyen, Maria Jose Redondo. (2012) Therapeutics in pediatric diabetes: Insulin and non-insulin approaches. Pharmacological Research 65:1, 1-4
    CrossRef

  5. 5

    J. H. DeVries. (2012) Continuous glucose monitoring: coming of age?. European Journal of Endocrinology 166:1, 1-4
    CrossRef

  6. 6

    Moshe Phillip, Thomas Danne, Shlomit Shalitin, Bruce Buckingham, Lori Laffel, William Tamborlane, Tadej Battelino, . (2012) Use of continuous glucose monitoring in children and adolescents. Pediatric Diabetesn/a-n/a
    CrossRef

  7. 7

    Fernando Gómez Peralta, Cristina Abreu Padín. (2012) Monitorización continua de la glucosa en tiempo real. Indicaciones clínicas probadas. Avances en Diabetología
    CrossRef

  8. 8

    MSol Ruiz-de-Adana, Marta-Elena Domínguez-Lopez, Alberto Machado, Natalia Colomo, M. Teresa Anarte, Federico C-Soriguer. (2012) La monitorización continua de glucosa a tiempo real y los servicios nacionales de salud. ¿Hay suficiente evidencia científica para justificar su financiación pública?. Avances en Diabetología
    CrossRef

  9. 9

    Desmond Barry Keenan, John Joseph Mastrototaro, Howard Zisser, Kenneth A. Cooper, Gautham Raghavendhar, Scott W. Lee, Jonathan Yusi, Timothy Bailey, Ronald Leonard Brazg, Rajiv V. Shah. (2011) Accuracy of the Enlite 6-Day Glucose Sensor with Guardian and Veo Calibration Algorithms. Diabetes Technology & Therapeutics111206132933005
    CrossRef

  10. 10

    Stefano Zucchini, Mirella Scipione, Claudia Balsamo, Giulio Maltoni, Alessandra Rollo, Emanuela Molinari, Lorenza Mangoni, Alessandro Cicognani. (2011) Comparison Between Sensor-Augmented Insulin Therapy with Continuous Subcutaneous Insulin Infusion or Multiple Daily Injections in Everyday Life: 3-Day Analysis of Glucose Patterns and Sensor Accuracy in Children. Diabetes Technology & Therapeutics 13:12, 1187-1193
    CrossRef

  11. 11

    C. R. Alleyn, L. M. B. Laffel, L. K. Volkening, B. J. Anderson, T. R. Nansel, T. Wysocki, J Weissberg-Benchell. (2011) Comparison of longitudinal point-of-care and high-performance liquid chromatography HbA1c measurements in a multi-centre trial. Diabetic Medicine 28:12, 1525-1529
    CrossRef

  12. 12

    P. Schaepelynck, P. Darmon, L. Molines, M.F. Jannot-Lamotte, C. Treglia, D. Raccah. (2011) Advances in pump technology: insulin patch pumps, combined pumps and glucose sensors, and implanted pumps. Diabetes & Metabolism 37, S85-S93
    CrossRef

  13. 13

    Jaime M. Realsen, H. Peter Chase. (2011) Recent Advances in the Prevention of Hypoglycemia in Type 1 Diabetes. Diabetes Technology & Therapeutics 13:12, 1177-1186
    CrossRef

  14. 14

    A. Sola-Gazagnes, C. Vigeral. (2011) Emergent technologies applied to diabetes: What do we need to integrate continuous glucose monitoring into daily practice?. Diabetes & Metabolism 37, S65-S70
    CrossRef

  15. 15

    M. Joubert, Y. Reznik. (2011) Personal continuous glucose monitoring (CGM) in diabetes management: Review of the literature and implementation for practical use. Diabetes Research and Clinical Practice
    CrossRef

  16. 16

    Thomas Danne, Olga Kordonouri, Martin Holder, Holger Haberland, Sven Golembowski, Kerstin Remus, Sara Bläsig, Tanja Wadien, Susanne Zierow, Reinhard Hartmann, Andreas Thomas. (2011) Prevention of Hypoglycemia by Using Low Glucose Suspend Function in Sensor-Augmented Pump Therapy. Diabetes Technology & Therapeutics 13:11, 1129-1134
    CrossRef

  17. 17

    Kathryn M. Thrailkill, Cynthia S. Moreau, Christopher Swearingen, Mallik Rettiganti, Kathy Edwards, Alba E. Morales, Stephen F. Kemp, J. Paul Frindik, John L. Fowlkes. (2011) Insulin Pump Therapy Started at the Time of Diagnosis: Effects on Glycemic Control and Pancreatic β-Cell Function in Type 1 Diabetes. Diabetes Technology & Therapeutics 13:10, 1023-1030
    CrossRef

  18. 18

    Wojciech Fendler, Anna Iza Baranowska, Beata Mianowska, Agnieszka Szadkowska, Wojciech Mlynarski. (2011) Three-year comparison of subcutaneous insulin pump treatment with multi-daily injections on HbA1c, its variability and hospital burden of children with type 1 diabetes. Acta Diabetologica
    CrossRef

  19. 19

    David Kerr, Helen Partridge. (2011) Deus ex machina: The use of technology in type 1 diabetes. Primary Care Diabetes 5:3, 159-165
    CrossRef

  20. 20

    Rebecca Thompson. (2011) ‘Get going and stay going’ with pump therapy – educating adolescents about pumps. Paediatrics and Child Health 21:9, 431-433
    CrossRef

  21. 21

    M. Torres Lacruz, R. Barrio Castellanos, B. García Cuartero, A. Gómez Gila, I. González Casado, F. Hermoso López, C. Luzuriaga Tomás, M. Oyarzabal Irigoyen, I. Rica Etxebarria, M. Rodríguez Rigual. (2011) Estado actual y recomendaciones sobre la utilización de los sistemas de monitorización continua de glucosa en niños y adolescentes con diabetes mellitus tipo 1. Anales de Pediatría 75:2, 134.e1-134.e6
    CrossRef

  22. 22

    Andriani Vazeou. (2011) Continuous blood glucose monitoring in diabetes treatment. Diabetes Research and Clinical Practice 93, S125-S130
    CrossRef

  23. 23

    Robert H. Slover, John B. Welsh, Amy Criego, Stuart A. Weinzimer, Steven M. Willi, Michael A. Wood, William V. Tamborlane. (2011) Effectiveness of sensor-augmented pump therapy in children and adolescents with type 1 diabetes in the STAR 3 study. Pediatric Diabetesno-no
    CrossRef

  24. 24

    Włodzimierz Łuczyński, Agnieszka Szypowska, Barbara Głowińska-Olszewska, Artur Bossowski. (2011) Overweight, obesity and features of metabolic syndrome in children with diabetes treated with insulin pump therapy. European Journal of Pediatrics 170:7, 891-898
    CrossRef

  25. 25

    Ana María Gómez Medina, Alejandro Marín Sánchez. (2011) Monitorización continua de glucosa en tiempo real: imprescindible su uso combinado con infusión continua de insulina. Avances en Diabetología 27:4, 143-150
    CrossRef

  26. 26

    Eda Cengiz, Jennifer L Sherr, Stuart A Weinzimer, William V Tamborlane. (2011) New-generation diabetes management: glucose sensor-augmented insulin pump therapy. Expert Review of Medical Devices 8:4, 449-458
    CrossRef

  27. 27

    John B. Buse, George Dailey, Andrew A. Ahmann, Richard M. Bergenstal, Jennifer B. Green, Tim Peoples, Robert J. Tanenberg, Qingqing Yang. (2011) Baseline Predictors of A1C Reduction in Adults Using Sensor-Augmented Pump Therapy or Multiple Daily Injection Therapy: The STAR 3 Experience. Diabetes Technology & Therapeutics 13:6, 601-606
    CrossRef

  28. 28

    Alfred Penfornis, Estelle Personeni, Sophie Borot. (2011) Evolution of Devices in Diabetes Management. Diabetes Technology & Therapeutics 13:S1, S-93-S-102
    CrossRef

  29. 29

    A. Hartemann, O. Bourron. (2011) Pompes à insuline, systèmes de mesure continue de la glycémie: est-ce vraiment « ce qu’il y a de meilleur » ? Pour tous les patients ?. Médecine des Maladies Métaboliques 5:3, 303-307
    CrossRef

  30. 30

    Melissa Y. Tjota, Benjamin M. Kozak, Eric M. Chang, Kelly L. Close. (2011) Journal of Diabetes NEWS. Journal of Diabetes 3:2, 95-102
    CrossRef

  31. 31

    David Smith, Patrik Španěl, Anthony A Fryer, Fahmy Hanna, Gordon A A Ferns. (2011) Can volatile compounds in exhaled breath be used to monitor control in diabetes mellitus?. Journal of Breath Research 5:2, 022001
    CrossRef

  32. 32

    Fiona J. White, Simon A. Jones. (2011) The use of continuous glucose monitoring in the practical management of glycogen storage disorders. Journal of Inherited Metabolic Disease 34:3, 631-642
    CrossRef

  33. 33

    D. Albrecht, J. Puder, U. Keller, H. Zulewski. (2011) Potential of education-based insulin therapy for achievement of good metabolic control: a real-life experience. Diabetic Medicine 28:5, 539-542
    CrossRef

  34. 34

    Daniela Elleri, Janet M. Allen, Marianna Nodale, Malgorzata E. Wilinska, Jasdip S. Mangat, Anne Mette F. Larsen, Carlo L. Acerini, David B. Dunger, Roman Hovorka. (2011) Automated Overnight Closed-Loop Glucose Control in Young Children with Type 1 Diabetes. Diabetes Technology & Therapeutics 13:4, 419-424
    CrossRef

  35. 35

    HR Murphy, D Elleri, K Kumareswaran, R Hovorka. (2011) Technological advances in pregnancy complicated by type 1 diabetes. Practical Diabetes International 28:3, 104-105a
    CrossRef

  36. 36

    Jesús Moreno-Fernández, Pedro Benito-López, Álvaro García-Manzanares, Julia Silva, María López, Miguel Aguirre, Inés R Gómez García. (2011) Resultados del sistema Paradigm Real Time 722® en una serie de pacientes con diabetes mellitus tipo I. Avances en Diabetología 27:2, 42-46
    CrossRef

  37. 37

    Subhashni D. Singh Joy. (2011) Sensor-Augmented Insulin Pumps vs. Injections. AJN, American Journal of Nursing 111:3, 62
    CrossRef

  38. 38

    Roman Hovorka. (2011) Closed-loop insulin delivery: from bench to clinical practice. Nature Reviews Endocrinology 7:7, 385-395
    CrossRef

  39. 39

    2011. Type 1 Diabetes: Insulin Treatment. , 64-82.
    CrossRef

  40. 40

    S. Shalitin, H. Peter Chase. (2011) Diabetes technology and treatments in the paediatric age group. International Journal of Clinical Practice 65, 76-82
    CrossRef

  41. 41

    , P. Zimmer. (2011) Bedeutung von körperlicher Aktivität beim Typ-1-Diabetes. Der Diabetologe 7:1, 27-30
    CrossRef

  42. 42

    William V. Tamborlane. (2011) Severe hypoglycemia in youth with T1DM: going, going … but not yet gone. Pediatric Diabetes 12:1, 1-3
    CrossRef

  43. 43

    Daniela Elleri, David B Dunger, Roman Hovorka. (2011) Closed-loop insulin delivery for treatment of type 1 diabetes. BMC Medicine 9:1, 120
    CrossRef

  44. 44

    Alberto de Leiva-Hidalgo, Eulàlia Bruguès-Bruguès, Alejandra de Leiva-Pérez. (2011) From pancreatic extracts to artificial pancreas: History, science and controversies about the discovery of the pancreatic antidiabetic hormone. Avances en Diabetología 27:1, 27-38
    CrossRef

  45. 45

    Aaron Benjamin Skolnik, Michele Burns Ewald. (2010) Case Files of the Harvard Medical Toxicology Fellowship at Children’s Hospital Boston: An Insulin Overdose. Journal of Medical Toxicology 6:4, 413-419
    CrossRef

  46. 46

    Andrea E. Scaramuzza, Gian Vincenzo Zuccotti. (2010) Commentary on ‘Continuous subcutaneous insulin infusion (CSII) versus multiple insulin injections for type 1 diabetes mellitus’ with a response from the review authors. Evidence-Based Child Health: A Cochrane Review Journal 5:4, 1870-1872
    CrossRef

  47. 47

    O. Kordonouri, E. Pankowska, B. Rami, T. Kapellen, R. Coutant, R. Hartmann, K. Lange, M. Knip, T. Danne. (2010) Sensor-augmented pump therapy from the diagnosis of childhood type 1 diabetes: results of the Paediatric Onset Study (ONSET) after 12 months of treatment. Diabetologia 53:12, 2487-2495
    CrossRef

  48. 48

    (2010) Sensor-Augmented Insulin-Pump Therapy in Type 1 Diabetes. New England Journal of Medicine 363:21, 2070-2071
    Full Text

  49. 49

    Wolpert, Howard A., . (2010) Continuous Glucose Monitoring — Coming of Age. New England Journal of Medicine 363:4, 383-384
    Full Text

Letters