Cardiac Replacement with a Total Artificial Heart as a Bridge to Transplantation

Jack G. Copeland, M.D., Richard G. Smith, M.S.E.E., C.C.E., Francisco A. Arabia, M.D., Paul E. Nolan, Pharm.D., Gulshan K. Sethi, M.D., Pei H. Tsau, M.D., Douglas McClellan, B.S.M.E., and Marvin J. Slepian, M.D. for the CardioWest Total Artificial Heart Investigators

N Engl J Med 2004; 351:859-867August 26, 2004

Abstract

Background

The CardioWest Total Artificial Heart orthotopically replaces both native cardiac ventricles and all cardiac valves, thus eliminating problems commonly seen in the bridge to transplantation with left ventricular and biventricular assist devices, such as right heart failure, valvular regurgitation, cardiac arrhythmias, ventricular clots, intraventricular communications, and low blood flows.

Methods

We conducted a nonrandomized, prospective study in five centers with the use of historical controls. The purpose was to assess the safety and efficacy of the CardioWest Total Artificial Heart in transplant-eligible patients at risk for imminent death from irreversible biventricular cardiac failure. The primary end points included the rates of survival to heart transplantation and of survival after transplantation.

Results

Eighty-one patients received the artificial-heart device. The rate of survival to transplantation was 79 percent (95 percent confidence interval, 68 to 87 percent). Of the 35 control patients who met the same entry criteria but did not receive the artificial heart, 46 percent survived to transplantation (P<0.001). Overall, the one-year survival rate among the patients who received the artificial heart was 70 percent, as compared with 31 percent among the controls (P<0.001). One-year and five-year survival rates after transplantation among patients who had received a total artificial heart as a bridge to transplantation were 86 and 64 percent.

Conclusions

Implantation of the total artificial heart improved the rate of survival to cardiac transplantation and survival after transplantation. This device prevents death in critically ill patients who have irreversible biventricular failure and are candidates for cardiac transplantation.

Media in This Article

Figure 1The CardioWest Total Artificial Heart.
Figure 2Overall Survival Rate from the Time of Study Entry to the Termination of the Study among the Patients Who Received a Total Artificial Heart According to Protocol and the Controls.
Article

The incidence of congestive heart failure, the final common pathway of myocardial dysfunction in most forms of cardiac disease, has increased by a factor of five in the past three decades and is currently increasing at a rate of 550,000 new cases per year. The nearly 5 million Americans who have heart failure generate annual health care costs of $38.1 billion.1 Most of these patients are treated medically or with surgical therapy tailored to their cardiac disease. Unfortunately, even the most advanced pharmacologic and surgical treatments fail in some patients. Each year, nearly 300,000 Americans die of heart failure1; among them are 15 to 25 percent of those waiting for transplantation.2 It has been estimated that 30,000 to 70,000 Americans each year could benefit from cardiac replacement.3

For the most severe forms of cardiac dysfunction, which lead to irreversible biventricular failure, cardiac transplantation has emerged as the only effective therapy affording longevity and quality of life, with survival rates at one and five years of up to 94 percent and 78 percent, respectively.4 Unfortunately, demand for donor hearts has doubled in recent years, and the annual national supply has decreased from about 2500 hearts to about 2000.2 Since 1998, the waiting time before transplantation among potential heart-transplant recipients has increased to more than one year. Nearly 75 percent of recent recipients require continuous intravenous inotropic support, and 54 percent are on life-support machines before transplantation.2

Many patients with end-stage congestive heart failure who are awaiting transplantation receive mechanical circulatory-support devices that have been approved by the Food and Drug Administration (FDA). Commercially available mechanical circulatory support devices include left ventricular assist devices and paracorporeal devices that may be used to support one or both ventricles. With the use of these devices, the rate of survival from the time of implantation to the time of transplantation has ranged from 51 to 71 percent.5-9 Worldwide, more than 5000 patients have received these devices.

The CardioWest Total Artificial Heart is a biventricular, pneumatic, pulsatile blood pump that completely replaces the patient's native ventricles and all four cardiac valves orthotopically (Figure 1Figure 1The CardioWest Total Artificial Heart.).10 It weighs 160 g and displaces 400 ml of volume. It is lined with polyurethane and has a four-layer, pneumatically driven diaphragm. At maximal stroke volume (70 ml), it delivers a cardiac output of more than 9 liters per minute. In the United States, the device is powered by a large console on wheels that prevents hospital discharge; portable drivers permitting discharge from the hospital are currently used in Europe. The device may be useful in patients for whom left ventricular assist devices and biventricular assist devices are contraindicated, including those with aortic regurgitation, cardiac arrhythmias, a left ventricular thrombus, an aortic prosthesis, an acquired ventricular septal defect, or irreversible biventricular failure requiring high pump outputs.

Methods

We conducted a prospective, investigator-initiated trial of the CardioWest Total Artificial Heart under an investigational-device exemption granted by the FDA. The study was initiated in January 1993 and completed in September 2002. The device is manufactured by SynCardia Systems (Tucson, Ariz.). The institutional review boards of the five participating institutions approved the protocol, and written informed consent was obtained from all patients or their representatives. The authors wrote the manuscript.

Patients

There were 130 patients: 81 patients who met the protocol-inclusion criteria (the protocol group), 35 controls, and 14 patients who did not meet the protocol-inclusion criteria but in whom the device was implanted on a compassionate-use basis (the off-protocol group) (Table 1Table 1Selection Criteria.). The median rate of accrual into the protocol group was 9 patients per year (range, 2 to 18). The patients selected for the protocol group were candidates for cardiac transplantation who were at high risk of imminent death due to irreversible biventricular cardiac failure. Implantation of a left ventricular assist device had been ruled out for the following reasons: 15 could not be weaned from cardiopulmonary bypass (with 12 of the 15 having had a prebypass cardiac arrest), 51 had central venous pressures greater than 18 mm Hg, 11 had right ventricular ejection fractions of less than 20 percent, 2 had ventricular tachycardia, 1 had an aortic prosthesis, and 1 had right ventricular damage sustained at the time of sternotomy.

The controls were patients who had been matched with the patients in the protocol group according to inclusion and exclusion criteria but who had not received any mechanical circulatory support. Data from the controls were used only for survival comparisons. Twenty-two were found on retrospective review of the participating centers' records for the period from 1991 to 1993. Three were prospective patients who had refused device implantation. Ten were identified by retrospectively screening 635 patients whose United Network for Organ Sharing (UNOS) status was 1 (indicating the need for continuous intravenous inotropic support or mechanical support) from participating centers' lists for the period from 1994 to 2001.

The baseline demographic factors, risk factors, hemodynamic measurements, and laboratory values of the patients in the protocol group and the controls are shown in Table 2Table 2Characteristics of the Protocol and Control Groups at the Time of Entry into the Study.. The selection process, based solely on the criteria listed in Table 1, resulted in groups that differed significantly in 12 of 65 baseline characteristics. Consequently, the control group is not an exact match and provides only a rough approximation of the natural history of patients with cardiogenic shock. An ischemic cause of heart failure, a history of smoking, use of anticoagulation, prior cardiac surgery, and current use of an intraaortic balloon pump were more prevalent in the control group, whereas the use of cardiopulmonary bypass was more common in the protocol group (19 percent, vs. 0 percent in the control group). The controls had a lower systolic pressure, pulmonary-artery mean pressure, central venous pressure, total bilirubin level, partial pressure of carbon dioxide, and bicarbonate level than the patients in the protocol group. Use of inotropic support was greater in the protocol group, in which 66 percent of the patients were receiving three or more inotropes; in contrast, 80 percent of the controls were receiving no more than two inotropes. The dosages of inotropes (dopamine, dobutamine, epinephrine, milrinone, and norepinephrine) did not differ significantly between the groups. The patients in the off-protocol group had distinctly different outcomes and are presented separately.

The total artificial heart was implanted with the use of simplified fitting, implantation, and explantation techniques, as previously described.10-12 Coagulation monitoring and anticoagulation were also performed as previously described.10

Major efficacy end points included the rates of survival to transplantation, overall survival, and survival after transplantation, as well as a composite end point called “treatment success.” Treatment was considered successful if the patient had the following characteristics 30 days after transplantation: he or she was alive, was in New York Heart Association class I or II, was ambulatory, was not dependent on a ventilator, and was not undergoing dialysis. Other efficacy end points included hemodynamic recovery, recovery of end-organ function (including the function of the kidneys and liver), the percentage of patients who were ambulatory, and the percentage of patients who could walk more than 100 ft (30.5 m).

Safety was measured by prospectively recording previously defined adverse events.13 Data from all 95 patients in whom a total artificial heart was implanted (the 81 patients in the protocol group and the 14 in the off-protocol group) were used in the analysis of adverse events.

Statistical Analysis

All the data generated by this study were collected and maintained by an independent monitoring group (Research Technology Management, Scottsdale, Ariz.). Statistical analyses were carried out by Synteract (Carlsbad, Calif.). For dichotomous variables, Fisher's exact test was used to test the null hypothesis (i.e., the hypothesis that there was no difference between the protocol group and the control group). P values of less than 0.05 were considered to indicate statistical significance. Kaplan–Meier curves were used to display time-to-event distributions.

Results

Survival

Survival to transplantation was achieved in 79 percent of the patients who received a total artificial heart according to protocol (95 percent confidence interval, 68 to 87 percent), as compared with 46 percent of the controls (P<0.001). The mean time from entry into the study to transplantation or death was 79.1 days among all the patients who received an implant and 8.5 days among the controls (P<0.001).

The overall survival rate at one year was 70 percent (95 percent confidence interval, 63 to 77 percent) among the patients who received an implant according to protocol, as compared with 31 percent among the controls (P<0.001) (Figure 2Figure 2Overall Survival Rate from the Time of Study Entry to the Termination of the Study among the Patients Who Received a Total Artificial Heart According to Protocol and the Controls.). Survival rates one and five years after transplantation among the controls were 69 percent and 34 percent, respectively, and among patients in the protocol group they were 86 percent and 64 percent, respectively; in comparison, contemporary UNOS data2 from more than 4000 patients include one- and five-year survival rates of 84.7 percent and 69.8 percent, respectively. Treatment success was achieved in 69 percent of the patients who received an implant, as compared with 37 percent of the controls (P=0.002).

Other End Points

In the protocol group, patients' hemodynamic status improved immediately after implantation of the total artificial heart, with a sustained increase in the cardiac index from a baseline value of 1.9 to 3.2 liters per minute per square meter of body-surface area. Other signs of hemodynamic recovery that occurred immediately after implantation and that persisted included the mean systolic arterial pressure, which rose from 93 to 122 mm Hg; the mean central venous pressure, which fell from 20 to 14 mm Hg; and the organ-perfusion pressure (mean arterial pressure minus central venous pressure), which rose from 49 to 68 mm Hg. Renal and hepatic function and the levels of blood urea nitrogen, creatinine, bilirubin, and liver enzymes returned to normal within three weeks after implantation. Other laboratory values, such as electrolyte levels, the platelet count, and the white-cell count, also returned to normal at three to four weeks.

The quality of life among the patients who received an implant according to protocol improved significantly. One week after implantation, 75 percent of these patients were out of bed. Mobility, defined as the ability to walk more than 100 ft two weeks after entry into the study, was observed in 60.5 percent.

Fourteen patients received a total artificial heart on an off-protocol basis. Reasons for off-protocol implantation included inability to document fulfillment of inclusion criteria (in four cases), the presence of a ventricular assist device (in three), conditions likely to preclude survival (in two), dialysis (in two), contraindications to heart transplantation (in two), and a previous heart transplantation (in one). The rate of survival to transplantation among these patients was 50 percent. Causes of deaths during the period when the total artificial heart was in use (all of which occurred within 20 days after the procedure) were pulmonary edema (in two cases), bleeding (in two), multiorgan failure (in two), and sepsis (in one).

Adverse Events

Table 3Table 3Adverse Events, Including Those That Affected Outcomes, from the Time of Study Entry to 30 Days after Transplantation. lists the adverse events among all 95 patients who received a total artificial heart and among the subgroup of 81 patients who received the implant according to protocol. The implant was used for a total of 7515 days (20.6 years) in the overall group and 6407 days (17.6 years) in the protocol group.

There were 102 bleeding events, 55 of which occurred during implantation. These 55 events were described as mediastinal bleeding or tamponade necessitating repeated surgery (23 events in 21 patients [21 percent]), the need for 3 units of blood within a 24-hour period after the first 48 postoperative hours (18 events), the need for 8 or more units of blood during surgery (13 events), and abdominal bleeding necessitating surgery (1 event). All but one of the repeated operations took place within the first 21 days after implantation. Two patients died from bleeding, one at the time of device implantation and one at the time of transplantation.

One serious device malfunction — a perforation in one of the four layers of the device's left ventricular diaphragm — resulted in the death of a patient on day 124 after implantation. No other serious device malfunctions occurred during more than 12,000 patient-days of use of the device. A complication of fitting occurred in five patients. In three of these cases the fit was corrected by repositioning or repeated surgery. A poor fit contributed to death in two patients.

The cardiac index fell to less than 2 liters per minute per square meter for four or more hours in association with eight events during the first 11 days after implantation. These events included one fitting problem, four episodes of hypovolemia, one case of pneumothorax, one of tamponade, and one of catheter entrapment. The five other instances in which the cardiac index fell were attributed to a single patient on the 117th day and were the result of a device malfunction. The systolic arterial pressure fell to less than 90 mm Hg for four or more hours 27 times in 18 patients as a result of sepsis (12 events), hypovolemia (12), an antihypertensive medication (1), aspiration (1), and pneumothorax (1).

There were 125 infections during use of the total artificial heart: 50 were respiratory infections, 28 were genitourinary infections, 17 involved the drive line, 12 were gastrointestinal infections, 7 were blood infections, 6 involved the infusion catheter, and 5 were mediastinal infections. In 68 of the 81 patients in the protocol group (84 percent), infection neither delayed transplantation nor contributed to death. Infection delayed transplantation in five patients in this group (6 percent) (two of whom had drive-line infections and one each of whom had a blood, respiratory, or mediastinal infection). Infection contributed to death in seven patients in the protocol group (9 percent) (five of whom had a respiratory infection, one of whom had an infection involving the infusion catheter, and one of whom had a mediastinal infection) and caused death in one patient (1 percent) (who had a respiratory infection). Sixty-two percent of the infections occurred during the first 28 days after the implantation procedure. There were 17 superficial drive-line infections, with none ascending to the mediastinum.

During the period that the total artificial heart was in use, there were 26 neurologic events among patients in the protocol group, including stroke (11 events in 10 patients [12 percent]), transient ischemic attacks (4 events), anoxic encephalopathy (5), metabolic encephalopathy (1), seizure (4), and syncope (1). The actuarial rate of freedom from any neurologic event among patients in the protocol group was 0.91 at day 7, 0.88 at day 14, 0.85 at day 28, and 0.67 at day 168. The curve remained flat (at 0.67) after three months. The linearized rate of stroke was 0.05 event per month. Six of the strokes had no residual effect after 48 hours, four had neurologic residual effects of a minor nature, and one caused a fixed deficit (hemiplegia) and delayed transplantation. There were no deaths from neurologic complications.

Technical or procedural complications occurred in three patients. In these patients, the mechanical tricuspid valve of the total artificial heart became obstructed as a result of migration of the central venous catheter, and the complication either caused or contributed to death. Central venous catheters in the right atrium of this total artificial heart are contraindicated.

Nineteen of the 35 control patients (54 percent) and 17 of the 81 patients in the protocol group (21 percent) died before transplantation. Causes of death before transplantation among the controls were cardiac arrest (in seven), heart failure (in seven), multiorgan failure (in three), graft failure or acute rejection (in one), and pulmonary edema (in one). Causes of death before transplantation in the protocol group were multiorgan failure (in seven patients), procedural or technical complications (in four), bleeding (in two), sepsis (in two), congestive heart failure (in one), and pulmonary edema (in one). After transplantation and before discharge in the control group, in which 16 patients remained at the time of transplantation, there were 2 deaths (1 due to acute rejection and 1 due to multiorgan failure). During the same interval in the protocol group, in which 64 patients remained at the time of transplantation, there were 6 deaths (3 due to graft failure, 1 due to sepsis, 1 due to procedural or technical complications, and 1 due to multiorgan failure).

Discussion

Implantation of the total artificial heart improved outcomes in dying patients by providing immediate hemodynamic restoration and clinical stabilization, leading to end-organ recovery and thus eventually allowing cardiac transplantation. Among the patients who received the device on a per-protocol basis, the rate of survival to transplantation was 79 percent. Previous studies of approved devices5-9 have reported rates of survival to transplantation of 51 to 71 percent with left ventricular assist devices and 58 to 61 percent with biventricular assist devices.

The overall survival rate was 70 percent one year after transplantation. The one- and five-year survival rates among transplant recipients were 86 percent and 64 percent, similar to those reported by UNOS.2 All end points with respect to survival and treatment success significantly favored the patients who received an implant as compared with the controls.

Since 1982, there have been over 430 implantations involving this type of device worldwide.10 Previous studies of the CardioWest device14-16 have summarized results for 219 patients, of whom 70 percent to 93 percent survived from implantation to cardiac transplantation; 93 percent to 100 percent of those patients survived from transplantation to hospital discharge. A retrospective comparison of the device and two of the currently approved devices in very sick patients with heart failure documented an advantage for the total artificial heart.17

Among the patients in this study who received a total artificial heart according to protocol, the cardiac index rose to and remained at 3 liters per minute per square meter or higher, systolic arterial pressure rose above 120 mm Hg, central venous pressure fell to less than 14 mm Hg, and organ-perfusion pressure rose to 68 mm Hg. As a consequence of hemodynamic restoration, serum creatinine, blood urea nitrogen, bilirubin, liver enzymes, and other laboratory values returned to normal within three weeks. Seventy-five percent of these patients were out of bed within one week.

Adverse events were broadly defined and thus frequently reported. However, the outcomes for most patients were good and were similar to those found with left ventricular and biventricular assist devices. There were no deaths from neurologic causes. During the implantation period, 12 percent of the patients had strokes; six of the strokes had no residual effect 48 hours after onset. Reported rates of stroke in patients with ventricular assist devices have varied: 0 to 12 percent for the HeartMate device,5,8,9 20 to 58 percent for the Novacor device,7-9 and 24 percent for the Thoratec device.8

In this study, bleeding caused death in 2 patients and necessitated repeated surgery in 21, for a combined frequency of 28 percent. Published studies5-9 have documented rates of similarly defined end points (death or repeated surgery) of 35 to 48 percent (for the HeartMate left ventricular assist device), 22 to 40 percent (for the Novacor device), and 23 to 33 percent (for the Thoratec biventricular assist device). Among the patients in the protocol group in this study, infection caused one death and contributed to seven others, but most infections neither delayed transplantation (84 percent, no delay) nor had a major clinical effect (81 percent, no clinical effect). Device malfunction, causing a perforated diaphragm, resulted in one death and (in a single patient) five episodes of a reduced cardiac index. Fitting complications, despite well-defined guidelines for fitting the device,10 contributed to two deaths. In three cases, the tricuspid valve of the device became obstructed by the central venous catheter, causing death in two of the cases and contributing to death in the third and thereby underscoring the contraindication to right atrial catheters in these patients. As expected in this group of very sick patients, most of the deaths before transplantation in the control group were due to end-stage heart disease, whereas most deaths in the protocol group were related to multiorgan failure, technical problems, sepsis, or bleeding during the first four weeks after implantation.

In conclusion, implantation of the CardioWest Total Artificial Heart was a successful bridge to cardiac transplantation in many patients with heart failure in whom inotropic therapy had failed and who were not candidates for the use of a left ventricular assist device. Implantation of the total artificial heart helped to restore hemodynamic function, promoted end-organ recovery and mobility, and was associated with a post-transplantation survival rate very similar to national survival rates five years after transplantation. For patients with biventricular cardiac failure whose condition had continued to deteriorate despite maximal inotropic support, the device helped to prevent death and afforded clinical stabilization. Its safety and efficacy have been documented. With this device, a considerable number of potential cardiac-transplant recipients who have no other suitable options may successfully await cardiac transplantation.

Supported from 1991 to 2001 by CardioWest Technologies and thereafter by SynCardia Systems with respect to the costs of data collection.

Dr. Copeland reports owning equity in SynCardia Systems, the manufacturer of the CardioWest Total Artificial Heart. Mr. Smith and Dr. Slepian report owning equity in SynCardia Systems and being paid for part-time employment by the company.

Source Information

From the Sarver Heart Center (J.G.C., R.G.S., F.A.A., P.E.N., G.K.S., P.H.T., M.J.S.), the Marshall Foundation Artificial Heart Program (R.G.S., D.M.), and the College of Pharmacy (P.E.N.), University of Arizona, Tucson.

Address reprint requests to Dr. Copeland at the University of Arizona Sarver Heart Center, 1501 N. Campbell Ave., No. 4402, Tucson, AZ 85724-5071, or at .

The CardioWest Total Artificial Heart Investigators are listed in the Appendix.

Appendix

The CardioWest Total Artificial Heart Investigators and their institutions are as follows: J.G. Copeland and F.A. Arabia (University Medical Center, Tucson, Ariz.), B. Foy (Loyola Medical Center, Chicago), J. Long and D. Doty (LDS Hospital, Salt Lake City), A. Tector (St. Luke's Medical Center, Milwaukee), and R. Kormos (University of Pittsburgh Medical Center, Pittsburgh).

References

References

  1. 1

    Hunt SA, Baker DW, Chin MH, et al. ACC/AHA guidelines for the evaluation and management of chronic heart failure in the adult: executive summary: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (Committee to Revise the 1995 Guidelines for the Evaluation and Management of Heart Failure): developed in collaboration with the International Society for Heart and Lung Transplantation: endorsed by the Heart Failure Society of America. Circulation 2001;104:2996-3007
    CrossRef | Web of Science | Medline

  2. 2

    2001 Annual report of the U.S. Organ Procurement and Transplantation Network and the Scientific Registry of Transplant Recipients. Vol. 1. Washington, D.C.: Department of Health and Human Services, 2001:439-83.

  3. 3

    Hogness JR, VanAntwerp M, eds. The artificial heart: prototypes, policies, and patients. Washington, D.C.: National Academy Press, 1991:1-13.

  4. 4

    Copeland JG, McCarthy M. Cardiac transplantation: changing patterns in selection and outcomes. In: Cecka JM, Terasaki PI, eds. Clinical transplants 2001. Los Angeles: UCLA Immunogenetics Center, 2002:203-7.

  5. 5

    Frazier OH, Rose EA, Oz MC, et al. Multicenter clinical evaluation of the HeartMate vented electric left ventricular assist system in patients awaiting heart transplantation. J Thorac Cardiovasc Surg 2001;122:1186-1195
    CrossRef | Web of Science | Medline

  6. 6

    Farrar DJ, Hill JD, Pennington DG, et al. Preoperative and postoperative comparison of patients with univentricular and biventricular support with the Thoratec ventricular assist device as a bridge to cardiac transplantation. J Thorac Cardiovasc Surg 1997;113:202-209
    CrossRef | Web of Science | Medline

  7. 7

    Di Bella I, Pagani F, Banfi C, et al. Results with the Novacor assist system and evaluation of long-term assistance. Eur J Cardiothorac Surg 2000;18:112-116
    CrossRef | Web of Science | Medline

  8. 8

    Minami K, El-Banayosy A, Sezai A, et al. Morbidity and outcome after mechanical ventricular support using Thoratec, Novacor, and HeartMate for bridging to heart transplantation. Artif Organs 2000;24:421-426
    CrossRef | Web of Science | Medline

  9. 9

    El-Banayosy A, Arusoglu L, Kizner L, et al. Novacor left ventricular assist system versus HeartMate vented electric left ventricular assist system as a long-term mechanical circulatory support device in bridging patients: a prospective study. J Thorac Cardiovasc Surg 2000;119:581-587
    CrossRef | Web of Science | Medline

  10. 10

    Copeland JG, Arabia FA, Tsau PH, et al. Total artificial hearts: bridge to transplantation. Cardiol Clin 2003;21:101-113
    CrossRef | Medline

  11. 11

    Arabia FA, Copeland JG, Pavie A, Smith RG. Implantation technique for the CardioWest Total Artificial Heart. Ann Thorac Surg 1999;68:698-704
    CrossRef | Web of Science | Medline

  12. 12

    Copeland JG, Arabia FA, Smith RG, Covington D. Synthetic membrane neo-pericardium facilitates total artificial heart explantation. J Heart Lung Transplant 2001;20:654-656
    CrossRef | Web of Science | Medline

  13. 13

    Copeland JG III, Arabia FA, Banchy ME, et al. The CardioWest Total Artificial Heart bridge to transplantation: 1993 to 1996 national trial. Ann Thorac Surg 1998;66:1662-1669
    CrossRef | Web of Science | Medline

  14. 14

    Copeland JG, Pavie A, Duveau D, et al. Bridge to transplantation with the CardioWest Total Artificial Heart: the international experience 1993 to 1995. J Heart Lung Transplant 1996;15:94-99
    Web of Science | Medline

  15. 15

    Copeland JG, Arabia FA, Smith RG, Sethi GK, Nolan PE, Banchy ME. Arizona experience with CardioWest Total Artificial Heart bridge to transplantation. Ann Thorac Surg 1999;68:756-760
    CrossRef | Web of Science | Medline

  16. 16

    Leprince P, Pavie A, Bonnet N, Leger P, Gandjbakhch I. Fifteen years single center experience with the total artificial heart Jarvik-7: a bridge to the new era of electrical total artificial heart. J Heart Lung Transplant 2002;21:106-106
    CrossRef

  17. 17

    Copeland JG III, Smith RG, Arabia FA, et al. Comparison of the CardioWest Total Artificial Heart, the Novacor left ventricular assist system and the Thoratec ventricular assist system in bridge to transplantation. Ann Thorac Surg 2001;71:Suppl:S92-S97
    CrossRef | Web of Science | Medline

Citing Articles (94)

Citing Articles

  1. 1

    L. K. Murray, J. Gonzalez-Costello, S. N. Jonas, D. B. Sims, K. A. Morrison, P. C. Colombo, D. M. Mancini, S. W. Restaino, E. Joye, E. Horn, H. Takayama, C. C. Marboe, Y. Naka, U. P. Jorde, N. Uriel. (2012) Ventricular assist device support as a bridge to heart transplantation in patients with giant cell myocarditis. European Journal of Heart Failure
    CrossRef

  2. 2

    Stephen Westaby, Kyriakos Anastasiadis, George M. Wieselthaler. (2012) Cardiogenic shock in ACS. Part 2: role of mechanical circulatory support. Nature Reviews Cardiology
    CrossRef

  3. 3

    Matthias Kirsch, Jean-Philippe Mazzucotelli, Jean-Christian Roussel, Olivier Bouchot, Joseph N'Loga, Pascal Leprince, Pierre-Yves Litzler, André Vincentelli. (2012) Survival after biventricular mechanical circulatory support: Does the type of device matter?. The Journal of Heart and Lung Transplantation
    CrossRef

  4. 4

    Jack G. Copeland, Hannah Copeland, Monica Gustafson, Nicole Mineburg, Diane Covington, Richard G. Smith, Mark Friedman. (2012) Experience with more than 100 total artificial heart implants. The Journal of Thoracic and Cardiovascular Surgery
    CrossRef

  5. 5

    E. Flécher, P. Ménestret, P. Squara. (2011) Assistance circulatoire en réanimation. Réanimation
    CrossRef

  6. 6

    Keyur B. Shah, Daniel G. Tang, Richard H. Cooke, Michael C. Kontos, Neil P. Lewis, Gundars J. Katlaps, Michael L. Hess, Vigneshwar Kasirajan. (2011) Evaluating Heart Failure After Implantation of Mechanical Circulatory Support Devices. Current Heart Failure Reports
    CrossRef

  7. 7

    Armin Zittermann, Uwe Fuchs, Joachim Kuhn, Jens Dreier, Uwe Schulz, Jan F. Gummert, Jochen Börgermann. (2011) Parameters of Mineral Metabolism predict Midterm Clinical Outcome in End-Stage Heart Failure Patients. Scandinavian Cardiovascular Journal 45:6, 342-348
    CrossRef

  8. 8

    Harajeshwar S. Kohli, Justin Canada, Ross Arena, Daniel G. Tang, Mary Ann Peberdy, Suzanne Harton, Maureen Flattery, Kelly Doolin, Gundars J. Katlaps, Michael L. Hess, Vigneshwar Kasirajan, Keyur B. Shah. (2011) Exercise blood pressure response during assisted circulatory support: Comparison of the total artifical heart with a left ventricular assist device during rehabilitation. The Journal of Heart and Lung Transplantation 30:11, 1207-1213
    CrossRef

  9. 9

    Hannah Copeland, Paul E. Nolan, Diane Covington, Monica Gustafson, Richard Smith, Jack G. Copeland. (2011) A Method for Anticoagulation of Children on Mechanical Circulatory Support. Artificial Organs 35:11, 1018-1023
    CrossRef

  10. 10

    T. Krabatsch, M. Schweiger, A. Stepanenko, T. Drews, E. Potapov, M. Pasic, Y. Weng, M. Huebler, R. Hetzer. (2011) Fortschritte bei implantierbaren mechanischen Kreislaufunterstüzungssystemen. Herz 36:7, 622-629
    CrossRef

  11. 11

    Satoshi Kainuma, Taichi Sakaguchi, Shunsuke Saito, Shigeru Miyagawa, Yasushi Yoshikawa, Takashi Yamauchi, Yasushi Sakata, Atsushi Takahashi, Takuya Uehata, Toru Kuratani, Yoshiki Sawa. (2011) Implantation of a Jarvik 2000 left ventricular assist device as a bridge to eligibility for refractory heart failure with renal dysfunction. Journal of Artificial Organs
    CrossRef

  12. 12

    André Pohlmann, Kay Hameyer. (2011) A study on permanent magnet topologies for hybrid bearings for medical drives applied in Ventricular Assist Devices. Archives of Electrical Engineering 60:3, 371-380
    CrossRef

  13. 13

    George M. Viola, Rabih O. Darouiche. (2011) Cardiovascular Implantable Device Infections. Current Infectious Disease Reports 13:4, 333-342
    CrossRef

  14. 14

    Margit Voegele-Kadletz, Ernst Wolner. (2011) Bio artificial surfaces - Blood surface interaction. Materials Science and Engineering: C 31:6, 1195-1200
    CrossRef

  15. 15

    Dawn E. Jaroszewski, Eric M. Anderson, Christopher N. Pierce, Francisco A. Arabia. (2011) The SynCardia freedom driver: A portable driver for discharge home with the total artificial heart. The Journal of Heart and Lung Transplantation 30:7, 844-845
    CrossRef

  16. 16

    Edwin C. McGee, Umraan Ahmad, Daniel Tamez, Michael Brown, Neil Voskoboynikov, Sukit Chris Malaisrie, Richard Lee, Patrick M. McCarthy. (2011) Biventricular Continuous Flow VADs Demonstrate Diurnal Flow Variation and Lead to End-Organ Recovery. The Annals of Thoracic Surgery 92:1, e1-e3
    CrossRef

  17. 17

    Brantley Dollar Gaitan, Christopher A. Thunberg, Lynn G. Stansbury, Dawn E. Jaroszewski, Francisco A. Arabia, Bartley P. Griffith, Alina M. Grigore. (2011) Development, Current Status, and Anesthetic Management of the Implanted Artificial Heart. Journal of Cardiothoracic and Vascular Anesthesia
    CrossRef

  18. 18

    Keyur B. Shah, Daniel G. Tang, Richard H. Cooke, Suzanne Harton, Maureen Flattery, Gundars J. Katlaps, Vigneshwar Kasirajan, Michael L. Hess. (2011) Implantable Mechanical Circulatory Support: Demystifying Patients With Ventricular Assist Devices and Artificial Hearts. Clinical Cardiology 34:3, 147-152
    CrossRef

  19. 19

    Maryl R Johnson, Walter G Kao, Elaine M Winkel, Joseph L Bobadilla, Takushi Kohmoto, Niloo M Edwards. 2011. Heart Transplantation. , 171-204.
    CrossRef

  20. 20

    G. Földes, S.E. Harding. 2011. Stem Cell Therapy to Treat Heart Failure. , 407-423.
    CrossRef

  21. 21

    Latif Arusoglu, Nils Reiss, Michiel Morshuis, Michael Schoenbrodt, Kavous Hakim-Meibodi, Jan Gummert. (2010) The Thoratec System Implanted as a Modified Total Artificial Heart: the Bad Oeynhausen Technique. The Heart Surgery Forum 13:6, E391-E393
    CrossRef

  22. 22

    John Lynn Jefferies, Jack F. Price, David L.S. Morales. (2010) Mechanical Support in Childhood Heart Failure. Heart Failure Clinics 6:4, 559-573
    CrossRef

  23. 23

    Adam J. Hansen, Jack G. Copeland. (2010) Combined heart–kidney transplant after CardioWest total artificial heart bridge. The Journal of Heart and Lung Transplantation 29:10, 1193-1195
    CrossRef

  24. 24

    Akira Shiose, Kathleen Nowak, David J. Horvath, Alex L. Massiello, Leonard A. R. Golding, Kiyotaka Fukamachi. (2010) Speed Modulation of the Continuous-Flow Total Artificial Heart to Simulate a Physiologic Arterial Pressure Waveform. ASAIO Journal 56:5, 403-409
    CrossRef

  25. 25

    Akihiko Homma, Yoshiyuki Taenaka, Eisuke Tatsumi, Yoshiaki Takewa, Toshihide Mizuno, Kyoko Shioya, Hwan Sung Lee, Tomonori Tsukiya, Yukihide Kakuta, Nobumasa Katagiri, Tomohiro Nishinaka, Kohji Koshiji. (2010) Development of an electrohydraulic total artificial heart system: Improvement of pump unit. Electronics and Communications in Japan 93:9, 34-46
    CrossRef

  26. 26

    Yasir Abu-Omar, Steven S. L. Tsui. (2010) Mechanical Circulatory Support for AMI and Cardiogenic Shock. Journal of Cardiac Surgery 25:4, 434-441
    CrossRef

  27. 27

    Yulong Guan, Tushar Karkhanis, Shigang Wang, Alan Rider, Steven C. Koenig, Mark S. Slaughter, Aly El Banayosy, Akif Ündar. (2010) Physiologic Benefits of Pulsatile Perfusion During Mechanical Circulatory Support for the Treatment of Acute and Chronic Heart Failure in Adults. Artificial Organs 34:7, 529-536
    CrossRef

  28. 28

    Gary S. Francis, Barry H. Greenberg, Daphne T. Hsu, Brian E. Jaski, Mariell Jessup, Martin M. LeWinter, Francis D. Pagani, Ileana L. Piña, Marc J. Semigran, Mary Norine Walsh, David H. Wiener, Clyde W. Yancy. (2010) ACCF/AHA/ACP/HFSA/ISHLT 2010 Clinical Competence Statement on Management of Patients With Advanced Heart Failure and Cardiac Transplant. Journal of the American College of Cardiology 56:5, 424-453
    CrossRef

  29. 29

    Heart Failure Society of America. (2010) Section 10: Surgical Approaches to the Treatment of Heart Failure. Journal of Cardiac Failure 16:6, e122-e125
    CrossRef

  30. 30

    M. Schönbrodt, C. Özpeker, M. Morshuis, J. Cantow, L. Arusoglu, J.F. Gummert. (2010) Kunstherzimplantation als ultima ratio bei fulminantem kardiogenem Schock. Zeitschrift für Herz-,Thorax- und Gefäßchirurgie 24:1, 22-27
    CrossRef

  31. 31

    Marius Berman, Jayan Parameshwar, David P. Jenkins, Kumud Dhital, Clive Lewis, Kirsty Dempster, Paul Lincoln, Catherine Sudarshan, Stephen R. Large, John Dunning, Steven S.L. Tsui. (2010) Thoratec implantable ventricular assist device: The Papworth experience. The Journal of Thoracic and Cardiovascular Surgery 139:2, 466-473
    CrossRef

  32. 32

    Hideyuki Fumoto, David J. Horvath, Santosh Rao, Alex L. Massiello, Tetsuya Horai, Tohru Takaseya, Yoko Arakawa, Nicole Mielke, Ji-Feng Chen, Raymond Dessoffy, Kiyotaka Fukamachi, Leonard A.R. Golding. (2010) In vivo acute performance of the Cleveland Clinic self-regulating, continuous-flow total artificial heart. The Journal of Heart and Lung Transplantation 29:1, 21-26
    CrossRef

  33. 33

    Barry A. Boilson, Eugenia Raichlin, Soon J. Park, Sudhir S. Kushwaha. (2010) Device Therapy and Cardiac Transplantation for End-Stage Heart Failure. Current Problems in Cardiology 35:1, 8-64
    CrossRef

  34. 34

    Eric Anderson, Dawn Jaroszewski, Christopher Pierce, Patrick DeValeria, Francisco Arabia. (2009) Parallel Application of Extracorporeal Membrane Oxygenation and the CardioWest Total Artificial Heart as a Bridge to Transplant. The Annals of Thoracic Surgery 88:5, 1676-1678
    CrossRef

  35. 35

    Jan Spillner, Rüdger Kopp, Thomas Finocchiaro, Mehdi Behbahani, Rolf Rossaint, Ulrich Steinseifer, Marek Behr, Rüdiger Autschbach. (2009) Assistierte Zirkulation: ein Überblick aus klinischer Sicht / Assisted circulation: an overview from a clinical perspective. Biomedizinische Technik/Biomedical Engineering 54:5, 255-267
    CrossRef

  36. 36

    Marco Ricci, Anthony L. Panos, Fotios M. Andreopoulos, Paolo Rusconi, Eliot Rosenkranz, Michael Barron, Si M. Pham. (2009) Successful Use of a Custom-Made Paracorporeal Total Artificial Heart as a Bridge to Retransplantation in Adult and Adolescent Patients. The Journal of Heart and Lung Transplantation 28:8, 834-837
    CrossRef

  37. 37

    Bodh I. Jugdutt. (2009) Current and novel cardiac support therapies. Current Heart Failure Reports 6:1, 19-27
    CrossRef

  38. 38

    Rob Phillips, Peter Lichtenthal, Julie Sloniger, Darryl Burstow, Malcolm West, Jack Copeland. (2009) Noninvasive Cardiac Output Measurement in Heart Failure Subjects on Circulatory Support. Anesthesia & Analgesia 108:3, 881-886
    CrossRef

  39. 39

    Dong-Choon Sin, Ho-Lun Kei, Xigeng Miao. (2009) Surface coatings for ventricular assist devices. Expert Review of Medical Devices 6:1, 51-60
    CrossRef

  40. 40

    Chisato Nojiri. (2009) Implantable Left Ventricular Assist System. Circulation Journal 73:SupplementA, A48-A54
    CrossRef

  41. 41

    A. Combes. 2009. Assistance circulatoire. , 435-448.
    CrossRef

  42. 42

    Akihiko Homma, Yoshiyuki Taenaka, Eisuke Tatsumi, Eiki Akagawa, Hwansung Lee, Tomohiro Nishinaka, Yoshiaki Takewa, Toshihide Mizuno, Tomonori Tsukiya, Yukihide Kakuta, Nobumasa Katagiri, Isao Shimosaki, Shigeru Hamada, Hiroshi Mukaibayashi, Wataru Iwaoka. (2008) Development of a compact wearable pneumatic drive unit for a ventricular assist device. Journal of Artificial Organs 11:4, 182-190
    CrossRef

  43. 43

    Firas Zahr, Yoshio Ootaki, Randall C. Starling, Nicholas G. Smedira, Mohamad Yamani, Lucy Thuita, Kiyotaka Fukamachi. (2008) Preoperative Risk Factors for Mortality After Biventricular Assist Device Implantation. Journal of Cardiac Failure 14:10, 844-849
    CrossRef

  44. 44

    Michael A Crouch, Vigeshwar Kasirajan, William Cahoon, Gundars J Katlaps, Kyle J Gunnerson. (2008) Successful Use and Dosing of Bivalirudin After Temporary Total Artificial Heart Implantation: A Case Series. Pharmacotherapy 28:11, 1413-1420
    CrossRef

  45. 45

    Lauren Canter, Elan A. Howell, Rohinton Morris, Drew A. Torigian. (2008) Chest Radiographic and Computed Tomographic Findings of the Temporary Total Artificial Heart (TAH-t). Journal of Thoracic Imaging 23:4, 269-271
    CrossRef

  46. 46

    Rohinton J. Morris. (2008) Total Artificial Heart—Concepts and Clinical Use. Seminars in Thoracic and Cardiovascular Surgery 20:3, 247-254
    CrossRef

  47. 47

    Joseph C. Cleveland, Frederick L. Grover, David A. Fullerton, David N. Campbell, Max B. Mitchell, JoAnn Lindenfeld, Eugene E. Wolfel, Brian D. Lowes, Simon F. Shakar, Andreas Brieke, Anne Cannon, Alastair D. Robertson. (2008) Left ventricular assist device as bridge to transplantation does not adversely affect one-year heart transplantation survival. The Journal of Thoracic and Cardiovascular Surgery 136:3, 774-777
    CrossRef

  48. 48

    Harvinder Singh Arora, Pranav Loyalka, Biswajit Kar, Reynolds Delgado. (2008) Devices for Heart Failure: The Future Is Now. Congestive Heart Failure 14:3, 141-148
    CrossRef

  49. 49

    Andreas Brieke, Joseph Cleveland, JoAnn Lindenfeld. (2008) Mechanical support in acute and chronic heart failure. Current Cardiology Reports 10:3, 168-175
    CrossRef

  50. 50

    Jack G. Copeland, Richard G. Smith, Raj K. Bose, Pei H. Tsau, Paul E. Nolan, Marvin J. Slepian. (2008) Risk Factor Analysis for Bridge to Transplantation With the CardioWest Total Artificial Heart. The Annals of Thoracic Surgery 85:5, 1639-1644
    CrossRef

  51. 51

    Jan D. Schmitto, Philipp Ortmann, Mustafa Akdis, Haidar Alekuzei, Katja Steinke, Philipp Kolat, Aron F. Popov, Oliver J. Liakopoulos, Regina Waldmann-Beushausen, Masoud Mirzaie, Marius Grossmann, Ralf Seipelt, Friedrich A. Schöndube. (2008) Miniaturized HIA Microdiagonal Pump as Left Ventricular Assist Device in a Sheep Model. ASAIO Journal 54:3, 233-236
    CrossRef

  52. 52

    Gorav Ailawadi, Irving L Kron. (2008) New strategies for surgical management of ischemic cardiomyopathy. Expert Review of Cardiovascular Therapy 6:4, 521-530
    CrossRef

  53. 53

    Markus Kondruweit, Thomas Strecker, Timo Seitz, Michael Uder, Michael Weyand, Rene Tandler. (2008) Stenting of the superior vena cava in a patient with a total artificial heart. The Journal of Thoracic and Cardiovascular Surgery 135:2, 455-456
    CrossRef

  54. 54

    Julia Hutchinson, David A Scott, Andrew J Clegg, Emma Loveman, Pam Royle, Jackie Bryant, Jill L Colquitt. (2008) Cost–effectiveness of left ventricular-assist devices in end-stage heart failure. Expert Review of Cardiovascular Therapy 6:2, 175-185
    CrossRef

  55. 55

    Marvin J Slepian, Jack G Copeland. (2008) The total artificial heart in refractory cardiogenic shock: saving the patient versus saving the heart. Nature Clinical Practice Cardiovascular Medicine 5:2, 64-65
    CrossRef

  56. 56

    D. Hoefer, H. Antretter, G. Laufer. (2007) Klinische Indikationskriterien für mechanische Kreislaufunterstützung. Zeitschrift für Herz-,Thorax- und Gefäßchirurgie 21:6, 273-279
    CrossRef

  57. 57

    M. Yacoub, R. Nerem. (2007) Introduction. Bioengineering the heart. Philosophical Transactions of the Royal Society B: Biological Sciences 362:1484, 1253-1255
    CrossRef

  58. 58

    Manuel Prinz von Bayern, Martin Cadeiras, Mario C. Deng. (2007) Destination Therapy: Does Progress Depend on Left Ventricular Assist Device Development?. Heart Failure Clinics 3:3, 349-367
    CrossRef

  59. 59

    Saleem Haj-Yahia, Emma J. Birks, Paula Rogers, Christopher Bowles, Mandy Hipkins, Robert George, Mohammed Amrani, Mario Petrou, John Pepper, Gilles Dreyfus, Asghar Khaghani. (2007) Midterm experience with the Jarvik 2000 axial flow left ventricular assist device. The Journal of Thoracic and Cardiovascular Surgery 134:1, 199-203
    CrossRef

  60. 60

    Mark S. Slaughter, Steven S. Tsui, Aly El-Banayosy, Benjamin C. Sun, Robert L. Kormos, Dale K. Mueller, H. Todd Massey, Timothy B. Icenogle, David J. Farrar, J. Donald Hill. (2007) Results of a multicenter clinical trial with the Thoratec Implantable Ventricular Assist Device. The Journal of Thoracic and Cardiovascular Surgery 133:6, 1573-1580.e2
    CrossRef

  61. 61

    Stavros G Drakos, Efstratios I Charitos, Serafim N Nanas, John N Nanas. (2007) Ventricular-assist devices for the treatment of chronic heart failure. Expert Review of Cardiovascular Therapy 5:3, 571-584
    CrossRef

  62. 62

    Tomasz Timek, Kai Ihnken. (2007) Surgical therapy for heart failure. Expert Review of Cardiovascular Therapy 5:2, 251-263
    CrossRef

  63. 63

    Pradeep S. Arumugham, Christopher M. O’Connor. (2007) Nonpharmacologic therapy in heart failure: An overview. Current Heart Failure Reports 4:1, 33-38
    CrossRef

  64. 64

    James D. Neaton, Sharon-Lise Normand, Annetine Gelijns, Randall C. Starling, Douglas L. Mann, Marvin A. Konstam. (2007) Designs for Mechanical Circulatory Support Device Studies. Journal of Cardiac Failure 13:1, 63-74
    CrossRef

  65. 65

    E. Flecher, T. Joudinaud. (2007) Les résultats du cœur artificiel. Annales de Cardiologie et d'Angéiologie 56:1, 54-59
    CrossRef

  66. 66

    John Beca, Timothy Willcox, Roger M.O. Hall. 2007. Mechanical Cardiac Support. , 342-364.
    CrossRef

  67. 67

    Wolf Sapirstein, Eric Chen, Julie Swain, Bram Zuckerman. (2006) US FDA perspective on regulatory issues affecting circulatory assist devices. Expert Review of Medical Devices 3:6, 749-753
    CrossRef

  68. 68

    John P. Boehmer, Eric Popjes. (2006) Cardiac failure: Mechanical support strategies. Critical Care Medicine 34:Suppl, S268-S277
    CrossRef

  69. 69

    Susan L. Zickmund, Julie A. Blasiole, Venture Brase, Robert M. Arnold. (2006) Congestive Heart Failure Patients Report Conflict With Their Physicians. Journal of Cardiac Failure 12:7, 546-553
    CrossRef

  70. 70

    Craig H. Selzman, Rajendra S. Bhati, Brett C. Sheridan, William E. Stansfield, Michael R. Mill. (2006) Surgical Therapy for Heart Failure. Journal of the American College of Surgeons 203:2, 226-239
    CrossRef

  71. 71

    Keiji Kamohara, Stephan Weber, Ryan S. Klatte, Yoshio Ootaki, Masatoshi Akiyama, Michael W. Kopcak, Viviane Luangphakdy, Christine R. Flick, Ji-Feng Chen, Jose L. Navia, William A. Smith, Kiyotaka Fukamachi. (2006) Replacement of the Left-Side Valves of an Implanted Total Artificial Heart. ASAIO Journal 52:4, 368-372
    CrossRef

  72. 72

    Norbert Lameire, Wim Van Biesen, Raymond Vanholder. (2006) The changing epidemiology of acute renal failure. Nature Clinical Practice Nephrology 2:7, 364-377
    CrossRef

  73. 73

    Gilbert H. Mudge, James C. Fang, Colleen Smith, Gregory Couper. (2006) The physiologic basis for the management of ventricular assist devices. Clinical Cardiology 29:7, 285-289
    CrossRef

  74. 74

    Daniel J. Goldstein, Douglas Smego, Robert E. Michler. (2006) Surgical aspects of congestive heart failure. Heart Failure Reviews 11:2, 171-192
    CrossRef

  75. 75

    Conrad M. Zapanta. 2006. Heart, Artificial. .
    CrossRef

  76. 76

    Marvin J. Slepian, Richard G Smith, Jack G Copeland. 2006. The SynCardia CardioWest™ Total Artificial Heart. , 473-490.
    CrossRef

  77. 77

    James K Kirklin, William L Holman. (2006) Mechanical circulatory support therapy as a bridge to transplant or recovery (new advances). Current Opinion in Cardiology 21:2, 120-126
    CrossRef

  78. 78

    D. G. Healy, M. T. Akbar, N. Baktiari, J. J. Egan, N. Mahon, D. Veerasingam, J. McCarthy, J. Hurley, M. Neligan, A. E. Wood. (2006) The first 20 years of heart transplantation in Ireland. Irish Journal of Medical Science 175:1, 5-9
    CrossRef

  79. 79

    Soren Schenk, Latif Arusoglu, Michiel Morshuis, Kazutomo Minami, Peter Sarnowski, Reiner Koerfer, Aly El-Banayosy. (2006) Triple Bridge-to-Transplant in a Case of Giant Cell Myocarditis Complicated by Human Leukocyte Antigen Sensitization and Heparin-Induced Thrombocytopenia Type II. The Annals of Thoracic Surgery 81:3, 1107-1109
    CrossRef

  80. 80

    O H. Frazier, Egemen Tuzun, William E. Cohn, Jeffrey L. Conger, Kamuran A. Kadipasaoglu. (2006) Total Heart Replacement Using Dual Intracorporeal Continuous-Flow Pumps in a Chronic Bovine Model: A Feasibility Study. ASAIO Journal 52:2, 145-149
    CrossRef

  81. 81

    Heart Failure Society of America. (2006) Section 10: Surgical Approaches to the Treatment of Heart Failure. Journal of Cardiac Failure 12:1, e76-e79
    CrossRef

  82. 82

    Ayesha S. Bryant, William L. Holman, Navin C. Nanda, Srinivas Vengala, Margaret S. Blood, Salpy V. Pamboukian, James K. Kirklin. (2006) Native Aortic Valve Insufficiency in Patients With Left Ventricular Assist Devices. The Annals of Thoracic Surgery 81:2, e6-e8
    CrossRef

  83. 83

    Manuel F. Jiménez Navarro, Javier Díez Martínez, Juan F. Delgado Jiménez, María G. Crespo Leiro. (2006) La insuficiencia cardíaca en el año 2005. Revista Española de Cardiología 59, 55-65
    CrossRef

  84. 84

    Pascal Leprince, Nicolas Bonnet, Shaida Varnous, Actar Rama, Philippe Léger, Alexandre Ouattara, Marc Landi, Jacques Szefner, Iradj Gandjbakhch, Alain Pavie. (2005) Patients With a Body Surface Area Less Than 1.7 m2 Have a Good Outcome With the CardioWest Total Artificial Heart. The Journal of Heart and Lung Transplantation 24:10, 1501-1505
    CrossRef

  85. 85

    James D. Neaton, Gerry Gray, Bram D. Zuckerman, Marvin A. Konstam. (2005) Key Issues in End Point Selection for Heart Failure Trials: Composite End Points. Journal of Cardiac Failure 11:8, 567-575
    CrossRef

  86. 86

    Aly El-Banayosy, Latif Arusoglu, Michel Morshuis, Lukasz Kizner, Gero Tenderich, Peter Sarnowski, Hendrik Milting, Reiner Koerfer. (2005) CardioWest Total Artificial Heart: Bad Oeynhausen Experience. The Annals of Thoracic Surgery 80:2, 548-552
    CrossRef

  87. 87

    Sonna M. Patel, Amy L. Throckmorton, Alexandrina Untaroiu, Paul E. Allaire, Houston G. Wood, Don B. Olsen. (2005) The Status of Failure and Reliability Testing of Artificial Blood Pumps. ASAIO Journal 51:4, 440-451
    CrossRef

  88. 88

    M.G. Crespo Leiro, R. Marzoa Rivas, M.J. Paniagua Martín, A. Castro Beiras. (2005) Tratamiento de la insuficiencia cardíaca aguda. Medicine - Programa de Formación Médica Continuada Acreditado 9:35, 2308-2317
    CrossRef

  89. 89

    Katherine Lietz, Leslie W Miller. (2005) Will left-ventricular assist device therapy replace heart transplantation in the foreseeable future?. Current Opinion in Internal Medicine 4:3, 291-296
    CrossRef

  90. 90

    John JV McMurray, Marc A Pfeffer. (2005) Heart failure. The Lancet 365:9474, 1877-1889
    CrossRef

  91. 91

    M. Cascalho, J. L. Platt. (2005) New Technologies for Organ Replacement and Augmentation. Mayo Clinic Proceedings 80:3, 370-378
    CrossRef

  92. 92

    Jonathan Leor, Yoram Amsalem, Smadar Cohen. (2005) Cells, scaffolds, and molecules for myocardial tissue engineering. Pharmacology & Therapeutics 105:2, 151-163
    CrossRef

  93. 93

    (2004) A Bridge to Heart Transplantation. New England Journal of Medicine 351:24, 2552-2553
    Full Text

  94. 94

    Renlund, Dale G., . (2004) Building a Bridge to Heart Transplantation. New England Journal of Medicine 351:9, 849-851
    Full Text

Letters