An Entirely Subcutaneous Implantable Cardioverter–Defibrillator

Gust H. Bardy, M.D., Warren M. Smith, M.B., Margaret A. Hood, M.B., Ian G. Crozier, M.B., Iain C. Melton, M.B., Luc Jordaens, M.D., Ph.D., Dominic Theuns, Ph.D., Robert E. Park, M.B., David J. Wright, M.D., Derek T. Connelly, M.D., Simon P. Fynn, M.D., Francis D. Murgatroyd, M.D., Johannes Sperzel, M.D., Jörg Neuzner, M.D., Stefan G. Spitzer, M.D., Andrey V. Ardashev, M.D., Ph.D., Amo Oduro, M.B., B.S., Lucas Boersma, M.D., Ph.D., Alexander H. Maass, M.D., Isabelle C. Van Gelder, M.D., Ph.D., Arthur A. Wilde, M.D., Ph.D., Pascal F. van Dessel, M.D., Reinoud E. Knops, M.D., Craig S. Barr, M.B., Pierpaolo Lupo, M.D., Riccardo Cappato, M.D., and Andrew A. Grace, M.B., Ph.D.

N Engl J Med 2010; 363:36-44July 1, 2010

Abstract

Background

Implantable cardioverter–defibrillators (ICDs) prevent sudden death from cardiac causes in selected patients but require the use of transvenous lead systems. To eliminate the need for venous access, we designed and tested an entirely subcutaneous ICD system.

Methods

First, we conducted two short-term clinical trials to identify a suitable device configuration and assess energy requirements. We evaluated four subcutaneous ICD configurations in 78 patients who were candidates for ICD implantation and subsequently tested the best configuration in 49 additional patients to determine the subcutaneous defibrillation threshold in comparison with that of the standard transvenous ICD. Then we evaluated the long-term use of subcutaneous ICDs in a pilot study, involving 6 patients, which was followed by a trial involving 55 patients.

Results

The best device configuration consisted of a parasternal electrode and a left lateral thoracic pulse generator. This configuration was as effective as a transvenous ICD for terminating induced ventricular fibrillation, albeit with a significantly higher mean (±SD) energy requirement (36.6±19.8 J vs. 11.1±8.5 J). Among patients who received a permanent subcutaneous ICD, ventricular fibrillation was successfully detected in 100% of 137 induced episodes. Induced ventricular fibrillation was converted twice in 58 of 59 patients (98%) with the delivery of 65-J shocks in two consecutive tests. Clinically significant adverse events included two pocket infections and four lead revisions. After a mean of 10±1 months, the device had successfully detected and treated all 12 episodes of spontaneous, sustained ventricular tachyarrhythmia.

Conclusions

In small, nonrandomized studies, an entirely subcutaneous ICD consistently detected and converted ventricular fibrillation induced during electrophysiological testing. The device also successfully detected and treated all 12 episodes of spontaneous, sustained ventricular tachyarrhythmia. (ClinicalTrials.gov numbers, NCT00399217 and NCT00853645.)

Media in This Article

Figure 1Four Configurations of a Subcutaneous Implantable Cardioverter–Defibrillator.
Figure 2Locations of the Components of a Subcutaneous Implantable Cardioverter–Defibrillator In Situ.
Article

The use of implantable cardioverter–defibrillators (ICDs) is an established therapy for the prevention of death from ventricular arrhythmia.1-5 However, conventional ICDs rely on transvenous leads for cardiac sensing and defibrillation. Complications of defibrillator implantation have been associated mainly with transvenous lead insertion and have included pneumothorax, hemothorax, and cardiac tamponade.6-10 Difficulties in achieving venous access can prolong the procedure and occasionally result in failed ICD implantation.11-13 In the long term, lead failure remains a major limitation in the use of ICDs, despite decades of innovations in lead design.12-22 Lead failure either generates inappropriate shocks or impedes appropriate therapy.20-23 Moreover, failed leads often require removal, a procedure that is associated with substantial morbidity and mortality.24-36 If cardiac pacing is not necessary, there may be a clinical advantage in avoiding the use of transvenous electrodes. In this report, we describe the initial evaluation of an entirely subcutaneous ICD system designed to avoid the need for the placement of sensing and therapy electrodes within or on the heart.

Methods

Study Design

We report the results of two short-term trials of a temporarily inserted subcutaneous ICD electrode system, followed by two trials of long-term subcutaneous ICD implantation of a fully functional system. All the studies were sponsored by the manufacturer of the subcutaneous ICD, Cameron Health, and were designed by six of the academic investigators. The protocols were approved by the ethics committee at each participating institution and associated national and local regulatory agencies. All study participants satisfied standard criteria for ICD implantation37 and provided written informed consent. Study data were collected by all the authors; device data were provided by engineers employed by the sponsor. The original manuscript was written by the first author with review and revision by all coauthors. All authors vouch for the accuracy and completeness of the data and the analyses.

Evaluation of Lead Configuration

From September 2001 through February 2004, we conducted the first short-term defibrillation trial to identify the best electrode configuration among those tested for the subcutaneous ICD. Four electrode configurations were selected on the basis of the use of specific anatomical landmarks: a left lateral pulse generator with an 8-cm coil electrode positioned at the left parasternal margin, a left pectoral pulse generator with a 4-cm coil electrode at the left inferior sternum, a left pectoral pulse generator with an 8-cm coil electrode curving from the left inferior sternum across to the inferior margin of the left sixth rib, and a left lateral pulse generator with a left parasternal 5-cm2 oval disk (Figure 1Figure 1Four Configurations of a Subcutaneous Implantable Cardioverter–Defibrillator.).

A total of 78 patients participated in this trial. Each patient underwent temporary subcutaneous implantation of one or more of the four device configurations evaluated and testing of the defibrillation threshold. The details of the protocol for defibrillation-threshold testing are described in the Supplementary Appendix, available with the full text of this article at NEJM.org. Testing was conducted in an interleaved fashion with the use of a Latin square design; the data were evaluated by means of analysis of variance.38,39 After completion of the study, all temporary subcutaneous devices were explanted, and each patient underwent implantation of a conventional transvenous ICD.

Comparison of Temporary Subcutaneous ICD with Transvenous ICD

From April 2004 through June 2005, in a second short-term trial involving 49 patients, we compared the best of the tested subcutaneous ICD systems in the first short-term trial (Figure 1A) with a transvenous ICD system. For each patient, both the subcutaneous and transvenous devices were implanted during the same procedure. Defibrillation thresholds were compared after both systems were in position and both surgical pockets had been closed. The system that was tested first was selected randomly. The protocol for defibrillation-threshold testing of the subcutaneous ICD was identical to that used in the first short-term trial, as described in the Supplementary Appendix. The statistical comparison of the defibrillation thresholds for the devices was performed with the use of a paired t-test. After completion of the study, the subcutaneous device was explanted.

Permanent Implantation

After the two short-term trials, we performed two trials of permanent subcutaneous ICD implantation: a pilot trial involving 6 patients who underwent implantation in July 2008 in New Zealand, followed by a trial involving 55 patients who underwent implantation in New Zealand and Europe between December 2008 and February 2009. We identified candidates for subcutaneous ICD implantation among the patients who were referred for ICD implantation at each participating center. The inclusion criterion was a class I, IIa, or IIb indication for ICD therapy.37 Exclusion criteria were an estimated glomerular filtration rate of less than 30 ml per minute, a requirement for antibradycardia pacing, a history of ventricular tachycardia at rates slower than 170 beats per minute, and documented ventricular tachycardia known to be reliably terminated with antitachycardia pacing. The primary end point was successful immediate conversion of two consecutive episodes of induced ventricular fibrillation, each with a single 65-J shock.

Subcutaneous ICD System

The subcutaneous ICD system that we tested in these studies consists of a 3-mm tripolar parasternal electrode (polycarbonate urethane 55D), which is connected to an electrically active pulse generator. The electrode is positioned parallel to and 1 to 2 cm to the left of the sternal midline, and the pulse generator is positioned over the sixth rib between the midaxillary line and the anterior axillary line (Figure 2Figure 2Locations of the Components of a Subcutaneous Implantable Cardioverter–Defibrillator In Situ.). The electrode has an 8-cm shocking coil, flanked by two sensing electrodes. The distal sensing electrode is positioned adjacent to the manubriosternal junction, and the proximal sensing electrode is positioned adjacent to the xiphoid process. The insertion of the subcutaneous ICD is guided exclusively by anatomical landmarks; no fluoroscopy is required. The surgical procedure and the device-testing protocol during implantation are described in the Supplementary Appendix.

During device operation, the cardiac rhythm is detected by the two sensing electrodes or by either of the sensing electrodes and the pulse generator. The subcutaneous ICD system automatically selects an appropriate vector for rhythm detection and for avoiding double QRS counting and T-wave oversensing. Once signals have been validated as free of noise and double detection, feature analysis and rate detection are used to sort rhythm type and determine the need for therapy. A conditional discrimination zone incorporating a feature-extraction technique can be programmed between rates of 170 and 240 beats per minute to distinguish supraventricular tachycardia from ventricular tachycardia and avoid inappropriate treatment of the former. Reconfirmation of ventricular tachyarrhythmia follows capacitor charging to avoid the delivery of shocks for nonsustained ventricular tachyarrhythmias. Testing of the device during implantation is done with the use of 65-J shocks to ensure a margin of safety. However, after the device has been implanted, it delivers only 80-J shocks. It can also reverse shock polarity automatically if the initial shock is not successful. In addition, demand pacing at 50 beats per minute is available for 30 seconds after a shock, with the use of a 200-mA biphasic transthoracic pulse. Pacing is activated only after more than 3.5 seconds of post-shock asystole.

All device settings are automated except for shock therapy (on/off), pacing after a shock (on/off), conditional discrimination of supraventricular tachycardia (on/off), and the upper-rate cutoff for the conditional shock zone (between 170 and 240 beats per minute). Data storage includes pre-event electrograms and rhythm markers through event termination. Up to 24 treated episodes can be stored, with up to 120 seconds of data per episode.

Results

Evaluation of Lead Configuration

In the study comparing four lead configurations, the mean (±SD) age of the 78 patients was 61±11 years (range, 31 to 80), and 72 of the patients were men. The average weight was 82.4±15.2 kg (range, 53.0 to 143.5 [182±34 lb; range, 117 to 316]). The mean ejection fraction was 0.35±0.14 (range, 0.10 to 0.69). The mean defibrillation thresholds were 32.5±17.0 J (95% confidence interval [CI], 27.8 to 37.3) for configuration 1, 40.4±13.7 J (95% CI, 35.4 to 45.4) for configuration 2, 40.1±14.9 J (95% CI, 33.7 to 46.5) for configuration 3, and 34.3±12.1 J (95% CI, 28.8 to 39.8) for configuration 4 (Figure 3AFigure 3Energy Delivered with the Subcutaneous Implantable Cardioverter–Defibrillator (ICD), According to Lead Configuration, and a Comparison of Defibrillation Thresholds in Transvenous and Subcutaneous ICDs.). The left lateral pulse generator with the 8-cm parasternal coil electrode (Figure 1A) had the lowest mean defibrillation threshold, although the differences among the configurations were not significant (P=0.07 for all comparisons by analysis of variance).

Temporary Subcutaneous ICD versus Transvenous ICD

In the study comparing the subcutaneous ICD with the transvenous ICD, the mean age of the 49 patients was 64±11 years (range, 42 to 79), and 47 of the patients were men. The average weight was 85.3±12.8 kg (range, 61.0 to 114.0 [188±28 lb; range, 134 to 251]). The mean ejection fraction was 0.37±0.13 (range, 0.19 to 0.70). The mean defibrillation threshold was 11.1±8.5 J (95% CI, 8.6 to 13.5) with the transvenous ICD and 36.6±19.8 J (95% CI, 31.1 to 42.5) with the subcutaneous ICD (P<0.001) (Figure 3B). The transvenous device in one patient and the subcutaneous device in another patient failed to terminate induced ventricular fibrillation at maximum device output. In the patient whose subcutaneous ICD failed defibrillation testing, the parasternal electrode had been incorrectly positioned approximately 6 cm to the left of the sternum, beyond the left lateral margin of the heart.

Permanent Subcutaneous ICD Pilot Study

Six patients requiring ICD therapy underwent permanent subcutaneous device implantation in the pilot study. The mean age of the patients was 60±11 years (range, 46 to 72), with a mean weight of 99.0±12.0 kg (range, 87.0 to 114.0 [218±26 lb; range, 192 to 251]). All the patients were men. The mean ejection fraction was 0.23±0.07 (range, 0.15 to 0.35). Five of the patients had coronary disease, and one had nonischemic cardiomyopathy. Two had undergone previous cardiac surgery. One patient had a secondary-prevention indication, and five had a primary-prevention indication.

All six patients underwent successful implantation of the subcutaneous ICD, and in all the patients, defibrillation with 65-J submaximal shocks was successful during two consecutive episodes of induced ventricular fibrillation. A total of 18 episodes of ventricular fibrillation were induced (with one patient having multiple episodes of nonsustained ventricular fibrillation terminating before shock delivery); all the episodes were appropriately detected. After 488±2 days of follow-up (95 patient-months of subcutaneous ICD therapy), no spontaneous episodes of ventricular tachycardia or ventricular fibrillation had occurred in the six patients, and all were well, with no device-related complications or inappropriate shocks. Figure 4Figure 4Chest Radiographs and an Electrocardiogram in a Patient Who Underwent Placement and Testing of a Subcutaneous Implantable Cardioverter–Defibrillator (ICD). shows data from one patient, with postoperative chest radiographs and an electrogram from one episode of induction and termination of ventricular fibrillation.

European Clinical Trial

For the European single-group clinical trial, 65 patients who presented for ICD implantation satisfied the enrollment criteria. Eight patients declined to participate in the study, and in two cases, the patient's physician opted for implantation of a transvenous ICD. Thus, 55 patients were enrolled in the trial, and all received a subcutaneous ICD. The clinical characteristics of the patients are shown in Table 1Table 1Clinical Characteristics of 55 Patients in the European Clinical Trial of a Subcutaneous Implantable Cardioverter–Defibrillator (ICD)..

Defibrillation testing was not possible in two patients because of intraoperative hemodynamic instability in one patient and failure to induce ventricular fibrillation in the other. Therefore, 53 patients were evaluated for sensing and defibrillation during implantation. Of 137 episodes of induced ventricular fibrillation, 100% were detected by the subcutaneous ICD. In 52 of the 53 patients who were tested (98%), two consecutive episodes of induced arrhythmia were successfully converted at 65 J. Among these 52 patients, conversion was achieved with standard polarity in 50 patients and with reverse polarity in 2 patients. In the 53rd patient, defibrillation at 65 J was achieved during the first induction but not during the second induction. As specified by the protocol, this patient received a transvenous ICD. In another patient, it was necessary to reposition the electrode, which was initially inserted in an inappropriate location 6 cm from the midline. The mean time to delivery of a shock was 14.0±2.5 seconds. The mean duration of the procedure, which was performed for the first time by most of the practitioners, including device insertion and at least two induction and termination tests, was 67±33 minutes. The procedure time was reduced to 55±23 minutes for practitioners who completed at least three implantations.

After 10±1 months and 46 patient-years of follow-up, 54 of 55 patients (98%) were alive. One death from renal failure occurred 6 months after device implantation in an 84-year-old patient. Eighteen days before he died, the patient requested that his subcutaneous ICD be turned off. A pocket infection developed in two patients; pocket revision was performed in one patient, and the other elected to discontinue defibrillator therapy. There were no cases of pocket erosion. No lead fractures developed in any patient, and no generator migration occurred. Minor lead migration was noted during follow-up in two patients. Another three patients had parasternal lead dislodgment due to inadequate anchoring of the distal tip of the electrode. In each of these patients, lead repositioning was required within a week after surgery. In another patient, lead dislodgment occurred at 6 months, during vigorous physical activity.

In one patient, oversensing occurred because of inadequate placement of the electrode in the header block, resulting in inadequate contact with one of the sensing electrodes. This problem was addressed the day after surgery with reprogramming of the detection vector rather than reoperation. Inappropriate sensing due to muscle noise occurred in three patients; the problem was addressed with device reprogramming in all three cases. One patient who had received a new drug therapy had inappropriate sensing (double detection) after the narrow QRS complex in sinus rhythm changed to a right bundle-branch block during sinus tachycardia (150 beats per minute). During 3 months of further follow-up after the detection algorithm was revised, no inappropriate shocks occurred. No shocks were delivered inappropriately for atrial fibrillation, sinus tachycardia, or supraventricular tachycardia at any time during the study when such episodes occurred at rates of more than 170 beats per minute.

A total of 12 episodes of spontaneous ventricular tachycardia were detected and successfully treated in three patients, including 1 episode after the above-mentioned software revisions (see the figure in the Supplementary Appendix). All patients were treated before the onset of syncope, and there were no adverse events. One of these patients was successfully treated for repetitive ventricular tachycardia (“VT storm”), including seven successive episodes of ventricular tachycardia.

Discussion

In the studies reported here, we describe the initial evaluation of an entirely subcutaneous ICD system. We identified a suitable device configuration, assessed the defibrillation threshold of the device in comparison to that of the standard transvenous ICD, and conducted two small, single-group trials of permanent device implantation. In the permanent-implantation studies, the subcutaneous ICD successfully and consistently detected and converted ventricular fibrillation that was induced during electrophysiological testing. The device also successfully detected and treated 12 episodes of spontaneous ventricular tachyarrhythmia that occurred in patients who were enrolled in the European clinical trial.

The goal of developing a subcutaneous ICD was to overcome some of the problems that are associated with transvenous leads in conventional ICDs.3,6-11,14,15,40 Such a device could potentially reduce or eliminate problems such as failure to achieve vascular access, intravascular injury, and lead failure requiring difficult procedures for extraction and replacement. Additional potential benefits of such a device include the preservation of venous access for other uses and the avoidance of radiation exposure during fluoroscopy, which is required for transvenous ICD implantation. These benefits would be especially important for young patients, in whom leads may fail during the decades that therapy is needed.41

The need for ICD systems that avoid the use of transvenous leads has been recognized previously,42-49 and some earlier exploratory efforts led to the present work.50,51 Some physicians have adapted existing technology to treat children with limited venous access. Leads have been inserted in unusual locations, and epicardial and subcutaneous leads have been used.44-46 Other subcutaneous defibrillation systems have been shown to work with an approximate tripling of the energy that is required by a transvenous system, findings that are consistent with the results of our study.47,48 However, such systems have not incorporated subcutaneous rhythm detection but instead have used transvenous or epicardial sensing electrodes for detection.

Our studies are preliminary, early-phase trials that were primarily intended to show the feasibility of an entirely subcutaneous ICD. They provide limited information regarding the detection and conversion of ventricular tachyarrhythmia in the clinical setting, despite the demonstration of consistent detection and termination of ventricular fibrillation at the time of implantation. These studies cannot show whether subcutaneous ICDs are superior to conventional transvenous ICDs with respect to such characteristics as lead stability or failure. Indeed, the initial experience includes several cases of problems such as lead migration, lead dislodgment, and inappropriate sensing. With respect to each of these issues, the subcutaneous ICD system was adjusted shortly after implantation in an effort to improve the system's reliability. Problems with lead dislodgment have led to the introduction of an anchoring sleeve and a new surgical technique. Inappropriate sensing has been addressed through software revision, with no incidents in the 3 months after revision. However, these modifications will require further testing in additional groups of patients. Ultimately, the relative benefit of subcutaneous ICDs, as compared with transvenous ICDs, will need to be shown in large, long-term, randomized, prospective, multicenter clinical trials.

In addition, there are inherent limitations of this device design. Although transient post-shock pacing is available, the subcutaneous ICD cannot provide long-term pacing. It is therefore not an alternative to transvenous ICDs when antibradycardia pacing is required. Also, the subcutaneous ICD is not designed to treat patients with ventricular tachycardia at rates slower than 170 beats per minute. The lack of capability to provide antitachycardia pacing may be a limitation in patients with frequent, recurrent, monomorphic ventricular tachycardia.

In conclusion, we found that in small, nonrandomized studies, an entirely subcutaneous ICD system successfully and consistently detected and converted episodes of ventricular fibrillation that were induced during electrophysiological testing. It also successfully detected and treated all 12 episodes of spontaneous, sustained ventricular tachyarrhythmia.

Supported by Cameron Health.

Disclosure forms provided by the authors are available with the full text of this article at NEJM.org.

This article (10.1056/NEJMoa0909545) was published on May 12, 2010, at NEJM.org.

We thank Jon Hunt, Ph.D., Linda Smith, B.A., Jay Warren, M.S.E.E., Stephen O'Connor, Ph.D., Rick Sanghera, B.S.E.E., Jill Anderson, B.S.N., and George Johnson, B.S.E.E., for their assistance.

Source Information

The authors' affiliations are listed in the Appendix.

Address reprint requests to Dr. Bardy at the Seattle Institute for Cardiac Research, 10115 NE 24th St., Bellevue, WA 98004, or at .

Appendix

The authors' affiliations are as follows: the Seattle Institute for Cardiac Research, Seattle (G.H.B.); Auckland City Hospital, Auckland (W.M.S., M.A.H.); and Christchurch Hospital, Christchurch (I.G.C., I.C.M., R.E.P.) — both in New Zealand; Liverpool Heart and Chest Hospital, Liverpool (D.J.W.); Glasgow Royal Infirmary, Glasgow (D.T.C.); Kings College Hospital, London (F.D.M.); the Departments of Cardiology (S.P.F., A.A.G.) and Anaesthetics and Intensive Care (A.O.), Papworth Hospital, Cambridge; Russell's Hall Hospital, Dudley (C.S.B.); and the School of Biological Sciences, University of Cambridge, Cambridge (A.A.G.) — all in the United Kingdom; Burdenko Hospital, Moscow (A.V.A.); I.R.C.C.S. Istituto Policlinico San Donato, Milan (P.L., R.C.); Kerckhoff-Klinik, Bad Nauheim (J.S.); Akademische Lehrpraxisklinik der TU Dresden, Dresden (S.G.S.); and Klinikum Kassel, Kassel (J.N.) — all in Germany; and Erasmus Medical Center, Rotterdam (L.J., D.T.); St. Antonius Hospital, Nieuwegein (L.B.); University Medical Center Groningen, University of Groningen, Groningen (A.H.M., I.C.V.G.); and Academic Medical Center, Amsterdam (A.A.W., P.F.D., R.E.K.) — all in the Netherlands.

References

References

  1. 1

    The Antiarrhythmics versus Implantable Defibrillators (AVID) Investigators. A comparison of antiarrhythmic-drug therapy with implantable defibrillators in patients resuscitated from near-fatal ventricular arrhythmias. N Engl J Med 1997;337:1576-1583
    Full Text | Web of Science | Medline

  2. 2

    Moss AJ, Zareba W, Hall WJ, et al. Prophylactic implantation of a defibrillator in patients with myocardial infarction and reduced ejection fraction. N Engl J Med 2002;346:877-883
    Full Text | Web of Science | Medline

  3. 3

    Bardy GH, Lee KL, Mark DB, et al. Amiodarone or an implantable cardioverter-defibrillator for congestive heart failure. N Engl J Med 2005;352:225-237[Erratum, N Engl J Med 2005;352:2146.]
    Full Text | Web of Science | Medline

  4. 4

    Myerburg RJ. Implantable cardioverter-defibrillators after myocardial infarction. N Engl J Med 2008;359:2245-2253
    Full Text | Web of Science | Medline

  5. 5

    Epstein AE. Benefits of the implantable cardioverter-defibrillator. J Am Coll Cardiol 2008;52:1122-1127
    CrossRef | Web of Science | Medline

  6. 6

    Grimm W, Menz V, Hoffmann J, et al. Complications of third-generation implantable cardioverter defibrillator therapy. Pacing Clin Electrophysiol 1999;22:206-211
    CrossRef | Web of Science | Medline

  7. 7

    van Rooden CJ, Molhoek SG, Rosendaal FR, Schalij MJ, Meinders AE, Huisman MV. Incidence and risk factors of early venous thrombosis associated with permanent pacemaker leads. J Cardiovasc Electrophysiol 2004;15:1258-1262
    CrossRef | Web of Science | Medline

  8. 8

    Alter P, Waldhans S, Plachta E, Moosdorf R, Grimm W. Complications of implantable cardioverter defibrillator therapy in 440 consecutive patients. Pacing Clin Electrophysiol 2005;28:926-932
    CrossRef | Web of Science | Medline

  9. 9

    Khairy P, Landzberg MJ, Gatzoulis MA, et al. Transvenous pacing leads and systemic thromboemboli in patients with intracardiac shunts: a multicenter study. Circulation 2006;113:2391-2397
    CrossRef | Web of Science | Medline

  10. 10

    Epstein AE, Baker JH II, Beau SL, Deering TF, Greenberg SM, Goldman DS. Performance of the St. Jude Medical Riata leads. Heart Rhythm 2009;6:204-209
    CrossRef | Web of Science | Medline

  11. 11

    Jones DG, Stiles MK, Stewart JT, Armstrong GP. Ultrasound-guided venous access for permanent pacemaker leads. Pacing Clin Electrophysiol 2006;29:852-857
    CrossRef | Web of Science | Medline

  12. 12

    Lelakowski J, Majewski J, Malecka B, Bednarek J, Stypula P, Szeglowski M. Retrospective analysis of reasons for failure of DDD pacemaker implantation in patients operated on between 1993 and 2005. Cardiol J 2007;14:155-159

  13. 13

    Sherrid MV, Daubert JP. Risks and challenges of implantable cardioverter-defibrillators in young adults. Prog Cardiovasc Dis 2008;51:237-263
    CrossRef | Web of Science | Medline

  14. 14

    Sohail MR, Uslan DZ, Khan AH, et al. Management and outcome of permanent pacemaker and implantable cardioverter-defibrillator infections. J Am Coll Cardiol 2007;49:1851-1859
    CrossRef | Web of Science | Medline

  15. 15

    Danik SB, Mansour M, Singh J, et al. Increased incidence of subacute lead perforation noted with one implantable cardioverter-defibrillator. Heart Rhythm 2007;4:439-442
    CrossRef | Web of Science | Medline

  16. 16

    Hauser RG, Almquist AK. Learning from our mistakes? Testing new ICD technology. N Engl J Med 2008;359:2517-2519
    Full Text | Web of Science | Medline

  17. 17

    Hauser RG, Maron BJ. Lessons from the failure and recall of an implantable cardioverter-defibrillator. Circulation 2005;112:2040-2042
    CrossRef | Web of Science | Medline

  18. 18

    Maisel WH, Moynahan M, Zuckerman BD, et al. Pacemaker and ICD generator malfunctions: analysis of Food and Drug Administration annual reports. JAMA 2006;295:1901-1906
    CrossRef | Web of Science | Medline

  19. 19

    Hauser RG, Kallinen LM, Almquist AK, Gornick CC, Katsiyiannis WT. Early failure of a small-diameter high-voltage implantable cardioverter-defibrillator lead. Heart Rhythm 2007;4:892-896
    CrossRef | Web of Science | Medline

  20. 20

    Dorwarth U, Frey B, Dugas M, et al. Transvenous defibrillation leads: high incidence of failure during long-term follow-up. J Cardiovasc Electrophysiol 2003;14:38-43
    CrossRef | Web of Science | Medline

  21. 21

    Ellenbogen KA, Wood MA, Shepard RK, et al. Detection and management of an implantable cardioverter defibrillator lead failure: incidence and clinical implications. J Am Coll Cardiol 2003;41:73-80
    CrossRef | Web of Science | Medline

  22. 22

    Kleemann T, Becker T, Doenges K, et al. Annual rate of transvenous defibrillation lead defects in implantable cardioverter-defibrillators over a period of >10 years. Circulation 2007;115:2474-2480
    CrossRef | Web of Science | Medline

  23. 23

    Tung R, Zimetbaum P, Josephson ME. A critical appraisal of implantable cardioverter-defibrillator therapy for the prevention of sudden cardiac death. J Am Coll Cardiol 2008;52:1111-1121
    CrossRef | Web of Science | Medline

  24. 24

    Byrd CL, Wilkoff BL, Love CJ, et al. Intravascular extraction of problematic or infected permanent pacemaker leads: 1994-1996: U.S. Extraction Database, MED Institute. Pacing Clin Electrophysiol 1999;22:1348-1357
    CrossRef | Web of Science | Medline

  25. 25

    Duray GZ, Schmitt J, Cicek-Hartvig S, Hohnloser SH, Israel CW. Complications leading to surgical revision in implantable cardioverter defibrillator patients: comparison of patients with single-chamber, dual-chamber, and biventricular devices. Europace 2009;11:297-302
    CrossRef | Web of Science | Medline

  26. 26

    Eckstein J, Koller MT, Zabel M, et al. Necessity from surgical revision of defibrillator leads implanted long-term: causes and management. Circulation 2008;117:2727-2733
    CrossRef | Web of Science | Medline

  27. 27

    Corrado A, Gasparini G, Raviele A. Lead malfunctions in implantable cardioverter defibrillators: where are we and where should we go? Europace 2009;11:276-277
    CrossRef | Web of Science | Medline

  28. 28

    Wilkoff BL. Lead failures: dealing with even less perfect. Heart Rhythm 2007;4:897-899
    CrossRef | Web of Science | Medline

  29. 29

    Maisel WH. Semper fidelis -- consumer protection for patients with implanted medical devices. N Engl J Med 2008;358:985-987
    Full Text | Web of Science | Medline

  30. 30

    Cannom DS, Fisher J. The Fidelis recall: how much pressure can the ICD world bear? Pacing Clin Electrophysiol 2008;31:1233-1235
    CrossRef | Web of Science | Medline

  31. 31

    Farwell D, Green MS, Lemery R, Gollob MH, Birnie DH. Accelerating risk of Fidelis lead fracture. Heart Rhythm 2008;5:1375-1379
    CrossRef | Web of Science | Medline

  32. 32

    Food and Drug Administration. Statement on Medtronic's voluntary market suspension of their Sprint Fidelis defibrillator leads. October 15, 2007. (Accessed April 27, 2010, at http://www.fda.gov/NewsEvents/Newsroom/PressAnnouncements/2007/ucm109007.htm.)

  33. 33

    Ellenbogen KA, Wood MA, Swerdlow CD. The Sprint Fidelis lead fracture story: what do we really know and where do we go from here? Heart Rhythm 2008;5:1380-1381
    CrossRef | Web of Science | Medline

  34. 34

    Gould PA, Gula LJ, Champagne J, et al. Outcome of advisory implantable cardioverter-defibrillator replacement: one-year follow-up. Heart Rhythm 2008;5:1675-1681
    CrossRef | Web of Science | Medline

  35. 35

    Maisel WH. Implantable cardioverter-defibrillator lead complication: when is an outbreak out of bounds? Heart Rhythm 2008;5:1673-1674
    CrossRef | Web of Science | Medline

  36. 36

    Schoenfeld MH. The “natural” history of implantable defibrillators under advisory. Heart Rhythm 2008;5:1682-1684
    CrossRef | Web of Science | Medline

  37. 37

    Epstein AE, DiMarco JP, Ellenbogen KA, et al. ACC/AHA/HRS 2008 guidelines for device-based therapy of cardiac rhythm abnormalities: executive summary. Heart Rhythm 2008;5:934-955[Erratum, Heart Rhythm 2009;6(1):e1.]
    CrossRef | Web of Science | Medline

  38. 38

    Armitage P, Berry G, Matthews JNS. Statistical methods in medical research. 4th ed. Malden, MA: Blackwell Science, 2002.

  39. 39

    Winer BJ. Statistical principles in experimental design. 2nd ed. New York: McGraw-Hill, 1971.

  40. 40

    Al-Khatib SM, Greiner MA, Peterson ED, Hernandez AF, Schulman KA, Curtis LH. Patient and implanting physician factors associated with mortality and complications after implantable cardioverter-defibrillator implantation, 2002-2005. Circ Arrhythmia Electrophysiol 2008;1:240-249
    CrossRef | Web of Science

  41. 41

    Sacher F, Probst V, Iesaka Y, et al. Outcome after implantation of a cardioverter-defibrillator in patients with Brugada syndrome: a multicenter study. Circulation 2006;114:2317-2324
    CrossRef | Web of Science | Medline

  42. 42

    Friedman RA, Garson A Jr. Implantable defibrillators in children: from whence to shock. J Cardiovasc Electrophysiol 2001;12:361-362
    CrossRef | Web of Science | Medline

  43. 43

    Alexander ME, Cecchin F, Walsh EP, Triedman JK, Bevilacqua LM, Berul CI. Implications of implantable cardioverter defibrillator therapy in congenital heart disease and pediatrics. J Cardiovasc Electrophysiol 2004;15:72-76
    CrossRef | Web of Science | Medline

  44. 44

    Rhee EK. Finite element modeling of novel ICD configurations in pediatric and congenital heart disease: validation of the MacGyver principle? Heart Rhythm 2008;5:573-574
    CrossRef | Web of Science | Medline

  45. 45

    Nanthakumar K. Trying to make “sense” of extra thoracic implantable defibrillator. Heart Rhythm 2007;5:35-36
    CrossRef | Web of Science | Medline

  46. 46

    Jolley M, Stinstra J, Pieper S, et al. A computer modeling tool for comparing novel ICD electrode orientations in children and adults. Heart Rhythm 2008;5:565-572
    CrossRef | Web of Science | Medline

  47. 47

    Gradaus R, Hammel D, Kotthoff S, Bocker D. Nonthoracotomy implantable cardioverter defibrillator placement in children: use of subcutaneous array leads and abdominally placed implantable cardioverter defibrillators in children. J Cardiovasc Electrophysiol 2001;12:356-360
    CrossRef | Web of Science | Medline

  48. 48

    Lieberman R, Havel WJ, Rashba E, DeGroot PJ, Stromberg K, Shorofsky SR. Acute defibrillation performance of a novel, non-transvenous shock pathway in adult ICD indicated patients. Heart Rhythm 2008;5:28-34
    CrossRef | Web of Science | Medline

  49. 49

    Burke MC, Coman JA, Cates AW, et al. Defibrillation energy requirements using a left anterior chest cutaneous to subcutaneous shocking vector: implications for a total subcutaneous implantable defibrillator. Heart Rhythm 2005;2:1332-1338
    CrossRef | Web of Science | Medline

  50. 50

    Cappato R, Bardy G. Sub-cutaneous electrical defibrillation in canines. J Am Coll Cardiol 2005;45:Suppl 1:106A-106A
    Web of Science

  51. 51

    Bardy GH, Cappato R, Smith WM, et al. The totally subcutaneous ICD system. Pacing Clin Electrophysiol 2002;25:Suppl:II-578

Citing Articles (41)

Citing Articles

  1. 1

    Gustavo Lopera, Gerald A. Beathard, Jose Exaire, Roger Carrillo. (2012) Cardiac implantable electronic devices in end-stage renal disease patients: preservation of central venous circulation. Journal of Interventional Cardiac Electrophysiology
    CrossRef

  2. 2

    Petra Opić, Elisabeth M. W. J. Utens, Philip Moons, Dominic A. M. J. Theuns, Arie P. J. Dijk, Elke S. Hoendermis, Hubert W. Vliegen, Natasja M. S. Groot, Maarten Witsenburg, Martin Schalij, Jolien W. Roos-Hesselink. (2012) Psychosocial impact of implantable cardioverter defibrillators (ICD) in young adults with Tetralogy of Fallot. Clinical Research in Cardiology
    CrossRef

  3. 3

    , J. C. Deharo, M. G. Bongiorni, A. Rozkovec, F. Bracke, P. Defaye, I. Fernandez-Lozano, P. G. Golzio, B. Hansky, C. Kennergren, A. S. Manolis, P. Mitkowski, E. S. Platou, , C. Love, B. Wilkoff. (2012) Pathways for training and accreditation for transvenous lead extraction: a European Heart Rhythm Association position paper. Europace 14:1, 124-134
    CrossRef

  4. 4

    P. A. Mehta, J. Bostock, C. A. Rinaldi. (2012) A modified subcutaneous implantable cardioverter-defibrillator implant in a patient with a previous left ventricular epicardial defibrillation patch. Europace 14:1, 149-150
    CrossRef

  5. 5

    José M. Guerr, Concepción Alonso-Martín, Enrique Rodríguez-Font, Xavier Viñolas. (2012) Novedades en electrofisiología cardiaca y arritmias. Revista Española de Cardiología 65, 73-81
    CrossRef

  6. 6

    Ian Crozier, Warren Smith. (2011) Modern Device Technologies. Heart, Lung and Circulation
    CrossRef

  7. 7

    Tim Smith, Dominic A.M.J. Theuns, Kadir Caliskan, Luc Jordaens. (2011) Long-Term Follow-Up of Prophylactic Implantable Cardioverter-Defibrillator-Only Therapy: Comparison of Ischemic and Nonischemic Heart Disease. Clinical Cardiology 34:12, 761-767
    CrossRef

  8. 8

    Stavros E. Mountantonakis, Mathew D. Hutchinson. (2011) Indications for Implantable Cardioverter-Defibrillator Placement in Ischemic Cardiomyopathy and after Myocardial Infarction. Current Heart Failure Reports 8:4, 252-259
    CrossRef

  9. 9

    S. Warren, G. A. Giridharan, R. D. Dowling, P. A. Spence, L. Tompkins, Eric Gratz, L. C. Sherwood, M. A. Sobieski, C. R. Bartoli, M. S. Slaughter, Robert S. Keynton, S. C. Koenig. (2011) Feasibility of Subcutaneous ECG Leads for Synchronized Timing of a Counterpulsation Device. Cardiovascular Engineering and Technology
    CrossRef

  10. 10

    BRUCE S. STAMBLER. (2011) ICD Arrhythmia Detection and Discrimination Algorithms: Whose is Best?. Journal of Cardiovascular Electrophysiologyno-no
    CrossRef

  11. 11

    Elia Maria, Lorenzo Bonetti, Giampiero Patrizi, John Scrivener, Alberto Andraghetti, Franco Gregorio, Alessio Montin, Giulia Zuccon, Stefano Cappelli. (2011) Implantation of a completely subcutaneous ICD system: case report of a patient with Brugada syndrome and state of the art. Journal of Interventional Cardiac Electrophysiology
    CrossRef

  12. 12

    MARC W. DEYELL, DAVID J. CALLANS. (2011) Between an ICD and a Hard Place. Journal of Cardiovascular Electrophysiology 22:10, 1120-1122
    CrossRef

  13. 13

    Ian Crozier. (2011) The subcutaneous defibrillator will replace the transvenous defibrillator. Journal of Interventional Cardiac Electrophysiology 32:1, 73-77
    CrossRef

  14. 14

    J. Köbe, S. Zumhagen, F. Reinke, E. Schulze-Bahr, L. Eckardt. (2011) Komplett subkutaner Kardioverter-Defibrillator (S-ICD®). Herz 36:7, 586-591
    CrossRef

  15. 15

    MICHAEL R. GOLD, DOMINIC A. THEUNS, BRADLEY P. KNIGHT, J. LACY STURDIVANT, RICK SANGHERA, KENNETH A. ELLENBOGEN, MARK A. WOOD, MARTIN C. BURKE. (2011) Head-To-Head Comparison of Arrhythmia Discrimination Performance of Subcutaneous and Transvenous ICD Arrhythmia Detection Algorithms: The START Study. Journal of Cardiovascular Electrophysiologyno-no
    CrossRef

  16. 16

    CHRISTINE TOMPKINS, RHONDALYN MCLEAN, ALAN CHENG, JEFFREY A. BRINKER, JOSEPH E. MARINE, SAMAN NAZARIAN, DAVID D. SPRAGG, SUNIL SINHA, HENRY HALPERIN, GORDON F. TOMASELLI, RONALD D. BERGER, HUGH CALKINS, CHARLES A. HENRIKSON. (2011) End-Stage Renal Disease Predicts Complications in Pacemaker and ICD Implants. Journal of Cardiovascular Electrophysiology 22:10, 1099-1104
    CrossRef

  17. 17

    Lara Dabiri Abkenari, Dominic A. M. J. Theuns, Suzanne D. A. Valk, Yves Belle, Natasja M. Groot, David Haitsma, Agnes Muskens-Heemskerk, Tamas Szili-Torok, Luc Jordaens. (2011) Clinical experience with a novel subcutaneous implantable defibrillator system in a single center. Clinical Research in Cardiology 100:9, 737-744
    CrossRef

  18. 18

    Robert F. Rea. (2011) Unusual ICD Access. Cardiac Electrophysiology Clinics 3:3, 441-450
    CrossRef

  19. 19

    Marwan Refaat, Samir Saba. (2011) Total Subcutaneous Defibrillators: The Current Status and Future Prospect of a New Technology. Cardiac Electrophysiology Clinics 3:3, 403-408
    CrossRef

  20. 20

    Genevieve Wanucha. (2011) New technologies aim to take shock out of treating irregular hearts. Nature Medicine 17:8, 906-906
    CrossRef

  21. 21

    Susan Currie, Elspeth B. Elliott, Godfrey L. Smith, Christopher M. Loughrey. (2011) Two candidates at the heart of dysfunction: The ryanodine receptor and calcium/calmodulin protein kinase II as potential targets for therapeutic intervention—An in vivo perspective. Pharmacology & Therapeutics 131:2, 204-220
    CrossRef

  22. 22

    Uri Elkayam. (2011) Clinical Characteristics of Peripartum Cardiomyopathy in the United States. Journal of the American College of Cardiology 58:7, 659-670
    CrossRef

  23. 23

    2011. Appendix. , 386-418.
    CrossRef

  24. 24

    John A Dodson, Mathew S Maurer. (2011) Changing nature of cardiac interventions in older adults. Aging Health 7:2, 283-295
    CrossRef

  25. 25

    K. Damman, A. H. Maass, P. van der Meer. (2011) Heart failure highlights in 2010. European Journal of Heart Failure 13:4, 358-363
    CrossRef

  26. 26

    PETR NEUZIL, JAN PETRU, VIVEK Y. REDDY. (2011) Subcutaneous ICD Implantation in the Setting of an Occluded Superior Vena Cava. Journal of Cardiovascular Electrophysiology 22:4, 475-476
    CrossRef

  27. 27

    Teun van der Bom, Paul Luijendijk, Berto J Bouma, Dave R Koolbergen, Joris R de Groot, Barbara JM Mulder. (2011) Treatment of congenital heart disease: risk-reducing measures in young adults. Future Cardiology 7:2, 227-240
    CrossRef

  28. 28

    SVEN ZUMHAGEN, ANDREW A. GRACE, STEPHEN O'CONNOR, ANDREAS LÖHER, JULIA KÖBE, LARS ECKARDT, ERIC SCHULZE-BAHR. (2011) Totally Subcutaneous Implantable Cardioverter Defibrillator with an Alternative, Right Parasternal, Electrode Placement. Pacing and Clinical Electrophysiologyno-no
    CrossRef

  29. 29

    James P. Daubert, Claudio Schuger. (2011) Percutaneous intravascular defibrillator: Preliminary data and many questions. Heart Rhythm 8:2, 293-294
    CrossRef

  30. 30

    Warren M. Smith, Gust Bardy, Margaret Hood, Andrew A. Grace. 2011. Subcutaneous Implantable Cardioverter-Defibrillators. , 428-433.
    CrossRef

  31. 31

    Peter H. Belott, Dwight W. Reynolds. 2011. Permanent Pacemaker and Implantable Cardioverter-Defibrillator Implantation. , 443-515.
    CrossRef

  32. 32

    Kathy L. Lee, Chu-Pak Lau. 2011. Leadless Pacing Concepts. , 434-440.
    CrossRef

  33. 33

    Shin-Hwa Lee, Jong-Sook Park, Choon-Sik Park. (2011) The Search for Genetic Variants and Epigenetics Related to Asthma. Allergy, Asthma and Immunology Research 3:4, 236
    CrossRef

  34. 34

    Antonio Hernández-Madrid, Roberto Matía Francés, Concepción Moro. (2011) Novedades en electrofisiología cardiaca y arritmias. Revista Española de Cardiología 64, 81-90
    CrossRef

  35. 35

    J. van Opstal, G. Geskes, L. Debie. (2011) A completely subcutaneous implantable cardioverter defibrillator system functioning simultaneously with an endocardial implantable cardioverter defibrillator programmed as pacemaker. Europace 13:1, 141-142
    CrossRef

  36. 36

    Zeliha Koyak, Joris R de Groot, Barbara JM Mulder. (2010) Interventional and surgical treatment of cardiac arrhythmias in adults with congenital heart disease. Expert Review of Cardiovascular Therapy 8:12, 1753-1766
    CrossRef

  37. 37

    H. Nagele. (2010) Implantable cardioverter defibrillator in paediatric patients. Europace 12:12, 1659-1659
    CrossRef

  38. 38

    Marcin Kowalski, Patricia Nicolato, Gautham Kalahasty, Vigneshwar Kasirajan, Mark A. Wood, Kenneth A. Ellenbogen, Richard K. Shepard. (2010) An alternative technique of implanting a nontransvenous implantable cardioverter-defibrillator system in adults with no or limited venous access to the heart. Heart Rhythm 7:11, 1572-1577
    CrossRef

  39. 39

    Gust H. Bardy. (2010) The end is near . . . . of tinkering that is. Heart Rhythm 7:11, 1578-1579
    CrossRef

  40. 40

    (2010) An Entirely Subcutaneous Implantable Cardioverter–Defibrillator. New England Journal of Medicine 363:16, 1577-1578
    Full Text

  41. 41

    Paolo Spirito, Giuseppe Boriani. (2010) Device therapy: A subcutaneous ICD—preliminary results. Nature Reviews Cardiology 7:10, 543-544
    CrossRef

Letters