Original Article

Anesthesia Awareness and the Bispectral Index

Michael S. Avidan, M.B., B.Ch., Lini Zhang, M.D., Beth A. Burnside, B.A., Kevin J. Finkel, M.D., Adam C. Searleman, B.S., Jacqueline A. Selvidge, B.S., Leif Saager, M.D., Michelle S. Turner, B.S., Srikar Rao, B.A., Michael Bottros, M.D., Charles Hantler, M.D., Eric Jacobsohn, M.B., Ch.B., and Alex S. Evers, M.D.

N Engl J Med 2008; 358:1097-1108March 13, 2008

Abstract

Background

Awareness during anesthesia is a serious complication with potential long-term psychological consequences. Use of the bispectral index (BIS), developed from a processed electroencephalogram, has been reported to decrease the incidence of anesthesia awareness when the BIS value is maintained below 60. In this trial, we sought to determine whether a BIS-based protocol is better than a protocol based on a measurement of end-tidal anesthetic gas (ETAG) for decreasing anesthesia awareness in patients at high risk for this complication.

Methods

We randomly assigned 2000 patients to BIS-guided anesthesia (target BIS range, 40 to 60) or ETAG-guided anesthesia (target ETAG range, 0.7 to 1.3 minimum alveolar concentration [MAC]). Postoperatively, patients were assessed for anesthesia awareness at three intervals (0 to 24 hours, 24 to 72 hours, and 30 days after extubation).

Results

We assessed 967 and 974 patients from the BIS and ETAG groups, respectively. Two cases of definite anesthesia awareness occurred in each group (absolute difference, 0%; 95% confidence interval [CI], −0.56 to 0.57%). The BIS value was greater than 60 in one case of definite anesthesia awareness, and the ETAG concentrations were less than 0.7 MAC in three cases. For all patients, the mean (±SD) time-averaged ETAG concentration was 0.81±0.25 MAC in the BIS group and 0.82±0.23 MAC in the ETAG group (P=0.10; 95% CI for the difference between the BIS and ETAG groups, −0.04 to 0.01 MAC).

Conclusions

We did not reproduce the results of previous studies that reported a lower incidence of anesthesia awareness with BIS monitoring, and the use of the BIS protocol was not associated with reduced administration of volatile anesthetic gases. Anesthesia awareness occurred even when BIS values and ETAG concentrations were within the target ranges. Our findings do not support routine BIS monitoring as part of standard practice. (ClinicalTrials.gov number, NCT00281489.)

Media in This Article

Figure 1Trial Enrollment.
Figure 2BIS Values and ETAG Concentrations over the Course of the Anesthesia in Patients with Definite or Possible Anesthesia Awareness in the BIS-Guided Group.
Article

Anesthesia awareness, also known as unintended intraoperative awareness, is the explicit recall of sensory perceptions during general anesthesia. Anesthesia awareness is rare,1,2 but the incidence may approach 1% in patients at high risk.3-5 Anesthesia awareness can lead to anxiety and post-traumatic stress disorder.6 The Joint Commission on Accreditation of Healthcare Organizations (JCAHO) has recommended that stringent efforts be made to prevent anesthesia awareness,7 and the American Society of Anesthesiologists (ASA) has published guidelines on the subject.8 According to a sentinel-event alert disseminated by the JCAHO, between 20,000 and 40,000 cases of anesthesia awareness may occur yearly in the United States.7

Of the 21 million patients who are given general anesthesia annually in the United States,7 the vast majority inhale an anesthetic gas. It is routine to monitor the concentration of exhaled anesthetic gases by analyzing the absorption of infrared light, which provides accurate identification and quantification of the exhaled gases.9 The minimum alveolar concentration (MAC) is the concentration of anesthetic gas required to prevent 50% of subjects from moving in response to a painful stimulus.10 The standard metric for assessing and comparing the potency of anesthetic-gas concentrations is the MAC equivalent; when anesthetic gases are used in combination, their MAC equivalents are additive. When the end-tidal anesthetic gas (ETAG) concentration is approximately one third of the MAC, 50% of subjects are not awake.10 Maintaining the ETAG concentration above 0.7 MAC may decrease the likelihood of anesthesia awareness.11,12

Several brain-function monitors based on the processed electroencephalogram or evoked potentials have been developed to assess anesthetic depth. The most widely used is the bispectral index (BIS) monitor (Aspect Medical Systems), which has been approved by the Food and Drug Administration. The BIS monitor processes a single frontal electroencephalographic signal to calculate a dimensionless number that provides a measure of the patient's level of consciousness. BIS values range from 100 to 0, reflecting the awake state and the absence of brain activity, respectively. Although the algorithm used to calculate the BIS number is proprietary, the electroencephalographic components of the algorithm have been identified: the frequency below which 95% of the power spectrum resides, the relative beta ratio, the relative synchrony of fast and slow waves, and the burst-suppression ratio.13 BIS values between 40 and 60 purportedly indicate adequate general anesthesia for surgery, and values below 40 indicate a deep hypnotic state.14 Targeting a range of BIS values between 40 and 60 is advocated to prevent anesthesia awareness while allowing a reduction in the administration of anesthetic agents.15,16

The B-Aware study showed an absolute risk reduction of 0.74% in anesthesia awareness among high-risk patients to whom anesthesia was administered according to a BIS-guided protocol, as compared with a control group who received no protocol-driven care.3 In a large, multicenter cohort study, BIS monitoring was used in about 40% of the patients without a specified BIS-guided protocol, and there was no reduction in anesthesia awareness associated with BIS monitoring.1 Despite the limited evidence, the BIS has been widely adopted, and many clinicians are administering anesthesia while aiming for BIS values below 60 in order to prevent anesthesia awareness.

We conducted the B-Unaware Trial to determine whether, in patients at high risk, the incidence of anesthesia awareness is reduced when clinicians follow a BIS-guided protocol rather than an ETAG-guided protocol.

Methods

Patients

The Washington University Human Research Protection Office approved the study. The Consolidated Standards of Reporting Trials guidelines were followed.17 We evaluated patients who were older than 18 years of age who were undergoing surgery at Barnes–Jewish Hospital for eligibility on the day before or the day of their surgery on the basis of preoperative assessment records. Patients were required to be at high risk for anesthesia awareness, to have general anesthesia with isoflurane, sevoflurane, or desflurane, and supplemental nitrous oxide was permitted. The criteria for identifying patients at high risk for anesthesia awareness were based on previous studies, reviews, and guidelines.3-5,7,8 The major criteria were preoperative long-term use of anticonvulsant agents, opiates, benzodiazepines, or cocaine; a cardiac ejection fraction less than 40%; a history of anesthesia awareness; a history of difficult intubation or anticipated difficult intubation; ASA physical status class 4 (those who have systemic disease that is a constant threat to life) or class 5 (those who are not expected to survive without the operation); aortic stenosis; end-stage lung disease; marginal exercise tolerance not resulting from musculoskeletal dysfunction; pulmonary hypertension; planned open-heart surgery; and daily alcohol consumption. The minor criteria were preoperative use of beta-blockers, chronic obstructive pulmonary disease, moderate exercise tolerance not resulting from musculoskeletal dysfunction, smoking two or more packs of cigarettes per day, and obesity, defined as a body-mass index (the weight in kilograms divided by the square of the height in meters) of more than 30. Patients at high risk were defined as those with at least one major criterion or two minor criteria. Patients were excluded if the surgical procedure or positioning of the patient prevented BIS monitoring or if the surgery required a wake-up test. Also excluded were patients who had dementia, who were unable to provide informed consent, or who had a history of stroke with residual neurologic deficits.

Study Design

The design was a single-center, prospective study, in which 2000 patients underwent prerandomization electronically in blocks of 100, with 50 patients assigned to a BIS-guided protocol and 50 to an ETAG-guided protocol. Eligible patients underwent randomization after providing written informed consent. The anesthesia practitioners were aware of the assignments of the patients, but the patients, the postoperative interviewers, the expert reviewers, and the statistician were not.

The manufacturer of the BIS monitor (Aspect Medical Systems) had no role in the study design, data collection, data analysis, or manuscript preparation. No study monitors or other means of support were provided by Aspect Medical Systems.

Procedures

A BIS Quatro Sensor (Aspect Medical Systems) was applied to the forehead of each patient. The anesthesia practitioners caring for the patients in the ETAG group used a monitor configuration that omitted the BIS number, so they were unaware of the BIS values. The practitioners in both groups could view the ETAG concentrations. In the BIS group, an audible alarm was set to indicate when the BIS value exceeded 60 or fell below 40; no ETAG alarms were set in the BIS group, and the practitioners were not instructed to maintain the ETAG concentration within any range.

In the ETAG group, an audible alarm was set to indicate when the ETAG concentration fell below 0.7 MAC or exceeded 1.3 MAC. Nitrous oxide was taken into account in the MAC calculations. During cardiopulmonary bypass, the anesthetic-gas concentration was measured from the effluent of the cardiopulmonary-bypass machine.18 A sign was affixed to the anesthesia machines reminding the practitioners to check the BIS value or ETAG concentration and to consider whether the patient might have intraoperative awareness. BIS values and ETAG concentrations were collected at 1-second intervals and were downloaded to a computer for subsequent analysis with TrendFace Solo software (ixellence). Manual records of anesthesia and digital photographs of monitor trends were used as alternatives in the event that the computer data were incomplete.

Anesthesia awareness was assessed with the use of the Brice questionnaire.19 The investigators were unaware of the patients' assignments and of previous assessments of their anesthesia awareness. Patients were contacted within 24 hours, between 24 and 72 hours, and at 30 days after extubation. Previous studies have shown that anesthesia awareness is not reliably detected with only a single interview.3,20 While the study was ongoing, the quality-assurance coordinator in the anesthesiology department contacted the patients who had reported memories of the period between “going to sleep” and “waking up” at any interview and obtained further information about their experiences. The coordinator offered all patients who reported such memories referral for counseling.

After all patients had completed the study, three experts who were unaware of the assignments of the patients independently reviewed the questionnaire responses and decided whether each patient definitely had had anesthesia awareness, might have had anesthesia awareness, or did not have anesthesia awareness. In assessing whether anesthesia awareness occurred, the experts were instructed to focus on memories of events that could occur only in the operating room during the anesthetic and surgical periods. The outcome was determined when at least two of the experts were in agreement. If two experts held opposing views, a fourth expert was asked to review the questionnaires.

On the basis of the accounts given by the patients and the information in the anesthesia records, an investigator who was unaware of the assignments identified a time window during which anesthesia awareness was likely to have occurred. When a specific time window could not be identified, the entire anesthetic period was considered.

To identify the maintenance period of anesthesia, researchers who were unaware of the treatment assignments systematically reviewed the BIS and ETAG traces from each patient who did not report anesthesia awareness. After the administration of the intravenous induction agent, there was a characteristic nadir in the BIS trace. The start of the maintenance period was defined as 3 minutes after the nadir unless the BIS rose above the upper bound of 60 during this time. Three minutes was predetermined to be adequate to account for the transition from an intravenous induction to a volatile gas-based general anesthetic. The end of the maintenance period and the start of emergence was defined as the beginning of a consistent and final decrease in the ETAG concentration to less than 0.2 MAC. If emergence did not occur in the operating room, the final data points were included in the maintenance period. Every trace was analyzed for sustained 30-second periods of BIS values above the threshold of 60 or ETAG concentrations below the threshold of 0.7 MAC during the maintenance period. Periods with missing data were excluded from the analysis.

Statistical Analysis

The primary outcome of the study was a decrease in definite anesthesia awareness in the BIS group as compared with the ETAG group. The anticipated incidence of anesthesia awareness was 1% for the ETAG group, on the basis of the incidence rates reported for patients at high risk for anesthesia awareness,3-5 and 0.1% for the BIS group, on the basis of previous studies.3,21 A total of 940 patients would be required in each group to detect this 0.9% difference with a one-tailed alpha of 0.05 and a power of 80% with the use of Fisher's exact test. Confidence intervals for absolute risk reduction were calculated with the use of Newcombe's method without continuity correction.22 There was no interim analysis. The chi-square test, Fisher's exact test, an unpaired t-test, and an unpaired Mann–Whitney test were used for other comparisons between groups. Intention-to-treat analysis was planned. Agreement among the experts who were assessing anesthesia awareness was quantified with the use of a two-way, random-effects, intraclass correlation coefficient for absolute agreement according to the following metric23: no=0, maybe=1, and yes=2. All reported P values are two-sided, and a P value less than 0.05 was considered to indicate statistical significance. Statistical analysis was performed with the R statistical environment (R Foundation for Statistical Computing).

Results

Of 2000 patients enrolled during a 14-month period (from September 2005 to October 2006), 1941 completed the study (Figure 1Figure 1Trial Enrollment.). Patients were excluded as a result of changes in the planned anesthetic technique or canceled surgery. No patients who were assigned to the ETAG group were treated with the BIS protocol, and no patients assigned to BIS were treated with the ETAG protocol. A total of 1754 patients (90.4%) completed all three interviews, 133 patients (6.9%) completed only two interviews, and 18 patients (0.9%) completed only one interview. Table 1Table 1Patient Characteristics in the BIS-Guided and ETAG-Guided Groups. shows the characteristics of the patients. There were significantly more patients with underlying neurologic disease in the ETAG group than in the BIS group (22.0% vs. 17.2%, P=0.008). The groups were similar with respect to other baseline characteristics. Overall, 71.4% of patients had an ASA physical status class of 3 (severe systemic disease), 4, or 5; 27.0% underwent open-heart surgery; 74.6% had three or more inclusion criteria; and 61.6% had two or more preexisting medical conditions. Nitrous oxide was administered to 350 patients in the BIS group and 351 patients in the ETAG group.

Table 2Table 2Patients Who Reported Memories of the Period between “Going to Sleep” and “Waking Up.” gives details for the 16 patients who, during at least one interview, reported memories of the period between “going to sleep” and “waking up” at the end of surgery. Experts who were unaware of the treatment assignments determined that four patients (Patients 1 through 4) had definite anesthesia awareness; two of these patients were in the BIS group and two were in the ETAG group (absolute risk reduction, 0%; 95% confidence interval [CI], −0.56 to 0.57). Two of these patients reported memories at all three interviews, one reported memories at only the last two interviews, and one reported memories at only the third interview. The overall incidence of definite anesthesia awareness was 0.21% (95% CI, 0.08 to 0.53).

Five other patients (Patients 5 through 9) had possible anesthesia awareness; four were in the BIS group and one was in the ETAG group. These patients reported memories at only one interview, and four reported memories at only the third interview. The incidence rates of definite or possible anesthesia awareness were 0.62% and 0.31% in the BIS and ETAG groups, respectively, with an absolute difference between the groups of 0.31% (95% CI, −0.36 to 1.07). The overall incidence of definite or possible anesthesia awareness was 0.46% (95% CI, 0.24 to 0.87). Table 2 shows the responses to interviews and the judgments of the expert reviewers, who had high agreement (intraclass correlation coefficient, 0.72; 95% CI, 0.49 to 0.88). There were no statistically significant differences between patients who had definite or possible anesthesia awareness and the rest of the patients in terms of the characteristics of the patients or the anesthetic drugs administered.

Figure 2Figure 2BIS Values and ETAG Concentrations over the Course of the Anesthesia in Patients with Definite or Possible Anesthesia Awareness in the BIS-Guided Group. shows the BIS values and ETAG concentrations for the six patients in the BIS-guided group with definite (Patients 3 and 4) or possible (Patients 5 through 8) anesthesia awareness. Figure 3Figure 3BIS Values and ETAG Concentrations over the Course of the Anesthesia in Patients with Definite or Possible Anesthesia Awareness in the ETAG-Guided Group. shows the corresponding data for the three patients in the ETAG-guided group with definite (Patients 1 and 2) or possible (Patient 9) anesthesia awareness. The BIS value was above 60 in one of the four patients with definite anesthesia awareness and in three of the nine patients with possible or definite anesthesia awareness. The ETAG concentration was below 0.7 MAC in three of the four patients with definite anesthesia awareness and in seven of the nine with possible or definite anesthesia awareness.

During the maintenance phase of anesthesia, BIS values were continuously less than 60 in 825 of 1834 patients (45.0%), and ETAG concentrations were continuously greater than 0.7 MAC in 438 of 1717 patients (25.5%). The mean time-averaged BIS during the maintenance phase was 43.1±9.2 in the BIS group and 43.4±9.8 in the ETAG group (P=0.51; 95% CI for the difference between the BIS and ETAG groups, −1.2 to 0.6). The mean (±SD) time-averaged ETAG concentration during the maintenance phase was 0.81±0.25 MAC in the BIS group and 0.82±0.23 MAC in the ETAG group (P=0.10; 95% CI for the difference between the BIS and ETAG groups, −0.04 to 0.01 MAC).

Discussion

In the B-Unaware Trial, a structured protocol based on the BIS resulted in neither a lower incidence of anesthesia awareness nor a reduced administration of volatile anesthetic gas as compared with a structured protocol based on ETAG. It has been suggested that when the BIS value is less than 60, anesthesia awareness is unlikely.3,15,21 In the majority of cases of definite or possible anesthesia awareness, BIS values were persistently under 60 during the period when anesthesia awareness was likely to have occurred (Figure 2 and Figure 3).

There were sustained periods when BIS values were greater than 60 in 55.0% of patients and ETAG concentrations were under 0.7 MAC in 74.5% of patients who did not have anesthesia awareness. The low mean BIS values in the BIS group could reflect the unwillingness of the anesthesia practitioners to decrease anesthetic administration solely on the basis of the BIS. The mean BIS value of the BIS group in the B-Aware study was low and almost identical to that in the present study.3 Both the BIS values and the ETAG concentrations were frequently outside the target ranges, possibly because of the difficulty of adhering to the prescribed protocols or reluctance on the part of the anesthesia practitioners to follow the protocols.

This trial has some important limitations. Although the trial did not demonstrate a reduction in anesthesia awareness, with 95% confidence intervals for absolute risk reduction of definite anesthesia awareness of −0.56 to 0.57%, the results remain consistent with a clinically significant number needed to treat in order to benefit of 179 and a clinically significant number needed to treat in order to harm of 175 with the BIS protocol. This study is also subject to some concerns common to all studies of anesthesia awareness: the diagnosis of anesthesia awareness may be subjective, the awareness interview may be invalid because repeated questioning may induce false memories, and it may be difficult to distinguish between memories of events in the operating room and events in the intensive care unit. It is encouraging that there was good agreement among the three assessors, who were unaware of the treatment assignments, and it was unnecessary to refer any decision to a fourth assessor.

It is important to emphasize that the results of this trial should not be extrapolated to patients receiving total intravenous anesthesia, which is considered to be a risk factor for anesthesia awareness.7 Indeed, BIS monitoring may be useful during total intravenous anesthesia, since it is not presently possible to monitor the blood concentrations of anesthetic agents continuously.

Anesthesia awareness cannot predictably be prevented in all patients with the BIS monitoring protocol used in this study. When a potent volatile anesthetic gas was administered, a structured protocol based on the BIS was not shown to be superior to a protocol based on ETAG concentrations for preventing anesthesia awareness. Reliance on BIS technology24 may provide patients and health care practitioners with a false sense of security about the reduction in the risk of anesthesia awareness. If BIS monitoring were routinely applied to all patients in the United States receiving general anesthesia,7 the cost of disposable electrodes alone would exceed $360 million annually. Our study was unable to demonstrate superiority of a BIS-guided protocol over an ETAG-guided protocol for preventing anesthesia awareness and does not provide support for the additional cost of BIS monitoring as part of standard practice.

Supported by a grant from the Barnes–Jewish Hospital Foundation (to Dr. Avidan).

No potential conflict of interest relevant to this article was reported.

We thank Jane Cueto, Dusica Stamenkovic, Elika Safarzadeh, Furqaan Sadiq, Barry Hong, Heiko Kaiser, Laura Veremakis, Brian Torres, Katy Dustman, Jeanine Riddle, Michael Rybak, Dia Tite, Nicholas Girdner, Bernadette Henrichs, and the many anesthesiologists, certified registered nurse anesthetists, and surgeons who facilitated the conduct of this study.

Source Information

From the Department of Anesthesiology, Washington University School of Medicine, St. Louis.

Address reprint requests to Dr. Avidan at Washington University School of Medicine, 660 S. Euclid Ave., Campus Box 8054, St. Louis, MO 63110, or at .

References

References

  1. 1

    Sebel PS, Bowdle TA, Ghoneim MM, et al. The incidence of awareness during anesthesia: a multicenter United States study. Anesth Analg 2004;99:833-839
    CrossRef | Web of Science | Medline

  2. 2

    Pollard RJ, Coyle JP, Gilbert RL, Beck JE. Intraoperative awareness in a regional medical system: a review of 3 years' data. Anesthesiology 2007;106:269-274
    CrossRef | Web of Science | Medline

  3. 3

    Myles PS, Leslie K, McNeil J, Forbes A, Chan MT. Bispectral index monitoring to prevent awareness during anaesthesia: the B-Aware randomised controlled trial. Lancet 2004;363:1757-1763
    CrossRef | Web of Science | Medline

  4. 4

    Ranta S, Jussila J, Hynynen M. Recall of awareness during cardiac anaesthesia: influence of feedback information to the anaesthesiologist. Acta Anaesthesiol Scand 1996;40:554-560
    CrossRef | Web of Science | Medline

  5. 5

    Phillips AA, McLean RF, Devitt JH, Harrington EM. Recall of intraoperative events after general anaesthesia and cardiopulmonary bypass. Can J Anaesth 1993;40:922-926
    CrossRef | Web of Science | Medline

  6. 6

    Lennmarken C, Bildfors K, Enlund G, Samuelsson P, Sandin R. Victims of awareness. Acta Anaesthesiol Scand 2002;46:229-231
    CrossRef | Web of Science | Medline

  7. 7

    Sentinel event alert: preventing, and managing the impact of, anesthesia awareness. Oakbrook Terrace, IL: Joint Commission, 2004. (Accessed February 19, 2008, at http://www.jointcommission.org/SentinelEvents/SentinelEventAlert/sea_32.htm.)

  8. 8

    Practice advisory for intraoperative awareness and brain function monitoring: a report by the American Society of Anesthesiologists Task Force on Intraoperative Awareness. Anesthesiology 2006;104:847-864
    CrossRef | Web of Science | Medline

  9. 9

    Walder B, Lauber R, Zbinden AM. Accuracy and cross-sensitivity of 10 different anesthetic gas monitors. J Clin Monit 1993;9:364-373
    CrossRef | Medline

  10. 10

    Eger EI II. Age, minimum alveolar anesthetic concentration, and minimum alveolar anesthetic concentration-awake. Anesth Analg 2001;93:947-953
    CrossRef | Web of Science | Medline

  11. 11

    Ghoneim MM. Awareness during anesthesia. Anesthesiology 2000;92:597-602
    CrossRef | Web of Science | Medline

  12. 12

    Gonsowski CT, Chortkoff BS, Eger EI II, Bennett HL, Weiskopf RB. Subanesthetic concentrations of desflurane and isoflurane suppress explicit and implicit learning. Anesth Analg 1995;80:568-572
    CrossRef | Web of Science | Medline

  13. 13

    Morimoto Y, Hagihira S, Koizumi Y, Ishida K, Matsumoto M, Sakabe T. The relationship between bispectral index and electroencephalographic parameters during isoflurane anesthesia. Anesth Analg 2004;98:1336-1340
    CrossRef | Web of Science | Medline

  14. 14

    BIS range and clinical state. Norwood, MA: Aspect Medical Systems. (Accessed February 19, 2008, at http://www.aspectmedical.com/assets/Documents/pdf/New_BIS_Range.pdf.)

  15. 15

    Punjasawadwong Y, Boonjeungmonkol N, Phongchiewboon A. Bispectral index for improving anaesthetic delivery and postoperative recovery. Cochrane Database Syst Rev 2007;4:CD003843-CD003843
    Medline

  16. 16

    Recart A, Gasanova I, White PF, et al. The effect of cerebral monitoring on recovery after general anesthesia: a comparison of the auditory evoked potential and bispectral index devices with standard clinical practice. Anesth Analg 2003;97:1667-1674
    CrossRef | Web of Science | Medline

  17. 17

    Moher D, Schulz KF, Altman D. The CONSORT statement: revised recommendations for improving the quality of reports of parallel-group randomized trials. JAMA 2001;285:1987-1991
    CrossRef | Web of Science | Medline

  18. 18

    Nussmeier NA, Moskowitz GJ, Weiskopf RB, Cohen NH, Fisher DM, Eger EI II. In vitro anesthetic washin and washout via bubble oxygenators: influence of anesthetic solubility and rates of carrier gas inflow and pump blood flow. Anesth Analg 1988;67:982-987
    CrossRef | Web of Science | Medline

  19. 19

    Brice DD, Hetherington RR, Utting JE. A simple study of awareness and dreaming during anaesthesia. Br J Anaesth 1970;42:535-542
    CrossRef | Web of Science | Medline

  20. 20

    Sandin RH, Enlund G, Samuelsson P, Lennmarken C. Awareness during anaesthesia: a prospective case study. Lancet 2000;355:707-711
    CrossRef | Web of Science | Medline

  21. 21

    Ekman A, Lindholm ML, Lennmarken C, Sandin R. Reduction in the incidence of awareness using BIS monitoring. Acta Anaesthesiol Scand 2004;48:20-26
    CrossRef | Web of Science | Medline

  22. 22

    Newcombe RG. Interval estimation for the difference between independent proportions: comparison of eleven methods. Stat Med 1998;17:873-890
    CrossRef | Web of Science | Medline

  23. 23

    Shrout PE. Measurement reliability and agreement in psychiatry. Stat Methods Med Res 1998;7:301-317
    CrossRef | Medline

  24. 24

    Corporate backgrounder. Norwood, MA: Aspect Medical Systems. (Accessed February 19, 2008, at http://www.aspectmedical.com/company/about/default.mspx.)

Citing Articles (99)

Citing Articles

  1. 1

    H. P. Grocott. (2012) Monitoring the brain in cardiac surgery - an evolving area for research. Anaesthesia 67:3, 216-219
    CrossRef

  2. 2

    M. Alkire, M. Avidan, E. Whitlock, H. A. Baghdoyan, G. Vanini, C. J. Watson, R. Lydic, M. I. Banks, S. M. Grady, H. L. Bennett, M. Boly, K. B. Domino, N. A. Metzger, G. A. Mashour, C. D. Kent, K. L. Posner, H. El Beheiry, P. Carlen, S. Freeman, N. Gokmen, O. Bayazit, A. Oniz, M. Ozgoren, C. Gonzalez, E. Jensen, K. Cardenas, S. Cortes, H. Litvan, S. Hameroff, T. Craddock, J. Tuszynski, H. C. Hemmings, V. Jantti, V. Jantti, P. Kauppinen, O. Vaisanen, K. Wendel, N. Subramaniyam, K. Kamata, D. Jordan, R. Ilg, A. Wohlschlager, E. F. Kochs, G. Schneider, E. Kahana, A. Raz, U. Livne, J. Resnik, R. Paz, M. Kelz, M. Kreuzer, S. Butovas, C. Schwarz, B. Antkowiak, G. Schneider, J. Kurata, S. B. Lazar, A. G. Garrity, G. A. Mashour, H. A. Baghdoyan, R. Lydic, D. T. J. Liley, X. Liu, B. D. Ward, S. Li, A. G. Hudetz, X. Liu, S. Pillay, P. Baracskay, A. G. Hudetz, G. A. Mashour, M. Maze, A. G. Messina, S. Munte, M. Ozgoren, O. Bayazit, N. Gokmen, S. Kocaaslan, A. Oniz, D. Pal, M. E. Walton, W. J. Lipinski, L. G. Koch, R. Lydic, S. L. Britton, G. A. Mashour, M. Perouansky, S. Pillay, J. A. Vizuete, J. B. McCallum, A. G. Hudetz, G. Plourde, R. Ramani, Q. Maolin, M. Rose, R. T. Todd, I. Rampil, A. Raz, D. Eimerl, H. Bergman, Z. Israel, F. C. Rodgers, M. Perouansky, R. A. Pearce, I. F. Russell, R. D. Sanders, G. Tononi, S. Laureys, J. Sleigh, C. Scheib, C. Scheib, G. Schneider, P. S. Sebel, J. K. Shin, U. C. Lee, S. W. Ku, S. Kim, G. A. Mashour, J. W. Sleigh, C. M. Scheib, T. A. Stekiel, J. B. McCallum, A. G. Hudetz, N. Terrando, C. Monaco, M. Feldmann, M. Maze, G. Untergehrer, D. Jordan, E. F. Kochs, G. Schneider, R. A. Veselis, J. A. Vizuete, S. Pillay, J. B. McCallum, K. M. Ropella, A. G. Hudetz, M. Wang, M. Wang, E. A. Trubshaw, S. Tordoff, N. Rich, L. Brittenden, S. Bowrey, J. Thompson. (2012) Abstracts presented at the 8th International Symposium on Memory and Awareness in Anesthesia (MAA8): Milwaukee, WI, USA, June 2-5, 2011. British Journal of Anaesthesia 108:2, 334-367
    CrossRef

  3. 3

    R Huhle, M Burghardt, S Zaunseder, N Wessel, T Koch, H Malberg, A R Heller. (2012) Effects of awareness and nociception on heart rate variability during general anaesthesia. Physiological Measurement 33:2, 207-217
    CrossRef

  4. 4

    Neville M. Gibbs, William M. Weightman. (2012) Beyond Effect Size. Anesthesia & Analgesia 114:2, 471-475
    CrossRef

  5. 5

    J. Rémi, S. Noachtar. (2012) Bedeutung des EEG in der Epilepsiediagnostik. Der Nervenarzt
    CrossRef

  6. 6

    C. Stoppe, D. Peters, A. V. Fahlenkamp, J. Cremer, S. Rex, G. Schalte, R. Rossaint, M. Coburn. (2012) aepEX monitor for the measurement of hypnotic depth in patients undergoing balanced xenon anaesthesia. British Journal of Anaesthesia 108:1, 80-88
    CrossRef

  7. 7

    Matthew T. V. Chan, Sin Shing Ho, Tony Gin. (2012) Performance of the Bispectral Index During Electrocautery. Journal of Neurosurgical Anesthesiology 24:1, 9-13
    CrossRef

  8. 8

    S. Bose, B. Lewis, E. Farag. (2012) Interference with BISTM values from a forced-air warming device. British Journal of Anaesthesia 108:1, 168-168
    CrossRef

  9. 9

    Gail D. Baura. 2012. Electroencephalographs. , 275-295.
    CrossRef

  10. 10

    Gregory Crosby, Deborah J. Culley. (2012) Processed Electroencephalogram and Depth of Anesthesia. Anesthesiology1
    CrossRef

  11. 11

    Won Joon Choi, Yun Hong Kim. (2012) How do you use bispectral index effectively for preventing re-awareness during general anesthesia?. Korean Journal of Anesthesiology 62:1, 1
    CrossRef

  12. 12

    L. David Hillis, Peter K. Smith, Jeffrey L. Anderson, John A. Bittl, Charles R. Bridges, John G. Byrne, Joaquin E. Cigarroa, Verdi J. DiSesa, Loren F. Hiratzka, Adolph M. Hutter, Michael E. Jessen, Ellen C. Keeley, Stephen J. Lahey, Richard A. Lange, Martin J. London, Michael J. Mack, Manesh R. Patel, John D. Puskas, Joseph F. Sabik, Ola Selnes, David M. Shahian, Jeffrey C. Trost, Michael D. Winniford. (2011) 2011 ACCF/AHA Guideline for Coronary Artery Bypass Graft Surgery. Journal of the American College of Cardiology 58:24, e123-e210
    CrossRef

  13. 13

    Shou-Zen Fan, Qin Wei, Pei-Fung Shi, Yu-Jen Chen, Quan Liu, Jiann-Shing Shieh. (2011) A comparison of patients’ heart rate variability and blood flow variability during surgery based on the Hilbert–Huang Transform. Biomedical Signal Processing and Control
    CrossRef

  14. 14

    Terri G. Monk, B. Craig Weldon. (2011) Does depth of anesthesia monitoring improve postoperative outcomes?. Current Opinion in Anaesthesiology 24:6, 665-670
    CrossRef

  15. 15

    Jaakko G. M. Klockars, Arja Hiller, Sinikka Münte, Mark J. van Gils, Tomi Taivainen. (2011) Spectral Entropy as a Measure of Hypnosis and Hypnotic Drug Effect of Total Intravenous Anesthesia in Children during Slow Induction and Maintenance. Anesthesiology1
    CrossRef

  16. 16

    (2011) Prevention of Intraoperative Awareness. New England Journal of Medicine 365:21, 2032-2035
    Full Text

  17. 17

    H. Babar-Craig, N. K. Rajani, P. Bailey, B. T. Kotecha. (2011) Validation of sleep nasendoscopy for assessment of snoring with bispectral index monitoring. European Archives of Oto-Rhino-Laryngology
    CrossRef

  18. 18

    Morgan Le Guen, Grégoire Trebbia, Edouard Sage, Charles Cerf, Marc Fischler. (2011) Intraoperative Cerebral Air Embolism During Lung Transplantation: Treatment with Early Hyperbaric Oxygen Therapy. Journal of Cardiothoracic and Vascular Anesthesia
    CrossRef

  19. 19

    Crosby, Gregory, . (2011) General Anesthesia — Minding the Mind during Surgery. New England Journal of Medicine 365:7, 660-661
    Full Text

  20. 20

    Avidan, Michael S., Jacobsohn, Eric, Glick, David, Burnside, Beth A., Zhang, Lini, Villafranca, Alex, Karl, Leah, Kamal, Saima, Torres, Brian, O'Connor, Michael, Evers, Alex S., Gradwohl, Stephen, Lin, Nan, Palanca, Ben J., Mashour, George A., . (2011) Prevention of Intraoperative Awareness in a High-Risk Surgical Population. New England Journal of Medicine 365:7, 591-600
    Full Text

  21. 21

    Stephen Napier. (2011) Out of the Frying Pan and Into the Fire. The American Journal of Bioethics 11:8, 60-61
    CrossRef

  22. 22

    Eugene A. Hessel. (2011) CNS Monitoring: The Current Weak State of the Evidence. Journal of Cardiothoracic and Vascular Anesthesia 25:4, e15-e16
    CrossRef

  23. 23

    Emery N. Brown, Patrick L. Purdon, Christa J. Van Dort. (2011) General Anesthesia and Altered States of Arousal: A Systems Neuroscience Analysis. Annual Review of Neuroscience 34:1, 601-628
    CrossRef

  24. 24

    David R. Drover, Clifford Schmiesing, Anthea F. Buchin, H. Rick Ortega, Jonathan W. Tanner, Joshua H. Atkins, Alex Macario. (2011) Titration of sevoflurane in elderly patients: blinded, randomized clinical trial, in non-cardiac surgery after beta-adrenergic blockade. Journal of Clinical Monitoring and Computing 25:3, 175-181
    CrossRef

  25. 25

    Toni Laurila, Tapio Sorvajärvi, Jaakko Saarela, Juha Toivonen, Daniel W. Wheeler, Luca Ciaffoni, Grant A. D. Ritchie, Clemens F. Kaminski. (2011) Optical Detection of the Anesthetic Agent Propofol in the Gas Phase. Analytical Chemistry 83:10, 3963-3967
    CrossRef

  26. 26

    George A. Mashour, Beverley A. Orser, Michael S. Avidan. (2011) Intraoperative Awareness. Anesthesiology 114:5, 1218-1233
    CrossRef

  27. 27

    George A Mashour, Michael S Avidan. (2011) Pharmacologic approaches to the prevention of intraoperative awareness. Expert Review of Neurotherapeutics 11:5, 611-613
    CrossRef

  28. 28

    &NA;. (2011) Pregnancy Does Not Enhance Volatile Anesthetic Sensitivity on the Brain. Survey of Anesthesiology 55:2, 82-83
    CrossRef

  29. 29

    Matthew A Klopman, Peter S Sebel. (2011) Cost-effectiveness of bispectral index monitoring. Current Opinion in Anaesthesiology 24:2, 177-181
    CrossRef

  30. 30

    Cor J. Kalkman, Linda M. Peelen, Karel G. Moons. (2011) Pick up the Pieces. Anesthesiology 114:3, 485-487
    CrossRef

  31. 31

    Miklos D. Kertai, Ben J. A. Palanca, Nirvik Pal, Beth A. Burnside, Lini Zhang, Furqaan Sadiq, Kevin J. Finkel, Michael S. Avidan. (2011) Bispectral Index Monitoring, Duration of Bispectral Index Below 45, Patient Risk Factors, and Intermediate-term Mortality after Noncardiac Surgery in the B-Unaware Trial. Anesthesiology 114:3, 545-556
    CrossRef

  32. 32

    Yasuhiro Morimoto, Yuko Nogami, Kaori Harada, Tsunehisa Tsubokawa, Kenichi Masui. (2011) Awareness during anesthesia: the results of a questionnaire survey in Japan. Journal of Anesthesia 25:1, 72-77
    CrossRef

  33. 33

    Hirobumi Okawa, Tomoko Ono, Eiji Hashiba, Toshihito Tsubo, Hironori Ishihara, Kazuyoshi Hirota. (2011) Use of bispectral index monitoring for a patient with hepatic encephalopathy requiring living donor liver transplantation: a case report. Journal of Anesthesia 25:1, 117-119
    CrossRef

  34. 34

    Shinji Matsuse, Yoshiki Hara, Takako Ohkura. (2011) The Possible Influence of Pulmonary Arterio-Venous Shunt and Hypoxic Pulmonary Vasoconstriction on Arterial Sevoflurane Concentration During One-Lung Ventilation. Anesthesia & Analgesia 112:2, 345-348
    CrossRef

  35. 35

    (2011) Therapeutic Hypothermia after Cardiac Arrest. New England Journal of Medicine 364:2, 185-188
    Full Text

  36. 36

    Richard Berkowitz, David Barinholtz. 2011. Anesthesia for Office-Based Pediatric Anesthesia. , 1077-1095.
    CrossRef

  37. 37

    Christopher M.B. Heard, James E. Fletcher. 2011. Sedation and Analgesia. , 1654-1681.
    CrossRef

  38. 38

    Hilary P Grocott, Sophie Davie, Christine Fedorow. (2010) Monitoring of brain function in anesthesia and intensive care. Current Opinion in Anaesthesiology 23:6, 759-764
    CrossRef

  39. 39

    TOM G. HANSEN, STEEN W. HENNEBERG, NEIL S. MORTON, KAARE CHRISTENSEN, ANDREW J. DAVIDSON, KATHERINE J. LEE, POLLYANNA HARDY, ANDREW WOLF. (2010) Pro-Con Debate: Pro-con debate: cohort studies vs the randomized clinical trial methodology in pediatric anesthesia. Pediatric Anesthesia 20:9, 880-894
    CrossRef

  40. 40

    A. V. Fahlenkamp, D. Peters, I. A. Biener, C. Billoet, C. C. Apfel, R. Rossaint, M. Coburn. (2010) Evaluation of bispectral index and auditory evoked potentials for hypnotic depth monitoring during balanced xenon anaesthesia compared with sevoflurane. British Journal of Anaesthesia 105:3, 334-341
    CrossRef

  41. 41

    Marion Leary, David A. Fried, David F. Gaieski, Raina M. Merchant, Barry D. Fuchs, Daniel M. Kolansky, Dana P. Edelson, Benjamin S. Abella. (2010) Neurologic prognostication and bispectral index monitoring after resuscitation from cardiac arrest. Resuscitation 81:9, 1133-1137
    CrossRef

  42. 42

    S. R. Springman, A.-C. Andrei, K. Willmann, D. A. Rusy, M. E. Warren, S. Han, M. Lee. (2010) A comparison of SNAP II© and bispectral index monitoring in patients undergoing sedation. Anaesthesia 65:8, 815-819
    CrossRef

  43. 43

    Kazuko Hayashi, Kenichiro Mita, Teiji Sawa. (2010) Electroencephalographic changes in the late cardiopulmonary bypass period are not reflected in the bispectral index. Clinical Neurophysiology 121:8, 1198-1204
    CrossRef

  44. 44

    M. ELHAKIM, D. ABDELHAMID, H. ABDELFATTACH, H. MAGDY, A. ELSAYED, M. ELSHAFEI. (2010) Effect of epidural dexmedetomidine on intraoperative awareness and post-operative pain after one-lung ventilation. Acta Anaesthesiologica Scandinavica 54:6, 703-709
    CrossRef

  45. 45

    Bruce H. Cohen. (2010) Pharmacologic effects on mitochondrial function. Developmental Disabilities Research Reviews 16:2, 189-199
    CrossRef

  46. 46

    Wakako Miyake, Yutaka Oda, Yuko Ikeda, Satoshi Hagihira, Hiroyoshi Iwaki, Akira Asada. (2010) Electroencephalographic response following midazolam-induced general anesthesia: relationship to plasma and effect-site midazolam concentrations. Journal of Anesthesia 24:3, 386-393
    CrossRef

  47. 47

    Miklos D. Kertai, Nirvik Pal, Ben J. A. Palanca, Nan Lin, Sylvia A. Searleman, Lini Zhang, Beth A. Burnside, Kevin J. Finkel, Michael S. Avidan. (2010) Association of Perioperative Risk Factors and Cumulative Duration of Low Bispectral Index with Intermediate-term Mortality after Cardiac Surgery in the B-Unaware Trial. Anesthesiology 112:5, 1116-1127
    CrossRef

  48. 48

    Alessandro E.P. Villa, Igor V. Tetko. (2010) Cross-frequency coupling in mesiotemporal EEG recordings of epileptic patients. Journal of Physiology-Paris 104:3-4, 197-202
    CrossRef

  49. 49

    Wakako Miyake, Yutaka Oda, Yuko Ikeda, Katsuaki Tanaka, Satoshi Hagihira, Hiroyoshi Iwaki, Akira Asada. (2010) Effect of remifentanil on cardiovascular and bispectral index responses following the induction of anesthesia with midazolam and subsequent tracheal intubation. Journal of Anesthesia 24:2, 161-167
    CrossRef

  50. 50

    Murali Chakravarthy, Srinivasa Holla, Vivek Jawali. (2010) Index of consciousness and bispectral index values are interchangeable during normotension and hypotension but not during non pulsatile flow state during cardiac surgical procedures: a prospective study. Journal of Clinical Monitoring and Computing 24:2, 83-91
    CrossRef

  51. 51

    Kay Robins, Gordon Lyons. (2010) Evidence Is Lacking for Interventions Proposed to Prevent Unintended Awareness During General Anesthesia for Cesarean Delivery. Anesthesia & Analgesia 110:3, 973-974
    CrossRef

  52. 52

    George A. Mashour, Roy K. Esaki, Kevin K. Tremper, David B. Glick, Michael OʼConnor, Michael S. Avidan. (2010) A Novel Classification Instrument for Intraoperative Awareness Events. Anesthesia & Analgesia 110:3, 813-815
    CrossRef

  53. 53

    Robert D. Sanders, Michael S. Avidan. (2010) Evidence Is Lacking for Interventions Proposed to Prevent Unintended Awareness During General Anesthesia for Cesarean Delivery. Anesthesia & Analgesia 110:3, 972-973
    CrossRef

  54. 54

    Jean Yves Martinez, Pierre François Wey, Christophe Lions, Andrei Cividjian, Muriel Rabilloud, Alvine Bissery, Lionel Bourdon, Marc Puidupin, Jacques Escarment, Luc Quintin. (2010) A Beat-by-Beat Cardiovascular Index, CARDEAN: A Prospective Randomized Assessment of Its Utility for the Reduction of Movement During Colonoscopy. Anesthesia & Analgesia 110:3, 765-772
    CrossRef

  55. 55

    Jiro Kurata. (2010) Deep Hypnosis as a Sign of “Imbalance” in Balanced Anesthesia. Anesthesia & Analgesia 110:3, 663-665
    CrossRef

  56. 56

    Christine Fedorow, Hilary P Grocott. (2010) Cerebral monitoring to optimize outcomes after cardiac surgery. Current Opinion in Anaesthesiology 23:1, 89-94
    CrossRef

  57. 57

    S. H. Kim, S. Y. Park, W. S. Chae, H. C. Jin, J. S. Lee, Y. I. Kim. (2010) Effect of desflurane at less than 1 MAC on QT interval prolongation induced by tracheal intubation. British Journal of Anaesthesia 104:2, 150-157
    CrossRef

  58. 58

    Leon Serfontein. (2010) Awareness in cardiac anesthesia. Current Opinion in Anaesthesiology 23:1, 103-108
    CrossRef

  59. 59

    Hedwig Schroeck, Karamarie Fecho, Kathleen Abode, Ann Bailey. (2010) Vocal cord function and bispectral index in pediatric bronchoscopy patients emerging from propofol anesthesia. Pediatric Pulmonologyn/a-n/a
    CrossRef

  60. 60

    R. W. D. Nickalls, R. P. Mahajan. (2010) Awareness and anaesthesia: think dose, think data. British Journal of Anaesthesia 104:1, 1-2
    CrossRef

  61. 61

    Eliseo Beldao, Karen J Blissitt, Juliet C Duncan, Francisco G Laredo, Mayte Escobar Gil de Montes, R Eddie Clutton. (2010) The bispectral index during recovery from halothane and sevoflurane anaesthesia in horses. Veterinary Anaesthesia and Analgesia 37:1, 25-34
    CrossRef

  62. 62

    David Green, Lise Paklet. (2010) Latest developments in peri-operative monitoring of the high-risk major surgery patient. International Journal of Surgery 8:2, 90-99
    CrossRef

  63. 63

    Christopher D Kent, Karen B Domino. (2009) Depth of anesthesia. Current Opinion in Anaesthesiology 22:6, 782-787
    CrossRef

  64. 64

    Michael R. Isley, Harvey L. Edmonds, Mark Stecker. (2009) Guidelines for intraoperative neuromonitoring using raw (analog or digital waveforms) and quantitative electroencephalography: a position statement by the American Society of Neurophysiological Monitoring. Journal of Clinical Monitoring and Computing 23:6, 369-390
    CrossRef

  65. 65

    T. Perl, E. Carstens, A. Hirn, M. Quintel, W. Vautz, J. Nolte, M. Junger. (2009) Determination of serum propofol concentrations by breath analysis using ion mobility spectrometry. British Journal of Anaesthesia 103:6, 822-827
    CrossRef

  66. 66

    B. Lortat-Jacob, V. Billard, W. Buschke, F. Servin. (2009) Assessing the clinical or pharmaco-economical benefit of target controlled desflurane delivery in surgical patients using the Zeus ® anaesthesia machine. Anaesthesia 64:11, 1229-1235
    CrossRef

  67. 67

    Ben Julian A Palanca, George A Mashour, Michael S Avidan. (2009) Processed electroencephalogram in depth of anesthesia monitoring. Current Opinion in Anaesthesiology 22:5, 553-559
    CrossRef

  68. 68

    George A. Mashour, Roy K. Esaki, John C. Vandervest, Amy Shanks, Sachin Kheterpal. (2009) A novel electronic algorithm for detecting potentially insufficient anesthesia: implications for the prevention of intraoperative awareness. Journal of Clinical Monitoring and Computing 23:5, 273-277
    CrossRef

  69. 69

    Andrea Kopp Lugli, Charles Spencer Yost, Christoph H Kindler. (2009) Anaesthetic mechanisms: update on the challenge of unravelling the mystery of anaesthesia. European Journal of Anaesthesiology 26:10, 807-820
    CrossRef

  70. 70

    C. Lysakowski, N. Elia, C. Czarnetzki, L. Dumont, G. Haller, C. Combescure, M. R. Tramer. (2009) Bispectral and spectral entropy indices at propofol-induced loss of consciousness in young and elderly patients. British Journal of Anaesthesia 103:3, 387-393
    CrossRef

  71. 71

    Kay Robins, Gordon Lyons. (2009) Intraoperative Awareness During General Anesthesia for Cesarean Delivery. Anesthesia & Analgesia 109:3, 886-890
    CrossRef

  72. 72

    R. Mihai, S. Scott, T. M. Cook. (2009) Litigation related to inadequate anaesthesia: an analysis of claims against the NHS in England 1995-2007. Anaesthesia 64:8, 829-835
    CrossRef

  73. 73

    DaiWai M. Olson, Suzanne M. Thoyre, Eric D. Peterson, Carmelo Graffagnino. (2009) A Randomized Evaluation of Bispectral Index-Augmented Sedation Assessment in Neurological Patients. Neurocritical Care 11:1, 20-27
    CrossRef

  74. 74

    Cambell Bennett, Logan J. Voss, John P. M. Barnard, James W. Sleigh. (2009) Practical Use of the Raw Electroencephalogram Waveform During General Anesthesia: The Art and Science. Anesthesia & Analgesia 109:2, 539-550
    CrossRef

  75. 75

    Mohammed A Qadeer, John J Vargo. (2009) BIS: Hit or Miss?. The American Journal of Gastroenterology 104:7, 1850-1850
    CrossRef

  76. 76

    Andrew B. Leibowitz, Marc A. Rozner. (2009) PRO: Manometry Should Routinely Be Used During Central Venous Catheterization. Anesthesia & Analgesia 109:1, 3-5
    CrossRef

  77. 77

    Goneppanavar Umesh, Kaur Jasvinder, Nanda Shetty. (2009) Low minimum alveolar concentration alarm: a standard for prevention of awareness during general anaesthesia maintained by inhalational anaesthetics. Journal of Clinical Monitoring and Computing 23:3, 185-186
    CrossRef

  78. 78

    Mark Quigg, Derek Leiner. (2009) Engineering Aspects of the Quantified Amplitude-Integrated Electroencephalogram in Neonatal Cerebral Monitoring. Journal of Clinical Neurophysiology 26:3, 145-149
    CrossRef

  79. 79

    JERROLD LERMAN, MARTIN JÖHR. (2009) Inhalational anesthesia vs total intravenous anesthesia (TIVA) for pediatric anesthesia. Pediatric Anesthesia 19:5, 521-534
    CrossRef

  80. 80

    Alejandro E. Delfino, Guillermo Lema. (2009) Two or Three Interviews?. Anesthesiology 110:4, 950
    CrossRef

  81. 81

    Andrew J. Davidson. (2009) Two or Three Interviews?. Anesthesiology 110:4, 950-951
    CrossRef

  82. 82

    Rohit Juneja, Matthew Hacking, Tim Wigmore. (2009) Peri-operative management of the oesphagectomy patient. Current Anaesthesia & Critical Care 20:1, 13-17
    CrossRef

  83. 83

    Mohamed M. Ghoneim, Robert I. Block, Mary Haffarnan, Maya J. Mathews. (2009) Awareness During Anesthesia: Risk Factors, Causes and Sequelae: A Review of Reported Cases in the Literature. Anesthesia & Analgesia 108:2, 527-535
    CrossRef

  84. 84

    Martin Savard, Eyad Al Thenayan, Loretta Norton, Michael D. Sharpe, Bryan Young. (2009) Continuous EEG Monitoring in Severe Guillain-Barré Syndrome Patients. Journal of Clinical Neurophysiology 26:1, 21-23
    CrossRef

  85. 85

    Peter S. Sebel. (2009) To Sleep, Perchance To Dream? NOT!. Anesthesia & Analgesia 108:2, 391-392
    CrossRef

  86. 86

    William Heegaard, Ryan Charles Fringer, R. J. Frascone, Greg Pippert, James Miner. (2009) Bispectral Index Monitoring in Helicopter Emergency Medical Services Patients. Prehospital Emergency Care 13:2, 193-197
    CrossRef

  87. 87

    Dragan Pavlovic. (2009) Two problems of empirical biomedical science: example of awareness monitoring with bispectral analysis. European Journal of Anaesthesiology 26:1, 87-89
    CrossRef

  88. 88

    K. Y. Yoo, C. W. Jeong, S. J. Kim, S. T. Jeong, M. H. Shin, J. Lee. (2009) Cardiovascular and arousal responses to laryngoscopy and tracheal intubation in patients with complete spinal cord injury. British Journal of Anaesthesia 102:1, 69-75
    CrossRef

  89. 89

    George A. Mashour, Eric LaRock. (2008) Inverse zombies, anesthesia awareness, and the hard problem of unconsciousness. Consciousness and Cognition 17:4, 1163-1168
    CrossRef

  90. 90

    Brett L. Foster, Ingo Bojak, David T. J. Liley. (2008) Population based models of cortical drug response: insights from anaesthesia. Cognitive Neurodynamics 2:4, 283-296
    CrossRef

  91. 91

    Yasuhiro Morimoto. (2008) Usefulness of electroencephalogramic monitoring during general anesthesia. Journal of Anesthesia 22:4, 498-501
    CrossRef

  92. 92

    Sebastian Rehberg, Thomas P. Weber, Hugo Van Aken, Marc Theisen, Christian Ertmer, Katrin Bröking, Tobias Schulte, Nani Osada, Dominique Asemann, Viola Bullmann. (2008) Sleep Disturbances after Posterior Scoliosis Surgery with an Intraoperative Wake-up Test Using Remifentanil. Anesthesiology 109:4, 629-641
    CrossRef

  93. 93

    Wolfram Miekisch, Patricia Fuchs, Svend Kamysek, Christine Neumann, Jochen K. Schubert. (2008) Assessment of propofol concentrations in human breath and blood by means of HS-SPME–GC–MS. Clinica Chimica Acta 395:1-2, 32-37
    CrossRef

  94. 94

    S. C. Høymork. (2008) Antinociceptive monitors: tools or fools?. Acta Anaesthesiologica Scandinavica 52:8, 1035-1037
    CrossRef

  95. 95

    J. Ponte, G. Lynch, I. Grant, A. Morley, J. R. Sneyd. (2008) Memory and awareness during anaesthesia. British Journal of Anaesthesia 101:5, 738-740
    CrossRef

  96. 96

    (2008) Anesthesia Awareness and the Bispectral Index. New England Journal of Medicine 359:4, 427-431
    Full Text

  97. 97

    Orser, Beverley A., . (2008) Depth-of-Anesthesia Monitor and the Frequency of Intraoperative Awareness. New England Journal of Medicine 358:11, 1189-1191
    Full Text

  98. 98

    O. Simanski, M. Janda, A. Schubert, J. Bajorat, R. Hofmockel, B. Lampe. (2008) Progress of automatic drug delivery in anaesthesia-The ‘Rostock assistant system for anaesthesia control (RAN)’. International Journal of Adaptive Control and Signal Processingn/a-n/a
    CrossRef

  99. 99

    Yodying Punjasawadwong, Aram Phongchiewboon, Nutchanart Bunchungmongkol, Yodying Punjasawadwong. 2007. Bispectral index for improving anaesthetic delivery and postoperative recovery. .
    CrossRef

Letters