Prognostic Importance of Defibrillator Shocks in Patients with Heart Failure

Jeanne E. Poole, M.D., George W. Johnson, B.S.E.E., Anne S. Hellkamp, M.S., Jill Anderson, R.N., David J. Callans, M.D., Merritt H. Raitt, M.D., Ramakota K. Reddy, M.D., Francis E. Marchlinski, M.D., Raymond Yee, M.D., Thomas Guarnieri, M.D., Mario Talajic, M.D., David J. Wilber, M.D., Daniel P. Fishbein, M.D., Douglas L. Packer, M.D., Daniel B. Mark, M.D., M.P.H., Kerry L. Lee, Ph.D., and Gust H. Bardy, M.D.

N Engl J Med 2008; 359:1009-1017September 4, 2008

Abstract

Background

Patients with heart failure who receive an implantable cardioverter–defibrillator (ICD) for primary prevention (i.e., prevention of a first life-threatening arrhythmic event) may later receive therapeutic shocks from the ICD. Information about long-term prognosis after ICD therapy in such patients is limited.

Methods

Of 829 patients with heart failure who were randomly assigned to ICD therapy, we implanted the ICD in 811. ICD shocks that followed the onset of ventricular tachycardia or ventricular fibrillation were considered to be appropriate. All other ICD shocks were considered to be inappropriate.

Results

Over a median follow-up period of 45.5 months, 269 patients (33.2%) received at least one ICD shock, with 128 patients receiving only appropriate shocks, 87 receiving only inappropriate shocks, and 54 receiving both types of shock. In a Cox proportional-hazards model adjusted for baseline prognostic factors, an appropriate ICD shock, as compared with no appropriate shock, was associated with a significant increase in the subsequent risk of death from all causes (hazard ratio, 5.68; 95% confidence interval [CI], 3.97 to 8.12; P<0.001). An inappropriate ICD shock, as compared with no inappropriate shock, was also associated with a significant increase in the risk of death (hazard ratio, 1.98; 95% CI, 1.29 to 3.05; P=0.002). For patients who survived longer than 24 hours after an appropriate ICD shock, the risk of death remained elevated (hazard ratio, 2.99; 95% CI, 2.04 to 4.37; P<0.001). The most common cause of death among patients who received any ICD shock was progressive heart failure.

Conclusions

Among patients with heart failure in whom an ICD is implanted for primary prevention, those who receive shocks for any arrhythmia have a substantially higher risk of death than similar patients who do not receive such shocks.

Media in This Article

Figure 1Hazard Ratios for the Association of ICD Shock with the Risk of Death, According to Shock Type.
Figure 2Hazard Ratios for the Risk of Death among Patients Who Survived at Least 24 Hours after a First ICD Shock.
Article

The Sudden Cardiac Death in Heart Failure Trial (SCD-HeFT) (ClinicalTrials.gov number, NCT00000609) and the Multicenter Automatic Defibrillator Implantation Trial II (MADIT II) both showed that therapy with implantable cardioverter–defibrillators (ICDs) improves survival among patients who are at risk for sudden cardiac death but who have not previously had a sustained ventricular arrhythmia.1-3 On the basis of these data, recent clinical guidelines consider the implantation of an ICD for “primary prevention” (i.e., prevention of a first life-threatening arrhythmic event) to be standard high-quality care for patients who meet the entry criteria for these trials.4

One result of this broader use of ICDs is that the natural history of the disease in these patients is modified as a consequence of the delivery of ICD therapies. Patients in whom an ICD is implanted for primary prevention and who subsequently receive an appropriate defibrillator shock may have a different prognosis from similar patients who do not receive a shock. Although data from MADIT II previously showed that patients who received appropriate ICD therapy had a risk of death that was increased by a factor of 3,5 there is little evidence from current practice that the importance of these events is recognized by physicians who care for these patients. We therefore assessed the long-term prognostic significance of both appropriate and inappropriate ICD shocks in SCD-HeFT.

Methods

Study Overview and Population

The study design, methods, and primary results of SCD-HeFT have been reported previously.1,2 SCD-HeFT was a multicenter clinical trial in which 2521 patients with New York Heart Association (NYHA) class II or III heart failure and a left ventricular ejection fraction of 35% or less, but no previous sustained ventricular arrhythmia, were randomly assigned in equal proportions to receive an ICD, amiodarone therapy, or placebo. The cause of the heart failure was ischemic in 52.0% of the patients and nonischemic in 48.0%. The use of evidence-based medical therapies was at high levels both at baseline and at the final follow-up assessment.2 After a median follow-up of 45.5 months, ICD therapy, as compared with medical therapy alone, was associated with a 23.0% decrease in the risk of death.

The protocol was approved by the institutional review board at each participating organization, and each patient provided written informed consent before enrollment. Sponsorship and oversight of the trial were provided by the National Heart, Lung, and Blood Institute. Study drugs and ICDs were provided free of charge by the manufacturers (Wyeth–Ayerst Laboratories [now Wyeth Pharmaceuticals] and Medtronic, respectively).

The authors designed this study, collected and analyzed the data, and made the decision to submit the results for publication. Dr. Poole wrote the manuscript, which was reviewed by all the authors. The corporate sponsors of the trial had no role in the design, analysis, or interpretation of the study. ICD-shock data and clinical data were maintained in confidential, secure databases (the ICD Core Laboratory at the Seattle Institute for Cardiac Research and the Duke Clinical Research Institute, respectively).

ICD Therapy

Each patient in the ICD-therapy group was assigned to receive a single-lead ICD (Medtronic Micro Jewel II model 7223Cx). This device was implanted in 804 of 811 patients (99.1%); 4 patients received other Medtronic single-lead models, and 3 patients received Medtronic dual-lead models, owing to deviations from the protocol. The protocol for the programming of the ICD was deliberately conservative, specifying that the ICD should intervene only for rapid, life-threatening ventricular tachycardia or ventricular fibrillation. A single zone of therapy was used, and an episode of tachycardia was defined as at least 18 of 24 beats at a rate of 188 beats per minute or more (≤320 msec). Up to six shocks could be delivered per episode. Neither antitachycardia pacing nor a second zone of therapy was allowed, since patients with known sustained ventricular tachycardia were excluded from the study. Antibradycardia pacing was programmed at 50 beats per minute but would not be triggered to pace unless the heart rate dropped below 34 beats per minute (a process defined as hysteresis).

Follow-up

Downloading of the information from the ICD was performed at the time of implantation, every 3 months thereafter, and after ICD shocks were delivered. Reports of the downloaded information were electronically sent or saved to a disk and mailed to the ICD Core Laboratory. Permission to obtain data on the ICD after the death of a patient was requested from the patient's next of kin. The cause of death for all patients was determined according to preset criteria by the events committee, whose members were unaware of the therapy group to which the patient had been assigned.

Electrogram Classification

All electrograms showing events that triggered ICD shocks were assigned to two independent members of the Electrogram Review Committee (see the Appendix), who classified the events according to predetermined diagnostic criteria for cardiac rhythms. To discriminate ventricular from supraventricular rhythms, onset characteristics, electrograms recorded before detection of the arrhythmia and after delivery of the shock, and plots of RR intervals were analyzed. Shocks were considered to be appropriate if the triggering rhythm was determined to be ventricular fibrillation or ventricular tachycardia. Inappropriate triggers of ICD shocks included supraventricular tachycardias, oversensing of P or T waves as R waves, double counting of R waves, and an artifact from lead fractures or electromagnetic interference. The delivery of a shock after the spontaneous termination of nonsustained ventricular tachycardia was also considered to be inappropriate. The term “ICD shock” in this analysis refers to ICD therapy that was triggered for a single rhythm event, regardless of the total number of actual shocks that were required to satisfy the criteria for termination of tachycardia by the ICD.

Statistical Analysis

Standard descriptive statistics were used, including percentages for discrete variables and medians and quartiles for continuous variables. The relationship of ICD-shock therapy to death from any cause was examined with the use of Cox proportional-hazards models,6 with adjustment for baseline prognostic factors measured in the trial. These included age, sex, cause of heart failure, NYHA class, time since the diagnosis of heart failure, left ventricular ejection fraction, distance covered on a 6-minute walk, systolic blood pressure, presence or absence of diabetes, use or nonuse of angiotensin-converting–enzyme inhibitors, use or nonuse of digoxin, presence or absence of mitral regurgitation, renal sufficiency or insufficiency, presence or absence of a history of substance abuse, baseline electrocardiographic intervals, and score on the Duke Activity Status Index.7 Risk relationships were characterized as hazard ratios and 95% confidence intervals, generated with the use of the Cox model.

Appropriate and inappropriate shocks were treated as two separate time-dependent covariates, allowing the risk to change after the occurrence of a shock. Thus, a patient's follow-up was credited to the no-shock group until a shock occurred, at which time the patient became part of the risk group for that shock type (i.e., appropriate or inappropriate). “Both shock types” refers to the receipt of both an appropriate and an inappropriate shock during follow-up.

The Cox model was also used to assess whether there was an interaction between appropriate and inappropriate shocks — that is, whether the prognostic effect of an appropriate shock differed if patients also received an inappropriate shock (and vice versa). Interaction terms were also used to determine whether the association of each shock type with the risk of death differed according to the cause of heart failure or NYHA class. Data for patients who had their ICD removed during follow-up were censored at the time the device was removed.

Analyses were also performed to assess the prognostic effect of multiple episodes of the same type of shock (appropriate or inappropriate) in contrast to multiple episodes consisting of both appropriate and inappropriate shocks. For this assessment, additional time-dependent covariates reflecting the occurrence of two (or more) appropriate shocks and two (or more) inappropriate shocks were also considered in the Cox model.

In addition, separate analyses were performed that included only shocks that patients survived for more than 24 hours. In this analysis, shocks that occurred 24 hours or less before death were not considered — that is, at the time of death, a patient was classified as a member of the risk group that he or she had been in before the shock.

Kaplan–Meier survival rates 1 year after a shock were calculated to compare survival after shock therapy according to the type of ICD shock received and in selected clinical subgroups. Reported P values are two-sided and have not been adjusted for multiple testing.

Results

ICD Implantation and Programming

Of the 829 patients who were randomly assigned to ICD therapy, 811 patients had an ICD implanted, 17 declined to undergo implantation, and 1 died before implantation. In 31 of the patients in whom an ICD was implanted (3.8%), the ICD was subsequently removed and not replaced, owing to complications with the device (8 patients), other medical problems (3), or heart transplantation (20). Deviations from the protocol occurred in patients who received a dual-lead ICD at initial implantation (3 patients) or who received a dual-lead ICD (15) or biventricular ICD (7) as a replacement for the original ICD. Deviations from the protocol with respect to ICD programming included a second zone of therapy in 64 patients (as well as antitachycardia pacing in 58 of those patients). The pacing rate was increased in 66 patients. Patients whose only detected events over the course of the study were the result of ICD therapy delivered in deviation from the protocol (arrhythmia events detected at <188 beats per minute and antitachycardia pacing) were not included in this analysis.

ICD-Shock Therapy

During the follow-up period (median, 45.5 months), 269 of the 811 patients with devices (33.2%) received at least one episode of ICD-shock therapy, whereas 542 patients (66.8%) received no known ICD-shock therapy. Either ventricular tachycardia or ventricular fibrillation was the only rhythm trigger for shocks in 128 of the 269 patients who received shocks (47.6%), and inappropriate triggers were the only cause of shocks in 87 (32.3%); 54 (20.1%) received both appropriate and inappropriate shocks. In total, 182 of the 811 patients with ICDs (22.4%) received shocks for ventricular tachycardia or ventricular fibrillation, with ventricular tachycardia or ventricular fibrillation recurring in 97 of those patients (53.3%). A total of 141 patients (17.4%) received inappropriate shocks, with recurrent inappropriate shocks delivered in 61 of those patients (43.3%).

Selected Clinical Characteristics at Study Entry

Baseline characteristics of each shock group are shown in Table 1Table 1Selected Baseline Clinical Characteristics of the Study Groups.. Patients who received appropriate shocks had a lower ejection fraction, were in a higher NYHA class, and were more likely to have atrial fibrillation than those who did not receive appropriate shocks (Table 1). Patients who received inappropriate shocks were more likely to have atrial fibrillation and were less likely to have ischemic heart failure than those who did not receive inappropriate shocks (Table 1).

Deaths

A total of 182 patients died among the 829 patients who were randomly assigned to the ICD group. Of the 18 patients in the ICD group who did not receive an ICD, 9 died, and of the 811 patients in whom an ICD was implanted, 173 died. Ten of these deaths occurred among the 31 patients whose ICDs were removed during the study. Among the 182 patients who received one or more appropriate shocks, there were 67 deaths (36.8%) (54 of the 182 patients also received one or more inappropriate shocks). Ten additional deaths occurred among the 87 patients who received only one or more inappropriate shocks (11.5%). Finally, 86 deaths occurred among the 542 patients who were not recorded as having received an ICD shock (15.9%). Postmortem data on the ICD were available for only 64 of the 173 patients with ICDs who died (37.0%). According to postmortem review, 20 patients died within 24 hours after their first ICD shock for ventricular tachycardia or ventricular fibrillation; 19 of them had ischemic heart failure.

Risk of Death Associated with ICD Shocks

In a multivariable Cox model, both appropriate and inappropriate ICD shocks were significant predictors of death (Figure 1AFigure 1Hazard Ratios for the Association of ICD Shock with the Risk of Death, According to Shock Type.). An appropriate shock, as compared with no appropriate shock, was associated with a risk that was increased by a factor of more than 5 (hazard ratio, 5.68; 95% confidence interval [CI], 3.97 to 8.12; P<0.001), and an inappropriate shock, as compared with no inappropriate shock, was associated with a near doubling of the risk of death (hazard ratio, 1.98; 95% CI, 1.29 to 3.05; P=0.002). The relative risk of death associated with an appropriate shock in patients who had not received a previous inappropriate shock did not differ significantly from that in patients who had received a previous inappropriate shock (P=0.59 for the interaction). We therefore estimated that in patients who received both shock types, the risk of death was increased by a factor of 11, as compared with the risk in patients who received no shocks (hazard ratio, 11.27; 95% CI, 6.70 to 18.94; P<0.001).

Analyses performed to refine these results and account for multiple shocks of the same type (Figure 1B) showed that one appropriate shock was associated with a risk of death that was increased by a factor of approximately 4 (hazard ratio, 3.98; 95% CI, 2.52 to 6.30; P<0.001), whereas a second appropriate shock was associated with a further significant increase by a factor of 2 (P=0.005), resulting in an overall risk of death that was increased by a factor of 8 for patients with two or more appropriate shocks as compared with patients who did not receive any shock (hazard ratio, 8.23; 95% CI, 4.64 to 14.59; P<0.001). After adjustment for the occurrence of multiple appropriate shocks, an inappropriate shock continued to have important prognostic significance and was associated with a risk that was increased by a factor of approximately 2 (P=0.01). The net result is that for patients with two or more appropriate shocks plus an inappropriate shock, the risk of death was increased by a factor of more than 15 (hazard ratio, 15.89; 95% CI, 7.42 to 34.02; P<0.001). Additional inappropriate shocks were not associated with a further increase in risk (P=0.69).

Significant interactions between the receipt of a shock and the cause of heart failure were noted (P<0.001 for the interaction with appropriate shocks, and P=0.05 for the interaction with inappropriate shocks). The hazard ratios for an appropriate shock among patients with ischemic heart failure and among those with nonischemic heart failure were 8.72 (95% CI, 5.68 to 13.39) and 2.61 (95% CI, 1.42 to 4.78), respectively, and the hazard ratios for inappropriate shocks were 2.97 (95% CI, 1.73 to 5.10) and 1.22 (95% CI, 0.59 to 2.51), respectively. No significant interactions between the receipt of a shock and the NYHA class were observed (P=0.62 for the interaction with appropriate shocks, and P=0.46 for the interaction with inappropriate shocks).

ICD shocks remained an important predictor of the outcome among patients who survived for at least 24 hours after a first ICD shock of either type (Figure 2Figure 2Hazard Ratios for the Risk of Death among Patients Who Survived at Least 24 Hours after a First ICD Shock.). An appropriate shock was associated with a risk of death that was increased by a factor of 3 (hazard ratio, 2.99; 95% CI, 2.04 to 4.37; P<0.001). An inappropriate shock was associated with a trend toward an increased risk (hazard ratio, 1.57; 95% CI, 0.99 to 2.50; P=0.06). As was the case when all shocks were considered, there was no significant interaction between appropriate and inappropriate shock types when shocks that occurred 24 hours or less before death were not considered (P=0.60). We therefore estimated that patients who received both types of shocks and survived longer than 24 hours had a risk of death that was increased by a factor of nearly 5, as compared with patients who received no ICD shocks (hazard ratio, 4.70; 95% CI, 2.70 to 8.18; P<0.001). No significant interaction was found between the receipt of a shock and the cause of heart failure (P=0.18 for an appropriate shock, and P=0.24 for an inappropriate shock) or NYHA class (P=0.11 and P=0.12, respectively).

Survival after ICD Shocks

One-year Kaplan–Meier survival rates (±SE) after the first ICD shock were compared among patients with a first shock of any type (82.5±2.4%), all patients with one or more appropriate shocks (76.9±3.2%), and those with only one or more inappropriate shocks (94.9±2.5%) (Table 2Table 2Time from ICD Shock to Death among Patients Who Received at Least One Shock.). Median times from shock to death for patients who received any shock were 204 days (interquartile range, 1 to 630); for patients with one or more appropriate shocks, 168 days (interquartile range, 1 to 797); and for patients with only one or more inappropriate shocks, 294 days (interquartile range, 28 to 509) (Table 2). One-year survival was lower and the time from shock to death shorter for patients with appropriate shocks who were in NYHA class III than for those who were in class II, for those in whom the cause of heart failure was ischemic than for those in whom the cause was nonischemic (Table 2), and for those whose first shock was for ventricular fibrillation than for those whose first shock was for ventricular tachycardia.

Cause of Death in Patients with ICD Shocks

The most common cause of death among patients who had any ICD shock was progressive heart failure (accounting for 33 of 77 deaths [42.9%]) (Table 3Table 3Cause of Death According to Type of Shock.). However, sudden death from arrhythmia did occur among patients who received ICD shocks for ventricular tachycardia or ventricular fibrillation (14 of 67 deaths [20.9%]).

Discussion

This study showed that in patients with heart failure who received an ICD for primary prevention, the occurrence of an appropriate ICD shock was associated with a markedly increased risk of death. This relationship was independent of other covariates that are predictive of the outcome and was seen both in patients with ischemic heart disease and in those with nonischemic heart disease. Although the response by physicians to ICD shocks is commonly a sense of relief that sudden death was averted, our findings highlight the need for a more thoughtful consideration of this patient group, directed in particular at a reassessment of the therapeutic options that might modify the prognosis.

Our study also identified an increased risk for patients who received inappropriately triggered ICD shocks. The cause of this association is uncertain, but several possibilities may be considered. First, patients with heart failure in whom atrial fibrillation develops have been shown to be at an increased risk for death.8-10 Second, in certain vulnerable patients, it may be postulated that the negative inotropic consequences of the shock itself could increase the risk of death, especially when the patient receives multiple shocks owing to oversensing or ongoing supraventricular tachycardia.11-16

The hazard ratios for the risk of death were lower when only patients who survived longer than 24 hours after a shock were considered. In some cases, the firing of the device may occur during the patient's fatal arrhythmia; such ICD shocks are not predictive in the sense of providing information about the likelihood of a subsequent event. The data from analyses that included only patients who survived longer than 24 hours after a shock may therefore represent a more accurate picture of the consequences of surviving an arrhythmia.

The MADIT II investigators were the first to describe an adverse prognosis associated with ICD therapy that is used for primary prevention.5,17,18 In MADIT II, among 719 patients with ischemic heart disease, an ICD shock or antitachycardia pacing was reported to be appropriate in 23.5% and inappropriate in 13.9%. Absolute rates of ICD use in MADIT II were similar to those in our study, despite a shorter follow-up (21 months, vs. 45.5 months in our study), use of dual-chamber ICDs in 43.6% of the patients, ICD programming that included antitachycardia pacing in 59% of the patients, and inclusion of patients with NYHA class I disease.3,5 Further analyses by the MADIT II investigators identified a risk of death that was increased by a factor of more than 3 among patients who received ICD shocks or antitachycardia pacing for ventricular tachycardia or ventricular fibrillation,5 as well as increased risks of heart failure and of death related to heart failure.17 The risk of death with appropriate ICD shocks was higher in our study (increased by a factor of more than 5) than in their study. This difference may be due to the longer follow-up and the exclusion of patients with NYHA class I disease in our study. We also identified a risk of death with inappropriate shocks that was increased by a factor of 2, which was similar to that in MADIT II.18

It is important to emphasize that our results reflect our use of primarily single-lead ICDs, a single zone of therapy, and shock-only programming for high-rate arrhythmias that were most likely to be life-threatening. Although decreasing the rate of painful ICD shocks for ventricular tachycardia is an accepted reason to activate antitachycardia pacing, inappropriate intervention for self-terminating rhythms such as nonsustained ventricular tachycardia may occur when antitachycardia pacing is used empirically.19,20 In shock-only programming, the delay of 5 to 10 seconds during the charging of the capacitors for the delivery of the shock allows some rhythms to normalize spontaneously and minimizes the likelihood of pacing-induced acceleration of rhythms that are best left alone.19,20

There are several limitations of our study. The episodes of ICD shock that were analyzed in this trial may not represent all ICD-detected events. It was not possible to guarantee that all ICD-detected events were transmitted to the ICD Core Laboratory, despite careful review of follow-up data forms to identify any missed events.

A second limitation, which was inherent in the design of the study, was our inability to record sustained rhythms with rates of less than 188 beats per minute, except in the few cases of deviation from the protocol. The goal of this design feature was to minimize false positive interventions in a population that was not known to have sustained ventricular tachycardia at the time of enrollment. In addition, the number of postmortem reports that were collected was limited, since family members often failed to provide notification of a patient's death in time to gather the data. Therefore, some patients may have had shock events associated with death that were not identified. Finally, the limited memory capacity of the device resulted in overwritten data in patients who had repetitive arrhythmic events occurring over a short period of time.

In conclusion, in patients with heart failure who receive an ICD for primary prevention, both appropriate and inappropriate ICD shocks are associated with a marked increase in the subsequent risk of death, particularly death from progressive heart failure. Our results do not establish what further therapies, if any, might be effective in reducing this risk.

Supported by grants from the National Heart, Lung, and Blood Institute (UO1 HL55766, UO1 HL55297, and UO1 HL55496) and by Medtronic and Wyeth Pharmaceuticals.

Dr. Poole reports receiving lecture fees from Medtronic and Boston Scientific/Guidant, grant support from Biotronik, and consulting fees from Philips; Ms. Anderson, receiving lecture fees from Boston Scientific/Guidant; Dr. Callans, receiving lecture fees from Biosense Webster, Siemens Acuson, Boston Scientific, and Medtronic and consulting fees from St. Jude Medical; Dr. Reddy, receiving consulting fees and lecture fees from Boston Scientific/Guidant, Medtronic, and St. Jude and holding stock options in Cambridge Heart; Dr. Marchlinski, receiving consulting fees, lecture fees, and grant support from Biosense Webster, lecture fees and grant support from Boston Scientific and St. Jude Medical, lecture fees from Medtronic, and consulting fees from GE Healthcare; Dr. Yee, receiving consulting fees from Medtronic; Dr. Talajic, receiving consulting fees and lecture fees from Medtronic and lecture fees from St. Jude Medical; Dr. Wilber, receiving consulting fees and lecture fees from Medtronic and lecture fees from Boston Scientific and St. Jude Medical; Dr. Fishbein, receiving lecture fees from Medtronic; Dr. Packer, receiving consulting fees from Medtronic and Boston Scientific/Guidant, royalties from St. Jude Medical, and grant support from Boston Scientific; Dr. Mark, receiving consulting fees, lecture fees, and research grants from Medtronic and fees as editor of the American Heart Journal from Mosby; Dr. Lee, receiving grant support and consulting fees from Medtronic; and Dr. Bardy, receiving grant support from and having intellectual property rights with Medtronic, receiving consulting fees and grant support from Philips, receiving grant support from Laerdal and being a founder of and board member of and having equity and intellectual property rights with Cameron Health. No other potential conflict of interest relevant to this article was reported.

We thank R. Conklin for help in the preparation of the manuscript.

Source Information

From the University of Washington (J.E.P., D.P.F., G.H.B.); and the Seattle Institute for Cardiac Research (G.W.J., J.A., G.H.B.) — both in Seattle; Duke Clinical Research Institute, Durham, NC (A.S.H., D.B.M., K.L.L.); University of Pennsylvania, Philadelphia (D.J.C., F.E.M.); Portland Veterans Affairs Medical Center and Oregon Health Sciences University, Portland (M.H.R.); Oregon Cardiology Associates, Eugene (R.K.R.); University Hospital, London, ON, Canada (R.Y.); Johns Hopkins University, Baltimore (T.G.); Institut de Cardiologie de Montréal, Université de Montréal, Montreal (M.T.); Loyola University Medical Center, Maywood, IL (D.J.W.); and the Mayo Clinic, Rochester, MN (D.L.P.).

Address reprint requests to Dr. Poole at the Division of Cardiology, University of Washington School of Medicine, 1959 NE Pacific St., Box 356422, Seattle, WA 98195-6422, or at .

Appendix

The Sudden Cardiac Death in Heart Failure Trial (SCD-HeFT) ICD Electrogram Review Committee consisted of the following members: J.E. Poole (University of Washington, Seattle), M.H. Raitt (Portland Veterans Affairs Medical Center and Oregon Health Sciences University, Portland), R.K. Reddy (Oregon Cardiology Associates, Eugene), D. Callans and F.E. Marchlinski (University of Pennsylvania, Philadelphia), M. Talajic (Institut de Cardiologie de Montréal, Montreal), D.J. Wilber (Loyola University Medical Center, Maywood, IL), T. Guarnieri (Johns Hopkins University, Baltimore), R. Yee (University Hospital, London, ON, Canada).

References

References

  1. 1

    Bardy GH, Lee KL, Mark DB, Poole JE, Fishbein DP, SCD-HeFT Investigators. Sudden Cardiac Death-Heart Failure Trial (SCD-HeFT). In: Woosley RL, Singh SN, eds. Arrhythmia treatment and therapy. New York: Marcel Dekker, 2000:323-42.

  2. 2

    Bardy GH, Lee KL, Mark DB, et al. Amiodarone or an implantable cardioverter-defibrillator for congestive heart failure. N Engl J Med 2005;352:225-237[Erratum, N Engl J Med 2005;352:2146.]
    Full Text | Web of Science | Medline

  3. 3

    Moss AJ, Zareba W, Hall WJ, et al. Prophylactic implantation of a defibrillator in patients with myocardial infarction and reduced ejection fraction. N Engl J Med 2002;346:877-883
    Full Text | Web of Science | Medline

  4. 4

    ACC/AHA/HRS 2008 guidelines for device-based therapy of cardiac rhythm abnormalities: report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (Writing Committee to Revise the ACC/AHA/NASPE 2002 Guideline Update for Implantation of Cardiac Pacemakers and Antiarrhythmia Devices) developed in collaboration with the American Association for Thoracic Surgery and Society of Thoracic Surgeons. J Am Coll Cardiol 2008;51:2085-2105
    CrossRef | Web of Science

  5. 5

    Moss AJ, Greenberg H, Case RB, et al. Long-term clinical course of patients after termination of ventricular tachyarrhythmia by an implanted defibrillator. Circulation 2004;110:3760-3765
    CrossRef | Web of Science | Medline

  6. 6

    Cox DR. Regression models and life-tables. J R Stat Soc [B] 1972;34:187-220

  7. 7

    Hlatky MA, Boineau RE, Higginbotham MB, et al. A brief self-administered questionnaire to determine functional capacity (the Duke Activity Status Index). Am J Cardiol 1989;64:651-654
    CrossRef | Web of Science | Medline

  8. 8

    Swedberg K, Olsson LG, Charlesworth A, et al. Prognostic relevance of atrial fibrillation in patients with chronic heart failure on long-term treatment with beta-blockers: results from COMET. Eur Heart J 2005;26:1303-1308
    CrossRef | Web of Science | Medline

  9. 9

    Benza RL, Tallaj JA, Felker GM, et al. The impact of arrhythmias in acute heart failure. J Card Fail 2004;10:279-284
    CrossRef | Web of Science | Medline

  10. 10

    Olsson LG, Swedberg K, Ducharme A, et al. Atrial fibrillation and risk of clinical events in chronic heart failure with and without left ventricular systolic dysfunction: results from the Candesartan in Heart failure-Assessment of Reduction in Mortality and morbidity (CHARM) program. J Am Coll Cardiol 2006;47:1997-2004
    CrossRef | Web of Science | Medline

  11. 11

    Tokano T, Bach D, Chang J, et al. The effect of ventricular shock strength on cardiac hemodynamics. J Cardiovasc Electrophysiol 1998;9:791-797
    CrossRef | Web of Science | Medline

  12. 12

    Xie J, Weil MH, Sun S, et al. High-energy defibrillation increases the severity of postresuscitation myocardial dysfunction. Circulation 1997;96:683-688
    Web of Science | Medline

  13. 13

    Tang W, Weil MH, Sun S, et al. The effects of biphasic and conventional monophasic defibrillation on postresuscitation myocardial function. J Am Coll Cardiol 1999;34:815-822
    CrossRef | Web of Science | Medline

  14. 14

    Yamaguchi H, Weil M, Tang W, Kamohara T, Jin X, Bisera J. Myocardial dysfunction after electrical defibrillation. Resuscitation 2002;54:289-296
    CrossRef | Web of Science | Medline

  15. 15

    Weaver WD, Cobb LA, Copass MK, Hallstrom AP. Ventricular defibrillation -- a comparative trial using 175-J and 320-J shocks. N Engl J Med 1982;307:1101-1106
    Full Text | Web of Science | Medline

  16. 16

    Hasdemir C, Shah N, Rao AP, et al. Analysis of troponin I levels after spontaneous implantable cardioverter defibrillator shocks. J Cardiovasc Electrophysiol 2002;13:144-150
    CrossRef | Web of Science | Medline

  17. 17

    Goldenberg I, Moss AJ, Hall WJ, et al. Causes and consequences of heart failure after prophylactic implantation of a defibrillator in the Multicenter Automatic Defibrillator Implantation Trial II. Circulation 2006;113:2810-2817
    CrossRef | Web of Science | Medline

  18. 18

    Daubert JP, Zareba W, Cannom DS, et al. Inappropriate implantable cardioverter-defibrillator shocks in MADIT II: frequency, mechanisms, predictors, and survival impact. J Am Coll Cardiol 2008;51:1357-1365
    CrossRef | Web of Science | Medline

  19. 19

    Wathen MS, DeGroot PJ, Sweeney MO, et al. Prospective randomized multicenter trial of empirical antitachycardia pacing versus shocks for spontaneous rapid ventricular tachycardia in patients with implantable cardioverter-defibrillators: Pacing Fast Ventricular Tachycardia Reduces Shock Therapies (PainFREE Rx II) trial results. Circulation 2004;110:2591-2596
    CrossRef | Web of Science | Medline

  20. 20

    Wilkoff BL, Hess M, Young J, Abraham WT. Differences in tachyarrhythmia detection and implantable cardioverter defibrillator therapy by primary or secondary prevention indication in cardiac resynchronization therapy patients. J Cardiovasc Electrophysiol 2004;15:1002-1009
    CrossRef | Web of Science | Medline

Citing Articles (174)

Citing Articles

  1. 1

    Paban Saha, Jeffrey J. Goldberger. (2012) Risk Stratification for Prevention of Sudden Cardiac Death. Current Treatment Options in Cardiovascular Medicine 14:1, 81-90
    CrossRef

  2. 2

    THOMAS KLEEMANN, MATTHIAS HOCHADEL, MARGIT STRAUSS, ALEXANDROS SKARLOS, KARLHEINZ SEIDL, RALF ZAHN. (2012) Comparison Between Atrial Fibrillation-Triggered Implantable Cardioverter-Defibrillator (ICD) Shocks and Inappropriate Shocks Caused by Lead Failure:. Journal of Cardiovascular Electrophysiologyno-no
    CrossRef

  3. 3

    Anh L. Bui, Gregg C. Fonarow. (2012) Home Monitoring for Heart Failure Management. Journal of the American College of Cardiology 59:2, 97-104
    CrossRef

  4. 4

    Avi Fischer, Kevin T. Ousdigian, James W. Johnson, Jeffrey M. Gillberg, Bruce L. Wilkoff. (2012) The impact of atrial fibrillation with rapid ventricular rates and device programming on shocks in 106,513 ICD and CRT-D patients. Heart Rhythm 9:1, 24-31
    CrossRef

  5. 5

    N. Dagres, F. Cantu, P. Geelen, T. Lewalter, A. Proclemer, C. Blomstrom-Lundqvist. (2012) Current practice of ventricular tachycardia ablation in patients with implantable cardioverter-defibrillators. Europace 14:1, 135-137
    CrossRef

  6. 6

    J. B. van Rees, C. J. Willem Borleffs, J. Thijssen, M. K. de Bie, L. van Erven, S. C. Cannegieter, J. J. Bax, M. J. Schalij. (2012) Prophylactic implantable cardioverter-defibrillator treatment in the elderly: therapy, adverse events, and survival gain. Europace 14:1, 66-73
    CrossRef

  7. 7

    Sumeet K. Mainigi, Khalid Almuti, Vincent M. Figueredo, Nils A. Guttenplan, Asma Aouthmany, Jessica Smukler, Bernadette Sheeron, Bryan Meldrum, Agustina D. Saenz, Gia Tran, Allan M. Greenspan. (2012) Usefulness of Radiofrequency Ablation of Supraventricular Tachycardia to Decrease Inappropriate Shocks from Implantable Cardioverter–Defibrillators. The American Journal of Cardiology 109:2, 231-237
    CrossRef

  8. 8

    Larry R. Jackson, James P. Daubert, Kevin L. Thomas. (2012) Expanding the Benefits of Implantable Cardioverter-Defibrillator Therapy: “Is Less More”?. Progress in Cardiovascular Diseases 54:4, 372-378
    CrossRef

  9. 9

    Jason Bradfield, Noel G. Boyle, Ravi Mandapati, Kalyanam Shivkumar. (2011) The Emerging Role of Cardiac Resynchronization Therapy in Milder Heart Failure: Are We Implanting Too Late for Response?. Current Heart Failure Reports
    CrossRef

  10. 10

    Pasquale Santangeli, Luigi Di Biase, Amin Al-Ahmad, Henry Hsia, J. David Burkhardt, Javier Sanchez, Rong Bai, Michela Casella, Antonio Dello Russo, Claudio Tondo, Andrea Natale. (2011) Primary Ablation for Ventricular Tachycardia: When and How?. Cardiac Electrophysiology Clinics 3:4, 675-688
    CrossRef

  11. 11

    Larisa G. Tereshchenko, Raja N. Ghanem, Athula Abeyratne, Charles D. Swerdlow. (2011) Intracardiac QT integral on far-field ICD electrogram predicts sustained ventricular tachyarrhythmias in ICD patients. Heart Rhythm 8:12, 1889-1894
    CrossRef

  12. 12

    Shu-Fang Hu, Cheng-Yi Wang, Chuan-Chieh Liu, Wei-Lian Phan, Jen-Yu Wang. (2011) Implantable cardioverter-defibrillator storm: An unusual adverse effect of β2-agonists. International Journal of Cardiology 153:3, e65-e67
    CrossRef

  13. 13

    Moustafa Banna, Julia H. Indik. (2011) Risk Stratification and Prevention of Sudden Death in Patients with Heart Failure. Current Treatment Options in Cardiovascular Medicine 13:6, 517-527
    CrossRef

  14. 14

    Renee M. Sullivan, Brian Olshansky. (2011) Is a Dual-Chamber ICD Always Better in Patients Requiring an ICD?. Cardiac Electrophysiology Clinics 3:4, 577-592
    CrossRef

  15. 15

    Deepak Bhakta, Lynne D. Foreman. (2011) Implantable cardioverter-defibrillator shock reduction: The new paradigm. Heart Rhythm 8:12, 1887-1888
    CrossRef

  16. 16

    John Groarke, Dermot Phelan, David Burke, Peter Crean, Brendan Foley. (2011) Very late stent thrombosis immediately after recurrent inappropriate shock delivery by an implantable cardioverter defibrillator. Journal of Electrocardiology
    CrossRef

  17. 17

    Gail K. Larsen, John Evans, William E. Lambert, Yiyi Chen, Merritt H. Raitt. (2011) Shocks burden and increased mortality in implantable cardioverter-defibrillator patients. Heart Rhythm 8:12, 1881-1886
    CrossRef

  18. 18

    Tim Smith, Dominic A.M.J. Theuns, Kadir Caliskan, Luc Jordaens. (2011) Long-Term Follow-Up of Prophylactic Implantable Cardioverter-Defibrillator-Only Therapy: Comparison of Ischemic and Nonischemic Heart Disease. Clinical Cardiology 34:12, 761-767
    CrossRef

  19. 19

    Girish M. Nair, Jeffrey S. Healey, Syamkumar M. Divakaramenon, Stuart J. Connolly, Carlos A. Morillo. (2011) Antiarrhythmic Drug Therapy for New-Onset Ventricular Arrhythmia (VT/VF) in ICD Patients. Cardiac Electrophysiology Clinics 3:4, 651-661
    CrossRef

  20. 20

    J. Seegers, M. A. Vos, P. Flevari, R. Willems, C. Sohns, D. Vollmann, L. Luthje, D. T. Kremastinos, V. Flore, M. Meine, A. Tuinenburg, R. C. Myles, D. Simon, J. Brockmoller, T. Friede, G. Hasenfuss, S. E. Lehnart, M. Zabel, . (2011) Rationale, objectives, and design of the EUTrigTreat clinical study: a prospective observational study for arrhythmia risk stratification and assessment of interrelationships among repolarization markers and genotype. Europace
    CrossRef

  21. 21

    T. J. Bunch, A. Darby, H. T. May, M. Ragosta, D. S. Lim, A. M. Taylor, J. P. DiMarco, G. Ailawadi, J. R. Revenaugh, J. P. Weiss, S. Mahapatra. (2011) Efficacy and safety of ventricular tachycardia ablation with mechanical circulatory support compared with substrate-based ablation techniques. Europace
    CrossRef

  22. 22

    J. L. Sturdivant, M. R. Gold. (2011) Pre-discharge defibrillation testing: clinically important or obsolete?. Europace
    CrossRef

  23. 23

    Paul Dorian. (2011) Counterpoint: Controversy in electrophysiology—Implantable cardioverter-defibrillator shocks for ventricular tachyarrhythmias increase mortality. Heart Rhythm
    CrossRef

  24. 24

    Yang Pang, Qi Jin, Ning Zhang, Shujing Ren, Tianyou Ling, Ying Chen, Gang Gu, Yongchu Shen, Liqun Wu. (2011) High-energy defibrillation increases the dispersion of regional ventricular repolarization. Journal of Interventional Cardiac Electrophysiology 32:2, 81-86
    CrossRef

  25. 25

    BRUCE S. STAMBLER. (2011) ICD Arrhythmia Detection and Discrimination Algorithms: Whose is Best?. Journal of Cardiovascular Electrophysiologyno-no
    CrossRef

  26. 26

    ANDREW BRENYO, MOHAN RAO, SUSHMA KONERU, WILLIAM HALLINAN, SAMIT SHAH, H. T. MASSEY, LEWAY CHEN, BRONISLAVA POLONSKY, SCOTT MCNITT, DAVID T. HUANG, ILAN GOLDENBERG, MEHMET AKTAS. (2011) Risk of Mortality for Ventricular Arrhythmia in Ambulatory LVAD Patients. Journal of Cardiovascular Electrophysiologyno-no
    CrossRef

  27. 27

    ALON BARSHESHET, ILAN GOLDENBERG, ARTHUR J. MOSS, DAVID T. HUANG, WOJCIECH ZAREBA, SCOTT McNITT, HELMUT U. KLEIN, VICTOR GUETTA. (2011) Effect of Elapsed Time From Coronary Revascularization to Implantation of a Cardioverter Defibrillator on Long-Term Survival in the MADIT-II Trial. Journal of Cardiovascular Electrophysiology 22:11, 1237-1242
    CrossRef

  28. 28

    Michael O. Sweeney. (2011) Counterpoint: Implantable cardioverter-defibrillator shocks for ventricular tachyarrhythmias increase mortality. Heart Rhythm
    CrossRef

  29. 29

    Ian Crozier. (2011) The subcutaneous defibrillator will replace the transvenous defibrillator. Journal of Interventional Cardiac Electrophysiology 32:1, 73-77
    CrossRef

  30. 30

    Renee M. Sullivan, Andrea M. Russo, Kellie Chase Berg, Kira Q. Stolen, Milan Seth, David Perschbacher, John D. Day, Brian Olshansky. (2011) Arrhythmia rate distribution and tachyarrhythmia therapy in an ICD population: Results from the INTRINSIC RV trial. Heart Rhythm
    CrossRef

  31. 31

    DAVID S. FRANKEL, STAVROS E. MOUNTANTONAKIS, MELISSA R. ROBINSON, ERICA S. ZADO, DAVID J. CALLANS, FRANCIS E. MARCHLINSKI. (2011) Ventricular Tachycardia Ablation Remains Treatment of Last Resort in Structural Heart Disease: Argument for Earlier Intervention. Journal of Cardiovascular Electrophysiology 22:10, 1123-1128
    CrossRef

  32. 32

    MICHAEL R. GOLD, DOMINIC A. THEUNS, BRADLEY P. KNIGHT, J. LACY STURDIVANT, RICK SANGHERA, KENNETH A. ELLENBOGEN, MARK A. WOOD, MARTIN C. BURKE. (2011) Head-To-Head Comparison of Arrhythmia Discrimination Performance of Subcutaneous and Transvenous ICD Arrhythmia Detection Algorithms: The START Study. Journal of Cardiovascular Electrophysiologyno-no
    CrossRef

  33. 33

    Roy M. John, William G. Stevenson. (2011) Catheter-based Ablation for Ventricular Arrhythmias. Current Cardiology Reports 13:5, 399-406
    CrossRef

  34. 34

    Ioanna Kosmidou, Keiichi Inada, Jens Seiler, Bruce Koplan, William G. Stevenson, Usha B. Tedrow. (2011) Role of repeat procedures for catheter ablation of postinfarction ventricular tachycardia. Heart Rhythm 8:10, 1516-1522
    CrossRef

  35. 35

    Olivera Gojkovic, Etienne M. Aliot, Alessandro Capucci, Stuart J. Connolly, Harry Crijns, Stefan H. Hohnloser, Piotr Kulakowski, Denis Roy, David Radzik, Bramah N. Singh, Peter R. Kowey. (2011) Celivarone in patients with an implantable cardioverter-defibrillator: Adjunctive therapy for the reduction of ventricular arrhythmia-triggered implantable cardioverter-defibrillator interventions. Heart Rhythm
    CrossRef

  36. 36

    Michael R. Gold, Saleem Ahmad, Kevin Browne, Kellie Chase Berg, Lisa Thackeray, Ronald D. Berger. (2011) Prospective comparison of discrimination algorithms to prevent inappropriate ICD therapy: Primary results of the Rhythm ID Going Head to Head Trial. Heart Rhythm
    CrossRef

  37. 37

    ARTHUR SHIYOVICH, AMOS KATZ. (2011) The “Shock Factor”: ICD Configuration and Programming to Optimize Shock Treatment. Journal of Cardiovascular Electrophysiology 22:9, 1030-1033
    CrossRef

  38. 38

    F. ROOSEVELT GILLIAM, DAVID L. HAYES, JOHN P. BOEHMER, JOHN DAY, PAUL A. HEIDENREICH, MILAN SETH, PAUL W. JONES, KENNETH M. STEIN, LESLIE A. SAXON. (2011) Real World Evaluation of Dual-Zone ICD and CRT-D Programming Compared to Single-Zone Programming: The ALTITUDE REDUCES Study. Journal of Cardiovascular Electrophysiology 22:9, 1023-1029
    CrossRef

  39. 39

    John A. Schoenhard, Paul C. Zei. (2011) Reducing ICD Shocks for Ventricular Arrhythmias. Cardiac Electrophysiology Clinics 3:3, 493-502
    CrossRef

  40. 40

    Norman C. Wang, Sanjoy Bhattacharya, Mihai Gheorghiade. (2011) The potential role of cardiac resynchronization therapy in acute heart failure syndromes. Heart Failure Reviews 16:5, 481-490
    CrossRef

  41. 41

    Michael A. Samara, W. H. Wilson Tang. (2011) Device monitoring strategies in acute heart failure syndromes. Heart Failure Reviews 16:5, 491-502
    CrossRef

  42. 42

    Henry H. Hsia, Karin K.M. Chia, John C. Evans. (2011) Defibrillation Threshold: Testing, Upper Limit of Vulnerability, and Methods for Reduction. Cardiac Electrophysiology Clinics 3:3, 473-492
    CrossRef

  43. 43

    GISELLE A. BAQUERO, JAVIER E. BANCHS, SONDRA DEPALMA, SALLIE K. YOUNG, ERICA D. PENNY-PETERSON, SORAYA M. SAMII, DEBORAH L. WOLBRETTE, GERALD V. NACCARELLI, MARIO D. GONZALEZ. (2011) Dofetilide Reduces the Frequency of Ventricular Arrhythmias and Implantable Cardioverter Defibrillator Therapies. Journal of Cardiovascular Electrophysiologyno-no
    CrossRef

  44. 44

    Joseph J. Gard, Paul A. Friedman. (2011) Strategies to Reduce ICD Shocks: The Role of Supraventricular Tachycardia–Ventricular Tachycardia Discriminators. Cardiac Electrophysiology Clinics 3:3, 373-387
    CrossRef

  45. 45

    Jason Bradfield, Roderick Tung, Noel G. Boyle, Kalyanam Shivkumar. (2011) Managing patients with ICD shocks and programming tachycardia therapies during acute heart failure syndromes. Heart Failure Reviews 16:5, 449-456
    CrossRef

  46. 46

    Timir S. Baman, Thomas Crawford, Patricia Sovitch, Pascal Meier, Nathan Sovitch, Lindsey Gakenheimer, James Kirkpatrick, Brad Wasserman, George Samson, Hakan Oral, Kim A. Eagle. (2011) Feasibility of postmortem device acquisition for potential reuse in underserved nations. Heart Rhythm
    CrossRef

  47. 47

    Thomas Deneke, Bernd Lemke, Andreas Mügge, Dong-In Shin, Peter H Grewe, Marc Horlitz, Osman Balta, Leif Bösche, Thomas Lawo. (2011) Catheter ablation of electrical storm. Expert Review of Cardiovascular Therapy 9:8, 1051-1058
    CrossRef

  48. 48

    KADIR CALISKAN, TAMAS SZILI-TOROK, DOMINIC A.M.J. THEUNS, ATTILA KARDOS, MARCEL L. GELEIJNSE, AGGIE H.M.M. BALK, RON T. VAN DOMBURG, LUC JORDAENS, MAARTEN L. SIMOONS. (2011) Indications and Outcome of Implantable Cardioverter-Defibrillators for Primary and Secondary Prophylaxis in Patients with Noncompaction Cardiomyopathy. Journal of Cardiovascular Electrophysiology 22:8, 898-904
    CrossRef

  49. 49

    Alexander M. Clark, Tiny Jaarsma, Patricia Strachan, Patricia M. Davidson, Megan Jerke, James M. Beattie, Amanda S. Duncan, Chantal F. Ski, David R. Thompson. (2011) Effective communication and ethical consent in decisions related to ICDs. Nature Reviews Cardiology 8:12, 694-705
    CrossRef

  50. 50

    Jean-Claude Daubert, Erwan Donal, Cecilia Linde. (2011) A plea for the wider use of CRT-P in candidates for cardiac resynchronisation therapy. Heart Failure Reviews
    CrossRef

  51. 51

    Wolfgang Dichtl, Thomas Wolber, Ursula Paoli, Simon Brüllmann, Markus Stühlinger, Thomas Berger, Karin Spuller, Alexander Strasak, Otmar Pachinger, Laurent M. Haegeli, Firat Duru, Florian Hintringer. (2011) Appropriate Therapy But Not Inappropriate Shocks Predict Survival in Implantable Cardioverter Defibrillator Patients. Clinical Cardiology 34:7, 433-436
    CrossRef

  52. 52

    Sandhya Murthy, Hannah I. Lipman. (2011) Management of End-Stage Heart Failure. Primary Care: Clinics in Office Practice 38:2, 265-276
    CrossRef

  53. 53

    Nikolaos Dagres, Philipp Sommer, Maria Anastasiou-Nana, Gerhard Hindricks. (2011) Treating arrhythmias: an expert opinion. Expert Opinion on Pharmacotherapy 12:9, 1359-1367
    CrossRef

  54. 54

    Kurt C. Roberts-Thomson, Dennis H. Lau, Prashanthan Sanders. (2011) The diagnosis and management of ventricular arrhythmias. Nature Reviews Cardiology 8:6, 311-321
    CrossRef

  55. 55

    Michael W. Fong, Luanda Grazette, David Cesario, Michael Cao, Leslie Saxon. (2011) Treatment of Ventricular Tachycardia in Patients with Heart Failure. Current Cardiology Reports 13:3, 203-209
    CrossRef

  56. 56

    PETER STEMPNIEWICZ, ALAN CHENG, ALLISON CONNOLLY, XIN YUAN WANG, HUGH CALKINS, GORDON F. TOMASELLI, RONALD D. BERGER, LARISA G. TERESHCHENKO. (2011) Appropriate and Inappropriate Electrical Therapies Delivered by an Implantable Cardioverter-Defibrillator: Effect on Intracardiac Electrogram. Journal of Cardiovascular Electrophysiology 22:5, 554-560
    CrossRef

  57. 57

    Tawfiq Zeidan-Shwiri, Doron Aronson, Khalid Atalla, Miry Blich, Mahmoud Suleiman, Ibrahim Marai, Lior Gepstein, Lena Lavie, Peretz Lavie, Monther Boulos. (2011) Circadian pattern of life-threatening ventricular arrhythmia in patients with sleep-disordered breathing and implantable cardioverter-defibrillators. Heart Rhythm 8:5, 657-662
    CrossRef

  58. 58

    S. Mahapatra, D. J. LaPar, C. M. Bhamidipati, G. McDaniel, S. Kamath, T. J. Bunch, G. Ailawadi. (2011) Incidence, risk factors, and consequences of new-onset atrial fibrillation following epicardial ablation for ventricular tachycardia. Europace 13:4, 548-554
    CrossRef

  59. 59

    Jaya Mallidi, Girish N. Nadkarni, Ronald D. Berger, Hugh Calkins, Saman Nazarian. (2011) Meta-analysis of catheter ablation as an adjunct to medical therapy for treatment of ventricular tachycardia in patients with structural heart disease. Heart Rhythm 8:4, 503-510
    CrossRef

  60. 60

    Niraj Varma, Bruce Wilkoff. (2011) Device Features for Managing Patients with Heart Failure. Heart Failure Clinics 7:2, 215-225
    CrossRef

  61. 61

    Ganesh Venkataraman, S. Adam Strickberger. (2011) The Role of Ventricular Tachycardia Ablation in the Reduction of Implantable Defibrillator Shocks. Heart Failure Clinics 7:2, 207-213
    CrossRef

  62. 62

    Melissa R. Robinson, Andrew E. Epstein, David J. Callans. (2011) Secondary Prevention in Heart Failure. Heart Failure Clinics 7:2, 185-194
    CrossRef

  63. 63

    Ayesha Hasan, Benjamin Sun. (2011) Defibrillators and Cardiac Resynchronization Therapy as a Bridge to Cardiac Transplantation. Heart Failure Clinics 7:2, 227-239
    CrossRef

  64. 64

    Jonathan P. Piccini, Sana M. Al-Khatib, Anne S. Hellkamp, Kevin J. Anstrom, Jeanne E. Poole, Daniel B. Mark, Kerry L. Lee, Gust H. Bardy. (2011) Mortality benefits from implantable cardioverter-defibrillator therapy are not restricted to patients with remote myocardial infarction: an analysis from the Sudden Cardiac Death in Heart Failure Trial (SCD-HeFT). Heart Rhythm 8:3, 393-400
    CrossRef

  65. 65

    KENT J. VOLOSIN, DEREK V. EXNER, MARK S. WATHEN, LOU SHERFESEE, ANTHONY P. SCINICARIELLO, JEFFREY M. GILLBERG. (2011) Combining Shock Reduction Strategies to Enhance ICD Therapy: A Role for Computer Modeling. Journal of Cardiovascular Electrophysiology 22:3, 280-289
    CrossRef

  66. 66

    JENNIFER D. COHEN, J. PAUL MOUNSEY. (2011) “Virtual Research” May Point the Way Towards Future Device Programming Trials. Journal of Cardiovascular Electrophysiology 22:3, 290-292
    CrossRef

  67. 67

    G. H. van Welsenes, J. B. van Rees, C. J. W. Borleffs, S. C. Cannegieter, J. J. Bax, L. van Erven, M. J. Schalij. (2011) Long-term follow-up of primary and secondary prevention implantable cardioverter defibrillator patients. Europace 13:3, 389-394
    CrossRef

  68. 68

    Brian Clarke, Jonathan Howlett, John Sapp, Pantelis Andreou, Ratika Parkash. (2011) The Effect of Comorbidity on the Competing Risk of Sudden and Nonsudden Death in an Ambulatory Heart Failure Population. Canadian Journal of Cardiology 27:2, 254-261
    CrossRef

  69. 69

    Raymond Cutro, Michael W. Rich, Paul J. Hauptman. (2011) Device therapy in patients with heart failure and advanced age: Too much too late?. International Journal of Cardiology
    CrossRef

  70. 70

    D. M. Leistner, J. Schmitt, S. Palm, J. Klotsche, S. Estel, A. Fink, C. W. Israel, B. Assmus, G. Z. Duray, S. Dimmeler, S. H. Hohnloser, A. M. Zeiher. (2011) Intracoronary administration of bone marrow-derived mononuclear cells and arrhythmic events in patients with chronic heart failure. European Heart Journal 32:4, 485-491
    CrossRef

  71. 71

    WILLIAM H. SAUER, CHRISTOPHER M. LOWERY, ROYCE L. BARGAS, JOSEPH L. SCHULLER, CHRISTOPER S. STEES, RYAN G. ALEONG, PAUL D. VAROSY. (2011) Utility of Postoperative Testing of Implantable Cardioverter-Defibrillators. Pacing and Clinical Electrophysiology 34:2, 186-192
    CrossRef

  72. 72

    C. Perings, W. Rudolf Bauer, H.-J. Bondke, C. Mewis, M. James, D. Bocker, P. Broadhurst, T. Korte, E. Toft, F. Hintringer, J. Clementy, J. O. Schwab. (2011) Remote monitoring of implantable-cardioverter defibrillators: results from the Reliability of IEGM Online Interpretation (RIONI) study. Europace 13:2, 221-229
    CrossRef

  73. 73

    Chirag M. Sandesara, Renee M. Sullivan, Andrea M. Russo, WeiWei Li, Arthur Kendig, John D. Day, Christopher Mullin, Kira Stolen, Brian Olshansky. (2011) Older Persons with Diabetes Receive Fewer Inappropriate ICD Shocks: Results from the INTRINSIC RV Trial. Journal of Cardiovascular Translational Research 4:1, 27-34
    CrossRef

  74. 74

    Patrick J. Tchou, Mark J. Niebauer. 2011. Defibrillation Testing, Implant Testing, and Relation to Empiric ICD Programming. , 374-382.
    CrossRef

  75. 75

    D. A. M. J. Theuns, B. A. Schaer, O. I. I. Soliman, D. Altmann, C. Sticherling, M. L. Geleijnse, S. Osswald, L. Jordaens. (2011) The prognosis of implantable defibrillator patients treated with cardiac resynchronization therapy: comorbidity burden as predictor of mortality. Europace 13:1, 62-69
    CrossRef

  76. 76

    Akihiko Shimizu. (2011) Indication of Implantable Devices for the Prevention of Sudden Cardiac Death. Journal of Arrhythmia 27:1, 87-90
    CrossRef

  77. 77

    Akira Maesato, Satoshi Higa, Yenn-Jiang Lin, Ichiro Chinen, Sugako Ishigaki, Machiko Yajima, Hiroaki Masuzaki, Shih-Ann Chen. (2011) Impact of Pacing and High-Pass Filter Settings on Ventricular Bipolar Electrograms in Implantable Cardioverter Defibrillator Systems. Circulation Journal 75:9, 2095-2104
    CrossRef

  78. 78

    Ahmed El-Damaty, John L Sapp. (2011) The role of catheter ablation for ventricular tachycardia in patients with ischemic heart disease. Current Opinion in Cardiology 26:1, 30-39
    CrossRef

  79. 79

    Yukio Sekiguchi, Hiroshi Tada, Kentaro Yoshida, Yoshihiro Seo, Shelby Li, Tomoyuki Tejima, Morio Shoda, Shiro Kamakura, Kazutaka Aonuma. (2011) Significant Increase in the Incidence of Ventricular Arrhythmic Events After an Intrathoracic Impedance Change Measured With a Cardiac Resynchronization Therapy Defibrillator. Circulation Journal 75:11, 2614-2620
    CrossRef

  80. 80

    Niraj Varma. 2011. Follow-up Monitoring of Cardiac Implantable Electronic Devices. , 987-1003.
    CrossRef

  81. 81

    David G. Strauss, Jeanne E. Poole, Galen S. Wagner, Ronald H. Selvester, Julie M. Miller, Jill Anderson, George Johnson, Steven E. McNulty, Daniel B. Mark, Kerry L. Lee, Gust H. Bardy, Katherine C. Wu. (2011) An ECG index of myocardial scar enhances prediction of defibrillator shocks: An analysis of the Sudden Cardiac Death in Heart Failure Trial. Heart Rhythm 8:1, 38-45
    CrossRef

  82. 82

    Sharon R. Kaufman, Paul S. Mueller, Abigale L. Ottenberg, Barbara A. Koenig. (2011) Ironic technology: Old age and the implantable cardioverter defibrillator in US health care. Social Science & Medicine 72:1, 6-14
    CrossRef

  83. 83

    Andrew E. Epstein, Michael P. Riley. 2011. Troubleshooting of Implantable Cardioverter-Defibrillators. , 889-910.
    CrossRef

  84. 84

    Takashi Kurita. (2011) Japanese Journal of Electrocardiology 31:4, 376-386
    CrossRef

  85. 85

    SUSANNE S. PEDERSEN, KRISTA C. VAN DEN BROEK, MARTHA VAN DEN BERG, DOMINIC A. M. J. THEUNS. (2010) Shock as a Determinant of Poor Patient-Centered Outcomes in Implantable Cardioverter Defibrillator Patients: Is There More to It Than Meets the Eye?. Pacing and Clinical Electrophysiology 33:12, 1430-1436
    CrossRef

  86. 86

    F. Braunschweig, G. Boriani, A. Bauer, R. Hatala, C. Herrmann-Lingen, J. Kautzner, S. S. Pedersen, S. Pehrson, R. Ricci, M. J. Schalij. (2010) Management of patients receiving implantable cardiac defibrillator shocks: Recommendations for acute and long-term patient management. Europace 12:12, 1673-1690
    CrossRef

  87. 87

    Wilbert S Aronow. (2010) Implantable Cardioverter-Defibrillators. American Journal of Therapeutics 17:6, e208-e220
    CrossRef

  88. 88

    Wilbert S Aronow. (2010) Use of implantable cardioverter-defibrillators. Future Cardiology 6:6, 859-870
    CrossRef

  89. 89

    EDWARD SZE, ARTHUR J. MOSS, SCOTT MCNITT, ALON BARSHESHET, MARK L. ANDREWS, WOJCIECH ZAREBA, ILAN GOLDENBERG, . (2010) Risk Factors for Recurrent Heart Failure Events in the Multicenter Automatic Defibrillator Implantation Trial II (MADIT-II). Journal of Cardiovascular Electrophysiology 21:11, 1217-1223
    CrossRef

  90. 90

    Larisa G. Tereshchenko, Lichy Han, Alan Cheng, Joseph E. Marine, David D. Spragg, Sunil Sinha, Darshan Dalal, Hugh Calkins, Gordon F. Tomaselli, Ronald D. Berger. (2010) Beat-to-beat three-dimensional ECG variability predicts ventricular arrhythmia in ICD recipients. Heart Rhythm 7:11, 1606-1613
    CrossRef

  91. 91

    (2010) An Entirely Subcutaneous Implantable Cardioverter–Defibrillator. New England Journal of Medicine 363:16, 1577-1578
    Full Text

  92. 92

    Philip B. Adamson. 2010. Devices for Following Heart Failure Patients. , 235-254.
    CrossRef

  93. 93

    J. Seegers, M. Zabel, L. Luthje, D. Vollmann. (2010) Ventricular oversensing due to manufacturer-related differences in implantable cardioverter-defibrillator signal processing and sensing lead properties. Europace 12:10, 1460-1466
    CrossRef

  94. 94

    Jason T. Jacobson, Justin B. Weiner. (2010) REVIEW: Management of Ventricular Tachycardia in Patients with Structural Heart Disease. Cardiovascular Therapeutics 28:5, 255-263
    CrossRef

  95. 95

    S. S. Pedersen, K. C. van den Broek, R. A. M. Erdman, L. Jordaens, D. A. M. J. Theuns. (2010) Pre-implantation implantable cardioverter defibrillator concerns and Type D personality increase the risk of mortality in patients with an implantable cardioverter defibrillator. Europace 12:10, 1446-1452
    CrossRef

  96. 96

    Mark W. Kroll, Joerg O. Schwab. (2010) Achieving low defibrillation thresholds at implant: pharmacological influences, RV coil polarity and position, SVC coil usage and positioning, pulse width settings, and the azygous vein. Fundamental & Clinical Pharmacology 24:5, 561-573
    CrossRef

  97. 97

    J. Kreuz, O. Balta, M. Linhart, R. Fimmers, L. Lickfett, F. Mellert, G. Nickenig, J. O. Schwab. (2010) An impaired renal function and advanced heart failure represent independent predictors of the incidence of malignant ventricular arrhythmias in patients with an implantable cardioverter/defibrillator for primary prevention. Europace 12:10, 1439-1445
    CrossRef

  98. 98

    Rajdeep S. Gaitonde, Roopa Subbarao, Mark A. Michael, Gopi Dandamudi, Deepak Bhakta, Jo Mahenthiran, Mithilesh K. Das. (2010) Segmental wall-motion abnormalities of the left ventricle predict arrhythmic events in patients with nonischemic cardiomyopathy. Heart Rhythm 7:10, 1390-1395
    CrossRef

  99. 99

    Judith Z. Goldfinger, Eric D. Adler. (2010) End-of-Life Options for Patients with Advanced Heart Failure. Current Heart Failure Reports 7:3, 140-147
    CrossRef

  100. 100

    Justin A. Ezekowitz, Padma Kaul. (2010) The epidemiology and management of elderly patients with myocardial infarction or heart failure. Heart Failure Reviews 15:5, 407-413
    CrossRef

  101. 101

    J. Jimenez-Candil, J. Hernandez, A. Martin, M. Ruiz-Olgado, J. Herrero, C. Ledesma, J. Morinigo, C. Martin-Luengo. (2010) Influence of beta-blocker therapy on antitachycardia pacing effectiveness for monomorphic ventricular tachycardias occurring in implantable cardioverter-defibrillator patients: a dose-dependent effect. Europace 12:9, 1231-1238
    CrossRef

  102. 102

    Cara N. Pellegrini, Melvin M. Scheinman. (2010) Clinical Management of Ventricular Tachycardia. Current Problems in Cardiology 35:9, 453-504
    CrossRef

  103. 103

    CHRISTOF KOLB, STYLIANOS TZEIS, MARCIO STURMER, DOMINIQUE BABUTY, JÖRG O. SCHWAB, GIUSEPPE MANTOVANI, SABINE JANKO, EZIO AIMÉ, ROLF OCKLENBURG, PETER SICK. (2010) Rationale and Design of the OPTION Study: Optimal Antitachycardia Therapy in ICD Patients without Pacing Indications. Pacing and Clinical Electrophysiology 33:9, 1141-1148
    CrossRef

  104. 104

    Melanie Maytin, Charles J. Love, Avi Fischer, Roger G. Carrillo, Juan D. Garisto, Maria Grazia Bongiorni, Luca Segreti, Roy M. John, Gregory F. Michaud, Christine M. Albert, Laurence M. Epstein. (2010) Multicenter Experience With Extraction of the Sprint Fidelis Implantable Cardioverter-Defibrillator Lead. Journal of the American College of Cardiology 56:8, 646-650
    CrossRef

  105. 105

    Amrut V. Ambardekar, Larry A. Allen, JoAnn Lindenfeld, Christopher M. Lowery, Anne P. Cannon, Joseph C. Cleveland, Andreas Brieke, William H. Sauer. (2010) Implantable cardioverter-defibrillator shocks in patients with a left ventricular assist device. The Journal of Heart and Lung Transplantation 29:7, 771-776
    CrossRef

  106. 106

    Norman C Wang, Jonathan P Piccini, Marvin A Konstam, Aldo P Maggioni, Brian Traver, Karl Swedberg, James E Udelson, Faiez Zannad, Thomas Cook, Christopher M OʼConnor, Alan B Miller, Liliana Grinfeld, Mihai Gheorghiade. (2010) Implantable Cardioverter-Defibrillators in Patients Hospitalized for Heart Failure With Chronically Reduced Left Ventricular Ejection Fraction. American Journal of Therapeutics 17:4, e78-e87
    CrossRef

  107. 107

    Badri Chandrasekaran, Peter J Cowburn. (2010) Heart failure: the challenge of selecting patients for implantable cardioverter defibrillator therapy. Expert Review of Medical Devices 7:4, 461-467
    CrossRef

  108. 108

    Oussama Wazni, Bruce L. Wilkoff. (2010) Strategic choices to reduce implantable cardioverter-defibrillator-related morbidity. Nature Reviews Cardiology 7:7, 376-383
    CrossRef

  109. 109

    N.C. Ewertsen, G. Steinbeck, J. Senges, D. Andresen. (2010) ICD-Therapie zur Primärprävention. Herzschrittmachertherapie + Elektrophysiologie 21:2, 109-111
    CrossRef

  110. 110

    Keiichi Inada, Kurt C. Roberts-Thomson, Jens Seiler, Daniel Steven, Usha B. Tedrow, Bruce A. Koplan, William G. Stevenson. (2010) Mortality and safety of catheter ablation for antiarrhythmic drug-refractory ventricular tachycardia in elderly patients with coronary artery disease. Heart Rhythm 7:6, 740-744
    CrossRef

  111. 111

    (2010) Section 9: Electrophysiology Testing and the Use of Devices in Heart Failure. Journal of Cardiac Failure 16:6, e115-e121
    CrossRef

  112. 112

    BRETT D. ATWATER, JAMES P. DAUBERT. (2010) Can We Improve the Identification of Ventricular Scar to Guide Substrate-Based Ventricular Tachycardia Ablation?. Journal of Cardiovascular Electrophysiology 21:6, 685-687
    CrossRef

  113. 113

    K.H. Ladwig, J. Ronel, J. Baumert, C. Kolb. (2010) Psychische Komorbidität und Lebensqualität bei Patienten mit implantierbarem Kardioverter/Defibrillator (ICD). Herzschrittmachertherapie + Elektrophysiologie 21:2, 129-136
    CrossRef

  114. 114

    K. Seidl, M. Strauss, T. Kleemann. (2010) ICD-Therapie zur Sekundärprävention. Herzschrittmachertherapie + Elektrophysiologie 21:2, 96-101
    CrossRef

  115. 115

    T. JARED BUNCH, J. PETER WEISS, BRIAN G. CRANDALL, JOHN D. DAY, JOHN P. DIMARCO, JOHN D. FERGUSON, PAMELA K. MASON, GEORGE McDANIEL, JEFFREY S. OSBORN, DAVE WIGGINS, SRIJOY MAHAPATRA. (2010) Image Integration Using Intracardiac Ultrasound and 3D Reconstruction for Scar Mapping and Ablation of Ventricular Tachycardia. Journal of Cardiovascular Electrophysiology 21:6, 678-684
    CrossRef

  116. 116

    Sanjeev P. Bhavnani, Jeffrey Kluger, Craig I. Coleman, C. Michael White, Danette Guertin, Nabil A. Shafi, Ravi K. Yarlagadda, Christopher A. Clyne. (2010) The prognostic impact of shocks for clinical and induced arrhythmias on morbidity and mortality among patients with implantable cardioverter-defibrillators. Heart Rhythm 7:6, 755-760
    CrossRef

  117. 117

    C.W. Israel. (2010) ICD-Therapie zur Primärprävention des plötzlichen Herztods. Herzschrittmachertherapie + Elektrophysiologie 21:2, 102-108
    CrossRef

  118. 118

    MANOJ N. OBEYESEKERE, WILLIAM CHAN, DION STUB, S PRABHU, ELIZA P TEO, GEOFF TOOGOOD, JUSTIN MARIANI, ARCHER BROUGHTON, PETER M KISTLER. (2010) Left Ventricular Ejection Fraction and Absence of ACE Inhibitor/Angiotensin II Receptor Blocker Predicts Appropriate Defibrillator Therapy in the Primary Prevention Population. Pacing and Clinical Electrophysiology 33:6, 696-704
    CrossRef

  119. 119

    C. Mewis, H.-R. Neuberger, A. Buob. (2010) Ist die intraoperative ICD-Testung noch nötig?. Herzschrittmachertherapie + Elektrophysiologie 21:2, 123-128
    CrossRef

  120. 120

    Beata Sredniawa, Michal Mazurek, Radoslaw Lenarczyk, Oskar Kowalski, Jacek Kowalczyk, Zbigniew Kalarus. (2010) Early therapy following myocardial infarction: arguments for and against implantable cardioverter-defibrillators. Future Cardiology 6:3, 315-323
    CrossRef

  121. 121

    JAVIER JIMÉNEZ-CANDIL, MARÍA RUIZ, JESÚS HERRERO, VÍCTOR LEÓN, ANA MARTÍN, JOSÉ MORIÑIGO, CLAUDIO LEDESMA, CÁNDIDO MARTÍN-LUENGO. (2010) Relationship between the Duration of the Basal QRS Complex and Electrical Therapies for Ventricular Tachycardias among ICD Patients. Pacing and Clinical Electrophysiology 33:5, 596-604
    CrossRef

  122. 122

    Richard N Vest, Michael R Gold. (2010) Risk stratification of ventricular arrhythmias in patients with systolic heart failure. Current Opinion in Cardiology 25:3, 268-275
    CrossRef

  123. 123

    Cihan Cevik, Kenneth Nugent, Alejandro Perez-Verdia, R. David Fish. (2010) Prophylactic Implantation of Cardioverter Defibrillators in Idiopathic Nonischemic Cardiomyopathy for the Primary Prevention of Death: A Narrative Review. Clinical Cardiology 33:5, 254-260
    CrossRef

  124. 124

    H. Van Herendael, A. Pinter, K. Ahmad, V. Korley, I. Mangat, P. Dorian. (2010) Role of antiarrhythmic drugs in patients with implantable cardioverter defibrillators. Europace 12:5, 618-625
    CrossRef

  125. 125

    David J. Whellan, Kevin T. Ousdigian, Sana M. Al-Khatib, Wenji Pu, Shantanu Sarkar, Charles B. Porter, Behzad B. Pavri, Christopher M. O'Connor. (2010) Combined Heart Failure Device Diagnostics Identify Patients at Higher Risk of Subsequent Heart Failure Hospitalizations. Journal of the American College of Cardiology 55:17, 1803-1810
    CrossRef

  126. 126

    Usha B. Tedrow, William G. Stevenson. (2010) Arrhythmias: Catheter ablation for prevention of ventricular tachycardia. Nature Reviews Cardiology 7:4, 181-182
    CrossRef

  127. 127

    ROBERT G. TURCOTT, TODD J. PAVEK. (2010) Identification of Hemodynamically Unstable Arrhythmias Using Subcutaneous Photoplethysmography. Journal of Cardiovascular Electrophysiology 21:4, 448-454
    CrossRef

  128. 128

    Michael O. Sweeney, Lou Sherfesee, Paul J. DeGroot, Mark S. Wathen, Bruce L. Wilkoff. (2010) Differences in effects of electrical therapy type for ventricular arrhythmias on mortality in implantable cardioverter-defibrillator patients. Heart Rhythm 7:3, 353-360
    CrossRef

  129. 129

    HELMUT U. KLEIN, ULF MELTENDORF, SVEN REEK, JAN SMID, SEBASTIAN KUSS, IWONA CYGANKIEWICZ, CHRISTIAN JONS, STEVEN SZYMKIEWICZ, FRANK BUHTZ, ANKE WOLLBRUECK, WOJCIECH ZAREBA, ARTHUR J. MOSS. (2010) Bridging a Temporary High Risk of Sudden Arrhythmic Death. Experience with the Wearable Cardioverter Defibrillator (WCD). Pacing and Clinical Electrophysiology 33:3, 353-367
    CrossRef

  130. 130

    Bernhard Maisch, Sabine Pankuweit. (2010) Therapie der fortgeschrittenen Herz - insuffizienz: medikamentös, Resynchronisation (CRT), Operation. Herz 35:2, 94-101
    CrossRef

  131. 131

    Massimo Santini, Maurizio Lunati, Pascal Defaye, Johann Mermi, Alessandro Proclemer, Silvia Castillo-Arroys, Giulio Molon, Elisabetta Santi, Tiziana Santo, Xavier Navarro, Axel Kloppe. (2010) Prospective multicenter randomized trial of fast ventricular tachycardia termination by prolonged versus conventional anti-tachyarrhythmia burst pacing in implantable cardioverter-defibrillator patients-Atp DeliVery for pAiNless ICD thErapy (ADVANCE-D) Trial results. Journal of Interventional Cardiac Electrophysiology 27:2, 127-135
    CrossRef

  132. 132

    Jason Bradfield, Noel G. Boyle. (2010) The paradox of ICD shocks: Sudden cardiac death prevention—Heart failure death acceleration. Heart Rhythm 7:3, 361-362
    CrossRef

  133. 133

    C. Jan Willem Borleffs, Johannes B. van Rees, Guido H. van Welsenes, Enno T. van der Velde, Lieselot van Erven, Jeroen J. Bax, Martin J. Schalij. (2010) Prognostic Importance of Atrial Fibrillation in Implantable Cardioverter-Defibrillator Patients. Journal of the American College of Cardiology 55:9, 879-885
    CrossRef

  134. 134

    Douglas S. Lee, Andrew D. Krahn, Jeffrey S. Healey, David Birnie, Eugene Crystal, Paul Dorian, Christopher S. Simpson, Yaariv Khaykin, Douglas Cameron, Amir Janmohamed, Raymond Yee, Peter C. Austin, Zhongliang Chen, Judy Hardy, Jack V. Tu. (2010) Evaluation of Early Complications Related to De Novo Cardioverter Defibrillator Implantation. Journal of the American College of Cardiology 55:8, 774-782
    CrossRef

  135. 135

    Garrick C. Stewart, Joanne R. Weintraub, Parakash P. Pratibhu, Marc J. Semigran, Janice M. Camuso, Kimberly Brooks, Sui W. Tsang, Mary Susan Anello, Viviane T. Nguyen, Eldrin F. Lewis, Anju Nohria, Akshay S. Desai, Michael M. Givertz, Lynne W. Stevenson. (2010) Patient Expectations From Implantable Defibrillators to Prevent Death in Heart Failure. Journal of Cardiac Failure 16:2, 106-113
    CrossRef

  136. 136

    Miguel Á. Arias, Marta Pachón, Alberto Puchol, Eduardo Castellanos. (2010) Electrofisiología cardiaca y arritmias. Revista Española de Cardiología 63, 61-72
    CrossRef

  137. 137

    William G Stevenson, Usha Tedrow. (2010) Preventing ventricular tachycardia with catheter ablation. The Lancet 375:9708, 4-6
    CrossRef

  138. 138

    Yuji Okuyama. (2010) Tactics for the Reduction of Inappropriate Implantable Cardioverter Defibrillator Shocks. Circulation Journal 74:7, 1290-1291
    CrossRef

  139. 139

    Anthony N. DeMaria, Jeroen J. Bax, Ori Ben-Yehuda, Gregory K. Feld, Barry H. Greenberg, Jennifer Hall, Mark Hlatky, Wilbur Y.W. Lew, João A.C. Lima, Alan S. Maisel, Sanjiv M. Narayan, Steven Nissen, David J. Sahn, Sotirios Tsimikas. (2010) Highlights of the Year in JACC 2009. Journal of the American College of Cardiology 55:4, 380-407
    CrossRef

  140. 140

    Karl-Heinz Kuck, Anselm Schaumann, Lars Eckardt, Stephan Willems, Rodolfo Ventura, Etienne Delacrétaz, Heinz-Friedrich Pitschner, Josef Kautzner, Burghard Schumacher, Peter S Hansen. (2010) Catheter ablation of stable ventricular tachycardia before defibrillator implantation in patients with coronary heart disease (VTACH): a multicentre randomised controlled trial. The Lancet 375:9708, 31-40
    CrossRef

  141. 141

    Fred Kusumoto, Nora Goldschlager. (2010) Remote Monitoring of Patients With Implanted Cardiac Devices. Clinical Cardiology 33:1, 10-17
    CrossRef

  142. 142

    Michiro Kiryu, Shinichi Niwano, Jun Kishihara, Yuya Aoyama, Shoko Ishikawa, Masami Murakami, Sayaka Kurokawa, Yoshihiro Yumoto, Ryuta Imaki, Hidehira Fukaya, Hiroe Niwano, Tohru Izumi. (2010) Preventive Effect of Amiodarone on VT/VF Events in ICD Patients with Structural Heart Diseases. Journal of Arrhythmia 26:4, 250-258
    CrossRef

  143. 143

    Naomichi Matsumoto, Koichiro Kumagai, Masahiro Ogawa, Kunihiro Matsuo, Tomoo Yasuda, Hideo Takashima, Chiharu Mitsutake, Soichi Muraoka, Akira Matsunaga, Shin-ichiro Miura, Keijiro Saku. (2010) Efficacy of Additional Amiodarone Therapy in Patients with an Implantable Cardioverter-Defibrillator. Journal of Arrhythmia 26:2, 103-110
    CrossRef

  144. 144

    Mithilesh Kumar Das, Waddah Maskoun, Changyu Shen, Mark A. Michael, Hussam Suradi, Mona Desai, Roopa Subbarao, Deepak Bhakta. (2010) Fragmented QRS on twelve-lead electrocardiogram predicts arrhythmic events in patients with ischemic and nonischemic cardiomyopathy. Heart Rhythm 7:1, 74-80
    CrossRef

  145. 145

    C. Adlbrecht, M. Hlsmann, M. Gwechenberger, G. Strunk, C. Khazen, F. Wiesbauer, M. Elhenicky, S. Neuhold, T. Binder, G. Maurer, I. M. Lang, R. Pacher. (2009) Outcome after device implantation in chronic heart failure is dependent on concomitant medical treatment. European Journal of Clinical Investigation 39:12, 1073-1081
    CrossRef

  146. 146

    M. Gasparini, C. Menozzi, A. Proclemer, M. Landolina, S. Iacopino, A. Carboni, E. Lombardo, F. Regoli, M. Biffi, V. Burrone, A. Denaro, G. Boriani. (2009) A simplified biventricular defibrillator with fixed long detection intervals reduces implantable cardioverter defibrillator (ICD) interventions and heart failure hospitalizations in patients with non-ischaemic cardiomyopathy implanted for primary prevention: the RELEVANT [Role of long dEtection window programming in patients with LEft VentriculAr dysfunction, Non-ischemic eTiology in primary prevention treated with a biventricular ICD] study. European Heart Journal 30:22, 2758-2767
    CrossRef

  147. 147

    Jeffrey J. Goldberger, Rod Passman. (2009) Implantable Cardioverter-Defibrillator Therapy After Acute Myocardial Infarction. Journal of the American College of Cardiology 54:22, 2001-2005
    CrossRef

  148. 148

    Joseph D. Mishkin, Sherry J. Saxonhouse, Gregory W. Woo, Thomas A. Burkart, William M. Miles, Jamie B. Conti, Richard S. Schofield, Samuel F. Sears, Juan M. Aranda. (2009) Appropriate Evaluation and Treatment of Heart Failure Patients After Implantable Cardioverter-Defibrillator Discharge. Journal of the American College of Cardiology 54:22, 1993-2000
    CrossRef

  149. 149

    Steinbeck, Gerhard, Andresen, Dietrich, Seidl, Karlheinz, Brachmann, Johannes, Hoffmann, Ellen, Wojciechowski, Dariusz, Kornacewicz-Jach, Zdzisława, Sredniawa, Beata, Lupkovics, Géza, Hofgärtner, Franz, Lubinski, Andrzej, Rosenqvist, Mårten, Habets, Alphonsus, Wegscheider, Karl, Senges, Jochen, . (2009) Defibrillator Implantation Early after Myocardial Infarction. New England Journal of Medicine 361:15, 1427-1436
    Full Text

  150. 150

    Garber, Alan M., Hlatky, Mark A., . (2009) The Confirmatory Trial in Comparative-Effectiveness Research. New England Journal of Medicine 361:15, 1498-1499
    Full Text

  151. 151

    M. Brignole. (2009) Are complications of implantable defibrillators under-estimated and benefits over-estimated?. Europace 11:9, 1129-1133
    CrossRef

  152. 152

    PIETRO FRANCIA, CRISTINA BALLA, ARIANNA UCCELLINI, RICCARDO CAPPATO. (2009) Arrhythmia Detection in Single- and Dual-Chamber Implantable Cardioverter Defibrillators: The More Leads, the Better?. Journal of Cardiovascular Electrophysiology 20:9, 1077-1082
    CrossRef

  153. 153

    Andrew E. Epstein. (2009) Update on primary prevention implantable cardioverter-defibrillator therapy. Current Cardiology Reports 11:5, 335-342
    CrossRef

  154. 154

    Richard K. Shepard, Kenneth A. Ellenbogen. (2009) Predicting Outcome After Implantable Cardioverter-Defibrillator Therapy. Journal of the American College of Cardiology 54:9, 829-831
    CrossRef

  155. 155

    Robert J. Myerburg, Vivek Reddy, Agustin Castellanos. (2009) Indications for Implantable Cardioverter-Defibrillators Based on Evidence and Judgment. Journal of the American College of Cardiology 54:9, 747-763
    CrossRef

  156. 156

    , J. Vijgen, G. Botto, J. Camm, C.-J. Hoijer, W. Jung, J.-Y. Le Heuzey, A. Lubinski, T. M. Norekval, M. Santomauro, M. Schalij, J.-P. Schmid, P. Vardas. (2009) Consensus statement of the European Heart Rhythm Association: updated recommendations for driving by patients with implantable cardioverter defibrillators. Europace 11:8, 1097-1107
    CrossRef

  157. 157

    Larisa G. Tereshchenko, Mitchell N. Faddis, Barry J. Fetics, Karl E. Zelik, Igor R. Efimov, Ronald D. Berger. (2009) Transient Local Injury Current in Right Ventricular Electrogram After Implantable Cardioverter-Defibrillator Shock Predicts Heart Failure Progression. Journal of the American College of Cardiology 54:9, 822-828
    CrossRef

  158. 158

    Gregory M. Marcus, Edmund Keung, Melvin M. Scheinman. (2009) The Year in Review of Clinical Cardiac Electrophysiology. Journal of the American College of Cardiology 54:9, 777-787
    CrossRef

  159. 159

    C. W. Israel. (2009) Do some implant too many defibrillators or others too few?. Europace 11:8, 982-984
    CrossRef

  160. 160

    Giovanni Lucignani. (2009) Metaiodobenzylguanidine (mIBG) molecular imaging in implantable cardioverter defibrillator (ICD) therapy planning: a health technology assessment issue. European Journal of Nuclear Medicine and Molecular Imaging 36:8, 1346-1350
    CrossRef

  161. 161

    Akli Otmani, Ludovic Trinquart, Eloi Marijon, Thomas Lavergne, Xavier Waintraub, Antoine Lepillier, Gilles Chatellier, Jean-Yves Le Heuzey. (2009) Rates and predictors of appropriate implantable cardioverter-defibrillator therapy delivery: Results from the EVADEF cohort study. American Heart Journal 158:2, 230-237.e1
    CrossRef

  162. 162

    M. Meine, T. Smith, R. N.W. Hauer. (2009) The economical challenge in the treatment of chronic heart failure: is primary prophylactic ICD therapy cost-effective in Europe?. Europace 11:6, 689-691
    CrossRef

  163. 163

    Etienne M. Aliot, William G. Stevenson, Jesus Ma Almendral-Garrote, Frank Bogun, C. Hugh Calkins, Etienne Delacretaz, Paolo Della Bella, Gerhard Hindricks, Pierre Jaïs, Mark E. Josephson, Josef Kautzner, G. Neal Kay, Karl-Heinz Kuck, Bruce B. Lerman, Francis Marchlinski, Vivek Reddy, Martin-Jan Schalij, Richard Schilling, Kyoko Soejima, David Wilber. (2009) EHRA/HRS Expert Consensus on Catheter Ablation of Ventricular Arrhythmias. Heart Rhythm 6:6, 886-933
    CrossRef

  164. 164

    C. Cevik, A. Perez-Verdia, K. Nugent. (2009) Implantable cardioverter defibrillators and their role in heart failure progression. Europace 11:6, 710-715
    CrossRef

  165. 165

    E. M. Aliot, W. G. Stevenson, J. M. Almendral-Garrote, F. Bogun, C. H. Calkins, E. Delacretaz, P. D. Bella, G. Hindricks, P. Jais, M. E. Josephson, J. Kautzner, G. N. Kay, K.-H. Kuck, B. B. Lerman, F. Marchlinski, V. Reddy, M.-J. Schalij, R. Schilling, K. Soejima, D. Wilber. (2009) EHRA/HRS Expert Consensus on Catheter Ablation of Ventricular Arrhythmias: Developed in a partnership with the European Heart Rhythm Association (EHRA), a Registered Branch of the European Society of Cardiology (ESC), and the Heart Rhythm Society (HRS); in collaboration with the American College of Cardiology (ACC) and the American Heart Association (AHA). Europace 11:6, 771-817
    CrossRef

  166. 166

    MICHAEL R. GOLD, MARK W. KROLL, KENNETH A. ELLENBOGEN. (2009) Defibrillation Testing at ICD Implantation: Are We Asking the Wrong Question?. Pacing and Clinical Electrophysiology 32:5, 567-569
    CrossRef

  167. 167

    Katherine C. Wu. (2009) MRI with late gadolinium enhancement as a predictor of ventricular arrhythmias. Current Cardiovascular Imaging Reports 2:2, 116-123
    CrossRef

  168. 168

    (2009) Defibrillators in Heart Failure and Quality of Life. New England Journal of Medicine 360:2, 187-189
    Full Text

  169. 169

    Daniel R. Frisch. (2009) Value of implantable cardioverter-defibrillators in managing atrial fibrillation. Heart Rhythm 6:1, 9-10
    CrossRef

  170. 170

    (2009) Mortality risk and patients' quality of life on ICD therapy: insights from SCD-HeFT. Nature Clinical Practice Cardiovascular Medicine 6:1, 3-4
    CrossRef

  171. 171

    Takanori Arimoto, Hiroshi Tada, Yukio Sekiguchi, Takashi Koyama, Miyako Igarashi, Hiro Yamasaki, Takeshi Machino, Kenji Kuroki, Keisuke Kuga, Kazutaka Aonuma. (2009) Efficacy and Safety of Strict Voltage-based Substrate Mapping and Radiofrequency Catheter Ablation in Electrical Storms—Review of Substrate-mapping Guided Ablation in Frequent Appropriate Shocks. Journal of Arrhythmia 25:4, 193-202
    CrossRef

  172. 172

    Yukio Hosaka, Masaomi Chinushi, Kenichi Iijima, Akiko Sanada, Hiroshi Furushima, Yoshifusa Aizawa. (2009) Correlation between Surface and Intracardiac Electrocardiogram in a Patient with Inappropriate Defibrillation Shocks Due to Hyperkalemia. Internal Medicine 48:13, 1153-1156
    CrossRef

  173. 173

    Mark V. Sherrid, James P. Daubert. (2008) Risks and Challenges of Implantable Cardioverter-Defibrillators in Young Adults. Progress in Cardiovascular Diseases 51:3, 237-263
    CrossRef

  174. 174

    Healey, Jeff, Connolly, Stuart, . (2008) Life and Death after ICD Implantation. New England Journal of Medicine 359:10, 1058-1059
    Full Text

Letters