Quality of Life with Defibrillator Therapy or Amiodarone in Heart Failure

Daniel B. Mark, M.D., M.P.H., Kevin J. Anstrom, Ph.D., Jie L. Sun, M.S., Nancy E. Clapp-Channing, R.N., M.P.H., Anastasios A. Tsiatis, Ph.D., Linda Davidson-Ray, M.A., Kerry L. Lee, Ph.D., and Gust H. Bardy, M.D. for the Sudden Cardiac Death in Heart Failure Trial Investigators

N Engl J Med 2008; 359:999-1008September 4, 2008

Abstract

Background

Implantable cardioverter–defibrillator (ICD) therapy significantly prolongs life in patients at increased risk for sudden death from depressed left ventricular function. However, whether this increased longevity is accompanied by deterioration in the quality of life is unclear.

Methods

In a randomized trial, we compared ICD therapy or amiodarone with state-of-the-art medical therapy alone in 2521 patients who had stable heart failure with depressed left ventricular function. We prospectively measured quality of life at baseline and at months 3, 12, and 30; data collection was 93 to 98% complete. The Duke Activity Status Index (which measures cardiac physical functioning) and the Medical Outcomes Study 36-Item Short-Form Mental Health Inventory 5 (which measures psychological well-being) were prespecified primary outcomes. Multiple additional quality-of-life outcomes were also examined.

Results

Psychological well-being in the ICD group, as compared with medical therapy alone, was significantly improved at 3 months (P=0.01) and at 12 months (P=0.003) but not at 30 months. No clinically or statistically significant differences in physical functioning among the study groups were observed. Additional quality-of-life measures were improved in the ICD group at 3 months, 12 months, or both, but there was no significant difference at 30 months. ICD shocks in the month preceding a scheduled assessment were associated with a decreased quality of life in multiple domains. The use of amiodarone had no significant effects on the primary quality-of-life outcomes.

Conclusions

In a large primary-prevention population with moderately symptomatic heart failure, single-lead ICD therapy was not associated with any detectable adverse quality-of-life effects during 30 months of follow-up.

Media in This Article

Figure 1Enrollment and Completeness of Quality-of-Life (QOL) Data Collection.
Figure 2Scores on the DASI and MHI-5 in the Three Study Groups.
Article

Implantable cardioverter–defibrillators (ICDs) significantly extend survival among patients who are at high risk for sudden death because of the severity of their underlying heart disease.1,2 However, concern has emerged about the effects of ICD therapy on the quality of life. One concern is that the use of ICD therapy could trade a quick, relatively painless (albeit premature) death for a more unpleasant death from progressive deterioration of the underlying heart disease or a coexisting illness.3 Furthermore, in some previous studies, receipt of multiple ICD shocks has been associated with a reduced quality of life, although the causality of this relationship is unclear.

To date, two secondary-prevention trials (in which an ICD was implanted after a life-threatening arrhythmia to prevent future events) and one primary-prevention trial (in which an ICD was implanted in patients who were at increased risk for life-threatening arrhythmia but had not had such an event) have reported quality-of-life outcomes.4-6 Although data from these trials have not shown any consistent evidence of a worse quality of life with ICD therapy, the conclusions that could be derived from these studies were limited by methodologic problems and relatively short follow-up.

The issue of long-term quality of life is particularly important in primary prevention, in which the willingness to accept a potentially unpleasant therapy for an uncertain future benefit may be low. We therefore examined the effects of primary-prevention ICD therapy on health-related quality of life in the Sudden Cardiac Death in Heart Failure Trial (SCD-HeFT) (ClinicalTrials.gov number, NCT00000609).

Although amiodarone did not improve survival in our trial, findings from earlier studies suggested that it might improve functional status.7,8 Thus, we also report on the quality-of-life outcomes from the comparison between the use of amiodarone and placebo in this trial.

Methods

Patient Population and Study Overview

Between September 16, 1997, and July 18, 2001, SCD-HeFT enrolled 2521 patients who were at least 18 years of age and who had New York Heart Association chronic, stable class II or III congestive heart failure and a left ventricular ejection fraction of 35% or less. The study design, methods, and primary results of SCD-HeFT have been reported previously.1 The cause of heart failure was ischemic in 52% of the patients and nonischemic in 48%. Ninety percent of the study patients were enrolled in the United States, with the remainder in Canada and New Zealand.

Patients were randomly assigned to receive state-of-the-art medical therapy plus amiodarone (Cordarone, Wyeth–Ayerst Pharmaceuticals), an amiodarone placebo, or a conservatively programmed, single-chamber ICD (Medtronic, model 7223). Therapy was initiated on an outpatient basis according to the protocol. After a median follow-up of 45.5 months (range, 24 to 72), ICD therapy was associated with a 23% reduction in mortality, as compared with medical therapy, whereas amiodarone therapy was not associated with a significant effect on mortality.

Quality-of-Life Study

The quality-of-life portion of the trial was funded by the National Heart, Lung, and Blood Institute. The authors designed the study, collected and analyzed the data, and prepared the manuscript. The corporate sponsors who contributed to the support of the trial (Wyeth–Ayerst Laboratories, Knoll Pharmaceuticals, and Medtronic) had no role in the design, analysis, or interpretation of the quality-of-life study.

All patients provided written informed consent, and the study was conducted in cooperation with the National Heart, Lung, and Blood Institute. The institutional review board or ethics committee at each site approved the study protocol.

Data-Collection Methods

Quality of life was measured by means of structured interviews at baseline, at months 3 and 12, and at 30 months or at the end of the study follow-up. At each site, the study coordinator conducted interviews at the time of a scheduled clinic visit or by telephone if a clinic visit was missed. Specific training was provided to each site coordinator to ensure standardization of data collection. According to the protocol, baseline quality-of-life assessments were to be conducted after informed consent had been obtained and before randomization. Follow-up quality-of-life assessments were to be performed within 1 month before or after the scheduled contact. For patients who were too ill to complete the full questionnaire, had a language barrier, or were otherwise unable to participate in the full interview, a short proxy form was collected.

Quality-of-Life Measures

Two measures were prespecified as primary outcomes for the quality-of-life portion of the trial: the Duke Activity Status Index (DASI), reflecting cardiac-specific physical functioning, and the Medical Outcomes Study 36-Item Short-Form (SF-36) Mental Health Inventory 5 (MHI-5), reflecting psychological well-being. DASI was constructed to be a questionnaire-based analogue of the maximal exercise stress test used for patients with cardiac disease; it is scored from 0 to 58, with higher scores indicating better function and a difference of 4 points or more considered to be clinically significant.9 The MHI-5, used to assess psychological well-being, is scored from 0 to 100, with higher scores indicating better function.10 A clinically significant difference in the MHI-5 has not been formally defined but can be approximated as one quarter of 1 SD (5 points in this study).

Other scales from the SF-36 were used to assess role functioning (both physical and emotional limitations), general health perceptions, bodily pain, social functioning, and vitality. Like the MHI-5, these scales are scored from 0 to 100, with higher scores indicating better function and with one quarter of 1 SD representing a reasonable indication of a clinically significant difference.

The quality-of-life interviews also collected information on total numbers of “bed days,” which were defined as the number of days out of the last 42 days in which a patient was at home in bed for all or most of the day for health reasons, and “disability days,” which were defined as the number of days out of the last 42 days (not counting “bed days”) in which the patient had to cut down on usual activities for health reasons. Additional questions were asked to determine whether a patient could currently drive a car (yes or no) and manage money independently (yes or no). Employment details were obtained with the use of an abbreviated series of questions adapted from the Bypass Angioplasty Revascularization Investigation Substudy on Economics and Quality of Life.11

We assessed the effect of heart failure on the quality of life, using the Minnesota Living with Heart Failure scale.12 This scale is scored from 0 to 105, with higher scores indicating worse function, and a clinically significant difference considered to be approximately 5 points.13

The relative desirability (utility) of each patient's health status, on a scale from 0 (a state of health equivalent to being dead) to 1 (excellent health), was assessed with the time–trade-off technique.14 Patients were asked to assume that they would have a life expectancy of 5 years in their current state of health and then were asked in a series of questions to decide how many of those 5 years they would be willing to trade for excellent health in the remaining time. As a second, more intuitive global measure, patients were also asked to rate their health on a scale of 0 to 100, with 100 indicating excellent health and 0 a state of health equivalent to being dead. A 5-point difference in this scale (one quarter of 1 SD) approximated clinical significance.

Statistical Analysis

We used means and standard deviations, medians and 25th to 75th percentiles, or both to describe the distributions of continuous variables. Percentages were used to describe categorical variables. Univariate comparisons were performed with the use of Pearson's chi-square test for categorical variables and the Wilcoxon rank-sum test for continuous variables. Each active treatment was compared pairwise with the placebo group.

Patients with an ICD who received a shock from the device within the month preceding a scheduled quality-of-life assessment were compared with patients in the ICD group who did not receive a shock during the same period. These comparisons were based on the Wilcoxon rank-sum test for changes in the scores from the most recent quality-of-life measurements before the shock occurred. This analysis was repeated with the use of 2-month and 12-month time frames.

The primary outcome of the trial — death from any cause — differed significantly between the ICD group and the placebo group. This mortality difference resulted in a nonrandom subgroup of survivors whose quality of life could be compared according to the assigned treatments. To account for this potential bias, we applied an estimator for the survival average causal effect as a sensitivity analysis.15,16 These estimates are based on weighted averages of the observed quality-of-life data multiplied by survival estimates specific to the study group, with P values and 95% confidence intervals for the estimated survival average causal effect based on a nonparametric bootstrap procedure.17

All reported P values are two-sided. No adjustments were made for multiple testing.

Results

Patients and Baseline Characteristics

Of 2521 patients who underwent randomization, 2479 (98%) completed quality-of-life questionnaires at baseline (Figure 1Figure 1Enrollment and Completeness of Quality-of-Life (QOL) Data Collection.). The initial demographic and clinical characteristics of the patients were well balanced among the study groups (Table 1Table 1Baseline Characteristics of the Patients.).1 At each follow-up interval, questionnaires were collected from 93 to 95% of eligible patients. Overall, from a total of 9171 expected contacts with patients, 8747 quality-of-life questionnaires (95%) were collected. Only 1.2% of patients declined to complete the questionnaires, and only 1.4% of forms were judged to be incomplete. In 69 of 6268 follow-up interviews (1.1%), proxy forms were substituted for the full questionnaire.

Quality-of-Life Outcomes

In unadjusted comparisons, cardiac-specific physical functioning on DASI did not differ significantly between the ICD group and the placebo group at baseline (median score of 23 in both groups, P=0.76) and at months 3, 12, and 30 (median score of 24 in both groups for all three intervals, P>0.10) (Table 2Table 2Primary Quality-of-Life Measures. and Figure 2AFigure 2Scores on the DASI and MHI-5 in the Three Study Groups.). There were also no significant differences at any point between the amiodarone group and the placebo group on cardiac-specific physical functioning.

Psychological well-being did not differ significantly between the ICD group and the placebo group at baseline on MHI-5 (median scores of 76 in both groups, P=0.17) but was better in the ICD group than in the placebo group at 3 months (median scores, 80 and 76, respectively; P=0.01) and at 12 months (median scores, 80 and 76, respectively; P=0.003). At 30 months, there was no significant difference between the two study groups (median score of 76 in both groups, P=0.79) (Table 2 and Figure 2B). In the comparison between the amiodarone and placebo groups, we did not observe significant differences at any point during follow-up on the basis of MHI-5.

For each of the six other SF-36 scales, at least one interval comparison (i.e., at 3 months, 12 months, or both) showed significantly better scores in the ICD group. However, values were clinically similar and did not differ significantly at baseline or at 30 months on any of these scales (Table 1 of the Supplementary Appendix, available with the full text of this article at www.nejm.org). As compared with patients in the placebo group, those in the amiodarone group had significantly higher scores on the SF-36 pain index at all four time points.

At baseline, patients reported a mean of approximately 2 bed days and 8 to 9 disability days during the preceding 42 days. In addition, 86% of patients were able to drive a car, 92% could manage their finances independently, and 27% were employed outside the home. We were not able to detect an effect of ICD therapy as compared with placebo on the number of bed days or disability days or on the proportion of patients who were able to drive a car, manage their finances, or maintain employment during the follow-up period.

The Minnesota Living with Heart Failure scores were similar at baseline in the ICD group and the placebo group (median scores, 41 and 43, respectively; P=0.77). Scores were generally better (i.e., lower) in the ICD group than in the placebo group at 3 months (median scores, 30 and 36, respectively; P=0.006), at 12 months (median scores, 32 and 36, respectively; P=0.07), and at 30 months (median scores, 32 and 36, respectively; P=0.05).

The time–trade-off utility measure averaged 0.80 at baseline in all three study groups; there was a significant improvement in the ICD group over the placebo group at 3 months but not at any of the other time points. On a scale of 0 (worst) to 100 (best), the patients in the ICD group rated their overall health more highly than did those in the placebo group at 3 months (median scores, 75 and 70, respectively; P=0.002) and at 12 months (median scores, 75 and 70, respectively; P=0.05), but there was no significant difference at 30 months (median score of 70 in both groups, P=0.18).

Effect of ICD Shocks on Quality of Life

In the ICD group, 49 patients received a shock within 1 month before a scheduled quality-of-life assessment. As compared with patients in the ICD group who did not receive a shock, the quality of life of patients in the month after a shock was characterized by a significant decrease in perceived general health, physical and emotional functioning, social functioning, and self-rated health (all comparisons unadjusted) (Figure 3Figure 3Effect of ICD Shocks on Patients' Quality of Life.).

For the 66 patients who had received a shock within 2 months before a scheduled quality-of-life assessment, the pattern was the same but with smaller differences. When we compared the quality of life of 100 surviving patients who had received an ICD shock at any time during the first year of the study with that of 638 patients who did not receive a shock, no significant differences were evident. In addition, the number of ICD discharges above an arbitrary number, ranging from 2 to 5 or more, did not have a significant effect on the subsequent quality of life.

Survival-Adjusted Analyses

To account for the improved survival of the patients in the ICD group, we estimated the survival average causal effect for each quality-of-life variable. Overall, the results were not materially different from the unadjusted comparisons described above (Table 3Table 3Differences between Study Groups in Quality-of-Life Measures.).

Discussion

In our trial, single-lead ICD therapy enhanced survival and did not detectably diminish health-related quality of life for up to 30 months in patients with stable but moderately symptomatic heart failure. Although ICD therapy for primary prevention of sudden death was not expected to improve the quality of life, the possibility of harm from either psychological or physical complications of the therapy was of concern. In our overall comparisons, we found no statistically or clinically significant evidence of either psychological or physical harm. In a double-blind comparison of amiodarone with placebo, we did not detect an effect of amiodarone on either of the two primary quality-of-life measures.

Among patients in the ICD group who had received an ICD shock within the month preceding a scheduled follow-up visit, as compared with patients who had not received a shock, the quality of life was diminished in multiple domains. An analysis that included patients who had received shocks within 2 months before a quality-of-life assessment showed the same trends but with reductions in the magnitude and statistical significance of the differences. These trends were eliminated altogether when the window between the ICD shock and subsequent assessment was extended to 1 year. Although it may be plausible to assume that this association was causal, our analyses did not have sufficient statistical power to examine the relative contributions of the ICD shocks and concomitant deterioration in clinical status to these observations.

Among other primary-prevention trials of ICD therapy, only the Coronary Artery Bypass Graft (CABG)-Patch trial has reported on quality-of-life outcomes.18 At 6 months, the ICD group had significantly lower levels of psychological well-being than the control group. Furthermore, patients who had received at least one ICD shock had a reduced quality of life on several measures. Important differences between the CABG-Patch quality-of-life study and our study include the ICD technology (a large, bulky ICD in the CABG-Patch study vs. a small, low-profile ICD in our study), the method of ICD implantation (open-chest implantation with an abdominal pocket vs. outpatient transvenous implantation with a pectoral pocket), and the target population (patients referred for CABG who had an ejection fraction of ≤35% vs. those with stable heart failure and an ejection fraction of ≤35%, about 50% of whom had a nonischemic cause of heart failure).

Another large study of quality of life in patients who had undergone ICD therapy was the Antiarrhythmics Versus Implantable Defibrillators (AVID) trial, a secondary-prevention study that was stopped prematurely by the data and safety monitoring board because of the improved survival in the ICD group.4,19 In that study, the ICD group and the antiarrhythmic-therapy group had similar changes in scores on both the physical and mental components of the SF-36 questionnaire. Among patients in the ICD group with complete data on ICD shocks and follow-up quality-of-life data, the occurrence of one or more ICD shocks was significantly associated with subsequent reductions in both physical functioning and psychological well-being. Major differences between the AVID trial and our study include differences in the study populations (patients with life-threatening arrhythmias in the AVID trial vs. those with stable heart failure in our study) and a significantly lower rate of quality-of-life data collection in the AVID trial (83% at baseline and 61% at 1 year). Nonetheless, a reasonable conclusion from the AVID analysis is that in the absence of administered shock therapy, ICDs were well tolerated and did not diminish the quality of life.

The Canadian Implantable Defibrillator Study (CIDS), another secondary-prevention ICD trial, obtained quality-of-life data at 6 and 12 months of follow-up.5 Scores for emotional and physical health improved in the ICD group, as compared with the amiodarone group. Patients in the ICD group who received five or more shocks did not have improvement in these quality-of-life scales, as compared with patients who received fewer than five shocks. However, the quality-of-life scores for patients in the ICD group who received one to four shocks during follow-up did not differ significantly from those for patients who received no shocks.

All studies of the effects of ICD therapy on quality of life, including our study, are limited by an inability to mask the therapy. Thus, the perceived effects of ICDs that we show may reflect attitudes of the study doctors and nurses that were transmitted to the patients, as well as the beliefs and expectations of the patients themselves. Patients may view the ICD either as an electronic security blanket or as an unpredictable and uncontrollable source of physical and emotional discomfort. In this unblinded comparison, we observed small improvements in some domains of quality of life during the first year of follow-up in the ICD group. We have no direct means of determining whether these improvements reflect the effects of such biases.

Caution should be exercised in interpreting significant differences in quality-of-life measures among the study groups, given the numerous statistical tests performed in this study. P values that are shown were uncorrected for multiple comparisons. Since the quality of life was a secondary outcome, the study was not constructed to test formally for the noninferiority of ICD therapy with respect to these outcomes.

Our evaluation of the effects of ICD shocks on subsequent quality of life was limited by the lack of quality-of-life data linked to the delivery of ICD shocks (collection of such data was judged to be logistically infeasible) and by the relatively small number of patients with a quality-of-life assessment shortly after a shock episode. We did not have enough patients with multiple shock episodes during a 24-hour period (“ICD storm”) to determine the effects of that phenomenon on subsequent quality of life.

In conclusion, we evaluated the quality of life of patients with moderately symptomatic, stable heart failure. Random assignment to the ICD group was not associated with adverse effects on health-related quality of life during the first 30 months of follow-up.

Supported by grants (UO1-HL55496, UO1-HL55766, and UO1-HL55297) from the National Heart, Lung, and Blood Institute, and by Medtronic, Wyeth–Ayerst Laboratories, and Knoll Pharmaceuticals.

Dr. Mark reports receiving grant support and lecture fees from Medtronic; Dr. Anstrom, grant support from Medtronic Vascular; and Dr. Lee, grant support and consulting fees from Medtronic; Dr. Bardy reports receiving grant support from and having intellectual property rights with Medtronic, receiving consulting fees and grant support from Philips Medical, receiving grant support from Laerdal, and being a founder of and board member of and having equity and intellectual property rights with Cameron Health. No other potential conflict of interest relevant to this article was reported.

We thank Melanie R. Daniels for her editorial support, Judy Stafford for her assistance in data analysis, the site coordinators who collected the data for the study, and the patients who volunteered to participate in the study.

Source Information

From the Outcomes Research Group (D.B.M., K.J.A., N.E.C.-C., L.D.-R.), Duke Clinical Research Institute (J.L.S., K.L.L.), and the Departments of Medicine (D.B.M.) and Biostatistics and Bioinformatics (K.J.A., K.L.L.), Duke University Medical Center, Durham, NC; the Department of Statistics, North Carolina State University, Raleigh (A.A.T.); and the Seattle Institute for Cardiac Research, University of Washington, Seattle (G.H.B.).

References

References

  1. 1

    Bardy GH, Lee KL, Mark DB, et al. Amiodarone or an implantable cardioverter-defibrillator for congestive heart failure. N Engl J Med 2005;352:225-237[Erratum, N Engl J Med 2005;352:2146.]
    Full Text | Web of Science | Medline

  2. 2

    Moss AJ, Hall WJ, Cannom DS, et al. Improved survival with an implanted defibrillator in patients with coronary disease at high risk for ventricular arrhythmia. N Engl J Med 1996;335:1933-1940
    Full Text | Web of Science | Medline

  3. 3

    Kolata G. Extending life, defibrillators can prolong misery. New York Times. March 25, 2002.

  4. 4

    Schron EB, Exner DV, Yao Q, et al. Quality of life in the antiarrhythmics versus implantable defibrillators trial: impact of therapy and influence of adverse symptoms and defibrillator shocks. Circulation 2002;105:589-594
    CrossRef | Web of Science | Medline

  5. 5

    Irvine J, Dorian P, Baker B, et al. Quality of life in the Canadian Implantable Defibrillator Study (CIDS). Am Heart J 2002;144:282-289
    Web of Science | Medline

  6. 6

    Bigger JT Jr. Prophylactic use of implanted cardiac defibrillators in patients at high risk for ventricular arrhythmias after coronary-artery bypass graft surgery. N Engl J Med 1997;337:1569-1575
    Full Text | Web of Science | Medline

  7. 7

    Doval HC, Nul DR, Grancelli HO, Perrone SV, Bortman GR, Curiel R. Randomised trial of low-dose amiodarone in severe congestive heart failure. Lancet 1994;344:493-498
    CrossRef | Web of Science | Medline

  8. 8

    Singh SN, Fletcher RD, Fisher SG, et al. Amiodarone in patients with congestive heart failure and asymptomatic ventricular arrhythmia. N Engl J Med 1995;333:77-82
    Full Text | Web of Science | Medline

  9. 9

    Hlatky MA, Boineau RE, Higginbotham MB, et al. A brief self-administered questionnaire to determine functional capacity (the Duke Activity Status Index). Am J Cardiol 1989;64:651-654
    CrossRef | Web of Science | Medline

  10. 10

    Ware JE Jr, Snow KK, Kosinski M, Gandek B. SF-36 health survey: manual and interpretation guide. Boston: Health Institute, New England Medical Center, 1993.

  11. 11

    Hlatky MA, Rogers WJ, Johnstone I, et al. Medical care costs and quality of life after randomization to coronary angioplasty or coronary bypass surgery. N Engl J Med 1997;336:92-99
    Full Text | Web of Science | Medline

  12. 12

    Rector TS, Kubo SH, Cohn JN. Patients' self-assessment of their congestive heart failure. 2. Content, reliability and validity of a new measure, the Minnesota Living with Heart Failure Questionnaire. Heart Fail 1987;3:198-209

  13. 13

    Rector TS. Overview of the Minnesota Living with Heart Failure Questionnaire. 2005. (Accessed July 18, 2008, at http://www.mlhfq.org/_dnld/mlhfq_overview.pdf.)

  14. 14

    Torrance GW, Feeny D. Utilities and quality-adjusted life years. Int J Technol Assess Health Care 1989;5:559-575
    CrossRef | Medline

  15. 15

    Hayden D, Pauler DK, Schoenfeld D. An estimator for treatment comparisons among survivors in randomized trials. Biometrics 2005;61:305-310
    CrossRef | Web of Science | Medline

  16. 16

    Dawid AP. Causal inference without counterfactuals. J Am Stat Assoc 2000;95:407-448
    CrossRef | Web of Science

  17. 17

    Efron B, Tibshirani RJ. An introduction to the bootstrap. New York: Chapman & Hall, 1993.

  18. 18

    Namerow PB, Firth BR, Heywood GM, Windle JR, Parides MK. Quality-of-life six months after CABG surgery in patients randomized to ICD versus no ICD therapy: findings from the CABG Patch Trial. Pacing Clin Electrophysiol 1999;22:1305-1313
    CrossRef | Web of Science | Medline

  19. 19

    The Antiarrhythmics versus Implantable Defibrillators (AVID) Investigators. A comparison of antiarrhythmic-drug therapy with implantable defibrillators in patients resuscitated from near-fatal ventricular arrhythmias. N Engl J Med 1997;337:1576-1583
    Full Text | Web of Science | Medline

Citing Articles (54)

Citing Articles

  1. 1

    THOMAS KLEEMANN, MATTHIAS HOCHADEL, MARGIT STRAUSS, ALEXANDROS SKARLOS, KARLHEINZ SEIDL, RALF ZAHN. (2012) Comparison Between Atrial Fibrillation-Triggered Implantable Cardioverter-Defibrillator (ICD) Shocks and Inappropriate Shocks Caused by Lead Failure:. Journal of Cardiovascular Electrophysiologyno-no
    CrossRef

  2. 2

    Sony Jacob, Sidakpal S. Panaich, Sandip K. Zalawadiya, George McKelvey, George Abraham, Rajeev Aravindhakshan, Samuel F. Sears, Jamie B. Conti, H. Michael Marsh. (2011) Phantom shocks unmasked: clinical data and proposed mechanism of memory reactivation of past traumatic shocks in patients with implantable cardioverter defibrillators. Journal of Interventional Cardiac Electrophysiology
    CrossRef

  3. 3

    DAN D. MATLOCK, CAROLYN T. NOWELS, FREDERICK A. MASOUDI, WILLIAM H. SAUER, DAVID B. BEKELMAN, DEBORAH S. MAIN, JEAN S. KUTNER. (2011) Patient and Cardiologist Perceptions on Decision Making for Implantable Cardioverter-Defibrillators: A Qualitative Study. Pacing and Clinical Electrophysiology 34:12, 1634-1644
    CrossRef

  4. 4

    Girish M. Nair, Jeffrey S. Healey, Syamkumar M. Divakaramenon, Stuart J. Connolly, Carlos A. Morillo. (2011) Antiarrhythmic Drug Therapy for New-Onset Ventricular Arrhythmia (VT/VF) in ICD Patients. Cardiac Electrophysiology Clinics 3:4, 651-661
    CrossRef

  5. 5

    Susanne S. Pedersen, Madelein T. Hoogwegt, Luc Jordaens, Dominic A.M.J. Theuns. (2011) Pre implantation psychological functioning preserved in majority of implantable cardioverter defibrillator patients 12months post implantation. International Journal of Cardiology
    CrossRef

  6. 6

    A.-E. Baruteau, S. Fouchard, A. Behaghel, P. Mabo, E. Villain, J.-B. Thambo, F. Marcon, V. Gournay, F. Rouault, A. Chantepie, S. Guillaumont, F. Godart, C. Bonnet, A. Fraisse, J.-M. Schleich, J.-R. Lusson, Y. Dulac, C. Leclercq, J.-C. Daubert, J.-J. Schott, H. Le Marec, V. Probst. (2011) Characteristics and long-term outcome of non-immune isolated atrioventricular block diagnosed in utero or early childhood: a multicentre study. European Heart Journal
    CrossRef

  7. 7

    John A. Schoenhard, Paul C. Zei. (2011) Reducing ICD Shocks for Ventricular Arrhythmias. Cardiac Electrophysiology Clinics 3:3, 493-502
    CrossRef

  8. 8

    Domingo Palacios-Ceña, Marta E Losa, César Fernández-de-las-Peñas, Paloma Salvadores-Fuentes. (2011) Living with life insurance: a qualitative analysis of the experience of male implantable defibrillator recipients in Spain. Journal of Clinical Nursing 20:13-14, 2003-2013
    CrossRef

  9. 9

    V. Probst, D. Plassard-Kerdoncuf, J. Mansourati, P. Mabo, F. Sacher, C. Fruchet, D. Babuty, G. Lande, B. Guyomarc'h, H. Le Marec. (2011) The psychological impact of implantable cardioverter defibrillator implantation on Brugada syndrome patients. Europace 13:7, 1034-1039
    CrossRef

  10. 10

    SEK Y. CHAIR, CHUI K. LEE, KAI C. CHOI, SAMUEL F. SEARS. (2011) Quality of Life Outcomes in Chinese Patients with Implantable Cardioverter Defibrillators. Pacing and Clinical Electrophysiology 34:7, 858-867
    CrossRef

  11. 11

    JENNIFER D. COHEN, J. PAUL MOUNSEY. (2011) “Virtual Research” May Point the Way Towards Future Device Programming Trials. Journal of Cardiovascular Electrophysiology 22:3, 290-292
    CrossRef

  12. 12

    Eldrin F. Lewis. (2011) End of Life Care in Advanced Heart Failure. Current Treatment Options in Cardiovascular Medicine 13:1, 79-89
    CrossRef

  13. 13

    ARIEH KEREN, SAMUEL F. SEARS, PABLO NERY, JORDAN SHAW, MARTIN S. GREEN, ROBERT LEMERY, MICHAEL H. GOLLOB, BARBARA AMYOTTE, DAVID H. BIRNIE. (2011) Psychological Adjustment in ICD Patients Living With Advisory Fidelis Leads. Journal of Cardiovascular Electrophysiology 22:1, 57-63
    CrossRef

  14. 14

    SUSANNE S. PEDERSEN, KRISTA C. VAN DEN BROEK, MARTHA VAN DEN BERG, DOMINIC A. M. J. THEUNS. (2010) Shock as a Determinant of Poor Patient-Centered Outcomes in Implantable Cardioverter Defibrillator Patients: Is There More to It Than Meets the Eye?. Pacing and Clinical Electrophysiology 33:12, 1430-1436
    CrossRef

  15. 15

    I. Battipaglia, L. Barone, L. Mariani, F. Infusino, R. Remoli, G. Careri, G. Pinnacchio, P. Tarzia, G. A. Lanza, F. Crea. (2010) Relationship between cardiac autonomic function and sustained ventricular tachyarrhythmias in patients with an implantable cardioverter defibrillators. Europace 12:12, 1725-1731
    CrossRef

  16. 16

    SAMUEL F. SEARS, KARI KIRIAN. (2010) Shock and Patient-Centered Outcomes Research: Is an ICD Shock Still a Critical Event?. Pacing and Clinical Electrophysiology 33:12, 1437-1441
    CrossRef

  17. 17

    F. Braunschweig, G. Boriani, A. Bauer, R. Hatala, C. Herrmann-Lingen, J. Kautzner, S. S. Pedersen, S. Pehrson, R. Ricci, M. J. Schalij. (2010) Management of patients receiving implantable cardiac defibrillator shocks: Recommendations for acute and long-term patient management. Europace 12:12, 1673-1690
    CrossRef

  18. 18

    Kasim Oguz Coskun, Aron Frederik Popov, Jan Dieter Schmitto, Sinan Tolga Coskun, Ivo Brandes, Dieter Zenker, Ivan Melnychenko, Friedrich Albert Schoendube, Wolfgang Ruschewski. (2010) Feasibility of Implantable Cardioverter Defibrillator Treatment in Five Patients With Familial Friedreich's Ataxia-A Case Series. Artificial Organs 34:11, 1061-1065
    CrossRef

  19. 19

    J. Seegers, M. Zabel, L. Luthje, D. Vollmann. (2010) Ventricular oversensing due to manufacturer-related differences in implantable cardioverter-defibrillator signal processing and sensing lead properties. Europace 12:10, 1460-1466
    CrossRef

  20. 20

    Brian Feingold, Gaurav Arora, Steven A. Webber, Kenneth J. Smith. (2010) Cost-Effectiveness of Implantable Cardioverter-Defibrillators in Children With Dilated Cardiomyopathy. Journal of Cardiac Failure 16:9, 734-741
    CrossRef

  21. 21

    Judith Z. Goldfinger, Eric D. Adler. (2010) End-of-Life Options for Patients with Advanced Heart Failure. Current Heart Failure Reports 7:3, 140-147
    CrossRef

  22. 22

    LAUREN D. VAZQUEZ, JAMIE B. CONTI, SAMUEL F. SEARS. (2010) Female-Specific Education, Management, and Lifestyle Enhancement for Implantable Cardioverter Defibrillator Patients: The FEMALE-ICD Study. Pacing and Clinical Electrophysiology 33:9, 1131-1140
    CrossRef

  23. 23

    Gillian D. Sanders, Melissa H. Kong, Sana M. Al-Khatib, Eric D. Peterson. (2010) Cost-effectiveness of implantable cardioverter defibrillators in patients ≥65 years of age. American Heart Journal 160:1, 122-131
    CrossRef

  24. 24

    S. Ahmed, A. V. Ranchor, H. J. G. M. Crijns, D. J. Van Veldhuisen, I. C. Van Gelder, . (2010) Effect of continuous versus episodic amiodarone treatment on quality of life in persistent atrial fibrillation. Europace 12:6, 785-791
    CrossRef

  25. 25

    Juliana Rigol Chachamovich, Eduardo Chachamovich, Hélène Ezer, Marcelo P. Fleck, Daniela Knauth, Eduardo P. Passos. (2010) Investigating quality of life and health-related quality of life in infertility: a systematic review. Journal of Psychosomatic Obstetrics & Gynecology 31:2, 101-110
    CrossRef

  26. 26

    (2010) Section 9: Electrophysiology Testing and the Use of Devices in Heart Failure. Journal of Cardiac Failure 16:6, e115-e121
    CrossRef

  27. 27

    BRETT D. ATWATER, JAMES P. DAUBERT. (2010) Can We Improve the Identification of Ventricular Scar to Guide Substrate-Based Ventricular Tachycardia Ablation?. Journal of Cardiovascular Electrophysiology 21:6, 685-687
    CrossRef

  28. 28

    K.H. Ladwig, J. Ronel, J. Baumert, C. Kolb. (2010) Psychische Komorbidität und Lebensqualität bei Patienten mit implantierbarem Kardioverter/Defibrillator (ICD). Herzschrittmachertherapie + Elektrophysiologie 21:2, 129-136
    CrossRef

  29. 29

    SANA M. AL-KHATIB, JONATHAN P. PICCINI, DAVID KNIGHT, MARGARET STEWART, NANCY CLAPP-CHANNING, GILLIAN D. SANDERS. (2010) Remote Monitoring of Implantable Cardioverter Defibrillators versus Quarterly Device Interrogations in Clinic: Results from a Randomized Pilot Clinical Trial. Journal of Cardiovascular Electrophysiology 21:5, 545-550
    CrossRef

  30. 30

    Jenny Tagney. (2010) A literature review comparing the experiences and emergent needs of adult patients with permanent pacemakers (PPMs) and implantable cardioverter defibrillators (ICDs). Journal of Clinical Nursingno-no
    CrossRef

  31. 31

    Daniel B. Mark, Kevin J. Anstrom, Steven E. McNulty, Greg C. Flaker, Andrew M. Tonkin, Warren M. Smith, William D. Toff, Paul Dorian, Nancy E. Clapp-Channing, Jill Anderson, George Johnson, Eleanor B. Schron, Jeanne E. Poole, Kerry L. Lee, Gust H. Bardy. (2010) Quality of life effects of automatic external defibrillators in the home: Results from the Home Automatic External Defibrillator Trial (HAT). American Heart Journal 159:4, 627-634.e7
    CrossRef

  32. 32

    Rodrigo Antonini Ribeiro, Steffan Frosi Stella, Suzi Alves Camey, Leandro Ioschpe Zimerman, Maurício Pimentel, Luis Eduardo Rohde, Carísi Anne Polanczyk. (2010) Cost-Effectiveness of Implantable Cardioverter-Defibrillators in Brazil: Primary Prevention Analysis in the Public Sector. Value in Health 13:2, 160-168
    CrossRef

  33. 33

    C. Jan Willem Borleffs, Johannes B. van Rees, Guido H. van Welsenes, Enno T. van der Velde, Lieselot van Erven, Jeroen J. Bax, Martin J. Schalij. (2010) Prognostic Importance of Atrial Fibrillation in Implantable Cardioverter-Defibrillator Patients. Journal of the American College of Cardiology 55:9, 879-885
    CrossRef

  34. 34

    Garrick C. Stewart, Joanne R. Weintraub, Parakash P. Pratibhu, Marc J. Semigran, Janice M. Camuso, Kimberly Brooks, Sui W. Tsang, Mary Susan Anello, Viviane T. Nguyen, Eldrin F. Lewis, Anju Nohria, Akshay S. Desai, Michael M. Givertz, Lynne W. Stevenson. (2010) Patient Expectations From Implantable Defibrillators to Prevent Death in Heart Failure. Journal of Cardiac Failure 16:2, 106-113
    CrossRef

  35. 35

    McMurray, John J.V., . (2010) Systolic Heart Failure. New England Journal of Medicine 362:3, 228-238
    Full Text

  36. 36

    Miguel Á. Arias, Marta Pachón, Alberto Puchol, Eduardo Castellanos. (2010) Electrofisiología cardiaca y arritmias. Revista Española de Cardiología 63, 61-72
    CrossRef

  37. 37

    Michiro Kiryu, Shinichi Niwano, Jun Kishihara, Yuya Aoyama, Shoko Ishikawa, Masami Murakami, Sayaka Kurokawa, Yoshihiro Yumoto, Ryuta Imaki, Hidehira Fukaya, Hiroe Niwano, Tohru Izumi. (2010) Preventive Effect of Amiodarone on VT/VF Events in ICD Patients with Structural Heart Diseases. Journal of Arrhythmia 26:4, 250-258
    CrossRef

  38. 38

    Mark A. Hlatky, Paul Heidenreich. (2009) The Year in Epidemiology, Health Services Research, and Outcomes Research. Journal of the American College of Cardiology 54:25, 2343-2351
    CrossRef

  39. 39

    Joseph D. Mishkin, Sherry J. Saxonhouse, Gregory W. Woo, Thomas A. Burkart, William M. Miles, Jamie B. Conti, Richard S. Schofield, Samuel F. Sears, Juan M. Aranda. (2009) Appropriate Evaluation and Treatment of Heart Failure Patients After Implantable Cardioverter-Defibrillator Discharge. Journal of the American College of Cardiology 54:22, 1993-2000
    CrossRef

  40. 40

    SAMUEL F. SEARS, MELISSA MATCHETT, JAMIE B. CONTI. (2009) Effective Management of ICD Patient Psychosocial Issues and Patient Critical Events. Journal of Cardiovascular Electrophysiology 20:11, 1297-1304
    CrossRef

  41. 41

    Juan Carlos Claro, Roberto Candia, Gabriel Rada, Francisco Larrondo, Fernando Baraona, Luz M Letelier, Juan Carlos Claro. 2009. Amiodarone versus other pharmacological interventions for prevention of sudden cardiac death. .
    CrossRef

  42. 42

    SANDRA B. DUNBAR, JONATHAN J. LANGBERG, CAROLYN M. REILLY, BINDU VISWANATHAN, FRANCES MCCARTY, STEVEN D. CULLER, MARIAN C. O’BRIEN, WILLIAM S. WEINTRAUB. (2009) Effect of a Psychoeducational Intervention on Depression, Anxiety, and Health Resource Use in Implantable Cardioverter Defibrillator Patients. Pacing and Clinical Electrophysiology 32:10, 1259-1271
    CrossRef

  43. 43

    M. Brignole. (2009) Are complications of implantable defibrillators under-estimated and benefits over-estimated?. Europace 11:9, 1129-1133
    CrossRef

  44. 44

    Weiwei Li, Brian Olshansky. (2009) Electrophysiological device therapy (implantable defibrillators) in older adults. Current Cardiovascular Risk Reports 3:5, 347-354
    CrossRef

  45. 45

    E. M. Aliot, W. G. Stevenson, J. M. Almendral-Garrote, F. Bogun, C. H. Calkins, E. Delacretaz, P. D. Bella, G. Hindricks, P. Jais, M. E. Josephson, J. Kautzner, G. N. Kay, K.-H. Kuck, B. B. Lerman, F. Marchlinski, V. Reddy, M.-J. Schalij, R. Schilling, K. Soejima, D. Wilber. (2009) EHRA/HRS Expert Consensus on Catheter Ablation of Ventricular Arrhythmias: Developed in a partnership with the European Heart Rhythm Association (EHRA), a Registered Branch of the European Society of Cardiology (ESC), and the Heart Rhythm Society (HRS); in collaboration with the American College of Cardiology (ACC) and the American Heart Association (AHA). Europace 11:6, 771-817
    CrossRef

  46. 46

    C. Cevik, A. Perez-Verdia, K. Nugent. (2009) Implantable cardioverter defibrillators and their role in heart failure progression. Europace 11:6, 710-715
    CrossRef

  47. 47

    Etienne M. Aliot, William G. Stevenson, Jesus Ma Almendral-Garrote, Frank Bogun, C. Hugh Calkins, Etienne Delacretaz, Paolo Della Bella, Gerhard Hindricks, Pierre Jaïs, Mark E. Josephson, Josef Kautzner, G. Neal Kay, Karl-Heinz Kuck, Bruce B. Lerman, Francis Marchlinski, Vivek Reddy, Martin-Jan Schalij, Richard Schilling, Kyoko Soejima, David Wilber. (2009) EHRA/HRS Expert Consensus on Catheter Ablation of Ventricular Arrhythmias. Heart Rhythm 6:6, 886-933
    CrossRef

  48. 48

    Andrew E. Epstein. (2009) Reply. Journal of the American College of Cardiology 53:13, 1163
    CrossRef

  49. 49

    Subhashni D. Singh Joy. (2009) IMPLANTABLE DEFIBRILLATORS EXTEND SURVIVAL WITHOUT ADVERSE EFFECTS. AJN, American Journal of Nursing 109:3, 72
    CrossRef

  50. 50

    Mark, Daniel B., Pan, Wenqin, Clapp-Channing, Nancy E., Anstrom, Kevin J., Ross, John R., Fox, Rebecca S., Devlin, Gerard P., Martin, C. Edwin, Adlbrecht, Christopher, Cowper, Patricia A., Ray, Linda Davidson, Cohen, Eric A., Lamas, Gervasio A., Hochman, Judith S., . (2009) Quality of Life after Late Invasive Therapy for Occluded Arteries. New England Journal of Medicine 360:8, 774-783
    Full Text

  51. 51

    (2009) Defibrillators in Heart Failure and Quality of Life. New England Journal of Medicine 360:2, 187-189
    Full Text

  52. 52

    (2009) Mortality risk and patients' quality of life on ICD therapy: insights from SCD-HeFT. Nature Clinical Practice Cardiovascular Medicine 6:1, 3-4
    CrossRef

  53. 53

    Vadim V. Fedorov. (2008) Does the postrepolarization refractoriness play a role in amiodarone's antiarrhythmic efficacy?. Heart Rhythm 5:12, 1743-1744
    CrossRef

  54. 54

    Healey, Jeff, Connolly, Stuart, . (2008) Life and Death after ICD Implantation. New England Journal of Medicine 359:10, 1058-1059
    Full Text

Letters