Original Article

Continuous Glucose Monitoring and Intensive Treatment of Type 1 Diabetes

The Juvenile Diabetes Research Foundation Continuous Glucose Monitoring Study Group

N Engl J Med 2008; 359:1464-1476October 2, 2008

Abstract

Background

The value of continuous glucose monitoring in the management of type 1 diabetes mellitus has not been determined.

Methods

In a multicenter clinical trial, we randomly assigned 322 adults and children who were already receiving intensive therapy for type 1 diabetes to a group with continuous glucose monitoring or to a control group performing home monitoring with a blood glucose meter. All the patients were stratified into three groups according to age and had a glycated hemoglobin level of 7.0 to 10.0%. The primary outcome was the change in the glycated hemoglobin level at 26 weeks.

Results

The changes in glycated hemoglobin levels in the two study groups varied markedly according to age group (P=0.003), with a significant difference among patients 25 years of age or older that favored the continuous-monitoring group (mean difference in change, −0.53%; 95% confidence interval [CI], −0.71 to −0.35; P<0.001). The between-group difference was not significant among those who were 15 to 24 years of age (mean difference, 0.08; 95% CI, −0.17 to 0.33; P=0.52) or among those who were 8 to 14 years of age (mean difference, −0.13; 95% CI, −0.38 to 0.11; P=0.29). Secondary glycated hemoglobin outcomes were better in the continuous-monitoring group than in the control group among the oldest and youngest patients but not among those who were 15 to 24 years of age. The use of continuous glucose monitoring averaged 6.0 or more days per week for 83% of patients 25 years of age or older, 30% of those 15 to 24 years of age, and 50% of those 8 to 14 years of age. The rate of severe hypoglycemia was low and did not differ between the two study groups; however, the trial was not powered to detect such a difference.

Conclusions

Continuous glucose monitoring can be associated with improved glycemic control in adults with type 1 diabetes. Further work is needed to identify barriers to effectiveness of continuous monitoring in children and adolescents. (ClinicalTrials.gov number, NCT00406133.)

Media in This Article

Figure 1Glycated Hemoglobin Levels during 26-Week Study Period, According to Age.
Figure 2Cumulative Distribution of Glycated Hemoglobin Levels, According to Age.
Article

Despite the increased use of insulin pumps and multiple-injection regimens and the introduction of insulin analogues, intensive treatment of type 1 diabetes mellitus often does not achieve the target glycated hemoglobin levels recommended by the Diabetes Control and Complications Trial (DCCT) more than 15 years ago.1 Although self-monitoring of blood glucose plays an important role in achieving target glycated hemoglobin levels, few patients with type 1 diabetes measure glucose levels after meals or overnight. Consequently, postprandial hyperglycemia and asymptomatic nocturnal hypoglycemia are commonly seen, even in patients with well-controlled type 1 diabetes who measure blood glucose several times daily with a home glucose meter.2-4 In many patients, fear of hypoglycemia5 or development of hypoglycemia-associated autonomic failure6 hinders the successful implementation of intensive insulin therapy.

The availability of devices for continuous glucose monitoring permits the measurement of interstitial glucose in an ongoing fashion. However, the first-generation continuous monitors either provided data only for short-term retrospective analysis7 or were too difficult and uncomfortable for clinical use.8 Although not yet as accurate as blood glucose meters,9,10 newer real-time devices for continuous glucose monitoring provide improved accuracy and functionality, such as sounding alarms when the glucose trend projects future hypoglycemia, and are better tolerated by users.11 Although short-term or uncontrolled studies have suggested benefit,11-15 whether these systems help to produce a sustained lowering of glycated hemoglobin levels and reduce hypoglycemia in patients with type 1 diabetes has not been established. Therefore, in this randomized, multicenter clinical trial, we evaluated the efficacy and safety of continuous glucose monitoring in adults and children with type 1 diabetes.

Methods

Patients

All eligible patients were 8 years of age or older, had received a diagnosis of type 1 diabetes at least 1 year before randomization, either used an insulin pump or received at least three daily insulin injections, had a glycated hemoglobin level of 7.0 to 10.0%, and had not used continuous glucose monitoring at home in the 6 months leading up to the trial. Patients completed a run-in phase using a continuous glucose monitor that was modified so that the glucose values were recorded in the receiver but were not visible to the patient; we refer to this as a “blinded” continuous glucose monitor. Eligibility required that patients wear a sensor for at least 6 of 7 days before randomization, with a minimum of 96 hours of glucose values including at least 24 hours overnight, and that home blood glucose monitoring be performed at least three times daily.

Study Treatment

Patients meeting these criteria were randomly assigned to receive continuous glucose monitoring (continuous-monitoring group) or home monitoring with a blood glucose meter (control group) with the use of a permuted-block design stratified according to clinical center, age group (≥25 years, 15 to 24 years, and 8 to 14 years), and glycated hemoglobin level (≤8.0% and >8.0%). A total of 23 patients (1 in the ≥25-year group, 8 in the 15-to-24–year group, and 14 in the 8-to-14–year group) were screened for the study but were not enrolled, either because the run-in phase was not successfully completed or because the patient elected not to enter the study after using the blinded continuous glucose monitor.

Patients in the continuous-monitoring group were provided with one of the following devices: the DexCom Seven (DexCom), the MiniMed Paradigm Real-Time Insulin Pump and Continuous Glucose Monitoring System (Medtronic), or the FreeStyle Navigator (Abbott Diabetes Care). Each system consists of a glucose oxidase–based electrochemical sensor, which is placed subcutaneously and replaced every 3 to 7 days (depending on the type of device), along with a receiver to which interstitial glucose measurements are sent wirelessly and stored. Since the purpose of the study was to evaluate a treatment strategy using the technology of continuous glucose monitoring and not a specific device, a device was assigned to each patient by the clinical center on the basis of device features and the patients' preferences. Patients were instructed to use the device on a daily basis and to verify the accuracy of the glucose measurement with a home blood glucose meter (provided by the study) before making management decisions, according to the regulatory labeling of the devices. Patients in the control group were given blood glucose meters and test strips and asked to perform home blood glucose monitoring at least four times daily.

Patients in the two groups were provided with written instructions on how to use the data provided by continuous glucose monitoring and blood glucose meters to make real-time adjustments of insulin doses and on the use of computer software (for those with a home computer) to retrospectively review the glucose data to alter future insulin doses.16,17 The two study groups had the same target premeal glucose values (70 to 130 mg per deciliter [3.9 to 7.2 mmol per liter]), peak postprandial values (<180 mg per deciliter [10.0 mmol per liter]), and bedtime or overnight values (100 to 150 mg per deciliter [5.6 to 8.3 mmol per liter]). Instructions for the insulin regimen included the determination of a premeal bolus dose on the basis of the glucose level and the patient's insulin-to-carbohydrate ratio and guidelines for correcting high glucose levels outside the target range at other times. Patients using a continuous glucose monitor received additional instructions for modifying their insulin doses and treatment of hypoglycemia on the basis of the glucose trend.

Follow-up

The number of scheduled contacts with study staff was identical for both study groups. Visits were conducted at 1, 4, 8, 13, 19, and 26 weeks (±1 week), with one telephone contact between each visit, to review glucose data and adjust diabetes management. After the visits at 13 weeks and 26 weeks, the control group used a blinded continuous glucose monitor for 1 week, which was repeated if fewer than 96 hours of glucose values were obtained. Glycated hemoglobin was measured at baseline and at 13 and 26 weeks at a central laboratory at the University of Minnesota with the use of the Tosoh A1c 2.2 Plus Glycohemoglobin Analyzer method.18 Reportable adverse events included severe hypoglycemia (which was defined as an event that required assistance from another person to administer oral carbohydrate, glucagon, or other resuscitative actions1), hyperglycemia resulting in ketoacidosis, unexpected study-related or device-related events, and serious adverse events regardless of cause.

Study Design

The protocol was approved by the institutional review board at each of the 10 participating centers, which included academic, community, and managed care–based practices. Written informed consent was obtained both from adult patients and patients who were minors; the parents or guardians of minors also provided written consent. Details of the study protocol have been reported previously.16

The authors designed the study, collectively wrote the manuscript, and vouch for the completeness and accuracy of the data. Continuous glucose monitors and sensors were purchased at a bulk discount price from DexCom, Medtronic, and Abbott Diabetes Care. Home glucose meters and test strips were provided to the study by LifeScan and Abbott Diabetes Care. The manufacturers had no involvement in the study design, data accrual or analysis, or preparation of the manuscript.

Statistical Analysis

The primary outcome was the change in the mean glycated hemoglobin level from baseline to 26 weeks, as determined by a central laboratory. A sample size of 110 patients in each of three age groups (≥25 years, 15 to 24 years, and 8 to 14 years) was planned to have a power of 90% within each age group to detect a difference in the mean glycated hemoglobin level between study groups, assuming a population difference of 0.5%, a standard deviation of 0.9 at 26 weeks, a correlation between baseline and 26-week values of 0.58, an alpha level of 0.05, and a loss to follow-up of no more than 15%. An interim estimation of sample size that was based on only the observed variance in the change in glycated hemoglobin levels indicated that 86 patients would be needed in each age group to provide sufficient statistical power.

All analyses were performed according to the intention-to-treat principle. The primary analysis was a comparison between the two study groups of the change in the glycated hemoglobin levels from baseline to 26 weeks in analysis of covariance (ANCOVA) models, conducted separately in each of the three age groups and adjusted for the baseline glycated hemoglobin level and clinical center. There was a highly significant interaction between study group and age group (P=0.003) in an ANCOVA model containing an age group–treatment group interaction term, a finding that confirmed the need to analyze outcomes separately for each age group. As a result, a P value of 0.0167 was considered the significance level for the primary analysis in each age group to maintain an overall type I error rate of 0.05.

Within each age group, in addition to the primary ANCOVA analysis, five prespecified binary outcomes for glycated hemoglobin at 26 weeks (a relative decrease of ≥10%, a 26-week level of <7.0%, an absolute decrease of ≥0.5%, a relative increase of ≥10%, and an absolute increase of ≥0.5%) were evaluated in logistic-regression models, adjusted for the baseline glycated hemoglobin level and clinical center. In a post hoc analysis, a binary outcome of a glycated hemoglobin level of less than 7.0% with no severe hypoglycemic events at 26 weeks was similarly analyzed. Prespecified exploratory analyses were conducted to assess the consistency of the treatment effect on the change in the glycated hemoglobin level from baseline to 26 weeks in subgroups that were based on the type of insulin delivery (pump or multiple daily injections) and the baseline glycated hemoglobin level (≤8.0% or >8.0%).

Data regarding continuous glucose monitoring in both groups after the 26-week visit (blinded monitors in the control group and unblinded monitors in the continuous-monitoring group) were used to estimate the amount of time per day the glucose level was hypoglycemic (≤70 mg per deciliter or ≤50 mg per deciliter [≤3.9 or ≤2.8 mmol per liter]), hyperglycemic (>180 mg per deciliter or >250 mg per deciliter [10.0 or 13.9 mmol per liter]), and in the target range (71 to 180 mg per deciliter [3.9 to 10.0 mmol per liter]). Variability in glucose levels was assessed by computing the absolute rate of change.19 Comparisons between study groups were performed with the use of ANCOVA models based on van der Waerden normal scores and adjusted for the corresponding baseline value, baseline glycated hemoglobin level, clinical center, and type of continuous glucose monitor. In the continuous-monitoring group, the association between age group and the amount of sensor use during the 26-week period was evaluated with the use of the Kruskal–Wallis test. The association between sensor use and the baseline glycated hemoglobin level was evaluated with Spearman's correlation coefficients.

The proportions of patients who had one or more severe hypoglycemic events in each study group were compared with the use of Fisher's exact test. Incidences of hypoglycemic events were compared and confidence intervals for the treatment group difference calculated with the use of permutation tests. Similar analyses were performed for the subgroup of hypoglycemic events associated with seizure or coma.

Analyses were conducted with the use of SAS software, version 9.1 (SAS Institute). All P values are two-sided. Adjustment for imbalances between baseline factors and imputation for missing data with the use of Rubin's method20 did not alter the results (data not shown). Parallel analyses were performed on data regarding glycated hemoglobin levels and continuous glucose monitoring obtained at 13 weeks.

Results

Patients

In 2007, between February and December, 322 patients underwent randomization, with 165 patients assigned to the continuous-monitoring group and 157 to the control group. Of those patients, 98 patients were 25 years of age or older, 110 patients were 15 to 24 years old, and 114 patients were 8 to 14 years old. The majority of patients were non-Hispanic white, were using insulin pumps, were measuring glucose levels more than five times per day with a home glucose meter, and had a mean glycated hemoglobin level of 8.0% or less (Table 1Table 1Baseline Characteristics of the Patients, According to Age.).

Across the three age groups, the rate of completion of visits in the two study groups was 95 to 100% (Figure 1 in the Supplementary Appendix, available with the full text of this article at www.nejm.org). For protocol-specified telephone contacts, the rate was 93 to 98%. The 26-week visit was completed by all but two patients in the continuous-monitoring group who were 25 years of age or older, by all but one patient in the continuous-monitoring group and two patients in the control group who were 15 to 24 years of age, and by all patients who were 8 to 14 years of age. Two patients in the control group (both 8 to 14 years of age) initiated the use of continuous glucose monitoring before completing the 26-week visit.

Glycemic Control

In the primary analysis, a significant between-group difference in the change in glycated hemoglobin levels from baseline to 26 weeks was seen in patients who were 25 years of age or older favoring the continuous-monitoring group (mean difference in change, −0.53%; 95% confidence interval [CI], −0.71 to −0.35; P<0.001) but not in those 15 to 24 years of age (0.08; 95% CI, −0.17 to 0.33; P=0.52) nor in those 8 to 14 years of age (−0.13; 95% CI, −0.38 to 0.11; P=0.29).

At 26 weeks, among patients in the continuous-monitoring group who were 25 years of age or older, there were improvements in virtually all measures of glycemic control, as compared with the control group (Table 2Table 2Glycemic Outcomes at 26 Weeks, According to Age. and Figure 1Figure 1Glycated Hemoglobin Levels during 26-Week Study Period, According to Age. and Figure 2Figure 2Cumulative Distribution of Glycated Hemoglobin Levels, According to Age.). In the secondary analyses, more patients in the continuous-monitoring group had a relative reduction of 10% or more in the mean glycated hemoglobin level, as compared with baseline (P=0.003), and more achieved the target glycated hemoglobin level of less than 7.0% (P=0.005), as recommended for adults by the American Diabetes Association.21 The frequency of the combined outcome of a 26-week glycated hemoglobin level of less than 7.0% and no severe hypoglycemic events was 30% in the continuous-monitoring group and 7% in the control group (P=0.006). The amount of time per day within the target glucose range of 71 to 180 mg per deciliter was significantly greater in the continuous-monitoring group than in the control group (P<0.001).

Among patients who were 15 to 24 years of age, the mean decrease in glycated hemoglobin levels from baseline to 26 weeks was approximately 0.2% in both study groups. There were no significant differences between groups on any of the secondary glycemic measures.

Among patients who were 8 to 14 years of age, the mean decrease in glycated hemoglobin levels was 0.37% in the continuous-monitoring group, which did not differ significantly from the decrease of 0.22% in the control group. However, secondary indexes of glycemic control were improved in the continuous-monitoring group — namely, more patients had a relative reduction of 10% or more in the glycated hemoglobin level from baseline (P=0.04) and more patients had glycated hemoglobin levels of less than 7.0% (P=0.01) (Table 2 and Figure 1 and Figure 2). At 26 weeks, there were no significant differences between the two study group in outcomes as measured by continuous glucose monitoring.

Within the three age groups, the 13-week results were similar to the 26-week results (Table 1 in the Supplementary Appendix). At 26 weeks, results were consistent in subgroups based on the type of insulin delivery (pump vs. multiple daily injections) and baseline glycated hemoglobin levels (≤8.0% vs. >8.0%) (Table 2 in the Supplementary Appendix).

Hypoglycemia and Other Adverse Events

Severe hypoglycemic events were infrequent in the two study groups. In the six age-stratified groups, 5 to 10% of patients had at least one severe hypoglycemic event, with no significant differences between the two study groups on the basis of age (Table 3Table 3Hypoglycemia and Other Adverse Events, According to Age.). Likewise, there were no significant differences in the incidence of severe hypoglycemic events between study groups according to age. One patient in the continuous-monitoring group who was 30 years of age had six events involving either seizure or coma during a period in which he reported not using any long-acting insulin; for four of the six events, he reported that he had not used short-acting insulin on the day of the event.

Biochemical hypoglycemia (blood glucose, ≤70 mg per deciliter), which was evaluated by collecting data from continuous glucose monitoring in both study groups after the 26-week visit, was present for only a small portion of the day and did not differ significantly between the two groups in each age group. There were few other adverse events (Table 3).

Frequency of Sensor Use

In the continuous-monitoring group, the use of sensors was greater among patients who were 25 years of age or older than in the other two age groups (P<0.001). The use of sensors was consistently high during the 26-week period among patients in the oldest age group but declined over time in the other two age groups (Figure 3Figure 3Use of Continuous Glucose Monitors in the Continuous-Monitoring Group, According to Age.). Excluding the three patients who dropped out, only one patient who was 25 years of age or older, seven who were 15 to 24 years of age, and two who were 8 to 14 years of age discontinued continuous glucose monitoring before completing the 26-week visit. At least 6.0 days of sensor use per week was the average for 83% of patients who were 25 years of age or older, 30% who were 15 to 24 years of age, and 50% who were 8 to 14 years of age. Sensor use was not associated with the baseline glycated hemoglobin level (Spearman's correlation coefficients, 0.08, −0.02, and 0.03 in the three groups, respectively).

Discussion

In this randomized, controlled trial, we observed that the benefit associated with continuous glucose monitoring was strongly related to age. A significant difference in the primary analysis of the change in glycated hemoglobin levels from baseline to 26 weeks was seen in the predefined group of patients who were 25 years of age or older but not in the two groups of patients who were younger. In patients 25 years of age or older, substantially tighter glycemic control was evident in the continuous-monitoring group in both glycated hemoglobin levels and sensor glucose results, without a significant increase in biochemical hypoglycemia (time per day with values of ≤70 mg per deciliter on continuous glucose monitoring). More patients in the continuous-monitoring group than in the control group had a glycated hemoglobin level of less than 7.0% without having a severe hypoglycemic event.

In contrast, comparisons between study groups showed less benefit of continuous glucose monitoring among patients who were 8 to 14 years of age and no benefit among those who were 15 to 24 years of age. Among those 8 to 14 years of age, the between-group difference in mean glycated hemoglobin levels did not achieve statistical significance. However, in this age group, in secondary analyses, more patients in the continuous-monitoring group than in the control group had glycated hemoglobin levels of less than 7.0% and more had a relative reduction of 10% or more in glycated hemoglobin levels from baseline values. The latter observation is important, since the DCCT showed that a relative reduction of 10% in glycated hemoglobin levels is associated with a reduction of more than 40% in the rate of development and progression of early diabetic retinopathy.22,23

The observed age effect may be related to substantially greater use of sensors in the adults than in patients in the two younger groups. Imperfect adherence with many aspects of diabetes management has long been recognized as an obstacle to successful intensive treatment in adolescents and young adults with type 1 diabetes.24-27 Greater parental involvement could be the reason that patients in the continuous-monitoring group between the ages of 8 and 14 years had greater sensor use than the patients between the ages of 15 and 24 years. In adolescents, the transition from parental assistance with management of diabetes to patient-only management is often accompanied by deterioration of glycemic control.25,28

Severe hypoglycemic events were infrequent and did not differ significantly according to study group. This finding must be interpreted with caution, since the trial was not powered to detect a between-group difference in such events. In the two groups, the rate of severe hypoglycemia was much lower than that reported in the DCCT.1,29 The use of a blinded continuous glucose monitor in the control group also provided the opportunity to compare the exposure of the two groups to biochemical hypoglycemia, which is often asymptomatic. In this regard, among patients 25 years of age or older, it was noteworthy that the decrease in glycated hemoglobin levels was not associated with an increase in hypoglycemia. This finding is in direct contrast to that of the DCCT, which showed that the rate of hypoglycemic events increased in patients who lowered their glycated hemoglobin levels.1,29

With respect to the generalizability of the results, it is important to recognize that before the study, patients were receiving intensive insulin therapy with either an insulin pump or multiple daily injections and frequent home blood glucose monitoring, and most had better-than-average glycated hemoglobin levels.30,31 In addition, to be eligible for the study, patients needed to show the ability to wear a sensor and insert a new sensor at home. Therefore, the results do not shed light on the use of such devices in a less well controlled, less motivated population of patients with type 1 diabetes. Although the results in patients using multiple daily injections were similar to the results in those using an insulin pump, the number of patients using multiple daily injections was too small for a definitive assessment.

The results of our study indicate that continuous glucose monitoring improves glycated hemoglobin levels and may enhance the management of type 1 diabetes in adults who have the motivation to use this technology and the capability to incorporate it into their own daily diabetes management. Further work is needed to identify and address the lack of effectiveness of continuous glucose monitoring in children and adolescents.

Supported by grants (22-2006-1107, 22-2006-1117, 22-2006-1112, 22-2006-1123, and 01-2006-8031) from the Juvenile Diabetes Research Foundation.

Dr. Tamborlane reports receiving consulting fees from Abbott Diabetes Care and Lifescan and consulting and lecture fees and grant support from Medtronic; Dr. Bode, receiving consulting and lecture fees, travel reimbursement, and grant support from Abbott Diabetes Care and Medtronic and grant support from DexCom; Dr. Buckingham, receiving lecture fees and grant support from Abbott Diabetes Care, lecture and consulting fees and grant support from Medtronic, and consulting fees from Novo Nordisk and serving on an advisory board for Lifescan; Dr. Chase, receiving lecture fees from Abbott Diabetes Care and Sanofi-Aventis and grant support from Symlin; Dr. Fiallo-Scharer, receiving supplies for research from Abbott Diabetes Care and Medtronic; Dr. Fox, receiving supplies for research from Abbott Diabetes Care; Dr. Hirsch, receiving consulting fees and travel reimbursement from Abbott Diabetes Care and grant support from Medtronic; Dr. Laffel, receiving consulting fees from Lifescan and Roche, consulting and lecture fees from Abbott Diabetes Care, consulting fees and grant support from Medtronic, and grant support from the Charles Hood Foundation; Dr. Mauras, receiving grant support from AstraZeneca; Dr. Weinzimer, receiving lecture fees and travel reimbursement from Medtronic; Dr. Wilson, receiving equipment and software from Abbott Diabetes Care and Medtronic and grant support from Medtronic; and Dr. Wolpert, receiving consulting fees from Abbott Diabetes Care and grant support from Medtronic. No other potential conflict of interest relevant to this article was reported.

The affiliations of the members of the writing committee and those of investigators in the Juvenile Diabetes Research Foundation Continuous Glucose Monitoring Study Group are listed in the Appendix.

The members of the writing committee (William V. Tamborlane, M.D., Roy W. Beck, M.D., Ph.D., Bruce W. Bode, M.D., Bruce Buckingham, M.D., H. Peter Chase, M.D., Robert Clemons, M.D., Rosanna Fiallo-Scharer, M.D., Larry A. Fox, M.D., Lisa K. Gilliam, M.D., Ph.D., Irl B. Hirsch, M.D., Elbert S. Huang, M.D., M.P.H., Craig Kollman, Ph.D., Aaron J. Kowalski, Ph.D., Lori Laffel, M.D., M.P.H., Jean M. Lawrence, Sc.D., M.P.H., M.S.S.A., Joyce Lee, M.D., M.P.H., Nelly Mauras, M.D., Michael O'Grady, Ph.D., Katrina J. Ruedy, M.S.P.H., Michael Tansey, M.D., Eva Tsalikian, M.D., Stuart Weinzimer, M.D., Darrell M. Wilson, M.D., Howard Wolpert, M.D., Tim Wysocki, Ph.D., and Dongyuan Xing, M.P.H.) assume responsibility for the overall content and integrity of the article.

This article (10.1056/NEJMoa0805017) was published at www.nejm.org on September 8, 2008.

Source Information

Address reprint requests to Dr. Beck at the Jaeb Center for Health Research, 15310 Amberly Dr., #350, Tampa, FL 33647, or at .

Appendix

The following members of the writing committee and study investigators are listed in order according to the number of patients enrolled at each clinical center: Clinical Center: Barbara Davis Center for Childhood Diabetes, University of Colorado, Denver: H.P. Chase (principal investigator), R. Fiallo-Scharer, L. Messer, V. Gage, P. Burdick; Pediatric Adolescent, and Young Adult Section, Joslin Diabetes Center, Boston: L. Laffel (principal investigator), K. Milaszewski, K. Pratt, E. Bismuth, J. Keady, M. Lawlor; Stanford University, Stanford, CA: B. Buckingham (principal investigator), D.M. Wilson, J. Block, K. Benassi; University of Iowa Carver College of Medicine, Iowa City: E. Tsalikian (principal investigator), M. Tansey, D. Kucera, J. Coffey, J. Cabbage; Adult Section, Joslin Diabetes Center, Boston: H. Wolpert (principal investigator), G. Shetty, A. Atakov-Castillo, J. Giusti, S. O'Donnell, S. Ghiloni; University of Washington, Seattle: I.B. Hirsch (principal investigator), L.K. Gilliam, K. Fitzpatrick, D. Khakpour; Nemours Children's Clinic, Jacksonville, FL: T. Wysocki (coprincipal investigator), L.A. Fox (coprincipal investigator), N. Mauras, K. Englert, J. Permuy; Atlanta Diabetes Associates, Atlanta: B.W. Bode (principal investigator), K. O'Neil, L. Tolbert; Kaiser Permanente, San Diego, CA: J.M. Lawrence (coprincipal investigator), R. Clemons (coprincipal investigator), M. Maeva, B. Sattler; Yale University School of Medicine, New Haven, CT: S. Weinzimer (principal investigator), W.V. Tamborlane, B. Ives, J. Bosson-Heenan. Coordinating Center: Jaeb Center for Health Research, Tampa, FL: R.W. Beck, K.J. Ruedy, C. Kollman, D. Xing, J. Jackson. Central Laboratory: University of Minnesota Central Laboratory: M. Steffes, J.M. Bucksa, M.L. Nowicki, C. Van Hale, V. Makky. Cost-Effectiveness Investigators: National Opinion Research Center, University of Chicago, Chicago: M. O'Grady, E. Huang, A. Basu, D.O. Meltzer, L. Zhao; University of Michigan: J. Lee. Juvenile Diabetes Research Foundation: A.J. Kowalski. Operations Committee: L. Laffel (cochair), W.V. Tamborlane (cochair), R.W. Beck, A.J. Kowalski, K.J. Ruedy. Data and Safety Monitoring Board: R.S. Weinstock (chair), B.J. Anderson, D. Kruger, L. LaVange, H. Rodriguez.

References

References

  1. 1

    The Diabetes Control and Complications Trial Research Group. The effect of intensive treatment of diabetes on the development and progression of long-term complications in insulin-dependent diabetes mellitus. N Engl J Med 1993;329:977-986
    Full Text | Web of Science | Medline

  2. 2

    Boland E, Monsod T, Delucia M, Brandt CA, Fernando S, Tamborlane WV. Limitations of conventional methods of self-monitoring of blood glucose: lessons learned from 3 days of continuous glucose sensing in pediatric patients with type 1 diabetes. Diabetes Care 2001;24:1858-1862
    CrossRef | Web of Science | Medline

  3. 3

    Kaufman FR, Austin J, Neinstein A, et al. Nocturnal hypoglycemia detected with the Continuous Glucose Monitoring System in pediatric patients with type 1 diabetes. J Pediatr 2002;141:625-630
    CrossRef | Web of Science | Medline

  4. 4

    Bode BW, Schwartz S, Stubbs HA, Block JE. Glycemic characteristics in continuously monitored patients with type 1 and type 2 diabetes: normative values. Diabetes Care 2005;28:2361-2366
    CrossRef | Web of Science | Medline

  5. 5

    Thompson CJ, Cummings JF, Chalmers J, Gould C, Newton RW. How have patients reacted to the implications of the DCCT? Diabetes Care 1996;19:876-879
    CrossRef | Web of Science | Medline

  6. 6

    Cryer PE. Diverse causes of hypoglycemia-associated autonomic failure in diabetes. N Engl J Med 2004;350:2272-2279
    Full Text | Web of Science | Medline

  7. 7

    Klonoff DC. Continuous glucose monitoring: roadmap for 21st century diabetes therapy. Diabetes Care 2005;28:1231-1239
    CrossRef | Web of Science | Medline

  8. 8

    Diabetes Research in Children Network (DirecNet) Study Group. A randomized multicenter trial comparing the GlucoWatch Biographer with standard glucose monitoring in children with type 1 diabetes. Diabetes Care 2005;28:1101-1106
    CrossRef | Web of Science | Medline

  9. 9

    Wilson DM, Beck RW, Tamborlane WV, et al. The accuracy of the FreeStyle Navigator continuous glucose monitoring system in children with type 1 diabetes. Diabetes Care 2007;30:59-64
    CrossRef | Web of Science | Medline

  10. 10

    Kovatchev B, Anderson S, Heinemann L, Clarke W. Comparison of the numerical and clinical accuracy of four continuous glucose monitors. Diabetes Care 2008;31:1160-1164
    CrossRef | Web of Science | Medline

  11. 11

    Diabetes Research in Children Network (DirecNet) Study Group. Continuous glucose monitoring in children with type 1 diabetes. J Pediatr 2007;151:388-393
    CrossRef | Web of Science | Medline

  12. 12

    Garg S, Zisser H, Schwartz S, et al. Improvement in glycemic excursions with a transcutaneous, real-time continuous glucose sensor: a randomized controlled trial. Diabetes Care 2006;29:44-50
    CrossRef | Web of Science | Medline

  13. 13

    Halvorson M, Carpenter S, Kaiserman K, Kaufman FR. A pilot trial in pediatrics with the sensor-augmented pump: combining real-time continuous glucose monitoring with the insulin pump. J Pediatr 2007;150:103-105
    CrossRef | Web of Science | Medline

  14. 14

    Diabetes Research in Children Network (DirecNet) Study Group. FreeStyle Navigator continuous glucose monitoring system use in children with type 1 diabetes using glargine-based multiple daily dose regimens: results of a pilot trial. Diabetes Care 2008;31:525-527
    CrossRef | Web of Science | Medline

  15. 15

    Deiss D, Bolinder J, Riveline J-P, et al. Improved glycemic control in poorly controlled patients with type 1 diabetes using real-time continuous glucose monitoring. Diabetes Care 2006;29:2730-2732
    CrossRef | Web of Science | Medline

  16. 16

    JDRF CGM Study Group. JDRF randomized clinical trial to assess the efficacy of real-time continuous glucose monitoring in the management of type 1 diabetes: research design and methods. Diabetes Technol Ther 2008;10:310-321
    CrossRef | Web of Science | Medline

  17. 17

    Diabetes Research in Children Network (DirecNet) Study Group. Use of the DirecNet Applied Treatment Algorithm (DATA) for diabetes management with a real-time continuous glucose monitor (the Freestyle Navigator). Pediatr Diabetes 2008;9:142-147
    CrossRef | Web of Science | Medline

  18. 18

    Gibb I, Parnham A, Fonfrede M, Lecock F. Multicenter evaluation of Tosoh glycohemoglobin analyzer. Clin Chem 1999;45:1833-1841
    Web of Science | Medline

  19. 19

    Kovatchev BP, Clarke WL, Breton M, Brayman K, McCall A. Quantifying temporal glucose variability in diabetes via continuous glucose monitoring: mathematical methods and clinical application. Diabetes Technol Ther 2005;7:849-862
    CrossRef | Medline

  20. 20

    Little RJA, Rubin DB. Statistical analysis with missing data. New York: John Wiley, 1987.

  21. 21

    American Diabetes Association. Standards of medical care in diabetes -- 2008. Diabetes Care 2008;31:Suppl 1:S12-S54
    CrossRef | Web of Science | Medline

  22. 22

    The Diabetes Control and Complications Trial Research Group. The relationship of glycemic exposure (HbA1c) to the risk of development and progression of retinopathy in the Diabetes Control and Complications Trial. Diabetes 1995;44:968-983
    CrossRef | Web of Science | Medline

  23. 23

    Lachin JM, Genuth S, Nathan DM, Zinman B, Rutledge BN, DCCT/EDIC Research Group. Effect of glycemic exposure on the risk of microvascular complications in the Diabetes Control and Complications Trial -- revisited. Diabetes 2008;57:995-1001
    CrossRef | Web of Science | Medline

  24. 24

    Anderson BJ, Vangsness L, Connell A, Butler D, Goebel-Fabbri A, Laffel LMB. Family conflict, adherence, and glycaemic control in youth with short duration Type 1 diabetes. Diabet Med 2002;19:635-642
    CrossRef | Web of Science | Medline

  25. 25

    Insabella G, Grey M, Knafl G, Tamborlane W. The transition to young adulthood in youth with type 1 diabetes on intensive treatment. Pediatr Diabetes 2007;8:228-234
    CrossRef | Web of Science | Medline

  26. 26

    Laffel LM, Vangsness L, Connell A, Goebel-Fabbri A, Butler D, Anderson BJ. Impact of ambulatory, family-focused teamwork intervention on glycemic control in youth with type 1 diabetes. J Pediatr 2003;142:409-416
    CrossRef | Web of Science | Medline

  27. 27

    Wysocki T, Iannotti R, Weissberg-Benchell J, et al. Diabetes problem solving by youths with type 1 diabetes and their caregivers: measurement, validation, and longitudinal associations with glycemic control. J Pediatr Psychol 2008;33:875-884
    CrossRef | Web of Science | Medline

  28. 28

    Moreland EC, Tovar A, Zuehlke JB, Butler DA, Milaszewski K, Laffel LM. The impact of physiological, therapeutic and psychosocial variables on glycemic control in youth with type 1 diabetes mellitus. J Pediatr Endocrinol Metab 2004;17:1533-1544
    CrossRef | Web of Science | Medline

  29. 29

    The DCCT Research Group. Epidemiology of severe hypoglycemia in the Diabetes Control and Complications Trial. Am J Med 1991;90:450-459
    Web of Science | Medline

  30. 30

    The Diabetes Control and Complications Trial/Epidemiology of Diabetes Interventions and Complications (DCCT/EDIC) Study Research Group. Intensive diabetes treatment and cardiovascular disease in patients with type 1 diabetes. N Engl J Med 2005;353:2643-2653
    Full Text | Web of Science | Medline

  31. 31

    Petitti DB, Imperatore G, Palla SL, et al. Serum lipids and glucose control: the SEARCH for Diabetes in Youth study. Arch Pediatr Adolesc Med 2007;161:159-165
    CrossRef | Web of Science | Medline

Citing Articles (158)

Citing Articles

  1. 1

    A. L. Secher, A. B. Madsen, L. Ringholm, C. Barfred, E. Stage, H. U. Andersen, P. Damm, E. R. Mathiesen. (2012) Patient satisfaction and barriers to initiating real-time continuous glucose monitoring in early pregnancy in women with diabetes. Diabetic Medicine 29:2, 272-277
    CrossRef

  2. 2

    Richard R. Rubin, Mark Peyrot. (2012) Health-Related Quality of Life and Treatment Satisfaction in the Sensor-Augmented Pump Therapy for A1C Reduction 3 (STAR 3) Trial. Diabetes Technology & Therapeutics 14:2, 143-151
    CrossRef

  3. 3

    Katherine E. Iscoe, Raymond J. Davey, Paul A. Fournier. (2012) Is the Response of Continuous Glucose Monitors to Physiological Changes in Blood Glucose Levels Affected by Sensor Life?. Diabetes Technology & Therapeutics 14:2, 135-142
    CrossRef

  4. 4

    Lisa T. Meade. (2012) The Use of Continuous Glucose Monitoring in Patients with Type 2 Diabetes. Diabetes Technology & Therapeutics 14:2, 190-195
    CrossRef

  5. 5

    Shlomit Shalitin, H. Peter Chase. (2012) Diabetes technology and treatment in the paediatric age group. International Journal of Clinical Practice 66, 72-78
    CrossRef

  6. 6

    Miranda W. Langendam, Yoeri M Luijf, Lotty Hooft, J Hans DeVries, Aart H Mudde, Rob JPM Scholten, Miranda W. Langendam. 2012. Continuous glucose monitoring systems for type 1 diabetes mellitus. .
    CrossRef

  7. 7

    J. H. DeVries. (2012) Continuous glucose monitoring: coming of age?. European Journal of Endocrinology 166:1, 1-4
    CrossRef

  8. 8

    Fátima Barceló-Rico, Jorge Bondia, José Luis Díez, Paolo Rossetti. (2012) A Multiple Local Models Approach to Accuracy Improvement in Continuous Glucose Monitoring. Diabetes Technology & Therapeutics 14:1, 74-82
    CrossRef

  9. 9

    Moshe Phillip, Thomas Danne, Shlomit Shalitin, Bruce Buckingham, Lori Laffel, William Tamborlane, Tadej Battelino, . (2012) Use of continuous glucose monitoring in children and adolescents. Pediatric Diabetesn/a-n/a
    CrossRef

  10. 10

    Fernando Gómez Peralta, Cristina Abreu Padín. (2012) Monitorización continua de la glucosa en tiempo real. Indicaciones clínicas probadas. Avances en Diabetología
    CrossRef

  11. 11

    Matthew Leinung, Shannon Thompson, Elizabeth Nardacci. (2012) Benefits of Continuous Glucose Monitor Use in Clinical Practice. Endocrine Practice 1:-1, 1-14
    CrossRef

  12. 12

    Desmond Barry Keenan, John Joseph Mastrototaro, Howard Zisser, Kenneth A. Cooper, Gautham Raghavendhar, Scott W. Lee, Jonathan Yusi, Timothy Bailey, Ronald Leonard Brazg, Rajiv V. Shah. (2011) Accuracy of the Enlite 6-Day Glucose Sensor with Guardian and Veo Calibration Algorithms. Diabetes Technology & Therapeutics111206132933005
    CrossRef

  13. 13

    Jaime M. Realsen, H. Peter Chase. (2011) Recent Advances in the Prevention of Hypoglycemia in Type 1 Diabetes. Diabetes Technology & Therapeutics 13:12, 1177-1186
    CrossRef

  14. 14

    L. Ratheau, N. Jeandidier, F. Moreau, S. Sigrist, M. Pinget. (2011) How technology has changed diabetes management and what it has failed to achieve. Diabetes & Metabolism 37, S57-S64
    CrossRef

  15. 15

    C. R. Alleyn, L. M. B. Laffel, L. K. Volkening, B. J. Anderson, T. R. Nansel, T. Wysocki, J Weissberg-Benchell. (2011) Comparison of longitudinal point-of-care and high-performance liquid chromatography HbA1c measurements in a multi-centre trial. Diabetic Medicine 28:12, 1525-1529
    CrossRef

  16. 16

    M. Joubert, Y. Reznik. (2011) Personal continuous glucose monitoring (CGM) in diabetes management: Review of the literature and implementation for practical use. Diabetes Research and Clinical Practice
    CrossRef

  17. 17

    Stefano Zucchini, Mirella Scipione, Claudia Balsamo, Giulio Maltoni, Alessandra Rollo, Emanuela Molinari, Lorenza Mangoni, Alessandro Cicognani. (2011) Comparison Between Sensor-Augmented Insulin Therapy with Continuous Subcutaneous Insulin Infusion or Multiple Daily Injections in Everyday Life: 3-Day Analysis of Glucose Patterns and Sensor Accuracy in Children. Diabetes Technology & Therapeutics 13:12, 1187-1193
    CrossRef

  18. 18

    Eva Tsalikian, Larry Fox, Stuart Weinzimer, Bruce Buckingham, Neil H White, Roy Beck, Craig Kollman, Dongyuan Xing, Katrina Ruedy, . (2011) Feasibility of prolonged continuous glucose monitoring in toddlers with type 1 diabetes. Pediatric Diabetesn/a-n/a
    CrossRef

  19. 19

    P. Schaepelynck, P. Darmon, L. Molines, M.F. Jannot-Lamotte, C. Treglia, D. Raccah. (2011) Advances in pump technology: insulin patch pumps, combined pumps and glucose sensors, and implanted pumps. Diabetes & Metabolism 37, S85-S93
    CrossRef

  20. 20

    Marilyn D. Ritholz, Elizabeth A. Beverly, Katie Weinger. (2011) Digging Deeper: The Role of Qualitative Research in Behavioral Diabetes. Current Diabetes Reports 11:6, 494-502
    CrossRef

  21. 21

    A. Sola-Gazagnes, C. Vigeral. (2011) Emergent technologies applied to diabetes: What do we need to integrate continuous glucose monitoring into daily practice?. Diabetes & Metabolism 37, S65-S70
    CrossRef

  22. 22

    J.-P. Riveline. (2011) Is continuous glucose monitoring (CGM) for everyone?. Diabetes & Metabolism 37, S80-S84
    CrossRef

  23. 23

    Beate Karges, Thomas Meissner, Andrea Icks, Thomas Kapellen, Reinhard W. Holl. (2011) Management of diabetes mellitus in infants. Nature Reviews Endocrinology
    CrossRef

  24. 24

    Goran Petrovski, Cedomir Dimitrovski, Milco Bogoev, Tatjana Milenkovic, Irfan Ahmeti, Iskra Bitovska. (2011) Is There a Difference in Pregnancy and Glycemic Outcome in Patients with Type 1 Diabetes on Insulin Pump with Constant or Intermittent Glucose Monitoring? A Pilot Study. Diabetes Technology & Therapeutics 13:11, 1109-1113
    CrossRef

  25. 25

    Pratik Choudhary, Kathryn Lonnen, Celia J. Emery, Jenny V. Freeman, Ken M. McLeod, Simon R. Heller. (2011) Relationship Between Interstitial and Blood Glucose During Hypoglycemia in Subjects with Type 2 Diabetes. Diabetes Technology & Therapeutics 13:11, 1121-1127
    CrossRef

  26. 26

    Carmen U Ludwig-Seibold, Martin Holder, Birgit Rami, Klemens Raile, Bettina Heidtmann, Reinhard W Holl, , , . (2011) Continuous glucose monitoring in children, adolescents, and adults with type 1 diabetes mellitus: analysis from the prospective DPV diabetes documentation and quality management system from Germany and Austria. Pediatric Diabetesn/a-n/a
    CrossRef

  27. 27

    Matthieu Raoux, Guilherme Bontorin, Yannick Bornat, Jochen Lang, Sylvie Renaud. 2011. Bioelectronic Sensing of Insulin Demand. , 191-202.
    CrossRef

  28. 28

    J. Hermanides, K. Nørgaard, D. Bruttomesso, C. Mathieu, A. Frid, C. M. Dayan, P. Diem, C. Fermon, I. M. E. Wentholt, J. B. L. Hoekstra, J. H. DeVries. (2011) Sensor-augmented pump therapy lowers HbA1c in suboptimally controlled Type 1 diabetes; a randomized controlled trial. Diabetic Medicine 28:10, 1158-1167
    CrossRef

  29. 29

    David Kerr, Helen Partridge. (2011) Deus ex machina: The use of technology in type 1 diabetes. Primary Care Diabetes 5:3, 159-165
    CrossRef

  30. 30

    Joyce M. Lee, Kirsten Rhee, Michael J. OʼGrady, Anirban Basu, Aaron Winn, Priya John, David O. Meltzer, Craig Kollman, Lori M. Laffel, Jean M. Lawrence, William V. Tamborlane, Tim Wysocki, Dongyuan Xing, Elbert S. Huang. (2011) Health Utilities for Children and Adults With Type 1 Diabetes. Medical Care 49:10, 924-931
    CrossRef

  31. 31

    Jae Hyuk Lee, Kyuseok Kim, You Hwan Jo, Joong Eui Rhee, Jung Chan Lee, Kyung Su Kim, Woon Yong Kwon, Gil Joon Suh, Hee Chan Kim, Ho Il Yoon, Sang Heon Park. (2011) Feasibility of Continuous Glucose Monitoring in Critically Ill Emergency Department Patients. The Journal of Emergency Medicine
    CrossRef

  32. 32

    Shahar Miller, Revital Nimri, Eran Atlas, Eli A. Grunberg, Moshe Phillip. (2011) Automatic Learning Algorithm for the MD-Logic Artificial Pancreas System. Diabetes Technology & Therapeutics 13:10, 983-990
    CrossRef

  33. 33

    Rebecca Thompson. (2011) ‘Get going and stay going’ with pump therapy – educating adolescents about pumps. Paediatrics and Child Health 21:9, 431-433
    CrossRef

  34. 34

    M. Tansey, L. Laffel, J. Cheng, R. Beck, J. Coffey, E. Huang, C. Kollman, J. Lawrence, J. Lee, K. Ruedy, W. Tamborlane, T. Wysocki, D. Xing, . (2011) Satisfaction with continuous glucose monitoring in adults and youths with Type 1 diabetes. Diabetic Medicine 28:9, 1118-1122
    CrossRef

  35. 35

    D. J. Lunn, C. Wei, R. Hovorka. (2011) Fitting dynamic models with forcing functions: Application to continuous glucose monitoring in insulin therapy. Statistics in Medicine 30:18, 2234-2250
    CrossRef

  36. 36

    Olga Kordonouri, Reinhard Hartmann, Thomas Danne. (2011) Treatment of type 1 diabetes in children and adolescents using modern insulin pumps. Diabetes Research and Clinical Practice 93, S118-S124
    CrossRef

  37. 37

    Daniela Elleri, Janet M. Allen, David B. Dunger, Roman Hovorka. (2011) Closed-loop in children with type 1 diabetes: Specific challenges. Diabetes Research and Clinical Practice 93, S131-S135
    CrossRef

  38. 38

    Andriani Vazeou. (2011) Continuous blood glucose monitoring in diabetes treatment. Diabetes Research and Clinical Practice 93, S125-S130
    CrossRef

  39. 39

    M. Torres Lacruz, R. Barrio Castellanos, B. García Cuartero, A. Gómez Gila, I. González Casado, F. Hermoso López, C. Luzuriaga Tomás, M. Oyarzabal Irigoyen, I. Rica Etxebarria, M. Rodríguez Rigual. (2011) Estado actual y recomendaciones sobre la utilización de los sistemas de monitorización continua de glucosa en niños y adolescentes con diabetes mellitus tipo 1. Anales de Pediatría 75:2, 134.e1-134.e6
    CrossRef

  40. 40

    Ana María Gómez Medina, Alejandro Marín Sánchez. (2011) Monitorización continua de glucosa en tiempo real: imprescindible su uso combinado con infusión continua de insulina. Avances en Diabetología 27:4, 143-150
    CrossRef

  41. 41

    Eda Cengiz, Jennifer L Sherr, Stuart A Weinzimer, William V Tamborlane. (2011) New-generation diabetes management: glucose sensor-augmented insulin pump therapy. Expert Review of Medical Devices 8:4, 449-458
    CrossRef

  42. 42

    A. Hartemann, O. Bourron. (2011) Pompes à insuline, systèmes de mesure continue de la glycémie: est-ce vraiment « ce qu’il y a de meilleur » ? Pour tous les patients ?. Médecine des Maladies Métaboliques 5:3, 303-307
    CrossRef

  43. 43

    Anna Lindholm Olinder, Kerstin Ternulf Nyhlin, Bibbi Smide. (2011) Reasons for missed meal-time insulin boluses from the perspective of adolescents using insulin pumps: ‘lost focus’. Pediatric Diabetes 12:4pt2, 402-409
    CrossRef

  44. 44

    Hui Zheng, David M. Nathan, David A. Schoenfeld. (2011) Using A Multi-Level B-Spline Model to Analyze and Compare Patient Glucose Profiles Based on Continuous Monitoring Data. Diabetes Technology & Therapeutics 13:6, 675-682
    CrossRef

  45. 45

    Alfred Penfornis, Estelle Personeni, Sophie Borot. (2011) Evolution of Devices in Diabetes Management. Diabetes Technology & Therapeutics 13:S1, S-93-S-102
    CrossRef

  46. 46

    Satish K. Garg, Emily G. Moser. (2011) How Basal Insulin Analogs Have Changed Diabetes Care. Diabetes Technology & Therapeutics 13:S1, S-1-S-4
    CrossRef

  47. 47

    Pantea P. Minnock, Carol J. Howe. (2011) Use of Continuous Glucose Monitoring Systems in Children with Type 1 Diabetes. Critical Care Nursing Clinics of North America 23:2, 273-290
    CrossRef

  48. 48

    D. Albrecht, J. Puder, U. Keller, H. Zulewski. (2011) Potential of education-based insulin therapy for achievement of good metabolic control: a real-life experience. Diabetic Medicine 28:5, 539-542
    CrossRef

  49. 49

    Kunjan R. Dave, Antonello Pileggi, Ami P. Raval. (2011) Recurrent hypoglycemia increases oxygen glucose deprivation-induced damage in hippocampal organotypic slices. Neuroscience Letters 496:1, 25-29
    CrossRef

  50. 50

    L. B. E. A. Hoeks, W. L. Greven, H. W. de Valk. (2011) Real-time continuous glucose monitoring system for treatment of diabetes: a systematic review. Diabetic Medicine 28:4, 386-394
    CrossRef

  51. 51

    Rami Almokayyad, Robert Cuddihy. 2011. What is the Role of Self-Monitoring in Diabetes? Is there a Role for Postprandial Glucose Monitoring? How does Continuous Glucose Monitoring Integrate into Clinical Practice?. , 41-55.
    CrossRef

  52. 52

    Jesús Moreno-Fernández, Pedro Benito-López, Álvaro García-Manzanares, Julia Silva, María López, Miguel Aguirre, Inés R Gómez García. (2011) Resultados del sistema Paradigm Real Time 722® en una serie de pacientes con diabetes mellitus tipo I. Avances en Diabetología 27:2, 42-46
    CrossRef

  53. 53

    N. R. Cunningham, A. T. Vesco, L. M. Dolan, K. K. Hood. (2011) From Caregiver Psychological Distress to Adolescent Glycemic Control: The Mediating Role of Perceived Burden around Diabetes Management. Journal of Pediatric Psychology 36:2, 196-205
    CrossRef

  54. 54

    Dongyuan Xing, Craig Kollman, Roy W. Beck, William V. Tamborlane, Lori Laffel, Bruce A. Buckingham, Darrell M. Wilson, Stuart Weinzimer, Rosanna Fiallo-Scharer, Katrina J. Ruedy for the Juvenile Diabetes Res. (2011) Optimal Sampling Intervals to Assess Long-Term Glycemic Control Using Continuous Glucose Monitoring. Diabetes Technology & Therapeutics 13:3, 351-358
    CrossRef

  55. 55

    Jean-Pierre Riveline. (2011) Évaluation des mesures continues du glucose associées à la pompe à insuline. Médecine des Maladies Métaboliques 5:1, H67-H68
    CrossRef

  56. 56

    Roman Hovorka. (2011) Closed-loop insulin delivery: from bench to clinical practice. Nature Reviews Endocrinology 7:7, 385-395
    CrossRef

  57. 57

    S. Shalitin, H. Peter Chase. (2011) Diabetes technology and treatments in the paediatric age group. International Journal of Clinical Practice 65, 76-82
    CrossRef

  58. 58

    A. Liberman, B. Buckingham, M. Phillip. (2011) Diabetes technology and the human factor. International Journal of Clinical Practice 65, 83-90
    CrossRef

  59. 59

    Alicia J. Jenkins, Balasubramanian Krishnamurthy, James D. Best, Fergus J. Cameron, Peter G. Colman, Peter S. Hamblin, Michele A. O'Connell, Christine Rodda, Helena Teede, David N. O'Neal. (2011) An Algorithm Guiding Patient Responses to Real-Time-Continuous Glucose Monitoring Improves Quality of Life. Diabetes Technology & Therapeutics 13:2, 105-109
    CrossRef

  60. 60

    Andrea E. Scaramuzza, Dario Iafusco, Ivana Rabbone, Riccardo Bonfanti, Fortunato Lombardo, Riccardo Schiaffini, Pietro Buono, Sonia Toni, Valentino Cherubini, Gian Vincenzo Zuccotti, for the Diabetes Study Gr. (2011) Use of Integrated Real-Time Continuous Glucose Monitoring/Insulin Pump System in Children and Adolescents with Type 1 Diabetes: A 3-Year Follow-Up Study. Diabetes Technology & Therapeutics 13:2, 99-103
    CrossRef

  61. 61

    Ignacio Conget, Tadej Battelino, Marga Giménez, Hannah Gough, J. Castañeda, Jan Bolinder. (2011) The SWITCH Study (Sensing With Insulin pump Therapy to Control HbA 1c ): Design and Methods of a Randomized Controlled Crossover Trial on Sensor-Augmented Insulin Pump Efficacy in Type 1 Diabetes Suboptimally Controlled with Pump Therapy. Diabetes Technology & Therapeutics 13:1, 49-54
    CrossRef

  62. 62

    Alberto de Leiva-Hidalgo, Eulàlia Bruguès-Bruguès, Alejandra de Leiva-Pérez. (2011) From pancreatic extracts to artificial pancreas: History, science and controversies about the discovery of the pancreatic antidiabetic hormone. Avances en Diabetología 27:1, 27-38
    CrossRef

  63. 63

    Tania Saxl, Faaizah Khan, Matteo Ferla, David Birch, John Pickup. (2011) A fluorescence lifetime-based fibre-optic glucose sensor using glucose/galactose-binding protein. The Analyst 136:5, 968
    CrossRef

  64. 64

    S. Braun. 2011. Encapsulation of Cells (Cellular Delivery) Using Sol–Gel Systems. , 529-543.
    CrossRef

  65. 65

    Yang Gyo Kang. (2011) Application of Continuous Glucose Monitoring System (CGMS) and Patient Education. Journal of Korean Diabetes 12:3, 159
    CrossRef

  66. 66

    J. K. Snell-Bergeon, R. Roman, D. Rodbard, S. Garg, D. M. Maahs, I. E. Schauer, B. C. Bergman, G. L. Kinney, M. Rewers. (2010) Glycaemic variability is associated with coronary artery calcium in men with Type 1 diabetes: the Coronary Artery Calcification in Type 1 Diabetes study. Diabetic Medicine 27:12, 1436-1442
    CrossRef

  67. 67

    O. Kordonouri, E. Pankowska, B. Rami, T. Kapellen, R. Coutant, R. Hartmann, K. Lange, M. Knip, T. Danne. (2010) Sensor-augmented pump therapy from the diagnosis of childhood type 1 diabetes: results of the Paediatric Onset Study (ONSET) after 12 months of treatment. Diabetologia 53:12, 2487-2495
    CrossRef

  68. 68

    Marion J. Franz, Margaret A. Powers, Carolyn Leontos, Lea Ann Holzmeister, Karmeen Kulkarni, Arlene Monk, Naomi Wedel, Erica Gradwell. (2010) The Evidence for Medical Nutrition Therapy for Type 1 and Type 2 Diabetes in Adults. Journal of the American Dietetic Association 110:12, 1852-1889
    CrossRef

  69. 69

    A. Michael Albisser, Francine Inhaber. (2010) Automation of the Consensus Guidelines in Diabetes Care: Potential Impact on Clinical Inertia. Endocrine Practice 16:6, 992-1002
    CrossRef

  70. 70

    Hye Seung Jung, Hong Il Kim, Min Joo Kim, Ji Won Yoon, Hwa Young Ahn, Young Min Cho, Kook-Hwan Oh, Kwon Wook Joo, Jung Geon Lee, Seong Yeon Kim, Kyong Soo Park. (2010) Analysis of Hemodialysis-Associated Hypoglycemia in Patients with Type 2 Diabetes Using a Continuous Glucose Monitoring System. Diabetes Technology & Therapeutics 12:10, 801-807
    CrossRef

  71. 71

    Bala Venkatesh, Thomas J. Morgan, Jeremy Cohen. (2010) Interstitium: The next diagnostic and therapeutic platform in critical illness. Critical Care Medicine 38, S630-S636
    CrossRef

  72. 72

    Philip E. Cryer. (2010) Hypoglycemia in Type 1 Diabetes Mellitus. Endocrinology & Metabolism Clinics of North America 39:3, 641-654
    CrossRef

  73. 73

    David M. Maahs, Nancy A. West, Jean M. Lawrence, Elizabeth J. Mayer-Davis. (2010) Epidemiology of Type 1 Diabetes. Endocrinology & Metabolism Clinics of North America 39:3, 481-497
    CrossRef

  74. 74

    M. D. Ritholz, A. Atakov-Castillo, M. Beste, E. A. Beverly, A. Leighton, K. Weinger, H. Wolpert. (2010) Psychosocial factors associated with use of continuous glucose monitoring. Diabetic Medicine 27:9, 1060-1065
    CrossRef

  75. 75

    (2010) Validation of Measures of Satisfaction with and Impact of Continuous and Conventional Glucose Monitoring. Diabetes Technology & Therapeutics 12:9, 679-684
    CrossRef

  76. 76

    Peter W Goss, Megan A Paterson, Jay Renalson. (2010) A ‘radical’ new rural model for pediatric diabetes care. Pediatric Diabetes 11:5, 296-304
    CrossRef

  77. 77

    Hyun Gu Kang, Diane F. Mahoney, Helen Hoenig, Victor A. Hirth, Paolo Bonato, Ihab Hajjar, Lewis A. Lipsitz, . (2010) In Situ Monitoring of Health in Older Adults: Technologies and Issues. Journal of the American Geriatrics Society 58:8, 1579-1586
    CrossRef

  78. 78

    Joshua S Borus, Lori Laffel. (2010) Adherence challenges in the management of type 1 diabetes in adolescents: prevention and intervention. Current Opinion in Pediatrics 22:4, 405-411
    CrossRef

  79. 79

    D. A. Gough, L. S. Kumosa, T. L. Routh, J. T. Lin, J. Y. Lucisano. (2010) Function of an Implanted Tissue Glucose Sensor for More than 1 Year in Animals. Science Translational Medicine 2:42, 42ra53-42ra53
    CrossRef

  80. 80

    Wolpert, Howard A., . (2010) Continuous Glucose Monitoring — Coming of Age. New England Journal of Medicine 363:4, 383-384
    Full Text

  81. 81

    Bergenstal, Richard M., Tamborlane, William V., Ahmann, Andrew, Buse, John B., Dailey, George, Davis, Stephen N., Joyce, Carol, Peoples, Tim, Perkins, Bruce A., Welsh, John B., Willi, Steven M., Wood, Michael A., . (2010) Effectiveness of Sensor-Augmented Insulin-Pump Therapy in Type 1 Diabetes. New England Journal of Medicine 363:4, 311-320
    Full Text

  82. 82

    Christopher M. Ryan. 2010. Psychologic Factors and Diabetes. , 807-834.
    CrossRef

  83. 83

    Howard Wolpert, Judy Shih. 2010. New Technologies for Insulin Administration and Glucose Monitoring. , 440-451.
    CrossRef

  84. 84

    Andrew J. Farmer. 2010. Monitoring Diabetes. , 399-409.
    CrossRef

  85. 85

    Kentaro Taki, Rimei Nishimura, Aya Morimoto, Daisuke Tsujino, Yumi Miyashita, Naoko Tajima. (2010) Analysis of 24-Hour Glycemic Excursions in Patients with Type 1 Diabetes by Using Continuous Glucose Monitoring. Diabetes Technology & Therapeutics 12:7, 523-528
    CrossRef

  86. 86

    Jeroen Hermanides, Annemarie E. Engström, Iris M.E. Wentholt, Krischan D. Sjauw, Joost B.L. Hoekstra, Jose P.S. Henriques, J. Hans DeVries. (2010) Sensor-Augmented Insulin Pump Therapy to Treat Hyperglycemia at the Coronary Care Unit: A Randomized Clinical Pilot Trial. Diabetes Technology & Therapeutics 12:7, 537-542
    CrossRef

  87. 87

    H. Peter Chase, Roy W. Beck, Dongyuan Xing, William V. Tamborlane, Julie Coffey, Larry A. Fox, Brett Ives, Joyce Keady, Craig Kollman, Lori Laffel, Katrina J. Ruedy. (2010) Continuous Glucose Monitoring in Youth with Type 1 Diabetes: 12-Month Follow-Up of the Juvenile Diabetes Research Foundation Continuous Glucose Monitoring Randomized Trial. Diabetes Technology & Therapeutics 12:7, 507-515
    CrossRef

  88. 88

    Marga Giménez, Mercè Lara, Ignacio Conget. (2010) Sustained Efficacy of Continuous Subcutaneous Insulin Infusion in Type 1 Diabetes Subjects with Recurrent Non-Severe and Severe Hypoglycemia and Hypoglycemia Unawareness: A Pilot Study. Diabetes Technology & Therapeutics 12:7, 517-521
    CrossRef

  89. 89

    David Kerr, Tolu Olateju. (2010) Pascal's Wager: Combining Continuous Glucose Monitoring and Continuous Subcutaneous Insulin Infusion. Diabetes Technology & Therapeutics 12:S1, S-43-S-50
    CrossRef

  90. 90

    Jay S. Skyler. (2010) Continuous Subcutaneous Insulin Infusion—An Historical Perspective. Diabetes Technology & Therapeutics 12:S1, S-5-S-9
    CrossRef

  91. 91

    David M. Maahs, Lauren A. Horton, H. Peter Chase. (2010) The Use of Insulin Pumps in Youth with Type 1 Diabetes. Diabetes Technology & Therapeutics 12:S1, S-59-S-65
    CrossRef

  92. 92

    Eric Renard. (2010) Insulin Pump Use in Europe. Diabetes Technology & Therapeutics 12:S1, S-29-S-32
    CrossRef

  93. 93

    Matthew Leinung, Shannon Thompson, Elizabeth Nardacci. (2010) Benefits of Continuous Glucose Monitor Use in Clinical Practice. Endocrine Practice 16:3, 371-375
    CrossRef

  94. 94

    IRIS SANCHEZ. (2010) The Use of Continuous Glucose Monitoring System in the Home Health Setting. Home Healthcare Nurse: The Journal for the Home Care and Hospice Professional 28:5, 291-295
    CrossRef

  95. 95

    Nancy M. Tofil, Priya Prabhakaran. (2010) What is the role for continuous glucose monitors in the pediatric intensive care unit?*. Pediatric Critical Care Medicine 11:3, 441-443
    CrossRef

  96. 96

    F. H. El-Khatib, S. J. Russell, D. M. Nathan, R. G. Sutherlin, E. R. Damiano. (2010) A Bihormonal Closed-Loop Artificial Pancreas for Type 1 Diabetes. Science Translational Medicine 2:27, 27ra27-27ra27
    CrossRef

  97. 97

    Stephen N. Davis, Edward S. Horton, Tadej Battelino, Richard R. Rubin, Kevin A. Schulman, William V. Tamborlane. (2010) STAR 3 Randomized Controlled Trial to Compare Sensor-Augmented Insulin Pump Therapy with Multiple Daily Injections in the Treatment of Type 1 Diabetes: Research Design, Methods, and Baseline Characteristics of Enrolled Subjects. Diabetes Technology & Therapeutics 12:4, 249-255
    CrossRef

  98. 98

    Claus B. Juhl, Kurt Højlund, Rasmus Elsborg, Mikael Kjær Poulsen, Peter E. Selmar, Jens Juul Holst, Claus Christiansen, Henning Beck-Nielsen. (2010) Automated detection of hypoglycemia-induced EEG changes recorded by subcutaneous electrodes in subjects with type 1 diabetes—The brain as a biosensor. Diabetes Research and Clinical Practice 88:1, 22-28
    CrossRef

  99. 99

    J. Hermanides, J. H. DeVries. (2010) Sense and nonsense in sensors. Diabetologia 53:4, 593-596
    CrossRef

  100. 100

    PJ Hammond, SA Amiel, CM Dayan, D Kerr, JC Pickup, JAM Shaw, FM Campbell, SA Greene, PC Hindmarsh, . (2010) ABCD position statement on continuous glucose monitoring: use of glucose sensing in outpatient clinical diabetes care. Practical Diabetes International 27:2, 66-68
    CrossRef

  101. 101

    Jean Halford, Claudia Harris. (2010) Determining Clinical and Psychological Benefits and Barriers with Continuous Glucose Monitoring Therapy. Diabetes Technology & Therapeutics 12:3, 201-205
    CrossRef

  102. 102

    J. K. Mader, H. Weinhandl, G. Köhler, J. Plank, G. Bock, S. Korsatko, M. Ratzer, D. Ikeoka, H. Köhler, T. R. Pieber, M. Ellmerer. (2010) Assessment of different techniques for subcutaneous glucose monitoring in Type 1 diabetic patients during ‘real-life’ glucose excursions. Diabetic Medicine 27:3, 332-338
    CrossRef

  103. 103

    Helen R. Murphy. (2010) Integrating educational and technological interventions to improve pregnancy outcomes in women with diabetes. Diabetes, Obesity and Metabolism 12:2, 97-104
    CrossRef

  104. 104

    Roman Hovorka, Janet M Allen, Daniela Elleri, Ludovic J Chassin, Julie Harris, Dongyuan Xing, Craig Kollman, Tomas Hovorka, Anne Mette F Larsen, Marianna Nodale, Alessandra De Palma, Malgorzata E Wilinska, Carlo L Acerini, David B Dunger. (2010) Manual closed-loop insulin delivery in children and adolescents with type 1 diabetes: a phase 2 randomised crossover trial. The Lancet 375:9716, 743-751
    CrossRef

  105. 105

    P. Aschner, E. Horton, L. A. Leiter, N. Munro, J. S. Skyler, . (2010) Practical steps to improving the management of type 1 diabetes: recommendations from the Global Partnership for Effective Diabetes Management. International Journal of Clinical Practice 64:3, 305-315
    CrossRef

  106. 106

    M. Hod, L. Jovanovič. (2010) Technology and pregnancy. International Journal of Clinical Practice 64, 47-52
    CrossRef

  107. 107

    C. Wei, D. J. Lunn, C. L. Acerini, J. M. Allen, A. M. Larsen, M. E. Wilinska, D. B. Dunger, R. Hovorka. (2010) Measurement delay associated with the Guardian® RT continuous glucose monitoring system. Diabetic Medicine 27:1, 117-122
    CrossRef

  108. 108

    Daniela Elleri, Carlo L. Acerini, Janet M. Allen, Josephine Hayes, Claire Pesterfield, Malgorzata E. Wilinska, David B. Dunger, Roman Hovorka. (2010) Parental Attitudes Towards Overnight Closed-Loop Glucose Control in Children with Type 1 Diabetes. Diabetes Technology & Therapeutics 12:1, 35-39
    CrossRef

  109. 109

    Timothy Bailey, Howard Zisser, Anna Chang. (2009) New Features and Performance of a Next-Generation SEVEN-Day Continuous Glucose Monitoring System with Short Lag Time. Diabetes Technology & Therapeutics 11:12, 749-755
    CrossRef

  110. 110

    David Rodbard, Lois Jovanovic, Satish K. Garg. (2009) Responses to Continuous Glucose Monitoring in Subjects with Type 1 Diabetes Using Continuous Subcutaneous Insulin Infusion or Multiple Daily Injections. Diabetes Technology & Therapeutics 11:12, 757-765
    CrossRef

  111. 111

    S. Borot. (2009) Quelle amélioration glycémique peut-on attendre de l’utilisation de la pompe à insuline couplée à une mesure continue du glucose par des patients diabétiques de type 1 non éduqués spécifiquement à l’interprétation de ces données ? Résultats d’une étude randomisée. Diabetologia Notes de lecture 1:3-4, 44-46
    CrossRef

  112. 112

    Ricardo G. Branco, Arjun Chavan, Robert C. Tasker. (2009) Pilot evaluation of continuous subcutaneous glucose monitoring in children with multiple organ dysfunction syndrome. Pediatric Critical Care Medicine1
    CrossRef

  113. 113

    Scott Drab. (2009) Translating clinical guidelines into clinical practice: Role of the pharmacist in type 2 diabetes management. Journal of the American Pharmacists Association 49:6, e152-e162
    CrossRef

  114. 114

    Margaret T. Ryan, Vincent W. Savarese, Brian Hipszer, Ismar Dizdarevic, Mark Joseph, Nathan Shively, Jeffrey I Joseph. (2009) Continuous Glucose Monitor Shows Potential for Early Hypoglycemia Detection in Hospitalized Patients. Diabetes Technology & Therapeutics 11:11, 745-747
    CrossRef

  115. 115

    David Rodbard, Timothy Bailey, Lois Jovanovic, Howard Zisser, Roy Kaplan, Satish K. Garg. (2009) Improved Quality of Glycemic Control and Reduced Glycemic Variability with Use of Continuous Glucose Monitoring. Diabetes Technology & Therapeutics 11:11, 717-723
    CrossRef

  116. 116

    Edmond A. Ryan, Judy Germsheid. (2009) Use of Continuous Glucose Monitoring System in the Management of Severe Hypoglycemia. Diabetes Technology & Therapeutics 11:10, 635-639
    CrossRef

  117. 117

    Marilyn Cox. (2009) An Overview of Continuous Glucose Monitoring Systems. Journal of Pediatric Health Care 23:5, 344-347
    CrossRef

  118. 118

    Cosimo Giannini, Angelika Mohn, Francesco Chiarelli. (2009) Technology and the issue of cost/benefit in diabetes. Diabetes/Metabolism Research and Reviews 25:S1, S34-S44
    CrossRef

  119. 119

    Marian Rewers, Catherine Pihoker, Kim Donaghue, Ragnar Hanas, Peter Swift, Georgeanna J Klingensmith. (2009) Assessment and monitoring of glycemic control in children and adolescents with diabetes. Pediatric Diabetes 10, 71-81
    CrossRef

  120. 120

    Ilene Fennoy. (2009) Pramlintide in Pediatric Type 1 Diabetes. The Journal of Pediatrics 155:3, 308-309
    CrossRef

  121. 121

    Hans-Jacob Bangstad, Thomas Danne, Larry Deeb, Przemyslawa Jarosz-Chobot, Tatsuhiko Urakami, Ragnar Hanas. (2009) Insulin treatment in children and adolescents with diabetes. Pediatric Diabetes 10, 82-99
    CrossRef

  122. 122

    J.-P. Riveline, J. Teynie, S. Belmouaz, S. Franc, D. Dardari, M. Bauwens, V. Caudwell, S. Ragot, F. Bridoux, G. Charpentier, R. Marechaud, S. Hadjadj. (2009) Glycaemic control in type 2 diabetic patients on chronic haemodialysis: use of a continuous glucose monitoring system. Nephrology Dialysis Transplantation 24:9, 2866-2871
    CrossRef

  123. 123

    Thomas Danne. (2009) Tailoring intensified insulin therapy in children. Pediatric Diabetes 10:5, 295-297
    CrossRef

  124. 124

    T. Danne, H. W. Valk, T. Kracht, K. Walte, R. Geldmacher, L. Sölter, W. Berge, Z. K. Welsh, J. R. Bugler, K. Lange, O. Kordonouri. (2009) Reducing glycaemic variability in type 1 diabetes self-management with a continuous glucose monitoring system based on wired enzyme technology. Diabetologia 52:8, 1496-1503
    CrossRef

  125. 125

    Peter Adolfsson, Hans Örnhagen, Johan Jendle. (2009) Accuracy and Reliability of Continuous Glucose Monitoring in Individuals with Type 1 Diabetes During Recreational Diving. Diabetes Technology & Therapeutics 11:8, 493-497
    CrossRef

  126. 126

    M. A. O’Connell, S. Donath, D. N. O’Neal, P. G. Colman, G. R. Ambler, T. W. Jones, E. A. Davis, F. J. Cameron. (2009) Glycaemic impact of patient-led use of sensor-guided pump therapy in type 1 diabetes: a randomised controlled trial. Diabetologia 52:7, 1250-1257
    CrossRef

  127. 127

    Alexander C. Wiseman. (2009) Simultaneous Pancreas Kidney Transplantation: A Critical Appraisal of the Risks and Benefits Compared With Other Treatment Alternatives. Advances in Chronic Kidney Disease 16:4, 278-287
    CrossRef

  128. 128

    Kavita Kumareswaran, Mark L Evans, Roman Hovorka. (2009) Artificial pancreas: an emerging approach to treat Type 1 diabetes. Expert Review of Medical Devices 6:4, 401-410
    CrossRef

  129. 129

    Louis Monnier, Claude Colette, David R. Owens. (2009) Integrating glycaemic variability in the glycaemic disorders of type 2 diabetes: a move towards a unified glucose tetrad concept. Diabetes/Metabolism Research and Reviews 25:5, 393-402
    CrossRef

  130. 130

    John Mastrototaro, Scott Lee. (2009) The Integrated MiniMed Paradigm Real-Time Insulin Pump and Glucose Monitoring System: Implications for Improved Patient Outcomes. Diabetes Technology & Therapeutics 11:s1, S-37-S-43
    CrossRef

  131. 131

    Geoffrey McGarraugh. (2009) The Chemistry of Commercial Continuous Glucose Monitors. Diabetes Technology & Therapeutics 11:s1, S-17-S-24
    CrossRef

  132. 132

    Anne Peters. (2009) Do We Really Need Continuous Glucose Monitoring?. Diabetes Technology & Therapeutics 11:s1, S-128-S-130
    CrossRef

  133. 133

    Jay S. Skyler. (2009) Continuous Glucose Monitoring: An Overview of Its Development. Diabetes Technology & Therapeutics 11:s1, S-5-S-10
    CrossRef

  134. 134

    Kristin Fabiato, John Buse, Michelle Duclos, Joseph Largay, Camille Izlar, Thomas O'Connell, Jennifer Stallings, Kathleen Dungan. (2009) Clinical Experience with Continuous Glucose Monitoring in Adults. Diabetes Technology & Therapeutics 11:s1, S-93-S-103
    CrossRef

  135. 135

    Lisa K. Gilliam, Irl B. Hirsch. (2009) Practical Aspects of Real-Time Continuous Glucose Monitoring. Diabetes Technology & Therapeutics 11:s1, S-76-S-82
    CrossRef

  136. 136

    R. Paul Wadwa, Rosanna Fiallo-Scharer, Brandon VanderWel, Laurel H. Messer, Erin Cobry, H. Peter Chase. (2009) Continuous Glucose Monitoring in Youth with Type 1 Diabetes. Diabetes Technology & Therapeutics 11:s1, S-83-S-91
    CrossRef

  137. 137

    Aaron J. Kowalski. (2009) Can We Really Close the Loop and How Soon? Accelerating the Availability of an Artificial Pancreas: A Roadmap to Better Diabetes Outcomes. Diabetes Technology & Therapeutics 11:s1, S-113-S-119
    CrossRef

  138. 138

    Irl B. Hirsch. (2009) Continuous Glucose Monitoring: Understanding Our Current Culture. Diabetes Technology & Therapeutics 11:s1, S-131-S-132
    CrossRef

  139. 139

    Satish K. Garg. (2009) The Future of Continuous Glucose Monitoring. Diabetes Technology & Therapeutics 11:s1, S-1-S-3
    CrossRef

  140. 140

    Veronique Guyomard, Phyo Kyaw Myint. (2009) Optimum control of blood glucose for prevention and treatment of ischemic and hemorrhagic stroke. Current Treatment Options in Cardiovascular Medicine 11:3, 201-211
    CrossRef

  141. 141

    Laurissa Kashmer, William Clarke, Matthew Gurka, Swati Elchuri, Maren Nyer, Linda Gonder-Frederick. (2009) Predictors of Parental Interest in Continuous Glucose Monitoring for Children with Type 1 Diabetes. Diabetes Technology & Therapeutics 11:6, 373-378
    CrossRef

  142. 142

    A. Liebl. (2009) Lösungsansätze bei frustraner Insulinpumpentherapie. Der Diabetologe 5:4, 275-282
    CrossRef

  143. 143

    William V Tamborlane, Roy W Beck. (2009) Continuous glucose monitoring in type 1 diabetes mellitus. The Lancet 373:9677, 1744-1746
    CrossRef

  144. 144

    D. Cooke, S. J. Hurel, A. Casbard, L. Steed, S. Walker, S. Meredith, A. J. Nunn, A. Manca, M. Sculpher, M. Barnard, D. Kerr, J. U. Weaver, J. Ahlquist, S. P. Newman. (2009) Randomized controlled trial to assess the impact of continuous glucose monitoring on HbA 1c in insulin-treated diabetes (MITRE Study). Diabetic Medicine 26:5, 540-547
    CrossRef

  145. 145

    Richard Kahn. (2009) Diabetes technology—now and in the future. The Lancet 373:9677, 1741-1743
    CrossRef

  146. 146

    Stephen N. Davis. (2009) Diabetes: Hypoglycemia—a new approach to an old problem. Nature Reviews Endocrinology 5:5, 243-245
    CrossRef

  147. 147

    G Hawthorne, J Waugh. (2009) The role of CGMS in the management of pregnant women with diabetes. Practical Diabetes International 26:3, 88-89
    CrossRef

  148. 148

    Martina Montagnana, Marco Caputo, Davide Giavarina, Giuseppe Lippi. (2009) Overview on self-monitoring of blood glucose. Clinica Chimica Acta 402:1-2, 7-13
    CrossRef

  149. 149

    Joseph I Wolfsdorf. (2009) How effective is continuous glucose monitoring in intensively treated type 1 diabetes mellitus?. Nature Clinical Practice Endocrinology &#38; Metabolism 5:3, 134-135
    CrossRef

  150. 150

    N. S. Oliver, C. Toumazou, A. E. G. Cass, D. G. Johnston. (2009) Glucose sensors: a review of current and emerging technology. Diabetic Medicine 26:3, 197-210
    CrossRef

  151. 151

    D. E. Flanagan. (2009) The future of glucose sensors-using the technology within a clinical service. Diabetic Medicine 26:3, 195-196
    CrossRef

  152. 152

    (2009) Prolonged use of continuous glucose monitoring: kids do not listen - says who?. Pediatric Diabetes 10:2, 89-90
    CrossRef

  153. 153

    Jay S. Skyler. (2009) CGM—A Technology in Evolution. Diabetes Technology & Therapeutics 11:2, 63-64
    CrossRef

  154. 154

    Frank Waldron-Lynch, Kevan C Herold. (2009) Continuous glucose monitoring: long live the revolution!. Nature Clinical Practice Endocrinology &#38; Metabolism 5:2, 82-83
    CrossRef

  155. 155

    Satish K. Garg, James Smith, Christie Beatson, Benita Lopez-Baca, Mary Voelmle, Peter A. Gottlieb. (2009) Comparison of Accuracy and Safety of the SEVEN and the Navigator Continuous Glucose Monitoring Systems. Diabetes Technology & Therapeutics 11:2, 65-72
    CrossRef

  156. 156

    S. L. Bratton. (2009) Early Insulin Therapy in Very-Low-Birthweight Infants. AAP Grand Rounds 21:2, 20-20
    CrossRef

  157. 157

    (2009) Continuous Glucose Monitoring and Type 1 Diabetes. New England Journal of Medicine 360:2, 190-192
    Full Text

  158. 158

    Joerg Martini, Peter Kiesel, Jeffrey Roe, Richard H. Bruce. (2009) Glucose concentration monitoring using a small Fabry-Pérot etalon. Journal of Biomedical Optics 14:3, 034029
    CrossRef

Letters