Use of a Continuous-Flow Device in Patients Awaiting Heart Transplantation

Leslie W. Miller, M.D., Francis D. Pagani, M.D., Ph.D., Stuart D. Russell, M.D., Ranjit John, M.D., Andrew J. Boyle, M.D., Keith D. Aaronson, M.D., John V. Conte, M.D., Yoshifumi Naka, M.D., Donna Mancini, M.D., Reynolds M. Delgado, M.D., Thomas E. MacGillivray, M.D., David J. Farrar, Ph.D., and O.H. Frazier, M.D. for the HeartMate II Clinical Investigators

N Engl J Med 2007; 357:885-896August 30, 2007

Abstract

Background

The use of left ventricular assist devices is an accepted therapy for patients with refractory heart failure, but current pulsatile volume-displacement devices have limitations (including large pump size and limited long-term mechanical durability) that have reduced widespread adoption of this technology. Continuous-flow pumps are newer types of left ventricular assist devices developed to overcome some of these limitations.

Methods

In a prospective, multicenter study without a concurrent control group, 133 patients with end-stage heart failure who were on a waiting list for heart transplantation underwent implantation of a continuous-flow pump. The principal outcomes were the proportions of patients who, at 180 days, had undergone transplantation, had cardiac recovery, or had ongoing mechanical support while remaining eligible for transplantation. We also assessed functional status and quality of life.

Results

The principal outcomes occurred in 100 patients (75%). The median duration of support was 126 days (range, 1 to 600). The survival rate during support was 75% at 6 months and 68% at 12 months. At 3 months, therapy was associated with significant improvement in functional status (according to the New York Heart Association class and results of a 6-minute walk test) and in quality of life (according to the Minnesota Living with Heart Failure and Kansas City Cardiomyopathy questionnaires). Major adverse events included postoperative bleeding, stroke, right heart failure, and percutaneous lead infection. Pump thrombosis occurred in two patients.

Conclusions

A continuous-flow left ventricular assist device can provide effective hemodynamic support for a period of at least 6 months in patients awaiting heart transplantation, with improved functional status and quality of life. (ClinicalTrials.gov number, NCT00121472.)

Media in This Article

Animation

Comparison of a Volume-Displacement Pump and an Axial-Flow Pump LVADs.

Comparison of a Volume-Displacement Pump and an Axial-Flow Pump LVADs.

Figure 1Components of the Continuous-Flow Left Ventricular Assist Device (LVAD).
Article

Animation

Comparison of a Volume-Displacement Pump and an Axial-Flow Pump LVADs.

Comparison of a Volume-Displacement Pump and an Axial-Flow Pump LVADs.

Therapy with a left ventricular assist device is an established form of treatment for patients with refractory heart failure.1 In the United States, most patients undergoing implantation of such a device as a bridge to heart transplantation have received support from pulsatile volume-displacement devices that fill with and eject blood in a cyclic fashion that is analogous to the systole and diastole of the native heart.2-7 These devices provide excellent hemodynamic support and improve survival but have substantial constraints, including the need for extensive surgical dissection, the requirement that the recipient have a large body habitus, the presence of a large-diameter percutaneous lead, audible pump operation, and limitations in long-term mechanical durability that frequently require subsequent operations for device exchange.8,9

More recently, several left ventricular assist devices have been developed with continuous-flow, rotary-pump technology (Figure 1Figure 1Components of the Continuous-Flow Left Ventricular Assist Device (LVAD).). One advantage of these newer pumps is a smaller device size, with the potential for extending therapy to underserved populations, including some women and adolescents.9,10 Another advantage is the potential for greater long-term mechanical reliability owing to a simplified design that requires only a single moving part, an internal rotor. Other benefits include less noise from the device and greater comfort for patients than with the typical pulsatile device.

Continuous-flow pumps — including the HeartMate II Left Ventricular Assist System (Thoratec),10 the MicroMed DeBakey Ventricular Assist Device (MicroMed),11 the Jarvik 2000 Heart (Jarvik Heart),12 and the VentrAssist Left Ventricular Assist System (Ventracor)13 — are the subject of ongoing clinical evaluation in the United States. We report on results from a large observational clinical study of a continuous-flow left ventricular assist device.

Methods

Study Design

The study was conducted at 26 centers in the United States between March 2005 and May 2006 and was supervised by the sponsor (Thoratec). Investigators in the clinical affairs and biostatistics departments at Thoratec designed the trial in ongoing consultation with the Food and Drug Administration (FDA) and the clinical investigators. Coordinators at each site collected all study data, which were then forwarded to the data analysis center of the sponsor. The academic authors had independent access to the data; they vouch for the completeness and accuracy of the data and the analyses.

A data and safety monitoring board, consisting of five independent physicians who were not investigators in the study, met routinely to review study compliance, adverse events, quality of life, and outcomes of patients. These five physicians were compensated for their time, but none have any financial interest in Thoratec or stand to gain financially from the outcome of the trial. A clinical events committee of four independent physicians reviewed, classified, and adjudicated the causes of death and all adverse events.

The study was conducted in compliance with FDA regulations for Good Clinical Practices. The protocol was approved by the FDA and the institutional review board at each participating center.

Study Subjects

Patients with end-stage heart failure who were on a waiting list for heart transplantation at each center were eligible for study enrollment. (Detailed inclusion and exclusion criteria are listed in the Supplementary Appendix, available with the full text of this article at www.nejm.org.) Patients were required to have symptoms of New York Heart Association (NYHA) class IV heart failure and to be ill enough to have high priority for transplantation (United Network for Organ Sharing status 1A or 1B). Exclusion criteria included severe renal, pulmonary, or hepatic dysfunction; active uncontrolled infection; a mechanical aortic valve; aortic insufficiency; an aortic aneurysm; the presence of other mechanical circulatory support, except for an intraaortic balloon pump; and technical obstacles thought by the investigator to pose an increased surgical risk. All participating patients provided written informed consent.

Baseline Assessment

We collected baseline data on all enrolled patients. Assessments included demographic characteristics, health history, NYHA functional class, surveys on quality of life (Minnesota Living with Heart Failure and Kansas City Cardiomyopathy questionnaires), blood chemical values, hematologic data, neurologic status, and concomitant medications.

Continuous-Flow Pump

The pump used in this study was the Heartmate II LVAD (Thoratec), which is a continuous-flow device consisting of an internal axial-flow blood pump with a percutaneous lead that connects the pump to an external system driver and power source (Figure 1).10 The pump contains an internal rotor with helical blades that curve around a central shaft. When the rotor spins on its axis, kinetic energy is imparted to the blood, which is drawn continuously from the left ventricular apex through the pump and into the ascending aorta. The pump has an implant volume of 63 ml and generates up to 10 liters per minute of flow at a mean pressure of 100 mm Hg. Details of the device's function and the approach to surgical implantation have been described elsewhere.10,14

Follow-up after Device Implantation

After implantation of the device, a standardized antithrombotic regimen was implemented with initiation of heparin followed by transition to warfarin as well as aspirin and dipyridamole (see the Supplementary Appendix). Postoperative medical care (including inotropic, antiarrhythmic, and heart-failure therapy) was managed according to each investigator's preference and usual practice. Data on performance of the device and hemodynamics of patients were recorded every 8 hours for 3 days, daily through day 14, and weekly through day 30 while the patient was hospitalized. The results of a physical assessment and laboratory tests and a list of medications were recorded on days 1, 3, 5, 7, 11, 14, 21, and 28 after implantation of the device while the patient was hospitalized. After 30 days, device measurements, laboratory evaluations, and physical assessments were required on a monthly basis.

After patients were discharged home, they were assessed over the telephone at least every 2 weeks; they returned to the investigational study site for follow-up, equipment review, and general status assessment weekly for the first 4 weeks and then monthly until the final outcome. Assessment of quality of life and a 6-minute walk test were completed at baseline and 1 month, 3 months, and 6 months after implantation of the device. Readmissions to the hospital and adverse events (including suspected device malfunction) were recorded throughout the study as they occurred with the use of standardized definitions (see the Supplementary Appendix). All deaths of patients and causes of death were determined at autopsy when possible or by examination of medical records or by interviews with family members. Final adjudication was determined by the clinical events committee.

Outcomes

The principal outcomes were the proportions of patients who, at 180 days, had undergone transplantation, had undergone explantation of the device because of recovery of ventricular function, or had ongoing mechanical support and remained eligible for transplantation (i.e., were not removed from the waiting list owing to irreversible complications or clinical deterioration). Secondary outcomes included overall survival, survival while receiving device support, survival after transplantation, frequency of adverse events, assessment of functional class by a 6-minute walk test, independent evaluation of NYHA functional class by a physician, and quality of life. Patients who had the original continuous-flow pump replaced with another identical device and survived to 180 days were included in the group meeting the principal outcomes, whereas the three patients who had the original pump replaced with a different type of device were not included.

Statistical Analysis

Differences between measures of hemodynamics and quality of life before and after implantation of the device were analyzed with the use of an independent-samples t-test. For a comparison between categorical variables, Fisher's exact test was used. The level of statistical significance was set at P<0.05. All statistical comparisons are two-sided. Biochemical and hemodynamic variables are presented as means (±SD), and medians and ranges were used where appropriate. Discrete variables are presented as percentages. Adverse events are presented both as the percentage of patients who had the event and as event rates per patient-year. Survival analysis for patients continuing on mechanical support was performed with the use of the Kaplan–Meier method with censoring for heart transplantation or cardiac recovery.

Results

Study Patients

A total of 133 patients who met study-entry criteria were enrolled in the study and underwent implantation of the continuous-flow pump as a bridge to cardiac transplantation. Most subjects were men with an average age of 50 years, and the primary cause of heart failure for the majority of patients was nonischemic cardiomyopathy (Table 1Table 1Baseline Characteristics of the 133 Patients.). Optimal oral medical therapy had failed in all patients, and all patients were receiving intravenous inotropic therapy, with 25% requiring more than one inotrope. Eleven percent of patients could not tolerate inotropes owing to cardiac arrhythmias. Forty-one percent of patients were on concomitant support with an intraaortic balloon pump.

Clinical Course

Five patients (4%) received temporary support from right ventricular assist devices for 3 to 93 days. These devices included two paracorporeal pneumatic devices and three short-term centrifugal pumps, two of which were converted to pneumatic devices. The median duration of postoperative inotropic support was 7 days. Seventeen patients (13%) required inotropic support for more than 14 days for right ventricular dysfunction. At 24 hours after implantation of the device, the cardiac index (liters per minute per square meter of body-surface area) increased from a mean (±SD) of 2.0±0.6 preoperatively to 2.8±0.7 (P<0.001); at the same time, pulmonary-capillary wedge pressure decreased from 26±8 to 16±5 mm Hg, and mean pulmonary-artery pressure decreased from 37±10 to 26±7 mm Hg (P<0.001 for both comparisons). The average pump flow index was 2.6±0.5 liters per minute per square meter on the first day at a mean pump speed of 9236±496 rpm and increased to 2.8±0.4 liters per minute per square meter at 1 month at a mean pump speed of 9502±525 rpm, with systolic and diastolic arterial blood pressures averaging 96±16 and 73±14 mm Hg, respectively. The international normalized ratio (INR) averaged 2.2±0.7 at 1 month. Values remained relatively stable throughout the support period.

From baseline to 3 months, renal and hepatic function improved during circulatory support, as evidenced in a paired analysis of 67 patients by reductions in levels of serum creatinine (from 1.4±0.5 to 1.1±0.5 mg per deciliter [124±44 to 97±44 μmol per liter], P<0.001), blood urea nitrogen (30.3±16.9 to 18.6±9.8 mg per deciliter [11±6 to 7±3 mmol per liter], P<0.001), and serum alanine aminotransferase (48±41 to 32±29 U per liter, P=0.006). Most patients who were evaluated at 3 months after device implantation had improvement in at least two NYHA functional classes and improvement in a 6-minute walk test by a distance of more than 200 m (Table 2Table 2Functional Status and Quality of Life.). Measures of quality of life significantly improved after device implantation on the basis of both survey instruments used (P<0.001).

Outcomes

All 133 patients were followed for at least 180 days or until either transplantation or death. Of these patients, 100 (75%) reached the principal outcomes of heart transplantation, cardiac recovery, or survival at 180 days with ongoing mechanical support and eligibility for transplantation (Table 3Table 3Outcomes of the 133 Patients.). Of these 100 patients, 56 underwent heart transplantation, 43 continued to receive support and were eligible for transplantation, and 1 did not need transplantation after recovery of cardiac function and explantation of the device (Figure 2AFigure 2Outcomes for 133 Patients after Implantation of the Continuous-Flow Left Ventricular Assist Device.). Of the 43 patients remaining on device support at 180 days, 32 were on the active list for heart transplantation, and 11 remained eligible for transplantation, including 4 who removed themselves from the transplantation list owing to a preference to continue mechanical support.

Among the 33 patients with unsuccessful outcomes were 25 patients who died before 180 days of support, with a median time to death of 38 days (range, 6 to 144). In addition, five patients became ineligible for transplantation during mechanical support owing to irreversible medical complications, and three patients underwent replacement of the continuous-flow pump with a different type of ventricular assist device (because of surgical complications that occurred shortly after pump implantation) and were withdrawn from the study. Two patients who underwent replacement of the continuous-flow pump with a second identical pump remained in the study, were alive on mechanical support at 216 and 367 days after the replacement, and are included as survivors in the actuarial survival curve. The overall rate of survival to transplantation, recovery, or continued support with no pump replacement was 75% at 180 days (Table 3).

Overall actuarial survival for patients continuing to receive pump support was 89% at 1 month, 75% at 6 months, and 68% at 12 months (Figure 2B). The median duration of support was 126 days (range, 1 to 600), with a mean of 168±148 days during a cumulative follow-up of 61.7 patient-years. The median time to transplantation was 97 days (range, 15 to 498), and the median time to cardiac recovery for three patients was 347 days (range, 161 to 380).

Twelve patients (9%) underwent transplantation during their initial hospital stay, and 18 patients (14%) died before discharge while receiving mechanical support. One hundred patients (75%) were discharged from the hospital while receiving mechanical support, with a median hospital stay after surgery of 25 days (range, 10 to 114). The median number of days out of hospital before transplantation, readmission, or death was 60 (range, 0 to 418). Fifty-four discharged patients required rehospitalization for complications, with a median duration of rehospitalization of 4 days (range, 0 to 57).

Adverse Events

The most common adverse event was bleeding, primarily in the early postoperative period (Table 4Table 4Adverse Events in the 133 Study Patients.). Eight patients (6%) had an ischemic stroke, and three (2%) had a hemorrhagic stroke. Five of these 11 events occurred within the first 2 days after device implantation. Five additional patients had transient ischemic attacks that were completely reversed. Nine patients were reported to have psychological symptoms. Eight patients had other neurologic events, six of which were completely reversed. Localized infection not related to device implantation occurred in 28% of patients, whereas device-related infection was observed in 14% of patients, with all infections involving the percutaneous lead and none involving the pump pocket. Five devices were replaced: two for pump thrombosis at 24 and 56 days after implantation and three for complications related to surgical implantation at 1, 15, and 32 days (Table 4).

The causes of death in the first 180 days after device implantation included sepsis (five patients), ischemic stroke (five), multisystem organ failure (four), hemorrhagic stroke (three), anoxic brain injury (two, one after a protamine reaction and one after a hemothorax with cardiac arrest), right heart failure (two), and miscellaneous other causes (four). There was one device-related death caused by an inflow graft that was accidentally twisted during implantation.

Discussion

In this study, we evaluated the use of a continuous-flow left ventricular assist device as a bridge to heart transplantation. We found that this device provided effective mechanical circulatory support in patients with refractory heart failure. Circulatory support with the continuous-flow pump significantly improved the hemodynamic status of patients and was associated with significant improvements in functional status, as assessed with a 6-minute walk test, and in NYHA functional class and quality of life, as measured by both the Minnesota Living with Heart Failure and Kansas City Cardiomyopathy questionnaires.

The use of a continuous-flow pump was not without the risk of complications. Significant adverse events included postoperative bleeding, stroke, right heart failure, and percutaneous-lead infection. At 6 months, 19% of patients had died while on device support, 4% had medical complications that precluded transplantation, and 2% had had their devices replaced by another type of left ventricular assist device. Although some of these events may be attributable to the severity of the patient's illness rather than to the device itself, they are indicative of the risks faced in this setting.

It has previously been shown that therapy with a left ventricular assist device substantially improves survival in patients with refractory heart failure. In a randomized comparison of patients ineligible for heart transplantation, the survival of patients assigned to ventricular assist support was 52% at 1 year, as compared with 25% for those assigned to medical therapy alone.15 Although randomized comparisons have not been performed in patients who are mechanically bridged to transplantation, the magnitude of benefit of device support appears to be similar in this population.2 We did not perform a direct comparison between patients receiving the continuous-flow pump either with those not receiving any device support or with those receiving a pulsatile-flow device. However, in previous reports involving patients who received pulsatile-flow pumps, approximately 70% of patients survived to transplantation or cardiac recovery.2,4,7 In our study, the overall survival of patients who underwent transplantation, recovered cardiac function, or continued to receive mechanical support while remaining a candidate for transplantation was estimated to be 70% at 1 year.

The current FDA-approved devices that have a pulsatile, volume-displacement design have significant limitations related to the size of the device, limited mechanical durability, and adverse events, such as infection. The effort to develop alternative approaches to ventricular assist device support has been in part motivated by these perceived shortcomings of the pulsatile pumps. In our study, adverse events per patient-year with the continuous-flow pump showed an acceptable risk profile, as compared with that reported for a pulsatile-flow pump,2 with respect to bleeding requiring surgery (0.78 vs. 1.47 events per patient-year), drive-line infection (0.37 vs. 3.49), stroke (0.19 vs. 0.44), other nonstroke neurologic events (0.26 vs. 0.67), and right heart failure requiring a right ventricular assist device (0.08 vs. 0.30). The higher incidence rate of infection with the pulsatile-flow pump in the previous study may have been related in part to the large diameter of the percutaneous lead (which is 50% larger than that of the continuous-flow pump used in our study) and the absence of a restraining device or belt to limit the movement of the percutaneous lead.

Although continuous-flow pumps may have some advantages over pulsatile pumps, as suggested by these comparisons, the devices also pose new or continuing challenges related to the treatment of patients. These issues include the risk of pump thrombosis and thromboembolism, with the requirement for higher levels of antithrombotic therapy than are required for some pulsatile devices and a consequent risk of bleeding. Infection remains a potential concern, as with all circulatory devices that have a percutaneous component. Mechanical failure may not be totally obviated by continuous-flow pump technology, although it appears to be less frequent than with some pulsatile pumps. Other issues include the need to determine the optimal pump-speed settings to provide sufficient blood flow without ventricular arrhythmias and difficulty in detecting vital signs in a systemic circulation with minimal pulsatility. A previous concern that diminished pulsatile pressure and flow might have unfavorable effects on major organ function has been dispelled,16 although few patients have received support for very extended periods.17

Several limitations of our study should be noted. As mentioned above, we did not perform a direct, randomized comparison of the continuous-flow pump with any other device or with medical management alone, and thus we cannot describe the comparative benefits of this form of therapy. We were not able to assess the functional status and quality of life of all patients in our study, which raises the concern that the estimates of typical benefit with respect to these end points may be subject to ascertainment bias. Finally, a comparison of our findings with those of other groups is difficult, in part because the appropriate criteria for selection of patients for ventricular assist remain somewhat subjective.16-18 The persisting rate of death of 20 to 25% before transplantation seen in our study is similar to previous reports with other pumps and suggests that the selection of patients and the presence of adverse risk factors at the time of device implantation contribute more to adverse outcome than the device used.16

In conclusion, we evaluated the efficacy of a continuous-flow pump in providing mechanical circulatory support as a bridge to heart transplantation. The results of this study show that effective hemodynamic support for periods of at least 6 months can be achieved with a continuous-flow left ventricular assist device, with improved functional status and quality of life.

Supported by Thoratec. Dr. Miller reports receiving consulting fees from Astellas and lecture fees and grant support from Thoratec; Dr. Russell, consulting fees from Thoratec; Dr. John, grant support from Bayer; Dr. Boyle, consulting and lecture fees from Thoratec; Dr. Conte, grant support from Paracor; Dr. Naka, consulting fees from Terumo Heart, Ventracor, Cardiomems, and Gerson Lehrman Group Councils (for investment advice related to ventricular assist devices) and lecture fees from Thoratec; Dr. Frazier, consulting and lecture fees from Thoratec, Terumo Heart, and Jarvik Heart; and Dr. Farrar, being an employee of Thoratec and having equity ownership in the company. No other potential conflict of interest relevant to this article was reported.

Drs. Miller and Pagani contributed equally to this article.

Source Information

From the University of Minnesota, Minneapolis (L.W.M., R.J., A.J.B.); the University of Michigan, Ann Arbor (F.D.P., K.D.A.); Johns Hopkins Hospital, Baltimore (S.D.R., J.V.C.); Columbia University, New York (Y.N., D.M.); Texas Heart Institute, Houston (R.M.D., O.H.F.); Massachusetts General Hospital, Boston (T.E.M.); and Thoratec, Pleasanton, CA (D.J.F.).

Address reprint requests to Dr. Miller at Washington Hospital Center, 110 Irving St. NW, Rm. 1F-1208, Washington, DC 20010-2975, or at .

Other investigators for the HeartMate II study are listed in the Appendix.

Appendix

In addition to the authors, the following surgeons, cardiologists, and study coordinators participated in this study: University of Michigan, Ann Arbor — J. Haft, T. Koelling, B. Dyke, E. Devaney, S. Wright; Johns Hopkins Hospital, Baltimore — D. Yuh, S. Ullrich; University of Minnesota, Minneapolis — L. Joyce, M. Colvin-Adams, E. Missov, C. Toninato; Columbia University, Presbyterian Hospital, New York — K. Idrissi, A. Stewart, E. Rose; Texas Heart Institute, Houston — B. Kar, B. Radovancevic, I. Gregoric, F. Smart, A. Civitello, E. Massin, C. Gemmato; Massachusetts General Hospital, Boston — A. Agnihotri, J. Madsen, G. Vlahakes, M. Semigran, S. Ennis; Barnes–Jewish Hospital, Washington University, St. Louis — N. Moazami, G. Ewald, K. Shelton; Shands Hospital, University of Florida, Gainesville — C. Klodell, J. Aranda, N. Staples; Sacred Heart Hospital, Spokane, WA — T. Icenogle, J. Everett, M. Pulhman; Duke University Medical Center, Durham, NC — C. Milano, J. Rogers, A. Lodge, L. Blue; Ohio State University, Columbus — B. Sun, D. Feldman, J. Sirak, S. Sudhaker, T. Yanssens; Medical City Hospital, Dallas — T. Dewey, M. Magee, M. Mack, A. Anderson, T. Worley; University of Washington, Seattle — E. Verrier, D. Fishbein, C. Salerno, G. Aldea, S. Andrus; University of Pittsburgh Medical Center, Pittsburgh — R. Kormos, D. McNamara, S. Weaver, K. Zehr; University of Rochester, Strong Memorial Hospital, Rochester, NY — T. Massey, L. Chen, W. Hallinan, V. Chiodo; Hospital of the University of Pennsylvania, Philadelphia — M. Acker, M. Jessup, R. Morris, S. Desai, M. O'Hara; Jewish Hospital, Louisville, KY — L. Gray, R. Dowling, S. Pagni, G. Bhat, P. Adkisson; Cleveland Clinic Foundation, Cleveland — N. Smedira, R. Starling, J. Navia, M. Banbury, R. Palumbo, T. Farillo; St. Luke's Medical Center, Milwaukee — A. Tector, J. Mendez, B. Pisani, J. Crouch, F. Downey, D. Kress, M. McDonald, D. O'Hair, M. Savitt, M. Miller; University of Alabama, Birmingham — J. Kirklin, R. Bourge, D. McGiffin, R. Benza, S. Pamboukian, B. Rayburn, J. Tallaj, S. Kinder; LDS Hospital, Salt Lake City — J. Long, S. Horton, D. Renland, J. Revenaugh, M. Eidson; Sharp Memorial Hospital, San Diego, CA — W. Dembitsky, B. Jaski, R. Adamson, S. Baradarian, S. Chillcott; Methodist Hospital–Clarian, Indianapolis — T. Wozniac, W. Ghumann, M. Turrentine, S. Becka; Henry Ford Hospital, Detroit — R. Brewer, B. Czerska, C. Williams, K. Leszczynski; Sentara Norfolk General Hospital, Norfolk, VA — J. Rich, J. Herre, L. Pine; and Baptist Memorial Hospital, Memphis — E. Garrett, T. Edwards, R. Carter, C. Porter.

Steering/Publication Committee — L. Miller, F. Pagani, O. Frazier, S. Russell, D. Farrar, Y. Naka, M. Slaughter; Data Safety and Monitoring Board — C. Yancy, S. Hunt, W. Holman, W. Richenbacher, D. Heitjan; Clinical Events Committee — S. Moore, V. Jeevanandam, C. Thomas, S. Gordon; Thoratec, Pleasanton, CA — L. Damme (study management), J. Heatley (biostatistics), S. Reichenbach (program management).

References

References

  1. 1

    Hunt SA, Abraham WT, Chin MH, et al. ACC/AHA 2005 guideline update for the diagnosis and management of chronic heart failure in the adult: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (Writing Committee to Update the 2001 Guidelines for the Evaluation and Management of Heart Failure): developed in collaboration with the American College of Chest Physicians and the International Society for Heart and Lung Transplantation: endorsed by the Heart Rhythm Society. Circulation 2005;112:e154-e235
    CrossRef | Web of Science | Medline

  2. 2

    Frazier OH, Rose EA, Oz MC, et al. Multicenter clinical evaluation of the HeartMate vented electric left ventricular assist system in patients awaiting heart transplantation. J Thorac Cardiovasc Surg 2001;122:1186-1195
    CrossRef | Web of Science | Medline

  3. 3

    Deng MC, Edwards LB, Hertz MI, et al. Mechanical circulatory support device database of the International Society for Heart and Lung Transplantation: third annual report -- 2005. J Heart Lung Transplant 2005;24:1182-1187
    CrossRef | Web of Science | Medline

  4. 4

    Frazier OH, Rose EA, McCarthy P, et al. Improved mortality and rehabilitation of transplant candidates treated with a long-term implantable left ventricular assist system. Ann Surg 1995;222:327-336
    CrossRef | Web of Science | Medline

  5. 5

    Farrar DJ, Hill JD, Pennington DG, et al. Preoperative and postoperative comparison of patients with univentricular and biventricular support with the Thoratec ventricular assist device as a bridge to heart transplantation. J Thorac Cardiovasc Surg 1997;113:202-209
    CrossRef | Web of Science | Medline

  6. 6

    Portner PM, Jansen PG, Oyer PE, Wheeldon DR, Ramasamy N. Improved outcomes with an implantable left ventricular assist system: a multicenter study. Ann Thorac Surg 2001;71:205-209
    CrossRef | Web of Science | Medline

  7. 7

    Slaughter MS, Tsui SS, El-Banayosy A, et al. Results of a multicenter clinical trial with the Thoratec implantable ventricular assist device. J Thorac Cardiovasc Surg 2007;133:1573-1580
    CrossRef | Web of Science | Medline

  8. 8

    Baldwin JT, Robbins RC. Executive summary for the National Heart, Lung, and Blood Institute Working Group on next generation ventricular assist devices for destination therapy. Semin Thorac Cardiovasc Surg 2005;17:369-371
    CrossRef | Medline

  9. 9

    Frazier OH, Kirklin JK, eds. Mechanical circulatory support. ISHLT monograph series. Vol. 1. Oxford, England: Elsevier, 2006.

  10. 10

    Griffith BP, Kormos RL, Borovetz HS, et al. HeartMate II left ventricular assist system: from concept to first clinical use. Ann Thorac Surg 2001;71:Suppl 3:S116-S120
    CrossRef | Web of Science | Medline

  11. 11

    Goldstein DJ. Worldwide experience with the MicroMed DeBakey ventricular assist device as a bridge to transplantation. Circulation 2003;108:Suppl 1:II-272
    CrossRef | Web of Science

  12. 12

    Frazier OH, Myers TJ, Westaby S, Gregoric ID. Clinical experience with an implantable, intracardiac, continuous flow circulatory support device: physiologic implications and their relationship to patient selection. Ann Thorac Surg 2004;77:133-142
    CrossRef | Web of Science | Medline

  13. 13

    Esmore DS, Kaye D, Salamonsen R, et al. First clinical implant of the VentrAssist left ventricular assist system as destination therapy for end-stage heart failure. J Heart Lung Transplant 2005;24:1150-1154
    CrossRef | Web of Science | Medline

  14. 14

    Goldstein D, Zucker M, Pagani FD, Frazier OH. Rotary ventricular assist devices. In: Frazier OH, Kirklin JK, eds. Mechanical circulatory support. ISHLT monograph series. Vol. 1. Oxford, England: Elsevier, 2006:77-104.

  15. 15

    Rose EA, Gelijns AC, Moskowitz AJ, et al. Long-term mechanical left ventricular assistance for end-stage heart failure. N Engl J Med 2001;345:1435-1443
    Full Text | Web of Science | Medline

  16. 16

    Thalmann M, Schima H, Wiesenthaler G, Wolner E. Physiology of continuous blood flow in recipients of rotary cardiac assist devices. J Heart Lung Transplant 2005;24:237-245
    CrossRef | Web of Science | Medline

  17. 17

    Westaby S, Frazier OH, Banning A, et al. Six years of continuous mechanical circulatory support. N Engl J Med 2006;355:325-327
    Full Text | Web of Science | Medline

  18. 18

    Lietz K, Long JW, Kfoury AG, et al. Outcomes of left ventricular assist device implantation as destination therapy in the post-REMATCH era: implications for patient selection. Circulation 2007;116:497-505
    CrossRef | Web of Science | Medline

Citing Articles (258)

Citing Articles

  1. 1

    S. E. A. Felix, J. R. Martina, J. H. Kirkels, C. Klopping, H. Nathoe, E. Sukkel, N. Hulstein, F. Z. Ramjankhan, P. A. F. M. Doevendans, J. R. Lahpor, N. de Jonge. (2012) Continuous-flow left ventricular assist device support in patients with advanced heart failure: points of interest for the daily management. European Journal of Heart Failure
    CrossRef

  2. 2

    Farouk Mookadam, Christopher B. Kendall, Raymond K. Wong, Anantharam Kalya, Tahlil Warsame, Francisco A. Arabia, Joan Lusk, Sherif Moustafa, Eric Steidley, Nishath Quader, Krishnaswamy Chandrasekaran. (2012) Left Ventricular Assist Devices: Physiologic Assessment using Echocardiography for Management and Optimization. Ultrasound in Medicine & Biology 38:2, 335-345
    CrossRef

  3. 3

    Robert J. Brewer, David E. Lanfear, Chittoor B. Sai-Sudhakar, Kartik S. Sundareswaran, Yazhini Ravi, David J. Farrar, Mark S. Slaughter. (2012) Extremes of body mass index do not impact mid-term survival after continuous-flow left ventricular assist device implantation. The Journal of Heart and Lung Transplantation 31:2, 167-172
    CrossRef

  4. 4

    L. K. Murray, J. Gonzalez-Costello, S. N. Jonas, D. B. Sims, K. A. Morrison, P. C. Colombo, D. M. Mancini, S. W. Restaino, E. Joye, E. Horn, H. Takayama, C. C. Marboe, Y. Naka, U. P. Jorde, N. Uriel. (2012) Ventricular assist device support as a bridge to heart transplantation in patients with giant cell myocarditis. European Journal of Heart Failure
    CrossRef

  5. 5

    Craig R. Butler, Bodh I. Jugdutt. (2012) Mechanical circulatory support for elderly heart failure patients. Heart Failure Reviews
    CrossRef

  6. 6

    Andrew Boyle. (2012) Arrhythmias in patients with ventricular assist devices. Current Opinion in Cardiology 27:1, 13-18
    CrossRef

  7. 7

    Scott R. Auerbach, Marc E. Richmond, Jonathan M. Chen, Ralph S. Mosca, Jan M. Quaegebeur, Linda J. Addonizio, Daphne T. Hsu, Jacqueline M. Lamour. (2012) Multiple Risk Factors Before Pediatric Cardiac Transplantation Are Associated With Increased Graft Loss. Pediatric Cardiology 33:1, 49-54
    CrossRef

  8. 8

    James R. Stone, Cristina Basso, Ulrik T. Baandrup, Patrick Bruneval, Jagdish Butany, Patrick J. Gallagher, Marc K. Halushka, Dylan V. Miller, Robert F. Padera, Stanley J. Radio, Mary N. Sheppard, Kim Suvarna, Carmela D. Tan, Gaetano Thiene, Allard C. van der Wal, John P. Veinot. (2012) Recommendations for processing cardiovascular surgical pathology specimens: a consensus statement from the Standards and Definitions Committee of the Society for Cardiovascular Pathology and the Association for European Cardiovascular Pathology. Cardiovascular Pathology 21:1, 2-16
    CrossRef

  9. 9

    Jens Garbade, Hartmuth Bruno Bittner, Sven Lehmann, Friedrich-Wilhelm Mohr, Markus Johannes Barten. (2012) Miniaturization of left ventricular assist devices: the ongoing trend. Expert Review of Medical Devices 9:1, 49-58
    CrossRef

  10. 10

    S J Park, S S Kushwaha, C G A McGregor. (2012) State-of-the-Art Implantable Cardiac Assist Device Therapy for Heart Failure: Bridge to Transplant and Destination Therapy. Clinical Pharmacology & Therapeutics 91:1, 94-100
    CrossRef

  11. 11

    Tal Hasin, Yan Topilsky, John A. Schirger, Zhuo Li, Yanjun Zhao, Barry A. Boilson, Alfredo L. Clavell, Richard J. Rodeheffer, Robert P. Frantz, Brooks S. Edwards, Naveen L. Pereira, Lyle Joyce, Richard Daly, Soon J. Park, Sudhir S. Kushwaha. (2012) Changes in Renal Function After Implantation of Continuous-Flow Left Ventricular Assist Devices. Journal of the American College of Cardiology 59:1, 26-36
    CrossRef

  12. 12

    Koichi Toda, Tomoyuki Fujita, Junjiro Kobayashi, Yusuke Shimahara, Soichiro Kitamura, Osamu Seguchi, Yoshihiro Murata, Masanobu Yanase, Takeshi Nakatani. (2012) Impact of Preoperative Percutaneous Cardiopulmonary Support on Outcome Following Left Ventricular Assist Device Implantation. Circulation Journal 76:1, 88-95
    CrossRef

  13. 13

    Daisuke Yoshioka, Taichi Sakaguchi, Shunsuke Saito, Shigeru Miyagawa, Hiroyuki Nishi, Yasushi Yoshikawa, Satsuki Fukushima, Takayoshi Ueno, Toru Kuratani, Yoshiki Sawa. (2012) Initial Experience of Conversion of Toyobo Paracorporeal Left Ventricular Assist Device to DuraHeart Left Ventricular Assist Device. Circulation Journal 76:2, 372-376
    CrossRef

  14. 14

    Mark Puhlman. (2012) Continuous-Flow Left Ventricular Assist Device and the Right Ventricle. AACN Advanced Critical Care 23:1, 86-90
    CrossRef

  15. 15

    O. Wever-Pinzon, J. Stehlik, A.G. Kfoury, J.V. Terrovitis, N.A. Diakos, C. Charitos, D.Y. Li, S.G. Drakos. (2012) Ventricular assist devices: Pharmacological aspects of a mechanical therapy. Pharmacology & Therapeutics
    CrossRef

  16. 16

    Ralph J. Petrucci, Joseph G. Rogers, Laura Blue, Colleen Gallagher, Stuart D. Russell, Dzifa Dordunoo, Brian E. Jaski, Suzanne Chillcott, Benjamin Sun, Tammy L. Yanssens, Antone Tatooles, Lalig Koundakjian, David J. Farrar, Mark S. Slaughter. (2012) Neurocognitive function in destination therapy patients receiving continuous-flow vs pulsatile-flow left ventricular assist device support. The Journal of Heart and Lung Transplantation 31:1, 27-36
    CrossRef

  17. 17

    Andres F. Osorio, Ruben Osorio, Andres Ceballos, Reginald Tran, William Clark, Eduardo A. Divo, I. Ricardo Argueta-Morales, Alain J. Kassab, William M. DeCampli. (2011) Computational fluid dynamics analysis of surgical adjustment of left ventricular assist device implantation to minimise stroke risk. Computer Methods in Biomechanics and Biomedical Engineering1-17
    CrossRef

  18. 18

    Keyur B. Shah, Daniel G. Tang, Richard H. Cooke, Michael C. Kontos, Neil P. Lewis, Gundars J. Katlaps, Michael L. Hess, Vigneshwar Kasirajan. (2011) Evaluating Heart Failure After Implantation of Mechanical Circulatory Support Devices. Current Heart Failure Reports
    CrossRef

  19. 19

    Brantley Dollar Gaitan, Christopher A. Thunberg, Lynn G. Stansbury, Dawn E. Jaroszewski, Francisco A. Arabia, Bartley P. Griffith, Alina M. Grigore. (2011) Development, Current Status, and Anesthetic Management of the Implanted Artificial Heart. Journal of Cardiothoracic and Vascular Anesthesia 25:6, 1179-1192
    CrossRef

  20. 20

    J.K. Vierecke, F. Kaufmann, M. Schweiger, A. Stepanenko, N. Dranishnikov, H. Lehmkuhl, E. Potapov, R. Hetzer, T. Krabatsch. (2011) Erste Maßnahmen und Verhalten bei Störungen eines Kunstherzsystems. Notfall + Rettungsmedizin 14:8, 635-646
    CrossRef

  21. 21

    STAVROS DIMOPOULOS, NIKOLAOS DIAKOS, ELENI TSELIOU, ATHANASIOS TASOULIS, ANTHI MPOUCHLA, CHRISTOS MANETOS, LAMBROS KATSAROS, STAVROS DRAKOS, JOHN TERROVITIS, SERAFIM NANAS. (2011) Chronotropic Incompetence and Abnormal Heart Rate Recovery Early after Left Ventricular Assist Device Implantation. Pacing and Clinical Electrophysiology 34:12, 1607-1614
    CrossRef

  22. 22

    Martin Stout, Roshan Ravindran, Christopher Miller, Keith Pearce. (2011) Preimplant Transthoracic Echocardiographic Assessment of Continuous Flow Left Ventricular Assist Device. Echocardiographyno-no
    CrossRef

  23. 23

    M. L. A. Haeck, G. E. Hoogslag, S. F. Rodrigo, D. E. Atsma, R. J. Klautz, E. E. Wall, M. J. Schalij, H. F. Verwey. (2011) Treatment options in end-stage heart failure: where to go from here?. Netherlands Heart Journal
    CrossRef

  24. 24

    Jesus M. Casida, Linda Marcuccilli, Rosalind M. Peters, Susan Wright. (2011) Lifestyle adjustments of adults with long-term implantable left ventricular assist devices: A phenomenologic inquiry. Heart & Lung: The Journal of Acute and Critical Care 40:6, 511-520
    CrossRef

  25. 25

    Carl A. Johnson, Venkat Shankarraman, Peter D. Wearden, Ergin Kocyildirim, Timothy M. Maul, John D. Marks, J. Scott Richardson, Barry N. Gellman, Harvey S. Borovetz, Kurt A. Dasse, William R. Wagner. (2011) Platelet Activation After Implantation of the Levitronix PediVAS in the Ovine Model. ASAIO Journal 57:6, 516-521
    CrossRef

  26. 26

    George J. Arnaoutakis, Timothy J. George, Arman Kilic, Eric S. Weiss, Stuart D. Russell, John V. Conte, Ashish S. Shah. (2011) Effect of sensitization in US heart transplant recipients bridged with a ventricular assist device: Update in a modern cohort. The Journal of Thoracic and Cardiovascular Surgery 142:5, 1236-1245.e1
    CrossRef

  27. 27

    Aaron H. Healy, Nathan O. Mason, M. Elizabeth Hammond, Bruce B. Reid, Stephen E. Clayson, Stavros G. Drakos, Abdallah G. Kfoury, Amit N. Patel, David A. Bull, Deborah Budge, Rami A. Alharethi, Feras M. Bader, Edward M. Gilbert, Josef Stehlik, Craig H. Selzman. (2011) Allograft Rejection in Patients Supported With Continuous-Flow Left Ventricular Assist Devices. The Annals of Thoracic Surgery 92:5, 1601-1607
    CrossRef

  28. 28

    B. Joseph Elmunzer, Kunjali T. Padhya, Jason J. Lewis, Amol S. Rangnekar, Sameer D. Saini, Shanti L. Eswaran, James M. Scheiman, Francis D. Pagani, Jonathan W. Haft, Akbar K. Waljee. (2011) Endoscopic Findings and Clinical Outcomes in Ventricular Assist Device Recipients with Gastrointestinal Bleeding. Digestive Diseases and Sciences 56:11, 3241-3246
    CrossRef

  29. 29

    Louis H. Stein, John A. Elefteriades. (2011) Can the Occurrence of Gastrointestinal Bleeding in Nonpulsatile Left Ventricular Assist Device Patients Provide Clues for the Reversal of Arteriosclerosis?. Cardiology Clinics 29:4, 641-645
    CrossRef

  30. 30

    Yoshio Iwashima, Masanobu Yanase, Takeshi Horio, Osamu Seguchi, Yoshihiro Murata, Tomoyuki Fujita, Kouichi Toda, Yuhei Kawano, Takeshi Nakatani. (2011) Effect of Pulsatile Left Ventricular Assist System Implantation on Doppler Measurements of Renal Hemodynamics in Patients With Advanced Heart Failure. Artificial Organsno-no
    CrossRef

  31. 31

    Ranjit John, Forum Kamdar, Peter Eckman, Monica Colvin-Adams, Andrew Boyle, Sara Shumway, Lyle Joyce, Kenneth Liao. (2011) Lessons Learned From Experience With Over 100 Consecutive HeartMate II Left Ventricular Assist Devices. The Annals of Thoracic Surgery 92:5, 1593-1600
    CrossRef

  32. 32

    Daniel Pereda, John V. Conte. (2011) Left Ventricular Assist Device Driveline Infections. Cardiology Clinics 29:4, 515-527
    CrossRef

  33. 33

    Benjamin Waters, Alanson Sample, Joshua Smith, Pramod Bonde. (2011) Toward Total Implantability Using Free-Range Resonant Electrical Energy Delivery System: Achieving Untethered Ventricular Assist Device Operation Over Large Distances. Cardiology Clinics 29:4, 609-625
    CrossRef

  34. 34

    Ming-Da Zhou, Chuan Yang, Zhiwen Liu, Joshua P. Cysyk, Si-Yang Zheng. (2011) An implantable Fabry-Pérot pressure sensor fabricated on left ventricular assist device for heart failure. Biomedical Microdevices
    CrossRef

  35. 35

    Tobias Bomholt, Claus Moser, Kaare Sander, Søren Boesgaard, Lars Køber, Peter Skov Olsen, Peter Bo Hansen, Svend-Aage Mortensen, Finn Gustafsson. (2011) Driveline infections in patients supported with a HeartMate II: Incidence, aetiology and outcome. Scandinavian Cardiovascular Journal 45:5, 273-278
    CrossRef

  36. 36

    Gregory Egnaczyk, Carmelo A. Milano, Joseph G. Rogers. (2011) Clinical Trials in Mechanical Circulatory Support. Heart Failure Clinics 7:4, 457-466
    CrossRef

  37. 37

    Jaime Moriguchi, Stephanie Davis, Rhodora Jocson, Fardad Esmailian, Abbas Ardehali, Hillel Laks, Murray Kwon, Michelle Kittleson, Jon Kobashigawa, Jignesh Patel, Daniel Marelli, Mark Plunkett, Ramin Beygui, Richard Shemin. (2011) Successful use of a pneumatic biventricular assist device as a bridge to transplantation in cardiogenic shock. The Journal of Heart and Lung Transplantation 30:10, 1143-1147
    CrossRef

  38. 38

    T. Krabatsch, M. Schweiger, A. Stepanenko, T. Drews, E. Potapov, M. Pasic, Y. Weng, M. Huebler, R. Hetzer. (2011) Fortschritte bei implantierbaren mechanischen Kreislaufunterstüzungssystemen. Herz 36:7, 622-629
    CrossRef

  39. 39

    Ranjit John, Yoshifumi Naka, Nicholas G. Smedira, Randall Starling, Ulrich Jorde, Peter Eckman, David J. Farrar, Francis D. Pagani. (2011) Continuous Flow Left Ventricular Assist Device Outcomes in Commercial Use Compared With the Prior Clinical Trial. The Annals of Thoracic Surgery 92:4, 1406-1413
    CrossRef

  40. 40

    Sanjay Maniar, Sreekanth Kondareddy, Veli K Topkara. (2011) Left ventricular assist device-related infections: past, present and future. Expert Review of Medical Devices 8:5, 627-634
    CrossRef

  41. 41

    Mark S. Slaughter, Guruprasad A. Giridharan, Dan Tamez, Jeff LaRose, Mike A. Sobieski, Leslie Sherwood, Steven C. Koenig. (2011) Transapical miniaturized ventricular assist device: Design and initial testing. The Journal of Thoracic and Cardiovascular Surgery 142:3, 668-674
    CrossRef

  42. 42

    Nader Moazami, Benjamin Sun, David Feldman. (2011) Stable patients on left ventricular assist device support have a disproportionate advantage: Time to re-evaluate the current UNOS policy. The Journal of Heart and Lung Transplantation 30:9, 971-974
    CrossRef

  43. 43

    Chetan B. Patel, Joseph G. Rogers. (2011) Durable Mechanical Circulatory Support Devices. Progress in Cardiovascular Diseases 54:2, 132-143
    CrossRef

  44. 44

    Prateeti Khazanie, Joseph G. Rogers. (2011) Patient Selection for Left Ventricular Assist Devices. Congestive Heart Failure 17:5, 227-234
    CrossRef

  45. 45

    Simon Maltais, Yan Topilsky, Vakhtang Tchantchaleishvili, Stephen H. McKellar, Lucian A. Durham, Lyle D. Joyce, Richard C. Daly, Soon J. Park. (2011) Surgical treatment of tricuspid valve insufficiency promotes early reverse remodeling in patients with axial-flow left ventricular assist devices. The Journal of Thoracic and Cardiovascular Surgery
    CrossRef

  46. 46

    Ralph J. Petrucci, Lynne A. Benish, Barbara L. Carrow, Lisa Prato, Shelley R. Hankins, Howard J. Eisen, John W. Entwistle. (2011) Ethical Considerations for Ventricular Assist Device Support: A 10-Point Model. ASAIO Journal 57:4, 268-273
    CrossRef

  47. 47

    Pere A. Ventura, Rami Alharethi, Deborah Budge, Bruce B. Reid, Benjamin D. Horne, Nathan O. Mason, Sandi Stoker, William T. Caine, Brad Rasmusson, John Doty, Stephen E. Clayson, Abdallah G. Kfoury. (2011) Differential impact on post-transplant outcomes between pulsatile- and continuous-flow left ventricular assist devices. Clinical Transplantation 25:4, E390-E395
    CrossRef

  48. 48

    Luc Jacquet, Olivier Vancaenegem, Agnès Pasquet, Pascal Matte, Alain Poncelet, Joel Price, Olivier Gurné, Philippe Noirhomme. (2011) Exercise Capacity in Patients Supported With Rotary Blood Pumps Is Improved by a Spontaneous Increase of Pump Flow at Constant Pump Speed and by a Rise in Native Cardiac Output. Artificial Organs 35:7, 682-690
    CrossRef

  49. 49

    Hussein S. Al-Amri, Abdulrahman M. Al-Moghairi, Rieda M. El Oakley. (2011) Surgical treatment of functional mitral regurgitation in dilated cardiomyopathy. Journal of the Saudi Heart Association 23:3, 125-134
    CrossRef

  50. 50

    DANIEL B. SIMS, GREGG ROSNER, NIR URIEL, JOSÉ GONZÁLEZ-COSTELLO, FREDERICK A. EHLERT, ULRICH P. JORDE. (2011) Twelve Hours of Sustained Ventricular Fibrillation Supported by a Continuous-Flow Left Ventricular Assist Device. Pacing and Clinical Electrophysiologyno-no
    CrossRef

  51. 51

    Robert M. Adamson, Marcia Stahovich, Suzanne Chillcott, Sam Baradarian, Joseph Chammas, Brian Jaski, Peter Hoagland, Walter Dembitsky. (2011) Clinical Strategies and Outcomes in Advanced Heart Failure Patients Older Than 70 Years of Age Receiving the HeartMate II Left Ventricular Assist Device. Journal of the American College of Cardiology 57:25, 2487-2495
    CrossRef

  52. 52

    Jonah Odim. (2011) Proteins caught in the proteome of the unloaded remodeling pump. The Journal of Heart and Lung Transplantation 30:5, 494-496
    CrossRef

  53. 53

    Matthew L. Williams, Jaimin R. Trivedi, Kelly C. McCants, Sumanth D. Prabhu, Emma J. Birks, Laurie Oliver, Mark S. Slaughter. (2011) Heart Transplant vs Left Ventricular Assist Device in Heart Transplant-Eligible Patients. The Annals of Thoracic Surgery 91:5, 1330-1334
    CrossRef

  54. 54

    Randall C. Starling, Yoshifumi Naka, Andrew J. Boyle, Gonzalo Gonzalez-Stawinski, Ranjit John, Ulrich Jorde, Stuart D. Russell, John V. Conte, Keith D. Aaronson, Edwin C. McGee, William G. Cotts, David DeNofrio, Duc Thinh Pham, David J. Farrar, Francis D. Pagani. (2011) Results of the Post-U.S. Food and Drug Administration-Approval Study With a Continuous Flow Left Ventricular Assist Device as a Bridge to Heart Transplantation. Journal of the American College of Cardiology 57:19, 1890-1898
    CrossRef

  55. 55

    Jose N. Nativi, Stavros G. Drakos, Anna Y. Kucheryavaya, Leah B. Edwards, Craig H. Selzman, David O. Taylor, Marshall I. Hertz, Abdallah G. Kfoury, Josef Stehlik. (2011) Changing outcomes in patients bridged to heart transplantation with continuous- versus pulsatile-flow ventricular assist devices: An analysis of the registry of the International Society for Heart and Lung Transplantation. The Journal of Heart and Lung Transplantation
    CrossRef

  56. 56

    Faustino A. Toba, Hirokazu Akashi, Carlos Arrecubieta, Franklin D. Lowy. (2011) Role of biofilm in Staphylococcus aureus and Staphylococcus epidermidis ventricular assist device driveline infections. The Journal of Thoracic and Cardiovascular Surgery 141:5, 1259-1264
    CrossRef

  57. 57

    Joseph C. Cleveland, David C. Naftel, T. Brett Reece, Margaret Murray, James Antaki, Francis D. Pagani, James K. Kirklin. (2011) Survival after biventricular assist device implantation: An analysis of the Interagency Registry for Mechanically Assisted Circulatory Support database. The Journal of Heart and Lung Transplantation
    CrossRef

  58. 58

    Roberta C. Bogaev, Salpy V. Pamboukian, Stephanie A. Moore, Leway Chen, Ranjit John, Andrew J. Boyle, Kartik S. Sundareswaran, David J. Farrar, O.H. Frazier. (2011) Comparison of outcomes in women versus men using a continuous-flow left ventricular assist device as a bridge to transplantation. The Journal of Heart and Lung Transplantation 30:5, 515-522
    CrossRef

  59. 59

    Mordechai Findler, Michael Findler, Ehud Rudis. (2011) Dental treatment of a patient with an implanted left ventricular assist device: expanding the frontiers. Oral Surgery, Oral Medicine, Oral Pathology, Oral Radiology, and Endodontology 111:5, e1-e4
    CrossRef

  60. 60

    Yoan Lamarche, Mark Kearns, Kiranbir Josan, Jamil Bashir, Andrew Ignaszewski, Annemarie Kaan, Jennifer Kealy, Robert Moss, Anson Cheung. (2011) Successful Weaning and Explantation of the Heartmate II Left Ventricular Assist Device. Canadian Journal of Cardiology 27:3, 358-362
    CrossRef

  61. 61

    Bart P. Meyns, Andre Simon, Stefan Klotz, Thorsten Wittwer, Christian Schlensak, Filip Rega, Daniel Burkhoff. (2011) Clinical benefits of partial circulatory support in New York Heart Association Class IIIB and Early Class IV patients. European Journal of Cardio-Thoracic Surgery 39:5, 693-698
    CrossRef

  62. 62

    Josef Stehlik, David S Feldman. (2011) 2011–2020: decade of the ventricular assist device. Current Opinion in Cardiology 26:3, 230-231
    CrossRef

  63. 63

    Robert M. Adamson, Walter P. Dembitsky, Sam Baradarian, Joseph Chammas, Karen May-Newman, Suzanne Chillcott, Marcia Stahovich, Vicki McCalmont, Kristi Ortiz, Peter Hoagland, Brian Jaski. (2011) Aortic valve closure associated with HeartMate left ventricular device support: Technical considerations and long-term results. The Journal of Heart and Lung Transplantation 30:5, 576-582
    CrossRef

  64. 64

    Jane MacIver, Vivek Rao, Heather J Ross. (2011) Quality of life for patients supported on a left ventricular assist device. Expert Review of Medical Devices 8:3, 325-337
    CrossRef

  65. 65

    Robert S. McKelvie, Gordon W. Moe, Anson Cheung, Jeannine Costigan, Anique Ducharme, Estrellita Estrella-Holder, Justin A. Ezekowitz, John Floras, Nadia Giannetti, Adam Grzeslo, Karen Harkness, George A. Heckman, Jonathan G. Howlett, Simon Kouz, Kori Leblanc, Elizabeth Mann, Eileen O'Meara, Miroslav Rajda, Vivek Rao, Jessica Simon, Elizabeth Swiggum, Shelley Zieroth, J. Malcolm O. Arnold, Tom Ashton, Michel D'Astous, Paul Dorian, Haissam Haddad, Debra L. Isaac, Marie-Hélène Leblanc, Peter Liu, Bruce Sussex, Heather J. Ross. (2011) The 2011 Canadian Cardiovascular Society Heart Failure Management Guidelines Update: Focus on Sleep Apnea, Renal Dysfunction, Mechanical Circulatory Support, and Palliative Care. Canadian Journal of Cardiology 27:3, 319-338
    CrossRef

  66. 66

    Ana Garcia-Alvarez, Leticia Fernandez-Friera, Joe F. Lau, Simonette T. Sawit, Jesus G. Mirelis, Javier G. Castillo, Sean Pinney, Anelechi C. Anyanwu, Valentin Fuster, Javier Sanz, Mario J. Garcia. (2011) Evaluation of right ventricular function and post-operative findings using cardiac computed tomography in patients with left ventricular assist devices. The Journal of Heart and Lung Transplantation
    CrossRef

  67. 67

    Melanie D. Everitt, Amy E. Donaldson, Josef Stehlik, Aditya K. Kaza, Deborah Budge, Rami Alharethi, Emily A. Bullock, Abdallah G. Kfoury, Anji T. Yetman. (2011) Would access to device therapies improve transplant outcomes for adults with congenital heart disease? Analysis of the United Network for Organ Sharing (UNOS). The Journal of Heart and Lung Transplantation 30:4, 395-401
    CrossRef

  68. 68

    Shunsuke Saito, Taichi Sakaguchi, Shigeru Miyagawa, Yasushi Yoshikawa, Takashi Yamauchi, Takayoshi Ueno, Toru Kuratani, Yoshiki Sawa. (2011) Biventricular support using implantable continuous-flow ventricular assist devices. The Journal of Heart and Lung Transplantation 30:4, 475-478
    CrossRef

  69. 69

    I. M. Hamour, A. Khaghani, P. K. Kanagala, A. G. Mitchell, N. R. Banner. (2011) Current outcome of heart transplantation: a 10-year single centre perspective and review. QJM 104:4, 335-343
    CrossRef

  70. 70

    Andrew J. Boyle, Deborah D. Ascheim, Mark J. Russo, Robert L. Kormos, Ranjit John, Yoshifumi Naka, Annetine C. Gelijns, Kimberly N. Hong, Jeffrey J. Teuteberg. (2011) Clinical outcomes for continuous-flow left ventricular assist device patients stratified by pre-operative INTERMACS classification. The Journal of Heart and Lung Transplantation 30:4, 402-407
    CrossRef

  71. 71

    T. Yagdi, E. Oguz, F. Ayik, S. Ertugay, S. Nalbantgil, C. Engin, M. Ozbaran. (2011) Ventricular Assist System Applications in End-stage Heart Failure. Transplantation Proceedings 43:3, 923-926
    CrossRef

  72. 72

    Jay Baumwol, Peter S. Macdonald, Anne M. Keogh, Eugene Kotlyar, Phillip Spratt, Paul Jansz, Christopher S. Hayward. (2011) Right heart failure and “failure to thrive” after left ventricular assist device: Clinical predictors and outcomes. The Journal of Heart and Lung Transplantation
    CrossRef

  73. 73

    Salpy V. Pamboukian. (2011) Mechanical Circulatory Support. Journal of the American College of Cardiology 57:12, 1383-1385
    CrossRef

  74. 74

    Justin M. Schaffer, George J. Arnaoutakis, Jeremiah G. Allen, Eric S. Weiss, Nishant D. Patel, Stuart D. Russell, Ashish S. Shah, John V. Conte. (2011) Bleeding Complications and Blood Product Utilization With Left Ventricular Assist Device Implantation. The Annals of Thoracic Surgery 91:3, 740-749
    CrossRef

  75. 75

    Jennifer Cowger, Matthew A Romano, John Stulak, Francis D Pagani, Keith D Aaronson. (2011) Left ventricular assist device management in patients chronically supported for advanced heart failure. Current Opinion in Cardiology 26:2, 149-154
    CrossRef

  76. 76

    Keyur B. Shah, Daniel G. Tang, Richard H. Cooke, Suzanne Harton, Maureen Flattery, Gundars J. Katlaps, Vigneshwar Kasirajan, Michael L. Hess. (2011) Implantable Mechanical Circulatory Support: Demystifying Patients With Ventricular Assist Devices and Artificial Hearts. Clinical Cardiology 34:3, 147-152
    CrossRef

  77. 77

    Yan Topilsky, Jae K. Oh, Dipesh K. Shah, Barry A. Boilson, John A. Schirger, Sudhir S. Kushwaha, Naveen L. Pereira, Soon J. Park. (2011) Echocardiographic Predictors of Adverse Outcomes After Continuous Left Ventricular Assist Device Implantation. JACC: Cardiovascular Imaging 4:3, 211-222
    CrossRef

  78. 78

    David H. Adams, Joanna Chikwe, Farzan Filsoufi, Anelechi C. Anyanwu. (2011) The Year in Cardiovascular Surgery. Journal of the American College of Cardiology 57:13, 1425-1444
    CrossRef

  79. 79

    Gautam V. Ramani, Mandeep R. Mehra. (2011) Evaluación del riesgo en pacientes en estado crítico a la espera de trasplante: un paso adelante. Revista Española de Cardiología 64:3, 175-176
    CrossRef

  80. 80

    Krishna K. Gaddam, Tara Jarreau, Hector Ventura. (2011) Developments in Heart Failure 2010. Congestive Heart Failure 17:2, 92-100
    CrossRef

  81. 81

    William S. Pierce. (2011) An Odyssey in Mechanical Circulatory Support. Artificial Organs 35:2, 102-104
    CrossRef

  82. 82

    Yan Topilsky, Jae K. Oh, Fawn W. Atchison, Dipesh K. Shah, Valentina M. Bichara, John A. Schirger, Sudhir S. Kushwaha, Naveen L. Pereira, Soon J. Park. (2011) Echocardiographic Findings in Stable Outpatients with Properly Functioning HeartMate II Left Ventricular Assist Devices. Journal of the American Society of Echocardiography 24:2, 157-169
    CrossRef

  83. 83

    Ravi Rasalingam, Stephanie N. Johnson, Kyle R. Bilhorn, Pei-Hsiu Huang, Majesh Makan, Nader Moazami, Julio E. Pérez. (2011) Transthoracic Echocardiographic Assessment of Continuous-Flow Left Ventricular Assist Devices. Journal of the American Society of Echocardiography 24:2, 135-148
    CrossRef

  84. 84

    Jennifer L. Hall, David R. Fermin, Emma J. Birks, Paul J.R. Barton, Mark Slaughter, Peter Eckman, Hideo A. Baba, Jeremias Wohlschlaeger, Leslie W. Miller. (2011) Clinical, Molecular, and Genomic Changes in Response to a Left Ventricular Assist Device. Journal of the American College of Cardiology 57:6, 641-652
    CrossRef

  85. 85

    Robert M. Weiss, Richard E. Kerber, Jennifer L. Goerbig-Campbell, Melissa K. Davis, Barry M. Cabuay, Guha Ashrith, Wassef Karrowni, James E. Davis, Frances L. Johnson. (2011) The Impact of Prolonged Rotary Ventricular Assist Device Support upon Ventricular Geometry and Flow Kinetics. Journal of the American Society of Echocardiography 24:2, 149-156
    CrossRef

  86. 86

    N. Uriel, Y. Naka, P. C. Colombo, M. Farr, S.-W. Pak, V. Cotarlan, J. B. Albu, D. Gallagher, D. Mancini, H. N. Ginsberg, U. P. Jorde. (2011) Improved diabetic control in advanced heart failure patients treated with left ventricular assist devices. European Journal of Heart Failure 13:2, 195-199
    CrossRef

  87. 87

    Koichiro Kinugawa. (2011) How to Treat Stage D Heart Failure?. Circulation Journal 75:9, 2038-2045
    CrossRef

  88. 88

    Sofie Gevaert, Yves Van Belleghem, Stefaan Bouchez, Ingrid Herck, Filip De Somer, Yasmina De Block, Fiona Tromp, Els Vandecasteele, Floor Martens, Michel De Pauw. (2011) Acute and critically ill peripartum cardiomyopathy and 'bridge to' therapeutic options: a single center experience with intra-aortic balloon pump, extra corporeal membrane oxygenation and continuous-flow left ventricular assist devices. Critical Care 15:2, R93
    CrossRef

  89. 89

    Farooq H Sheikh, Stuart D Russell. (2011) HeartMate ® II continuous-flow left ventricular assist system. Expert Review of Medical Devices 8:1, 11-21
    CrossRef

  90. 90

    Yan Topilsky, Simon Maltais, Jae K. Oh, Fawn W. Atchison, Louis P. Perrault, Michel Carrier, Soon J. Park. (2011) Focused Review on Transthoracic Echocardiographic Assessment of Patients with Continuous Axial Left Ventricular Assist Devices. Cardiology Research and Practice 2011, 1-11
    CrossRef

  91. 91

    A. Kumar, P.S. Khanwilkar. 2011. Long-Term Implantable Ventricular Assist Devices (VADs) and Total Artificial Hearts (TAHs). , 389-402.
    CrossRef

  92. 92

    Jens Garbade, Hartmuth B. Bittner, Markus J. Barten, Friedrich-Wilhelm Mohr. (2011) Current Trends in Implantable Left Ventricular Assist Devices. Cardiology Research and Practice 2011, 1-9
    CrossRef

  93. 93

    Ranjit John, Katie Mantz, Peter Eckman, Alan Rose, Karen May-Newman. (2010) Aortic valve pathophysiology during left ventricular assist device support. The Journal of Heart and Lung Transplantation 29:12, 1321-1329
    CrossRef

  94. 94

    Mark S. Slaughter. (2010) Hematologic Effects of Continuous Flow Left Ventricular Assist Devices. Journal of Cardiovascular Translational Research 3:6, 618-624
    CrossRef

  95. 95

    N. Boudou, B. Marcheix, C. Biendel-Piquet, N. Dumonteil, T. Lhermusier, H. Benhaoua, D. Carrié. (2010) Assistances circulatoires dans les chocs cardiogéniques. Annales de Cardiologie et d'Angéiologie 59:6, 362-366
    CrossRef

  96. 96

    Ranjit John, Sangjin Lee, Peter Eckman, Kenneth Liao. (2010) Right Ventricular Failure—A Continuing Problem in Patients with Left Ventricular Assist Device Support. Journal of Cardiovascular Translational Research 3:6, 604-611
    CrossRef

  97. 97

    Satya S. Shreenivas, J. Eduardo Rame, Mariell Jessup. (2010) Mechanical Circulatory Support as a Bridge to Transplant or for Destination Therapy. Current Heart Failure Reports 7:4, 159-166
    CrossRef

  98. 98

    Chetan B. Patel, Kevin M. Alexander, Joseph G. Rogers. (2010) Mechanical Circulatory Support for Advanced Heart Failure. Current Treatment Options in Cardiovascular Medicine 12:6, 549-565
    CrossRef

  99. 99

    Keith A Thompson, Parag Bharadwaj, Kiran J Philip, Ernst R Schwarz. (2010) Heart failure therapy: beyond the guidelines. Journal of Cardiovascular Medicine 11:12, 919-927
    CrossRef

  100. 100

    Patrick Cervini, Soon J. Park, Dipesh K. Shah, Irina E. Penev, Bradley D. Lewis. (2010) Carotid Doppler Ultrasound Findings in Patients With Left Ventricular Assist Devices. Ultrasound Quarterly 26:4, 255-261
    CrossRef

  101. 101

    David J. Ficke, Justin Lee, Mark A. Chaney, Heidi Bas, Marcos F. Vidal-Melo, Marc E. Stone. (2010) Case 6—2010 Noncardiac Surgery in Patients With a Left Ventricular Assist Device. Journal of Cardiothoracic and Vascular Anesthesia 24:6, 1002-1009
    CrossRef

  102. 102

    George V. Letsou, Thomas D. Pate, Jeffrey R. Gohean, Mark Kurusz, Raul G. Longoria, Larry Kaiser, Richard W. Smalling. (2010) Improved left ventricular unloading and circulatory support with synchronized pulsatile left ventricular assistance compared with continuous-flow left ventricular assistance in an acute porcine left ventricular failure model. The Journal of Thoracic and Cardiovascular Surgery 140:5, 1181-1188
    CrossRef

  103. 103

    Georg M. Wieselthaler, Gerry O′Driscoll, Paul Jansz, Asghar Khaghani, Martin Strueber. (2010) Initial clinical experience with a novel left ventricular assist device with a magnetically levitated rotor in a multi-institutional trial. The Journal of Heart and Lung Transplantation 29:11, 1218-1225
    CrossRef

  104. 104

    Ronald Freudenberger. 2010. Management of Advanced Heart Failure. , 357-383.
    CrossRef

  105. 105

    Veli K. Topkara, Sreekanth Kondareddy, Fardina Malik, I.-Wen Wang, Douglas L. Mann, Gregory A. Ewald, Nader Moazami. (2010) Infectious Complications in Patients With Left Ventricular Assist Device: Etiology and Outcomes in the Continuous-Flow Era. The Annals of Thoracic Surgery 90:4, 1270-1277
    CrossRef

  106. 106

    John Lynn Jefferies, Jack F. Price, David L.S. Morales. (2010) Mechanical Support in Childhood Heart Failure. Heart Failure Clinics 6:4, 559-573
    CrossRef

  107. 107

    Sang-Woo Pak, Nir Uriel, Hiroo Takayama, Sarah Cappleman, Robert Song, Paolo C. Colombo, Sandy Charles, Donna Mancini, Linda Gillam, Yoshifumi Naka, Ulrich P. Jorde. (2010) Prevalence of de novo aortic insufficiency during long-term support with left ventricular assist devices. The Journal of Heart and Lung Transplantation 29:10, 1172-1176
    CrossRef

  108. 108

    Nir Uriel, Sang-Woo Pak, Ulrich P. Jorde, Brigitte Jude, Sophie Susen, Andre Vincentelli, Pierre-Vladimir Ennezat, Sarah Cappleman, Yoshifumi Naka, Donna Mancini. (2010) Acquired von Willebrand Syndrome After Continuous-Flow Mechanical Device Support Contributes to a High Prevalence of Bleeding During Long-Term Support and at the Time of Transplantation. Journal of the American College of Cardiology 56:15, 1207-1213
    CrossRef

  109. 109

    Leslie W. Miller. (2010) The Development of the von Willebrand Syndrome With the Use of Continuous Flow Left Ventricular Assist Devices. Journal of the American College of Cardiology 56:15, 1214-1215
    CrossRef

  110. 110

    Jerry D. Estep, Raymond F. Stainback, Stephen H. Little, Guillermo Torre, William A. Zoghbi. (2010) The Role of Echocardiography and Other Imaging Modalities in Patients With Left Ventricular Assist Devices. JACC: Cardiovascular Imaging 3:10, 1049-1064
    CrossRef

  111. 111

    Matthew A. Romano, Jennifer Cowger, Keith D. Aaronson, Francis D. Pagani. (2010) Diagnosis and Management of Right-Sided Heart Failure in Subjects Supported With Left Ventricular Assist Devices. Current Treatment Options in Cardiovascular Medicine 12:5, 420-430
    CrossRef

  112. 112

    Andreas Arndt, Peter Nüsser, Bernhard Lampe. (2010) Fully Autonomous Preload-Sensitive Control of Implantable Rotary Blood Pumps. Artificial Organs 34:9, 726-735
    CrossRef

  113. 113

    George S. Chrysant, Douglas A. Horstmanshof, Trevor Snyder, John S. Chaffin, C. Craig Elkins, Paul J. Kanaly, James W. Long. (2010) Successful Percutaneous Management of Acute Left Ventricular Assist Device Stoppage. ASAIO Journal 56:5, 483-485
    CrossRef

  114. 114

    Elizabeth A. Genovese, Mary Amanda Dew, Jeffrey J. Teuteberg, Marc A. Simon, Jay K. Bhama, Christian A. Bermudez, Kathleen L. Lockard, Steve Winowich, Robert L. Kormos. (2010) Early adverse events as predictors of 1-year mortality during mechanical circulatory support. The Journal of Heart and Lung Transplantation 29:9, 981-988
    CrossRef

  115. 115

    Takeshi Komoda, Thorsten Drews, Roland Hetzer, Hans B. Lehmkuhl. (2010) New prioritization of heart transplant candidates on mechanical circulatory support in an era of severe donor shortage. The Journal of Heart and Lung Transplantation 29:9, 989-996
    CrossRef

  116. 116

    T. Drews, A. Stepanenko, M. Dandel, S. Buz, H. B. Lehmkuhl, R. Hetzer. (2010) Mechanical circulatory support in patients of advanced age. European Journal of Heart Failure 12:9, 990-994
    CrossRef

  117. 117

    Eric S. Ketchum, Alec J. Moorman, Daniel P. Fishbein, Nahush A. Mokadam, Edward D. Verrier, Gabriel S. Aldea, Shauna Andrus, Kenneth W. Kenyon, Wayne C. Levy. (2010) Predictive value of the Seattle Heart Failure Model in patients undergoing left ventricular assist device placement. The Journal of Heart and Lung Transplantation 29:9, 1021-1025
    CrossRef

  118. 118

    Sheri Crow, Carmelo Milano, Lyle Joyce, Dong Chen, Gowthami Arepally, Dawn Bowles, William Thomas, Nestor Villamizar Ortiz. (2010) Comparative Analysis of von Willebrand Factor Profiles in Pulsatile and Continuous Left Ventricular Assist Device Recipients. ASAIO Journal 56:5, 441-445
    CrossRef

  119. 119

    Ranjit John, Kenneth Liao, Forum Kamdar, Peter Eckman, Andrew Boyle, Monica Colvin-Adams. (2010) Effects on pre- and posttransplant pulmonary hemodynamics in patients with continuous-flow left ventricular assist devices. The Journal of Thoracic and Cardiovascular Surgery 140:2, 447-452
    CrossRef

  120. 120

    Christopher A. Thunberg, Brantley Dollar Gaitan, Francisco A. Arabia, Daniel J. Cole, Alina M. Grigore. (2010) Ventricular Assist Devices Today and Tomorrow. Journal of Cardiothoracic and Vascular Anesthesia 24:4, 656-680
    CrossRef

  121. 121

    Anson Cheung, Jamil Bashir, Annemarie Kaan, Jennifer Kealy, Robert Moss, Hossein Shayan. (2010) Minimally invasive, off-pump explant of a continuous-flow left ventricular assist device. The Journal of Heart and Lung Transplantation 29:7, 808-810
    CrossRef

  122. 122

    Katherine Lietz. (2010) Destination Therapy: Patient Selection and Current Outcomes. Journal of Cardiac Surgery 25:4, 462-471
    CrossRef

  123. 123

    Sanjeev Aggarwal, Francis D. Pagani. (2010) Bridge to Transplantation: Current Outcomes. Journal of Cardiac Surgery 25:4, 455-461
    CrossRef

  124. 124

    William L. Holman, Salpy V. Pamboukian, David C. McGiffin, Jose A. Tallaj, Martin Cadeiras, James K. Kirklin. (2010) Device Related Infections: Are We Making Progress?. Journal of Cardiac Surgery 25:4, 478-483
    CrossRef

  125. 125

    Ranjit John, Francis D. Pagani, Yoshifumi Naka, Andrew Boyle, John V. Conte, Stuart D. Russell, Charles T. Klodell, Carmelo A. Milano, Joseph Rogers, David J. Farrar, O. Howard Frazier. (2010) Post–cardiac transplant survival after support with a continuous-flow left ventricular assist device: Impact of duration of left ventricular assist device support and other variables. The Journal of Thoracic and Cardiovascular Surgery 140:1, 174-181
    CrossRef

  126. 126

    Duc Q. Nguyen, Vinod H. Thourani. (2010) Third-Generation Continuous Flow Left Ventricular Assist Devices. Innovations: Technology and Techniques in Cardiothoracic and Vascular Surgery 5:4, 250-258
    CrossRef

  127. 127

    Andrew J. Boyle. (2010) Are Ventricular Assist Devices Underutilized?. Journal of Cardiac Surgery 25:4, 421-424
    CrossRef

  128. 128

    Amrut V. Ambardekar, Larry A. Allen, JoAnn Lindenfeld, Christopher M. Lowery, Anne P. Cannon, Joseph C. Cleveland, Andreas Brieke, William H. Sauer. (2010) Implantable cardioverter-defibrillator shocks in patients with a left ventricular assist device. The Journal of Heart and Lung Transplantation 29:7, 771-776
    CrossRef

  129. 129

    Gary S. Francis, Barry H. Greenberg, Daphne T. Hsu, Brian E. Jaski, Mariell Jessup, Martin M. LeWinter, Francis D. Pagani, Ileana L. Piña, Marc J. Semigran, Mary Norine Walsh, David H. Wiener, Clyde W. Yancy. (2010) ACCF/AHA/ACP/HFSA/ISHLT 2010 Clinical Competence Statement on Management of Patients With Advanced Heart Failure and Cardiac Transplant. Journal of the American College of Cardiology 56:5, 424-453
    CrossRef

  130. 130

    Mark S. Slaughter, Yoshifumi Naka, Ranjit John, Andrew Boyle, John V. Conte, Stuart D. Russell, Keith D. Aaronson, Kartik S. Sundareswaran, David J. Farrar, Francis D. Pagani. (2010) Post-operative heparin may not be required for transitioning patients with a HeartMate II left ventricular assist system to long-term warfarin therapy. The Journal of Heart and Lung Transplantation 29:6, 616-624
    CrossRef

  131. 131

    William L. Holman, James K. Kirklin, David C. Naftel, Robert L. Kormos, Patricia Desvign-Nickens, Margarita T. Camacho, Deborah D. Ascheim. (2010) Infection after implantation of pulsatile mechanical circulatory support devices. The Journal of Thoracic and Cardiovascular Surgery 139:6, 1632-1636.e2
    CrossRef

  132. 132

    Heart Failure Society of America. (2010) Section 10: Surgical Approaches to the Treatment of Heart Failure. Journal of Cardiac Failure 16:6, e122-e125
    CrossRef

  133. 133

    Magdi H. Yacoub. (2010) Intervention: LVADs—from an imaginative leap to clinical decision making. Nature Reviews Cardiology 7:5, 246-248
    CrossRef

  134. 134

    Andreas Arndt, Peter Nüsser, Bernhard P. Lampe. (2010) Regelung von rotierenden Blutpumpen zur Linksherzunterstützung. at - Automatisierungstechnik 58:05, 241-250
    CrossRef

  135. 135

    Robert L. Kormos, Jeffrey J. Teuteberg, Francis D. Pagani, Stuart D. Russell, Ranjit John, Leslie W. Miller, Todd Massey, Carmelo A. Milano, Nader Moazami, Kartik S. Sundareswaran, David J. Farrar. (2010) Right ventricular failure in patients with the HeartMate II continuous-flow left ventricular assist device: Incidence, risk factors, and effect on outcomes. The Journal of Thoracic and Cardiovascular Surgery 139:5, 1316-1324
    CrossRef

  136. 136

    Mosi K. Bennett, Celine A. Roberts, Dzifa Dordunoo, Ashish Shah, Stuart D. Russell. (2010) Ideal methodology to assess systemic blood pressure in patients with continuous-flow left ventricular assist devices. The Journal of Heart and Lung Transplantation 29:5, 593-594
    CrossRef

  137. 137

    David R. Stern, Jacob Kazam, Pauline Edwards, Simon Maybaum, Ricardo A. Bello, David A. D’Alessandro, Daniel J. Goldstein. (2010) Increased Incidence of Gastrointestinal Bleeding Following Implantation of the HeartMate II LVAD. Journal of Cardiac Surgery 25:3, 352-356
    CrossRef

  138. 138

    L. H. Lund, J. Matthews, K. Aaronson. (2010) Patient selection for left ventricular assist devices. European Journal of Heart Failure 12:5, 434-443
    CrossRef

  139. 139

    Mark S. Slaughter, Francis D. Pagani, Joseph G. Rogers, Leslie W. Miller, Benjamin Sun, Stuart D. Russell, Randall C. Starling, Leway Chen, Andrew J. Boyle, Suzanne Chillcott, Robert M. Adamson, Margaret S. Blood, Margarita T. Camacho, Katherine A. Idrissi, Michael Petty, Michael Sobieski, Susan Wright, Timothy J. Myers, David J. Farrar. (2010) Clinical management of continuous-flow left ventricular assist devices in advanced heart failure. The Journal of Heart and Lung Transplantation 29:4, S1-S39
    CrossRef

  140. 140

    Akshay Desai, Frederic S. Resnic. (2010) Highlights from American Heart Association Annual Scientific Sessions 2009: Orlando, Florida, November 14–18, 2009. The Journal of Thoracic and Cardiovascular Surgery 139:4, 1090-1092
    CrossRef

  141. 141

    Randall C. Starling. (2010) Improved Quantity and Quality of Life. Journal of the American College of Cardiology 55:17, 1835-1836
    CrossRef

  142. 142

    M. R. Johnson, K. H. Meyer, J. Haft, D. Kinder, S. A. Webber, D. B. Dyke. (2010) Heart Transplantation in the United States, 1999-2008. American Journal of Transplantation 10:4p2, 1035-1046
    CrossRef

  143. 143

    Barry Cabuay, Katarzyna Hryniewicz, Benjamin Sun, David Feldman. (2010) Door-to-VAD time: an expedient management strategy for cardiogenic shock. Current Opinion in Cardiology1
    CrossRef

  144. 144

    Joseph G. Rogers, Keith D. Aaronson, Andrew J. Boyle, Stuart D. Russell, Carmelo A. Milano, Francis D. Pagani, Brooks S. Edwards, Soon Park, Ranjit John, John V. Conte, David J. Farrar, Mark S. Slaughter. (2010) Continuous Flow Left Ventricular Assist Device Improves Functional Capacity and Quality of Life of Advanced Heart Failure Patients. Journal of the American College of Cardiology 55:17, 1826-1834
    CrossRef

  145. 145

    Amrut V. Ambardekar, Christopher M. Lowery, Larry A. Allen, Anne P. Cannon, Joseph C. Cleveland, Joann Lindenfeld, Andreas Brieke, William H. Sauer. (2010) Effect of Left Ventricular Assist Device Placement on Preexisting Implantable Cardioverter-defibrillator Leads. Journal of Cardiac Failure 16:4, 327-331
    CrossRef

  146. 146

    Michiel Morshuis, Michael Schoenbrodt, Chisato Nojiri, Daniela Roefe, Sebastian Schulte-Eistrup, Jochen Boergermann, Jan F Gummert, Latif Arusoglu. (2010) DuraHeart™ magnetically levitated centrifugal left ventricular assist system for advanced heart failure patients. Expert Review of Medical Devices 7:2, 173-183
    CrossRef

  147. 147

    Jeremiah G. Allen, Eric S. Weiss, Justin M. Schaffer, Nishant D. Patel, Susan L. Ullrich, Stuart D. Russell, Ashish S. Shah, John V. Conte. (2010) Quality of life and functional status in patients surviving 12 months after left ventricular assist device implantation. The Journal of Heart and Lung Transplantation 29:3, 278-285
    CrossRef

  148. 148

    Cesare M. Terracciano, Leslie W. Miller, Magdi H. Yacoub. (2010) Contemporary Use of Ventricular Assist Devices. Annual Review of Medicine 61:1, 255-270
    CrossRef

  149. 149

    Rajan Krishnamani, David DeNofrio, Marvin A. Konstam. (2010) Emerging ventricular assist devices for long-term cardiac support. Nature Reviews Cardiology 7:2, 71-76
    CrossRef

  150. 150

    Marius Berman, Jayan Parameshwar, David P. Jenkins, Kumud Dhital, Clive Lewis, Kirsty Dempster, Paul Lincoln, Catherine Sudarshan, Stephen R. Large, John Dunning, Steven S.L. Tsui. (2010) Thoratec implantable ventricular assist device: The Papworth experience. The Journal of Thoracic and Cardiovascular Surgery 139:2, 466-473
    CrossRef

  151. 151

    Sangjin Lee, Forum Kamdar, Richard Madlon-Kay, Andrew Boyle, Monica Colvin-Adams, Marc Pritzker, Ranjit John. (2010) Effects of the HeartMate II continuous-flow left ventricular assist device on right ventricular function. The Journal of Heart and Lung Transplantation 29:2, 209-215
    CrossRef

  152. 152

    Michael Kühne, Michaela Sakumura, Stephen Scott Reich, Jean-Francois Sarrazin, Darryl Wells, Nagib Chalfoun, Thomas Crawford, Warangkna Boonyapisit, Laura Horwood, Aman Chugh, Eric Good, Krit Jongnarangsin, Frank Bogun, Hakan Oral, Fred Morady, Francis Pagani, Frank Pelosi. (2010) Simultaneous Use of Implantable Cardioverter-Defibrillators and Left Ventricular Assist Devices in Patients With Severe Heart Failure. The American Journal of Cardiology 105:3, 378-382
    CrossRef

  153. 153

    Nanhi Mitter, Rosanne Sheinberg. (2010) Update on ventricular assist devices. Current Opinion in Anaesthesiology 23:1, 57-66
    CrossRef

  154. 154

    Kiyotaka Fukamachi, David J. Horvath, Alex L. Massiello, Hideyuki Fumoto, Tetsuya Horai, Santosh Rao, Leonard A.R. Golding. (2010) An innovative, sensorless, pulsatile, continuous-flow total artificial heart: Device design and initial in vitro study. The Journal of Heart and Lung Transplantation 29:1, 13-20
    CrossRef

  155. 155

    Ashley L. Raymond, Abdallah G. Kfoury, Corey J. Bishop, Erin S. Davis, Kimberly M. Goebel, Sandi Stoker, Craig H. Selzman, Stephen E. Clayson, Hildegard Smith, Cris G. Cowley, Rami Alharethi, Deborah Budge, Bruce B. Reid. (2010) Obesity and Left Ventricular Assist Device Driveline Exit Site Infection. ASAIO Journal 56:1, 57-60
    CrossRef

  156. 156

    L. Henry Edmunds. (2010) Managing Fibrinolysis Without Aprotinin. The Annals of Thoracic Surgery 89:1, 324-331
    CrossRef

  157. 157

    Valluvan Jeevanandam. (2010) The evolution of cardiac assist device technology. The Journal of Heart and Lung Transplantation 29:1, 11-12
    CrossRef

  158. 158

    Slaughter, Mark S., Rogers, Joseph G., Milano, Carmelo A., Russell, Stuart D., Conte, John V., Feldman, David, Sun, Benjamin, Tatooles, Antone J., Delgado, Reynolds M. III, Long, James W., Wozniak, Thomas C., Ghumman, Waqas, Farrar, David J., Frazier, O. Howard, . (2009) Advanced Heart Failure Treated with Continuous-Flow Left Ventricular Assist Device. New England Journal of Medicine 361:23, 2241-2251
    Full Text

  159. 159

    Fang, James C., . (2009) Rise of the Machines — Left Ventricular Assist Devices as Permanent Therapy for Advanced Heart Failure. New England Journal of Medicine 361:23, 2282-2285
    Full Text

  160. 160

    Mark J. Russo, Kimberly N. Hong, Ryan R. Davies, Jonathan M. Chen, Robert A. Sorabella, Deborah D. Ascheim, Mathew R. Williams, Annetine C. Gelijns, Allan S. Stewart, Michael Argenziano, Yoshifumi Naka. (2009) Posttransplant survival is not diminished in heart transplant recipients bridged with implantable left ventricular assist devices. The Journal of Thoracic and Cardiovascular Surgery 138:6, 1425-1432.e3
    CrossRef

  161. 161

    Vegard Tuseth, Jan Erik Nordrehaug. (2009) Role of percutaneous left ventricular assist devices in preventing cerebral ischemia. Interventional Cardiology 1:2, 197-208
    CrossRef

  162. 162

    Jay D. Pal, Valentino Piacentino, Angela D. Cuevas, Tim Depp, Mani A. Daneshmand, Adrian F. Hernandez, G. Michael Felker, Andrew J. Lodge, Joseph G. Rogers, Carmelo A. Milano. (2009) Impact of Left Ventricular Assist Device Bridging on Posttransplant Outcomes. The Annals of Thoracic Surgery 88:5, 1457-1461
    CrossRef

  163. 163

    Jane MacIver, Heather J. Ross, Diego H. Delgado, Robert J. Cusimano, Terrence M. Yau, Marnie Rodger, Stephen Harwood, Vivek Rao. (2009) Community support of patients with a left ventricular assist device: The Toronto General Hospital experience. Canadian Journal of Cardiology 25:11, e377-e381
    CrossRef

  164. 164

    Sean R. Wilson, Michael M. Givertz, Garrick C. Stewart, Gilbert H. Mudge. (2009) Ventricular Assist Devices. Journal of the American College of Cardiology 54:18, 1647-1659
    CrossRef

  165. 165

    Elizabeth A. Genovese, Mary Amanda Dew, Jeffrey J. Teuteberg, Marc A. Simon, Joy Kay, Michael P. Siegenthaler, Jay K. Bhama, Christian A. Bermudez, Kathleen L. Lockard, Steve Winowich, Robert L. Kormos. (2009) Incidence and Patterns of Adverse Event Onset During the First 60 Days After Ventricular Assist Device Implantation. The Annals of Thoracic Surgery 88:4, 1162-1170
    CrossRef

  166. 166

    JR Lahpor. (2009) State of the art: implantable ventricular assist devices. Current Opinion in Organ Transplantation 14:5, 554-559
    CrossRef

  167. 167

    Garrick C. Stewart, Kimberly Brooks, Parakash P. Pratibhu, Sui W. Tsang, Marc J. Semigran, Colleen M. Smith, Catherine Saniuk, Janice M. Camuso, James C. Fang, Gilbert H. Mudge, Gregory S. Couper, Kenneth L. Baughman, Lynne W. Stevenson. (2009) Thresholds of Physical Activity and Life Expectancy for Patients Considering Destination Ventricular Assist Devices. The Journal of Heart and Lung Transplantation 28:9, 863-869
    CrossRef

  168. 168

    Andrew J. Boyle, Stuart D. Russell, Jeffrey J. Teuteberg, Mark S. Slaughter, Nader Moazami, Francis D. Pagani, O. Howard Frazier, Gerald Heatley, David J. Farrar, Ranjit John. (2009) Low Thromboembolism and Pump Thrombosis With the HeartMate II Left Ventricular Assist Device: Analysis of Outpatient Anti-coagulation. The Journal of Heart and Lung Transplantation 28:9, 881-887
    CrossRef

  169. 169

    Kaitlyn My-Tu Lam, Stephanie Ennis, Gerry O'Driscoll, Jorge M. Solis, Thomas MacGillivray, Michael H. Picard. (2009) Observations From Non-Invasive Measures of Right Heart Hemodynamics in Left Ventricular Assist Device Patients. Journal of the American Society of Echocardiography 22:9, 1055-1062
    CrossRef

  170. 170

    Jonathan P. Martin, Jeremiah G. Allen, Eric S. Weiss, Luca A. Vricella, Stuart D. Russell, John V. Conte. (2009) Glenn Shunt Facilitated Weaning of Right Ventricular Mechanical Support. The Annals of Thoracic Surgery 88:3, e16-e17
    CrossRef

  171. 171

    Takeshi Komoda, Roland Hetzer, Hans B. Lehmkuhl. (2009) Who Should be Delisted from Urgent Heart Transplantation in Germany?. ASAIO Journal 55:5, 452-455
    CrossRef

  172. 172

    Corstiaan A. den Uil, Alexander P. Maat, Wim K. Lagrand, Martin van der Ent, Lucia S.D. Jewbali, Robert J. van Thiel, Peter E. Spronk, Maarten L. Simoons. (2009) Mechanical Circulatory Support Devices Improve Tissue Perfusion in Patients With End-Stage Heart Failure or Cardiogenic Shock. The Journal of Heart and Lung Transplantation 28:9, 906-911
    CrossRef

  173. 173

    Till Neumann, Georg Aidonides, Thomas Konorza, Peter Krings, Raimund Erbel. (2009) Neurohumoral response and clinical effectiveness of continuous aortic flow augmentation in patients with decompensated heart failure. Journal of Artificial Organs 12:3, 166-171
    CrossRef

  174. 174

    Jonathan W. Haft, Francis D. Pagani, Matthew A. Romano, Christina L. Leventhal, D. Bradley Dyke, Jennifer C. Matthews. (2009) Short- and Long-Term Survival of Patients Transferred to a Tertiary Care Center on Temporary Extracorporeal Circulatory Support. The Annals of Thoracic Surgery 88:3, 711-718
    CrossRef

  175. 175

    Mehdi Jafar, Igor D. Gregoric, Rajko Radovancevic, William E. Cohn, Nichole McGuire, O H. Frazier. (2009) Urgent Exchange of a HeartMate II Left Ventricular Assist Device After Percutaneous Lead Fracture. ASAIO Journal 55:5, 523-524
    CrossRef

  176. 176

    Ana C Alba, Diego H Delgado. (2009) The future is here: ventricular assist devices for the failing heart. Expert Review of Cardiovascular Therapy 7:9, 1067-1077
    CrossRef

  177. 177

    Deborah Meyers. (2009) Medical therapies for the management of cardiomyopathy and chronic congestive heart failure. Current Cardiovascular Risk Reports 3:5, 315-322
    CrossRef

  178. 178

    Mark Harrison, Anelechi Anyanwu, Sean P. Pinney. (2009) The Management of Stage D Heart Failure. Mount Sinai Journal of Medicine: A Journal of Translational and Personalized Medicine 76:4, 404-414
    CrossRef

  179. 179

    Francis D. Pagani, Leslie W. Miller, Stuart D. Russell, Keith D. Aaronson, Ranjit John, Andrew J. Boyle, John V. Conte, Roberta C. Bogaev, Thomas E. MacGillivray, Yoshifumi Naka, Donna Mancini, H. Todd Massey, Leway Chen, Charles T. Klodell, Juan M. Aranda, Nader Moazami, Gregory A. Ewald, David J. Farrar, O. Howard Frazier. (2009) Extended Mechanical Circulatory Support With a Continuous-Flow Rotary Left Ventricular Assist Device. Journal of the American College of Cardiology 54:4, 312-321
    CrossRef

  180. 180

    Mads Andersen, Regitze Videbæk, Søren Boesgaard, Kåre Sander, Peter B. Hansen, Finn Gustafsson. (2009) Incidence of Ventricular Arrhythmias in Patients on Long-term Support With a Continuous-flow Assist Device (HeartMate II). The Journal of Heart and Lung Transplantation 28:7, 733-735
    CrossRef

  181. 181

    O. Gjesdal, E. Gude, S. Arora, T. Leivestad, A. K. Andreassen, L. Gullestad, L. Aaberge, H. Brunvand, T. Edvardsen, O. R. Geiran, S. Simonsen. (2009) Intra-aortic balloon counterpulsation as a bridge to heart transplantation does not impair long-term survival. European Journal of Heart Failure 11:7, 709-714
    CrossRef

  182. 182

    Shmuel Shoham, Leslie W. Miller. (2009) Cardiac assist device infections. Current Infectious Disease Reports 11:4, 268-273
    CrossRef

  183. 183

    Karl A. Soderlund, Raghu R. Chivukula, Stuart D. Russell, John V. Conte, James O. Mudd, Marc K. Halushka. (2009) Prognostic value of left ventricular apical tissue removed for HeartMate II left ventricular assist device placement. Cardiovascular Pathology 18:4, 217-222
    CrossRef

  184. 184

    Howard J. Eisen, Shelley R. Hankins. (2009) Continuous Flow Rotary Left Ventricular Assist Device. Journal of the American College of Cardiology 54:4, 322-324
    CrossRef

  185. 185

    DAVID FOO, BRUCE D. WALKER, DENNIS L. KUCHAR, CHARLES W. THORBURN, ANDRE TAY, CHRISTOPHER S. HAYWARD, PETER MACDONALD, ANNE KEOGH, EUGENE KOTLYAR, PHILIP SPRATT, PAUL JANSZ. (2009) Left Ventricular Mechanical Assist Devices and Cardiac Device Interactions: An Observational Case Series. Pacing and Clinical Electrophysiology 32:7, 879-887
    CrossRef

  186. 186

    Ethan D. Miller, D. Eric Steidley, Francisco A. Arabia, G. Anton Decker. (2009) Gastric erosion associated with left ventricular assist device: new technology, new complication. Gastrointestinal Endoscopy 70:1, 181-183
    CrossRef

  187. 187

    Daniel G. Tang, Philip E. Oyer, Hari R. Mallidi. (2009) Ventricular Assist Devices: History, Patient Selection, and Timing of Therapy. Journal of Cardiovascular Translational Research 2:2, 159-167
    CrossRef

  188. 188

    Ranjit John, Andrew Boyle, Frank Pagani, Leslie Miller. (2009) Physiologic and Pathologic Changes in Patients with Continuous-Flow Ventricular Assist Devices. Journal of Cardiovascular Translational Research 2:2, 154-158
    CrossRef

  189. 189

    Ralph J. Petrucci, Susan Wright, Yoshifuma Naka, Kathy A. Idrissi, Stuart D. Russell, Dzifa Dordunoo, Brian Jaski, Suzanne Chillcott, David Feldman, Tammy Yanssens, Gerald Heatley, Lalig Koundakjian, David J. Farrar, Keith D. Aaronson. (2009) Neurocognitive Assessments in Advanced Heart Failure Patients Receiving Continuous-flow Left Ventricular Assist Devices. The Journal of Heart and Lung Transplantation 28:6, 542-549
    CrossRef

  190. 190

    Gabriel Sayer, Yoshifumi Naka, Ulrich P. Jorde. (2009) Ventricular Assist Device Therapy. Cardiovascular Therapeutics 27:2, 140-150
    CrossRef

  191. 191

    Satoru Osaki, Niloo M. Edwards, Maryl R. Johnson, Mauricio Velez, Alejandro Munoz, Lucian Lozonschi, Margaret A. Murray, Amy K. Proebstle, Takushi Kohmoto. (2009) Improved Survival After Heart Transplantation in Patients With Bridge to Transplant in the Recent Era: A 17-year Single-center Experience. The Journal of Heart and Lung Transplantation 28:6, 591-597
    CrossRef

  192. 192

    Bart Meyns, Stefan Klotz, Andre Simon, Walter Droogne, Filip Rega, Bartley Griffith, Robert Dowling, Mark J. Zucker, Daniel Burkhoff. (2009) Proof of Concept. Journal of the American College of Cardiology 54:1, 79-86
    CrossRef

  193. 193

    Jolanta Klovaite, Finn Gustafsson, Svend A. Mortensen, Kåre Sander, Lars B. Nielsen. (2009) Severely Impaired von Willebrand Factor-Dependent Platelet Aggregation in Patients With a Continuous-Flow Left Ventricular Assist Device (HeartMate II). Journal of the American College of Cardiology 53:23, 2162-2167
    CrossRef

  194. 194

    Michiel Morshuis, Aly El-Banayosy, Latif Arusoglu, Reiner Koerfer, Roland Hetzer, Georg Wieselthaler, Alain Pavie, Chisato Nojiri. (2009) European experience of DuraHeart™ magnetically levitated centrifugal left ventricular assist system. European Journal of Cardio-Thoracic Surgery 35:6, 1020-1028
    CrossRef

  195. 195

    Timothy J. Myers, Mathius Bolmers, Igor D. Gregoric, Biswajit Kar, O.H. Howard Frazier. (2009) Assessment of Arterial Blood Pressure During Support With an Axial Flow Left Ventricular Assist Device. The Journal of Heart and Lung Transplantation 28:5, 423-427
    CrossRef

  196. 196

    Katherine Lietz, Leslie W Miller. (2009) Patient selection for left-ventricular assist devices. Current Opinion in Cardiology 24:3, 246-251
    CrossRef

  197. 197

    Maria M Mountis, Randall C Starling. (2009) Management of left ventricular assist devices after surgery: bridge, destination, and recovery. Current Opinion in Cardiology 24:3, 252-256
    CrossRef

  198. 198

    Micha T. Maeder, Angeline Leet, Andrew Ross, Donald Esmore, David M. Kaye. (2009) Changes in Right Ventricular Function During Continuous-low Left Ventricular Assist Device Support. The Journal of Heart and Lung Transplantation 28:4, 360-366
    CrossRef

  199. 199

    Forum Kamdar, Andrew Boyle, Kenneth Liao, Monica Colvin-adams, Lyle Joyce, Ranjit John. (2009) Effects of Centrifugal, Axial, and Pulsatile Left Ventricular Assist Device Support on End-Organ Function in Heart Failure Patients. The Journal of Heart and Lung Transplantation 28:4, 352-359
    CrossRef

  200. 200

    Sigrid E. Sandner, Daniel Zimpfer, Philipp Zrunek, Angela Rajek, Heinrich Schima, Daniela Dunkler, Michael Grimm, Ernst Wolner, Georg M. Wieselthaler. (2009) Renal Function and Outcome After Continuous Flow Left Ventricular Assist Device Implantation. The Annals of Thoracic Surgery 87:4, 1072-1078
    CrossRef

  201. 201

    Sigrid E. Sandner, Daniel Zimpfer, Philipp Zrunek, Angela Rajek, Heinrich Schima, Daniela Dunkler, Andreas O. Zuckermann, Georg M. Wieselthaler. (2009) Age and Outcome After Continuous-Flow Left Ventricular Assist Device Implantation as Bridge to Transplantation. The Journal of Heart and Lung Transplantation 28:4, 367-372
    CrossRef

  202. 202

    J. D. Vega, J. Moore, S. Murray, J. M. Chen, M. R. Johnson, D. B. Dyke. (2009) Heart Transplantation in the United States, 1998-2007. American Journal of Transplantation 9:4p2, 932-941
    CrossRef

  203. 203

    Takeshi Komoda, Hans B. Lehmkuhl, Brigitte Stiller, Felix Berger, Roland Hetzer. (2009) Pediatric Heart Transplant Candidates With Failed Donor Heart Allocation After Eurotransplant Urgency Listing Profit From Pretransplant Mechanical Circulatory Support Bridging. Artificial Organs 33:4, 346-351
    CrossRef

  204. 204

    Satoru Osaki, Nancy K. Sweitzer, Peter S. Rahko, Margaret A. Murray, Jennifer A. Hoffmann, Maryl R. Johnson, Niloo M. Edwards, Takushi Kohmoto. (2009) To Explant or Not to Explant: An Invasive and Noninvasive Monitoring Protocol to Determine the Need of Continued Ventricular Assist Device Support. Congestive Heart Failure 15:2, 58-62
    CrossRef

  205. 205

    Mitesh V Badiwala, Vivek Rao. (2009) Left ventricular device as destination therapy: are we there yet?. Current Opinion in Cardiology 24:2, 184-189
    CrossRef

  206. 206

    Wayne C. Levy, Dariush Mozaffarian, David T. Linker, David J. Farrar, Leslie W. Miller. (2009) Can the Seattle Heart Failure Model Be Used to Risk-stratify Heart Failure Patients for Potential Left Ventricular Assist Device Therapy?. The Journal of Heart and Lung Transplantation 28:3, 231-236
    CrossRef

  207. 207

    Matthias Antz, Bettina Hullmann, Christian Neufert, Wolfgang Vocke. (2009) Antikoagulation bei Vorhofflimmern – Update. Herz Kardiovaskuläre Erkrankungen 34:2, 140-140
    CrossRef

  208. 208

    Andrew Boyle. (2009) Current status of cardiac transplantation and mechanical circulatory support. Current Heart Failure Reports 6:1, 28-33
    CrossRef

  209. 209

    Sheri Crow, Ranjit John, Andrew Boyle, Sara Shumway, Kenneth Liao, Monica Colvin-Adams, Carol Toninato, Emil Missov, Marc Pritzker, Cindy Martin, Daniel Garry, William Thomas, Lyle Joyce. (2009) Gastrointestinal bleeding rates in recipients of nonpulsatile and pulsatile left ventricular assist devices. The Journal of Thoracic and Cardiovascular Surgery 137:1, 208-215
    CrossRef

  210. 210

    Antonis A. Pitsis, Aikaterini Visouli. (2009) Mechanical circulatory support in the ICCU. Acute Cardiac Care 11:4, 204-215
    CrossRef

  211. 211

    Hans Granfeldt, Bengt Peterzén, Laila Hübbert, Kjell Jansson, Henrik Ahn. (2009) A single center experience with the HeartMate II™ Left Ventricular Assist Device (LVAD). Scandinavian Cardiovascular Journal 43:6, 360-365
    CrossRef

  212. 212

    Arnt E. Fiane. (2009) Long-term left ventricle assist device (LVAD) therapy with rotary pumps. Scandinavian Cardiovascular Journal 43:6, 357-359
    CrossRef

  213. 213

    Vishnu Patlolla, Richard D. Patten, David DeNofrio, Marvin A. Konstam, Rajan Krishnamani. (2009) The Effect of Ventricular Assist Devices on Post-Transplant Mortality. Journal of the American College of Cardiology 53:3, 264-271
    CrossRef

  214. 214

    Leslie Miller. (2009) The Impact of Mechanical Circulatory Support on Post-Transplant Survival. Journal of the American College of Cardiology 53:3, 272-274
    CrossRef

  215. 215

    William L. Holman, Robert L. Kormos, David C. Naftel, Marissa A. Miller, Frank D. Pagani, Elizabeth Blume, Timothy Cleeton, Steven C. Koenig, Leah Edwards, James K. Kirklin. (2009) Predictors of Death and Transplant in Patients With a Mechanical Circulatory Support Device: A Multi-institutional Study. The Journal of Heart and Lung Transplantation 28:1, 44-50
    CrossRef

  216. 216

    Bart Meyns, Filip Rega, Joris Ector, Walter Droogne, Johan Vanhaecke, Jan Van Hemelrijck, Bartley Griffith, Robert Dowling, Mark Zucker, Daniel Burkhoff. (2009) Partial left ventricular support implanted through minimal access surgery as a bridge to cardiac transplant. The Journal of Thoracic and Cardiovascular Surgery 137:1, 243-245
    CrossRef

  217. 217

    A. Combes. 2009. Assistance circulatoire. , 435-448.
    CrossRef

  218. 218

    Mariell Jessup, Iván J. Núñez-Gil. (2008) Insuficiencia cardiaca y asistencias ventriculares: nuevas respuestas para antiguas preguntas. Revista Española de Cardiología 61:12, 1231-1235
    CrossRef

  219. 219

    James O. Mudd, Jonathan D. Cuda, Marc Halushka, Karl A. Soderlund, John V. Conte, Stuart D. Russell. (2008) Fusion of Aortic Valve Commissures in Patients Supported by a Continuous Axial Flow Left Ventricular Assist Device. The Journal of Heart and Lung Transplantation 27:12, 1269-1274
    CrossRef

  220. 220

    (2008) Guía de práctica clínica de la Sociedad Europea de Cardiología (ESC) para el diagnóstico y tratamiento de la insuficiencia cardiaca aguda y crónica (2008). Revista Española de Cardiología 61:12, 1329.e1-1329.e70
    CrossRef

  221. 221

    James N. Kirkpatrick, Bradley P. Knight. (2008) The management of implantable cardiac devices at the end of life. Progress in Palliative Care 16:5, 250-256
    CrossRef

  222. 222

    Tofy Mussivand. (2008) Neurological Dysfunction Associated With Mechanical Circulatory Support: Complications That Still Need Attention. Artificial Organs 32:11, 831-834
    CrossRef

  223. 223

    Kiyotaka Fukamachi, Diyar Saeed, Alex L. Massiello, David J. Horvath, Hideyuki Fumoto, Tetsuya Horai, Roula Zahr, Shanaz Shalli, Tomohiro Anzai, Raymond Dessoffy, Jacquelyn Catanese, Ji-Feng Chen, Qun Zhou, Stephen Benefit, Sue Alfini, Leonard A. R. Golding. (2008) Development of DexAide Right Ventricular Assist Device: Update II. ASAIO Journal 54:6, 589-593
    CrossRef

  224. 224

    , K. Dickstein, A. Cohen-Solal, G. Filippatos, J. J.V. McMurray, P. Ponikowski, P. A. Poole-Wilson, A. Stromberg, D. J. van Veldhuisen, D. Atar, A. W. Hoes, A. Keren, A. Mebazaa, M. Nieminen, S. G. Priori, K. Swedberg, , A. Vahanian, J. Camm, R. De Caterina, V. Dean, K. Dickstein, G. Filippatos, C. Funck-Brentano, I. Hellemans, S. D. Kristensen, K. McGregor, U. Sechtem, S. Silber, M. Tendera, P. Widimsky, J. L. Zamorano, , M. Tendera, A. Auricchio, J. Bax, M. Bohm, U. Corra, P. della Bella, P. M. Elliott, F. Follath, M. Gheorghiade, Y. Hasin, A. Hernborg, T. Jaarsma, M. Komajda, R. Kornowski, M. Piepoli, B. Prendergast, L. Tavazzi, J.-L. Vachiery, F. W. A. Verheugt, J. L. Zamorano, F. Zannad. (2008) ESC Guidelines for the diagnosis and treatment of acute and chronic heart failure 2008: The Task Force for the Diagnosis and Treatment of Acute and Chronic Heart Failure 2008 of the European Society of Cardiology. Developed in collaboration with the Heart Failure Association of the ESC (HFA) and endorsed by the European Society of Intensive Care Medicine (ESICM). European Heart Journal 29:19, 2388-2442
    CrossRef

  225. 225

    Ranjit John, Forum Kamdar, Kenneth Liao, Monica Colvin-Adams, Andrew Boyle, Lyle Joyce. (2008) Improved Survival and Decreasing Incidence of Adverse Events With the HeartMate II Left Ventricular Assist Device as Bridge-to-Transplant Therapy. The Annals of Thoracic Surgery 86:4, 1227-1235
    CrossRef

  226. 226

    Antonis A Pitsis, Aikaterini N Visouli. (2008) Update on ventricular assist device management in the ICU. Current Opinion in Critical Care 14:5, 569-578
    CrossRef

  227. 227

    John Woodard. (2008) The Cusp of a New Era in Long-term Mechanical Circulatory Support. Artificial Organs 32:10, 757-760
    CrossRef

  228. 228

    Simon Maybaum, Gayathri Kamalakannan, Sandhya Murthy. (2008) Cardiac Recovery During Mechanical Assist Device Support. Seminars in Thoracic and Cardiovascular Surgery 20:3, 234-246
    CrossRef

  229. 229

    Francis D. Pagani. (2008) Continuous-Flow Rotary Left Ventricular Assist Devices with “3rd Generation” Design. Seminars in Thoracic and Cardiovascular Surgery 20:3, 255-263
    CrossRef

  230. 230

    Katherine Lietz, Leslie W. Miller. (2008) Destination Therapy: Current Results and Future Promise. Seminars in Thoracic and Cardiovascular Surgery 20:3, 225-233
    CrossRef

  231. 231

    Ranjit John. (2008) Current Axial-Flow Devices—the HeartMate II and Jarvik 2000 Left Ventricular Assist Devices. Seminars in Thoracic and Cardiovascular Surgery 20:3, 264-272
    CrossRef

  232. 232

    Nishant D. Patel, Eric S. Weiss, Justin Schaffer, Susan L. Ullrich, Dennis C. Rivard, Ashish S. Shah, Stuart D. Russell, John V. Conte. (2008) Right Heart Dysfunction After Left Ventricular Assist Device Implantation: A Comparison of the Pulsatile HeartMate I and Axial-Flow HeartMate II Devices. The Annals of Thoracic Surgery 86:3, 832-840
    CrossRef

  233. 233

    Stephen Westaby. (2008) Destination therapy: time for real progress. Nature Clinical Practice Cardiovascular Medicine 5:8, 477-483
    CrossRef

  234. 234

    Marc A. Simon, John Watson, J. Timothy Baldwin, William R. Wagner, Harvey S. Borovetz. (2008) Current and Future Considerations in the Use of Mechanical Circulatory Support Devices*. Annual Review of Biomedical Engineering 10:1, 59-84
    CrossRef

  235. 235

    Anna Lassia Meyer, Christian Kuehn, Clemens Gras, Doris Malehsa, Christoph Bara, Stefan Fischer, Axel Haverich, Martin Strüber. (2008) Implantation of a Left Ventricular Assist Device in a Patient With Primary Antiphospholipid Syndrome. The Annals of Thoracic Surgery 86:2, 639-640
    CrossRef

  236. 236

    Manuel Prinz von Bayern, Carlos Arrecubieta, Sim Oz, Hirokazu Akashi, Martin Cedeiras, Yoshifumi Naka, Mario C. Deng, Franklin D. Lowy. (2008) Development of A Murine Ventricular Assist Device Transcutaneous Drive-Line Model. The Journal of Heart and Lung Transplantation 27:7, 812-814
    CrossRef

  237. 237

    Juan M. Aranda, Tomas D. Martin, Dana D. Leach, Charles T. Klodell. (2008) Reoperative Cardiac Surgery After Long-term Ventricular Support Therapy. The Journal of Heart and Lung Transplantation 27:7, 789-791
    CrossRef

  238. 238

    Vance G. Nielsen, James K. Kirklin, William L. Holman, Brad L. Steenwyk, James F. George, Fen Zhou, Dale A. Parks, Truitt C. Ellis. (2008) Mechanical Circulatory Device Thrombosis: A New Paradigm Linking Hypercoagulation and Hypofibrinolysis. ASAIO Journal 54:4, 351-358
    CrossRef

  239. 239

    Michael M Koerner, Jama Jahanyar. (2008) Assist devices for circulatory support in therapy-refractory acute heart failure. Current Opinion in Cardiology 23:4, 399-406
    CrossRef

  240. 240

    J.-M. Casillas, V. Gremeaux, M. Labrunee, O. Troigros, Y. Laurent, G. Deley, J.-C. Eicher. (2008) Low-frequency electromyostimulation and chronic heart failure. Annales de Réadaptation et de Médecine Physique 51:6, 461-472
    CrossRef

  241. 241

    Farhana Latif, Martin Cadeiras, Manuel Prinz von Bayern, Khurram Shahzad, Anshu Sinha, Mario C Deng. (2008) Challenges of long-term mechanical circulatory support therapy. Expert Review of Medical Devices 5:4, 413-414
    CrossRef

  242. 242

    Donald Esmore, David Kaye, Phillip Spratt, Robert Larbalestier, Peter Ruygrok, Steven Tsui, Deborah Meyers, Arnt E. Fiane, John Woodard. (2008) A Prospective, Multicenter Trial of the VentrAssist Left Ventricular Assist Device for Bridge to Transplant: Safety and Efficacy. The Journal of Heart and Lung Transplantation 27:6, 579-588
    CrossRef

  243. 243

    G. Michael Felker, Kevin J. Anstrom, Joseph G. Rogers. (2008) A Global Ranking Approach to End Points in Trials of Mechanical Circulatory Support Devices. Journal of Cardiac Failure 14:5, 368-372
    CrossRef

  244. 244

    Stefan Klotz, A.H. Jan Danser, Daniel Burkhoff. (2008) Impact of left ventricular assist device (LVAD) support on the cardiac reverse remodeling process. Progress in Biophysics and Molecular Biology 97:2-3, 479-496
    CrossRef

  245. 245

    Andreas Brieke, Joseph Cleveland, JoAnn Lindenfeld. (2008) Mechanical support in acute and chronic heart failure. Current Cardiology Reports 10:3, 168-175
    CrossRef

  246. 246

    Larry A Allen, G Michael Felker. (2008) Advances in the surgical treatment of heart failure. Current Opinion in Cardiology 23:3, 249-253
    CrossRef

  247. 247

    Harvinder Singh Arora, Pranav Loyalka, Biswajit Kar, Reynolds Delgado. (2008) Devices for Heart Failure: The Future Is Now. Congestive Heart Failure 14:3, 141-148
    CrossRef

  248. 248

    L. Henry Edmunds, Robert W. Colman. (2008) Bleeding That Won’t Stop. The Annals of Thoracic Surgery 85:4, 1153-1154
    CrossRef

  249. 249

    Edgar G Chedrawy, Sunil M Prasad, Malek G Massad. (2008) Surgical management of congestive heart failure: translational research to clinical application – the future is bright!. Expert Review of Cardiovascular Therapy 6:4, 439-442
    CrossRef

  250. 250

    Gorav Ailawadi, Irving L Kron. (2008) New strategies for surgical management of ischemic cardiomyopathy. Expert Review of Cardiovascular Therapy 6:4, 521-530
    CrossRef

  251. 251

    Jalal K Ghali, JoAnn Lindenfeld. (2008) Sex differences in response to chronic heart failure therapies. Expert Review of Cardiovascular Therapy 6:4, 555-565
    CrossRef

  252. 252

    Santiago Garcia, Forum Kandar, Andrew Boyle, Monica Colvin-Adams, Kenneth Lliao, Lyle Joyce, Ranjit John. (2008) Effects of Pulsatile- and Continuous-flow Left Ventricular Assist Devices on Left Ventricular Unloading. The Journal of Heart and Lung Transplantation 27:3, 261-267
    CrossRef

  253. 253

    Soon J Park. (2008) Can a pulsatile left ventricular assist device provide destination therapy for patients with severe heart failure?. Nature Clinical Practice Cardiovascular Medicine 5:2, 82-83
    CrossRef

  254. 254

    Magdi H Yacoub, Leslie W Miller. (2008) Long-term left-ventricular-assist-device therapy is here to stay. Nature Clinical Practice Cardiovascular Medicine 5:2, 60-61
    CrossRef

  255. 255

    Eric A Rose. (2008) A new continuous-flow LV assist device for patients with end-stage heart failure. Nature Clinical Practice Cardiovascular Medicine 5:2, 80-81
    CrossRef

  256. 256

    Jin-Ho Choi, Eun Seok Jun. (2008) Treatment of Medically Intractable End-Stage Heart Failure. Journal of the Korean Medical Association 51:4, 306
    CrossRef

  257. 257

    (2007) Continuous-Flow Ventricular Assist Device. New England Journal of Medicine 357:22, 2305-2306
    Full Text

  258. 258

    Baughman, Kenneth L., Jarcho, John A., . (2007) Bridge to Life — Cardiac Mechanical Support. New England Journal of Medicine 357:9, 846-849
    Full Text

Letters