Original Article

Endoluminal Beta-Radiation Therapy for the Prevention of Coronary Restenosis after Balloon Angioplasty

Vitali Verin, M.D., Youri Popowski, M.D., Bernard de Bruyne, M.D., Dietrich Baumgart, M.D., Wolfgang Sauerwein, M.D., Markus Lins, M.D., György Kovacs, M.D., Martyn Thomas, M.D., Francis Calman, M.D., Clemens Disco, M.Sc., Patrick W. Serruys, M.D., William Wijns, M.D., Marleen Piessens, Ph.D., John Kurtz, M.D., Ruediger Simon, M.D., Patrice Delafontaine, M.D., and Raimund Erbel, M.D. for the Dose-Finding Study Group

N Engl J Med 2001; 344:243-249January 25, 2001

Abstract

Background

Beta radiation is effective in reducing vascular neointimal proliferation in animals after injury caused by balloon angioplasty. However, the lowest dose that can prevent restenosis after coronary angioplasty has yet to be determined.

Methods

After successful balloon angioplasty of a previously untreated coronary stenosis, 181 patients were randomly assigned to receive 9, 12, 15, or 18 Gy of radiation delivered by a centered yttrium-90 source. Adjunctive stenting was required in 28 percent of the patients. The primary end point was the minimal luminal diameter six months after treatment, as a function of the delivered dose of radiation.

Results

At the time of follow-up coronary angiography, the mean minimal luminal diameter was 1.67 mm in the 9-Gy group, 1.76 mm in the 12-Gy group, 1.83 mm in the 15-Gy group, and 1.97 mm in the 18-Gy group (P=0.06 for the comparison of 9 Gy with 18 Gy), resulting in restenosis rates of 29 percent, 21 percent, 16 percent, and 15 percent, respectively (P=0.14 for the comparison of 9 Gy with 18 Gy). At that time, 86 percent of the patients had had no serious cardiac events. In 130 patients treated with balloon angioplasty without a stent, restenosis rates were 28 percent, 17 percent, 16 percent, and 4 percent, respectively (P= 0.02 for the comparison of 9 Gy with 18 Gy). Among these patients, there was a dose-dependent enlargement of the lumen in 28 percent, 50 percent, 45 percent, and 74 percent of patients, respectively (P<0.001 for the comparison of 9 Gy with 18 Gy). The rate of repeated revascularization was 18 percent with 9 Gy and 6 percent with 18 Gy (P=0.26).

Conclusions

Intracoronary beta-radiation therapy produces a significant dose-dependent decrease in the rate of restenosis after angioplasty. An 18-Gy dose not only prevents the renarrowing of the lumen typically observed after successful balloon angioplasty, but actually induces luminal enlargement.

Media in This Article

Figure 2Cumulative Frequency Distribution of the Loss in Minimal Luminal Diameter from Immediately after Angioplasty to Follow-up at Six Months in Patients Treated with Angioplasty and Irradiation with 9, 12, 15, and 18 Gy of Beta Radiation.
Figure 1Loss in Minimal Luminal Diameter per Vessel Segment from Immediately after Angioplasty to Follow-up in the 168 Patients for Whom Angiographic Follow-up Data Were Available at Six Months.
Article

Arterial renarrowing after angioplasty, or restenosis, occurs in 30 to 40 percent of patients and results from neointimal proliferation and constrictive remodeling of the injured artery.1 A number of trials that took pharmacologic approaches to the prevention of restenosis have evaluated more than 20 candidate drugs, but the results so far have been disappointing.2 Coronary stenting has led to a 30 to 50 percent decrease in the rate of restenosis, primarily by preventing the constrictive remodeling of the artery, but at the cost of an increase in neointimal proliferation.3-5 Thus, a new technique for treatment that could prevent both neointimal hyperplasia and constrictive remodeling would be of great clinical benefit.

Experimental evaluation of beta and gamma radiation in animals has shown an inhibitory effect on neointimal proliferation as well as on constrictive remodeling.6-9 Researchers applied a recently developed technique of treatment with endovascular beta radiation from an endoluminally centered, pure metallic yttrium-90 source.10-13 In hypercholesterolemic rabbits, the delivery of a 9-Gy dose of radiation 1 mm below the surface of the tissue markedly inhibited neointimal hyperplasia.7 However, in a study of safety and feasibility, the same dose used in 15 patients did not have a beneficial effect on restenosis.14 We hypothesized that higher doses of radiation were necessary, since the arterial wall in humans is thicker than that in rabbits. Thus, the primary objective of the present study was to determine the effect of 9, 12, 15, and 18 Gy of beta radiation at a tissue depth of 1 mm on the rate of restenosis after a first coronary angioplasty procedure. Secondary objectives were to evaluate the safety of the procedure and to assess the technical performance of the yttrium-90 system for intracoronary beta-radiation therapy.

Methods

Study Design and Objectives

The study was designed as a prospective, randomized, multicenter, dose-finding trial. Between October 1997 and February 1999, 183 patients were enrolled in the trial at five European centers and randomly assigned by computer, over the telephone, to receive 9 Gy, 12 Gy, 15 Gy, or 18 Gy of radiation. Patients more than 50 years old who had angina pectoris or silent ischemia were enrolled if they were suitable candidates for the dilation of a previously untreated native coronary stenosis. For the enrollment criteria to be met, the diameter of the vessel had to be between 2.5 and 4.0 mm, and the stenosis had to be shorter than 15 mm. Patients also had to be eligible for angiographic and clinical follow-up at six months. Patients were deemed ineligible if they had had a recent myocardial infarction with abnormal base-line levels of cardiac enzymes, had a life expectancy of less than six months, were pregnant, had had cancer within the previous five years, had previously received mediastinal irradiation, or were currently participating in another trial. Written informed consent was obtained from all participating patients.

Irradiation was performed after the completion of the balloon angioplasty. The physicians performing the angioplasty were discouraged from using multiple balloon inflations and balloon displacements, so that the length of the vessel segment injured would be limited. Beta irradiation was not performed if there was an urgent need for stent implantation, if glycoprotein IIb/IIIa–receptor blockers had been administered, or if there was abrupt vessel closure that remained unresolved. Complementary stent implantation was allowed in cases in which there were extensive dissections, symptomatic reductions in blood flow, or both. Initially, ticlopidine and aspirin were given for two months after the procedure; starting in November 1998, the duration of treatment with these drugs was extended to seven months.

The primary end point of the study was the minimal luminal diameter at six months within the vessel treated by balloon angioplasty, as measured by quantitative coronary angiography, as a function of the delivered dose of beta radiation. The secondary end points were the incidences in the entire study population of the following serious cardiac events: death, myocardial infarction, percutaneous intervention in the target vessel (defined as any additional intervention within the treated vessel), and coronary-artery bypass grafting. Myocardial infarction was diagnosed when two of the following occurred: chest discomfort lasting at least 30 minutes, the development of abnormal new Q waves, and an increase in the level of creatine kinase or MB isoenzymes to more than twice the upper limit of normal.

Procedure for Radiation Therapy

The system used for intraarterial beta-radiation therapy has been described previously10-13; it consists of the yttrium-90 beta-ray–emitting source (half-life, 64 hours; maximal energy, 2.284 MeV), a centering balloon, and an automated delivery device. The radioactive source consists of a 29-mm-long, flexible coil, secured at the end of a 0.035-cm thrust wire between distal and proximal 6-mm-long, radiopaque tungsten markers, which allow precise localization of the source under fluoroscopy. The effective length of the vessel segment being irradiated (the 90 percent isodose line) is 24 mm. The centering balloon, consisting of four interconnected compartments, is designed to position the source wire centrally inside the coronary lumen, thereby contributing to a more homogeneous distribution of the dose of radiation along the vessel wall. Three radiopaque markers are located between the balloon compartments and allow the device to be positioned, with the use of fluoroscopy, at the exact site of the previous angioplasty. The centering balloon is inflated with 5 ml of carbon dioxide up to a maximal pressure of 4 atm. As compared with the use of contrast medium, the use of carbon dioxide permits faster inflation and deflation of the balloon and a shorter treatment time (because there is less attenuation of the radiation).11,12

The use of an automated delivery device ensured the safe storage of the source, its easy insertion and withdrawal, its accurate positioning, and the instantaneous calculation and delivery of the dose. At the end of each treatment, a printed report was automatically generated. After the completion of balloon angioplasty, the centering balloon, which had the same diameter as the angioplasty balloon, was positioned so that it would cover fully the site dilated by the angioplasty balloon. After the successful advancement and retrieval of a nonradioactive test wire, the yttrium-90 source was automatically advanced to the same site. This procedure was performed by a team composed of a cardiologist, an oncologist, and a medical physicist, who collaborated according to local practices and regulations. All teams applied the measures normally used in the interventional suite for protection from radiation.14

Angiographic Analysis

Coronary angiograms were obtained in multiple views after patients had received an intracoronary injection of nitrates. An independent core laboratory (Cardialysis, Rotterdam, the Netherlands), whose personnel were unaware of the dose of radiation associated with each angiogram, analyzed the angiograms quantitatively using edge-detection techniques.15 Coronary luminal diameter and degree of stenosis (as a percentage of the diameter) were measured before dilation, at the end of the procedure, and during follow-up angiography six months later (or earlier if there were recurrent symptoms). Restenosis was defined as the presence of stenosis of more than 50 percent of the luminal diameter. The loss in luminal diameter was calculated as the difference between the luminal diameter measured immediately after the procedure and that measured at six months. The entire irradiated segment of the vessel (24 mm long) and the edges of that segment (5 to 6 mm on each side of the segment) were analyzed. The resulting segment (34 to 36 mm long) encompassed the initial site of stenosis, the 20-mm segment injured by the angioplasty device, the 30-mm area affected by the centering balloon, and the immediately adjacent proximal and distal segments of the vessel (each 2 to 3 mm long). This method of analysis has been used in previous trials of vascular brachytherapy.15

Statistical Analysis

For a study with a power (1–β) of 90 percent, a one-sided type I error (α) of 0.05, and an expected minimal luminal diameter at follow-up of 1.67 mm (that achieved in the Benestent II trial5 among patients who underwent balloon angioplasty alone), with a standard deviation of 27.5 percent, we calculated that we had to enroll 34 patients in each group in order to detect a 20 percent improvement in the minimal luminal diameter at six months (2.00 vs. 1.67 mm). In order to account for an expected 30 percent rate of stent implantation, we increased the study population to 180 patients (45 in each of four groups).

For the comparison of binary variables, a Fisher's exact test for two groups was used, or the chi-square test was used when applicable. For the comparison of continuous variables, two-tailed Student's t-tests were used. The data are expressed as means ±SE.

Results

Patient Population

Of 183 patients who underwent randomization, 181 (mean age, 64±0.6 years; 74 percent male) received the prescribed doses of beta radiation: 45 received 9 Gy, 45 received 12 Gy, 46 received 15 Gy, and 45 received 18 Gy. Radiation therapy was not administered in two cases (in one due to a technical failure, and in one because the angioplasty was unsuccessful). There were no significant differences among the dose groups in terms of the demographic characteristics of the patients or the characteristics of the lesions (Table 1Table 1Characteristics of the Patients.). Balloon angioplasty alone was performed in 130 patients (72 percent). Stent implantation following brachytherapy was required in 51 patients (in 47 percent of these patients because of residual stenosis of more than 50 percent of luminal diameter, and in the remainder because of major dissection).

Angiographic Results

The mean degree of stenosis before the angioplasty procedure, expressed as a percentage of the vessel diameter, was 65±1.5, 64±1.6, 66±1.3, and 65±1.7 in the 9-Gy, 12-Gy, 15-Gy, and 18-Gy groups, respectively. There were no significant differences among the groups in the mean degree of residual stenosis after balloon angioplasty or in the postprocedural minimal luminal diameter (Table 2Table 2Angiographic Results Immediately after Balloon Angioplasty and at Six Months.). The mean duration of radiation treatment was 1.81±0.10, 2.55±0.20, 3.01± 0.16, and 3.17±0.19 minutes, respectively, in the four groups. Ten patients who were asymptomatic at six months declined to undergo a second catheterization; one patient died after having an acute myocardial infarction; and lung cancer was diagnosed in another patient. The full complement of angiograms was thus available for analysis in 169 patients; all three angiograms — preprocedural, postprocedural, and follow-up — were suitable for quantitative analysis in 168 of these patients.

The largest loss in minimal luminal diameter (from the postprocedural result to follow-up) occurred within the irradiated segment of the vessel (Figure 1Figure 1Loss in Minimal Luminal Diameter per Vessel Segment from Immediately after Angioplasty to Follow-up in the 168 Patients for Whom Angiographic Follow-up Data Were Available at Six Months.). Six months after balloon angioplasty, a significant dose-dependent benefit of beta radiation was evident in the minimal luminal diameter (P=0.006 for the comparison of 9 Gy with 18 Gy); the mean diameter for patients who had balloon angioplasty alone was 1.67± 0.09, 1.82±0.13, 1.80±0.11, and 2.10±0.12 mm, respectively (Table 2). The loss in luminal diameter was 0.31±0.08 mm in the 9-Gy group, 0.12±0.09 mm in the 12-Gy group, and 0.09±0.10 mm in the 15-Gy group (Figure 2Figure 2Cumulative Frequency Distribution of the Loss in Minimal Luminal Diameter from Immediately after Angioplasty to Follow-up at Six Months in Patients Treated with Angioplasty and Irradiation with 9, 12, 15, and 18 Gy of Beta Radiation.). After irradiation at the 18-Gy dose, the luminal diameter actually increased by 0.04± 0.10 mm (P=0.008 for the comparison of 9 Gy with 18 Gy). Luminal enlargement occurred in 28 percent, 50 percent, 45 percent, and 74 percent of the patients in the 9-Gy, 12-Gy, 15-Gy, and 18-Gy groups, respectively (P<0.001 for the comparison of 9 Gy with 18 Gy). On follow-up angiographic examination, there was occlusion of the vessel in 4 of the 120 patients who had had balloon angioplasty alone (3.3 percent) and in 7 of the 49 who had also required stents (14.3 percent). Of the patients with occlusion, four (one treated with balloon angioplasty alone and three who received stents) presented with acute symptoms (at 7 days and at 10, 11, and 12 weeks after irradiation). None of these patients were receiving long-term ticlopidine treatment. As a consequence of the absence of ticlopidine therapy, the rates of angiographic restenosis in the patients who required stents were exceedingly high: 30 percent, 33 percent, 15 percent, and 38 percent, respectively (22 percent, 20 percent, 8 percent, and 20 percent, respectively, after the exclusion of the patients with stent thrombosis).

Clinical Results

The incidence of serious cardiac events was within the range of reported values and did not differ significantly among the dose groups (Table 3Table 3Incidence of Serious Cardiac Events at 210 Days.). The overall incidence of myocardial infarction was 3.3 percent. In addition, three patients had an isolated rise in the creatine phosphokinase level. Repeated percutaneous revascularization of the target vessel was required in 12 percent of the patients treated with balloon angioplasty alone and in 17.6 percent of the patients who required stents.

Discussion

This dose-finding study demonstrates a marked reduction in the rate of restenosis in nonstented arteries after beta-radiation therapy at the 18-Gy level (at a tissue depth of 1 mm). No device or pharmaceutical approach has yielded similarly low rates of restenosis after balloon angioplasty alone. The lowest rate of restenosis reported in the context of a randomized trial was 16 percent at six months with the use of coronary stents,5 which are currently implanted during more than 60 percent of coronary interventions.16 With the use of beta radiation, a restenosis rate of less than 5 percent was achieved in patients who had had good angiographic results after balloon angioplasty alone, suggesting that beta-radiation therapy should be evaluated as a first-line adjunct to angioplasty. The effect observed in our study results from the effect of irradiation on both remodeling and the formation of neointima, as suggested by intravascular ultrasound imaging.17,18

Several systems for endovascular irradiation with sources of gamma or beta radiation have been developed. Gamma radiation was shown to decrease the rates of restenosis after the percutaneous treatment of restenotic lesions, but its effectiveness at the time of the first angioplasty has not been evaluated.19-21 Moreover, the use of gamma radiation is less practical than that of beta radiation in the environment of a catheterization laboratory, because the medical staff must be exposed to greater radiation and the treatment times must be longer.

Results with other systems of beta radiation have not been as encouraging as our results. The initial analysis of the Beta Energy Restenosis Trial (BERT), which used a noncentered strontium-90–yttrium-90 source, found a 15 percent rate of restenosis when the angiographic analysis was limited to the stenotic segment.22 However, more recent reports using either strontium-90–yttrium-90 or phosphorus-32 found higher rates of restenosis (as high as 25 percent) when a longer segment of the vessel was analyzed, as in our study.23

Our data also indicate that caution should be exercised when combining radiation therapy with the implantation of stents. Historically, the occurrence of subacute stent thrombosis was reduced from 25 percent to less than 2 percent when patients received treatment with aspirin and ticlopidine for four weeks after implantation.24-26 After irradiation, there was abrupt thrombosis or late occlusion of vessels in 14.3 percent of patients with stents, a finding that is consistent with other reports.27,28 By delaying endothelialization after the implantation of a stent, radiation therapy may extend the risk of stent thrombosis beyond four weeks. Further studies addressing the safety of combining radiation therapy with stenting are necessary; the long-term use of antiplatelet drugs will presumably be required.

The use of radiation therapy for the treatment of a nonmalignant disease may cause concern about several potential problems. One is radiation-induced arteriopathy, an arterial narrowing that occurs after external fractionated radiation treatment for malignant diseases. This complication occurs after a mean of 16 years, and it occurs more frequently in younger patients in whom large volumes of tissue have been irradiated.29 It is unknown whether this adverse effect can occur several years after the irradiation of a small volume of coronary artery. Another concern is that remodeling of the vessel might ultimately result in the formation of coronary aneurysms. Despite experimental30 and clinical31 evidence that radiation may induce aneurysms, we did not observe this phenomenon. We believe that the aneurysmal dilatations observed in earlier studies were related to high-dose gamma radiation of large volumes of tissue,32 but longer follow-up is certainly warranted.

Another reason for concern is that radiation may induce a soft-tissue sarcoma in the arterial wall or the myocardium. Previous experience with radiation therapy for breast cancer33 suggests that the maximal risk of carcinogenesis during the 10-year period after intravascular radiation therapy is in the range of 1 case per 1 million patients treated, given the small volume of irradiated tissue. Radiation-induced cancers were not found in large groups of patients who had received radiation treatment for keloids or pterygia with focused doses of 12 and 18 Gy.34,35

Our principal findings have important implications for the field of interventional cardiology. Current practice follows the “bigger is better” paradigm, which holds that the bigger the luminal diameter achieved by angioplasty, the better the long-term result.36 A larger lumen can indeed be achieved safely and reliably with the implantation of a stent. In this respect, stents are superior to balloons, although they do not inhibit the neointimal proliferation that causes restenosis. Intracoronary beta radiation has the potential to change this treatment paradigm, not only because it reduces neointimal proliferation but also because of its beneficial effect on the healing process in the arterial wall. In a similar fashion to its effect on the healing surgical wound,37,38 beta radiation decreases chronic arterial constriction and induces luminal enlargement, as we have shown. Further randomized trials should therefore test the clinical efficacy of the combination of beta-radiation therapy and balloon angioplasty, as compared with the implantation of coronary stents, in improving long-term, event-free survival after percutaneous revascularization.

Supported in part by a grant from the G. and S. Prevot Foundation, Switzerland (to Dr. Popowski); by a grant from the Swiss National Science Foundation (3200-049849, to Dr. Verin and Dr. Popowski); and by the Wenckeback prize from the Dutch Heart Foundation (to Dr. Serruys).

Boston Scientific Schneider provided the delivery devices and the radioactive sources used in the study but was not involved in the analysis and interpretation of the data or in the preparation of the manuscript. Dr. Verin and Dr. Popowski are the inventors of the Beta-Med irradiation system and earn royalties from the sponsoring company.

Source Information

From the University Hospital, Geneva (V.V., Y.P.); the Cardiovascular Center, Onze-Lieve-Vrouw Ziekenhuis, Aalst, Belgium (B.B., W.W.); the University Hospital, Essen, Germany (D.B., W.S.); the University Hospital, Kiel, Germany (M.L., G.K.); King's College Hospital, London (M.T., F.C.); and Cardialysis and Erasmus University, Rotterdam, the Netherlands (C.D., P.W.S.).

Address reprint requests to Dr. Wijns at the Cardiovascular Center, OLV Hospital, B 9300 Aalst, Belgium, or at .

Other members of the Dose-Finding Study Group are listed in the Appendix.

Appendix

Other institutions and investigators that participated in the Dose-Finding Study Group are as follows (the numbers of patients enrolled are given in parentheses): University Hospital, Geneva (57) — I. Papirov, A. Sergey, P. Debruyne, and J. Ramos de Olival; Cardiovascular Center, Onze-Lieve-Vrouw Ziekenhuis, Aalst, Belgium (54) — G. Heyndrickx, L. Verbeke, and J. De Jans; University Hospital, Essen, Germany (26) — M. Haude, D. Flühs, U. Quast, A. Müller, K. Hidgeghty, and C. von Birgelen; University Hospital, Kiel, Germany (22) — M. Thomas, G. Herrmann, R. Wilhelm, and P. Kohl; King's College Hospital, London (22) — N. Lewis; Study Coordination (Boston Scientific) — T. Thaler; Critical Events Committee — J. Dekkers; Angiographic Core Laboratory and Data Analysis — Y. Teunissen, A. Spierings, C. van der Wiel, and G. Kloek.

References

References

  1. 1

    Bittl JA. Advances in coronary angioplasty. N Engl J Med 1996;335:1290-1302[Erratum, N Engl J Med 1997;336:670.]
    Full Text | Web of Science | Medline

  2. 2

    Mak KH, Topol EJ. Clinical trials to prevent restenosis after percutaneous coronary revascularization. Ann N Y Acad Sci 1997;811:255-284
    CrossRef | Web of Science | Medline

  3. 3

    Serruys PW, de Jaegere P, Kiemeneij F, et al. A comparison of balloon-expandable-stent implantation with balloon angioplasty in patients with coronary artery disease. N Engl J Med 1994;331:489-495
    Full Text | Web of Science | Medline

  4. 4

    Fischman DL, Leon MB, Baim DS, et al. A randomized comparison of coronary-stent placement and balloon angioplasty in the treatment of coronary artery disease. N Engl J Med 1994;331:496-501
    Full Text | Web of Science | Medline

  5. 5

    Serruys PW, van Hout B, Bonnier H, et al. Randomised comparison of implantation of heparin-coated stents with balloon angioplasty in selected patients with coronary artery disease. Lancet 1998;352:673-681
    CrossRef | Web of Science | Medline

  6. 6

    Wiedermann JG, Marboe C, Amols H, Schwartz A, Weinberger J. Intracoronary irradiation markedly reduces restenosis after balloon angioplasty in a porcine model. J Am Coll Cardiol 1994;23:1491-1498
    CrossRef | Web of Science | Medline

  7. 7

    Verin V, Popowski Y, Urban P, et al. Intraarterial beta irradiation prevents neointimal hyperplasia in a hypercholesterolemic rabbit restenosis model. Circulation 1995;92:2284-2290
    Web of Science | Medline

  8. 8

    Waksman R, Robinson KA, Crocker IR, et al. Intracoronary low-dose beta-irradiation inhibits neointima formation after coronary artery balloon injury in the swine restenosis model. Circulation 1995;92:3025-3031
    Web of Science | Medline

  9. 9

    Waksman R, Rodriguez JC, Robinson KA, et al. Effect of intravascular irradiation on cell proliferation, apoptosis and vascular remodeling after balloon overstretch injury of porcine coronary arteries. Circulation 1997;96:1944-1952
    Web of Science | Medline

  10. 10

    Popowski Y, Verin V, Urban P, et al. Intra-arterial yttrium-90 brachytherapy for restenosis prevention. In: Bruggmoser G, Mould RF, eds. Brachytherapy review. Freiburg oncology series monograph no. 1. Freiburg, Germany: Albert-Ludwigs-University, 1994:163-5.

  11. 11

    Popowski Y, Verin V, Papirov I, et al. High dose rate brachytherapy for prevention of restenosis after percutaneous transluminal coronary angioplasty: preliminary dosimetric tests of a new source presentation. Int J Radiat Oncol Biol Phys 1995;33:211-215
    CrossRef | Web of Science | Medline

  12. 12

    Popowski Y, Verin V, Papirov I, et al. Intra-arterial 90-Y brachytherapy: preliminary dosimetric study using a specially modified angioplasty balloon. Int J Radiat Oncol Biol Phys 1995;33:713-717
    CrossRef | Web of Science | Medline

  13. 13

    Popowski Y, Verin V, Schwager M, et al. A novel system for intracoronary beta-irradiation: description and dosimetric results. Int J Radiat Oncol Biol Phys 1996;36:923-931
    CrossRef | Web of Science | Medline

  14. 14

    Verin V, Urban P, Popowski Y, et al. Feasibility of intracoronary beta-irradiation to reduce restenosis after balloon angioplasty: a clinical pilot study. Circulation 1997;95:1138-1144
    Web of Science | Medline

  15. 15

    Sabate M, Costa MA, Kozuma K, et al. Methodological and clinical implications of the relocation of the minimal luminal diameter after intracoronary radiation therapy. J Am Coll Cardiol 2000;36:1536-1541
    CrossRef | Web of Science | Medline

  16. 16

    Topol EJ. Coronary-artery stents -- gauging, gorging, and gouging. N Engl J Med 1998;339:1702-1704
    Full Text | Web of Science | Medline

  17. 17

    Meerkin D, Tardif JC, Crocker IR, et al. Effects of intracoronary beta-radiation therapy after coronary angioplasty: an intravascular ultrasound study. Circulation 1999;99:1660-1665
    Web of Science | Medline

  18. 18

    Sabate M, Serruys PW, van der Giessen WJ, et al. Geometric vascular remodeling after balloon angioplasty and β-radiation therapy: a three-dimensional intravascular ultrasound study. Circulation 1999;100:1182-1188
    Web of Science | Medline

  19. 19

    Teirstein PS, Massullo V, Jani S, et al. Catheter-based radiotherapy to inhibit restenosis after coronary stenting. N Engl J Med 1997;336:1697-1703
    Full Text | Web of Science | Medline

  20. 20

    Waksman R, White RL, Chan RC, et al. Intracoronary gamma-radiation therapy after angioplasty inhibits recurrence in patients with in-stent restenosis. Circulation 2000;101:2165-2171
    Web of Science | Medline

  21. 21

    Leon MB, Teirstein PS, Lansky AJ, et al. Intracoronary gamma radiation to reduce in-stent restenosis: the Multicenter Gamma I Randomized Clinical Trial. J Am Coll Cardiol 1999;33:19A-19A abstract.

  22. 22

    King SB III, Williams DO, Chougule P, et al. Endovascular beta-radiation to reduce restenosis after coronary balloon angioplasty: results of the Beta Energy Restenosis Trial (BERT). Circulation 1998;97:2025-2030
    Web of Science | Medline

  23. 23

    Raizner AE, Oesterle SN, Waksman R, et al. Inhibition of restenosis with beta-emitting radiotherapy: report of the Proliferation Reduction with Vascular Energy Trial (PREVENT). Circulation 2000;102:951-958
    Web of Science | Medline

  24. 24

    Serruys PW, Strauss BH, Beatt KJ, et al. Angiographic follow-up after placement of a self-expanding coronary-artery stent. N Engl J Med 1991;324:13-17
    Full Text | Web of Science | Medline

  25. 25

    Colombo A, Hall P, Nakamura S, et al. Intracoronary stenting without anticoagulation accomplished with intravascular ultrasound guidance. Circulation 1995;91:1676-1688
    Web of Science | Medline

  26. 26

    Schomig A, Neumann F-J, Kastrati A, et al. A randomized comparison of antiplatelet and anticoagulant therapy after the placement of coronary-artery stents. N Engl J Med 1996;334:1084-1089
    Full Text | Web of Science | Medline

  27. 27

    Costa MA, Sabat M, van der Giessen WJ, et al. Late coronary occlusion after intracoronary brachytherapy. Circulation 1999;100:789-792
    Web of Science | Medline

  28. 28

    Waksman R, Bhargava B, Mintz GS, et al. Late total occlusion after intracoronary brachytherapy for patients with in-stent restenosis. J Am Coll Cardiol 2000;36:65-68
    CrossRef | Web of Science | Medline

  29. 29

    Stewart JR, Fajardo LF, Gillette SM, Constine LS. Radiation injury to the heart. Int J Radiat Oncol Biol Phys 1995;31:1205-1211
    CrossRef | Web of Science | Medline

  30. 30

    Gillette EL, Powers BE, McChesney SL, Park RD, Withrow SJ. Response of aorta and branch arteries to experimental intraoperative irradiation. Int J Radiat Oncol Biol Phys 1989;17:1247-1255
    CrossRef | Web of Science | Medline

  31. 31

    Condado JA, Waksman R, Gurdiel O, et al. Long-term angiographic and clinical outcome after percutaneous transluminal coronary angioplasty and intracoronary radiation therapy in humans. Circulation 1997;96:727-732
    Web of Science | Medline

  32. 32

    Serruys PW, Levendag PC. Intracoronary brachytherapy: the death knell of restenosis or just another episode of a never-ending story? Circulation 1997;96:709-712
    Web of Science | Medline

  33. 33

    Taghian A, Vathaire F, Terrier P, et al. Long-term risk of sarcoma following radiation treatment for breast cancer. Int J Radiat Oncol Biol Phys 1991;21:361-367
    CrossRef | Web of Science | Medline

  34. 34

    Kovalic JJ, Perez CA. Radiation therapy following keloidectomy: a 20-year experience. Int J Radiat Oncol Biol Phys 1989;17:77-80
    Web of Science | Medline

  35. 35

    Wilder RB, Buatti JM, Kittelson JM, et al. Pterygium treated with excision and postoperative beta irradiation. Int J Radiat Oncol Biol Phys 1992;23:533-537
    CrossRef | Web of Science | Medline

  36. 36

    Kuntz RE, Gibson CM, Nobuyoshi M, Baim DS. Generalized model of restenosis after conventional balloon angioplasty, stenting and directional atherectomy. J Am Coll Cardiol 1993;21:15-25
    CrossRef | Web of Science | Medline

  37. 37

    Gorodetsky R, McBride WH, Withers HR. Assay of radiation effects in mouse skin as expressed in wound healing. Radiat Res 1988;116:135-144
    CrossRef | Web of Science | Medline

  38. 38

    Vegesna V, Withers HR, Holly FE, McBride WH. The effect of local and systemic irradiation on impairment of wound healing in mice. Radiat Res 1993;135:431-433
    CrossRef | Web of Science | Medline

Citing Articles (84)

Citing Articles

  1. 1

    Nathan Steinle, Mark Barakat, Darius Moshfeghi, Peter K Kaiser. (2011) Radiation therapy in the treatment of exudative age-related macular degeneration. Expert Review of Ophthalmology 6:3, 323-337
    CrossRef

  2. 2

    Hee Hwa Ho, On Hing Kwok, Man Hong Jim, Chung Wah Siu, Vincent Pong, Wing Hing Chow. (2011) Long-term clinical outcomes after intravascular brachytherapy for instent restenosis and de novo coronary artery lesions in percutaneous coronary intervention. Cardiovascular Revascularization Medicine 12:3, 152-157
    CrossRef

  3. 3

    Jin Cheng, Yoshinobu Onuma, Jannet Eindhoven, Peter Levendag, Patrick Serruys, Ron van Domburg, Wim van der Giessen. (2011) Late outcome after intracoronary beta radiation brachytherapy: a matched-propensity controlled ten-year follow-up study. EuroIntervention 6:6, 695-702
    CrossRef

  4. 4

    Ron Waksman. (2011) Vascular brachytherapy: a landmark chapter in the field of interventional cardiology. EuroIntervention 6:6, 669-671
    CrossRef

  5. 5

    Martin Unverdorben, Franz X. Kleber, Hubertus Heuer, Hans-Reiner Figulla, Christian Vallbracht, Matthias Leschke, Bodo Cremers, Stefan Hardt, Michael Buerke, Hanns Ackermann, Michael Boxberger, Ralf Degenhardt, Bruno Scheller. (2010) Treatment of small coronary arteries with a paclitaxel-coated balloon catheter. Clinical Research in Cardiology 99:3, 165-174
    CrossRef

  6. 6

    Joseph W. Villard, Amit S. Paranjape, Danielle A. Victor, Marc D. Feldman. (2009) Applications of optical coherence tomography in cardiovascular medicine, Part 2. Journal of Nuclear Cardiology 16:4, 620-639
    CrossRef

  7. 7

    Michail I. Papafaklis, Christos V. Bourantas, Panagiotis E. Theodorakis, Christos S. Katsouras, Dimitrios I. Fotiadis, Lampros K. Michalis. (2009) Relationship of shear stress with in-stent restenosis: Bare metal stenting and the effect of brachytherapy. International Journal of Cardiology 134:1, 25-32
    CrossRef

  8. 8

    Daniel H. Steinberg, Michael A. Gaglia, Tina L. Pinto Slottow, Probal Roy, Laurent Bonello, Axel De Labriolle, Gilles Lemesle, Rebecca Torguson, Kimberly Kineshige, Zhenyi Xue, William O. Suddath, Kenneth M. Kent, Lowell F. Satler, Augusto D. Pichard, Joseph Lindsay, Ron Waksman. (2009) Outcome Differences With the Use of Drug-Eluting Stents for the Treatment of In-Stent Restenosis of Bare-Metal Stents Versus Drug-Eluting Stents. The American Journal of Cardiology 103:4, 491-495
    CrossRef

  9. 9

    Walter A. Wohlgemuth, Giesbert Leissner, Hermann Wengenmair, Klaus Bohndorf, Klaus Kirchhof. (2008) Endovascular Brachytherapy in the Femoropopliteal Segment Using 192Ir and 188Re. CardioVascular and Interventional Radiology 31:4, 698-708
    CrossRef

  10. 10

    Ikuo Katayama, Yuka Hotokezaka, Toshifumi Matsuyama, Tadateru Sumi, Takashi Nakamura. (2008) Ionizing Radiation Induces Macrophage Foam Cell Formation and Aggregation Through JNK-Dependent Activation of CD36 Scavenger Receptors. International Journal of Radiation Oncology*Biology*Physics 70:3, 835-846
    CrossRef

  11. 11

    Tae Kyoon Kim, Yong Joo Kim, Chan Seok Park, Hun-Jun Park, Dong-Bin Kim, Sung-Won Jang, Pum-Joon Kim, Hae-Ok Jung, Sang Hong Baek, Ki Bae Seung, Kyu-Bo Choi. (2008) Effect of High Dose External Irradiation on the Matrix Metalloprotease-2 Expression in a Rat Carotid Artery Injury Model. Korean Circulation Journal 38:4, 212
    CrossRef

  12. 12

    PRAMOD KUCHULAKANTI, RON WAKSMAN. 2008. Intracoronary brachytherapy. , 197-217.
    CrossRef

  13. 13

    Imad Sheiban, Gian Paolo Ballari, Claudio Moretti, Walter Grosso Marra, Emanuele Meliga, Pierluigi Omedè, Filippo Sciuto, Gian Paolo Trevi. (2007) Paclitaxel-eluting stents for the treatment of complex coronary lesions: immediate and 12-month results. Journal of Cardiovascular Medicine 8:8, 582-588
    CrossRef

  14. 14

    Ulrich Sigwart. (2007) Drug-Eluting Stents: Some Thoughts From Old Europe. The American Heart Hospital Journal 5:3, 135-137
    CrossRef

  15. 15

    Charles A. Herzog, Khalid Ashai. 2007. Percutaneous Coronary Revascularization in Patients with End-Stage Renal Disease. , 19-50.
    CrossRef

  16. 16

    Valeria Ferrero, Flavio Ribichini, Marleen Piessens, Guy R. Heyndrickx, Luc Verbeke, Bernard de Bruyne, Mauro Feola, Corrado Vassanelli, William Wijns. (2007) Intracoronary β-irradiation for the treatment of de novo lesions: 5-Year clinical follow-up of the BetAce randomized trial. American Heart Journal 153:3, 398-402
    CrossRef

  17. 17

    Dimitrios N. Nikas, John Kalef-Ezra, Christos S. Katsouras, Periklis Tsekeris, George Bozios, Christos Pappas, Katerina K. Naka, Anna Kotsia, Nikolaos Papamichael, Dimitrios A. Sideris, Lampros K. Michalis. (2007) Long-term clinical outcome of patients treated with β-brachytherapy in routine clinical practice. International Journal of Cardiology 115:2, 183-189
    CrossRef

  18. 18

    Karel K.J. Kuiper, Mohamed Salem, Svein Rotevatn, Joseph Mills, Jan Erik Nordrehaug. (2007) Implementing a best-treatment strategy with intracoronary brachytherapy for in-stent restenosis in patients at high risk for recurrence. Cardiovascular Revascularization Medicine 8:1, 9-14
    CrossRef

  19. 19

    Erik Van Limbergen, Prabhakar Trepuraneni. (2007) Is this the swan song of endovascular brachytherapy?. Radiotherapy and Oncology 82:1, 1-4
    CrossRef

  20. 20

    S Misra, R Bonan, T Pflederer, P Roy-Chaudhury. (2006) BRAVO I: A pilot study of vascular brachytherapy in polytetrafluoroethylene dialysis access grafts. Kidney International
    CrossRef

  21. 21

    HIDEAKI KANEDA, YASUHIRO HONDA, YOSHIHIRO MORINO, ALEXANDRA J. LANSKY, PAUL G. YOCK, RAOUL BONAN, PETER J. FITZGERALD. (2006) Predictors of Recurrent In-Stent Restenosis after Beta-Radiation: An Analysis from the START 40/20 Trial. Journal of Interventional Cardiology 19:5, 376-380
    CrossRef

  22. 22

    Eric Ducasse, Jacques Chevalier, Jean-Marc Cosset, Colette Creusy, François Eschwege, Francesco Speziale, Enrico Sbarigia, Dominique Midy, Jean-Claude Baste, Eric Lartigau. (2006) Ionizing Radiation to Prevent Arterial Intimal Hyperplasia At the Edges of the Stent: Induces Necrosis and Fibrosis. Journal of Surgical Research 135:2, 331-336
    CrossRef

  23. 23

    Klaus Reynen, Joachim Kropp, Ulrich K??ckeritz, Gerd Wunderlich, Furn F. (Russ) Knapp, Alexander Schmeisser, Ruth H. Strasser. (2006) Intracoronary radiotherapy with a 188Rhenium liquid-filled angioplasty balloon system in in-stent restenosis: a single-center, prospective, randomized, placebo-controlled, double-blind evaluation. Coronary Artery Disease 17:4, 371-377
    CrossRef

  24. 24

    Ulrich Sigwart. (2006) Drug-Eluting Stents Are Safe and Effective. Journal of the American College of Cardiology 47:7, 1361-1362
    CrossRef

  25. 25

    David de la Croix de Ravignan, Colette Creusy, Jacques Chevalier, Jean-Marc Cosset, François Eschwege, Paul Puppinck, Francesco Speziale, Paolo Fiorani, Eric Lartigau, Eric Ducasse. (2005) Chronological Changes in Morphometry and Histology in the Rabbit Vascular Wall after External Radiation for the Prevention of Intimal Hyperplasia. Journal of Surgical Research 128:1, 120-125
    CrossRef

  26. 26

    Raúl Moreno. (2005) Stents recubiertos y otros dispositivos antirreestenosis. Revista Española de Cardiología 58:7, 842-862
    CrossRef

  27. 27

    Pilar Jiménez-Quevedo, Manel Sabaté. (2005) Percutaneous coronary revascularization in diabetics: from balloon angioplasty to drug-eluting stents. Expert Review of Cardiovascular Therapy 3:4, 635-646
    CrossRef

  28. 28

    P. Wexberg, C. Kirisits, D. Berger, I. Sulzbacher, G. Maurer, R. Potter, D. Georg, D. Glogar. (2005) Quantification of dose perturbation by plaque in vascular brachytherapy. European Journal of Clinical Investigation 35:3, 180-185
    CrossRef

  29. 29

    (2004) REFERENCES. Journal of the ICRU 4:2, 165-175
    CrossRef

  30. 30

    Thomas M Schiele, Florian Krötz, Volker Klauss. (2004) Vascular restenosis – striving for therapy. Expert Opinion on Pharmacotherapy 5:11, 2221-2232
    CrossRef

  31. 31

    Ganesan Karthikeyan, Balram Bhargava. (2004) Prevention of restenosis after coronary angioplasty. Current Opinion in Cardiology 19:5, 500-509
    CrossRef

  32. 32

    Benjamin Thierry, Françoise M Winnik, Yahye Merhi, Jim Silver, Maryam Tabrizian. (2004) Radionuclides-hyaluronan-conjugate thromboresistant coatings to prevent in-stent restenosis. Biomaterials 25:17, 3895-3905
    CrossRef

  33. 33

    Ron Waksman. (2004) Vascular Brachytherapy and Coronary Stenting for De Novo Lesions. Journal of the American College of Cardiology 44:3, 538-540
    CrossRef

  34. 34

    Manel Sabaté, Gela Pimentel, Carlos Prieto, José María Corral, Camino Bañuelos, Dominick J. Angiolillo, Fernando Alfonso, Rosana Hernández-Antolín, Javier Escaned, Panayotis Fantidis, Cristina Fernández, Antonio Fernández-Ortiz, Raúl Moreno, Carlos Macaya. (2004) Intracoronary Brachytherapy After Stenting De Novo Lesions in Diabetic Patients. Journal of the American College of Cardiology 44:3, 520-527
    CrossRef

  35. 35

    Patrick W. Serruys, William Wijns, Georgios Sianos, Ivan de Scheerder, Paul A. van den Heuvel, Wolfgang Rutsch, Helmut D. Glogar, Carlos Macaya, Pierre H. Materne, Susan Veldhof, Heike Vonhausen, Patricia C. Otto-Terlouw, Wim J. van der Giessen. (2004) Direct Stenting Versus Direct Stenting Followed by Centered Beta-Radiation With Intravascular Ultrasound-Guided Dosimetry and Long-Term Anti-Platelet Treatment. Journal of the American College of Cardiology 44:3, 528-537
    CrossRef

  36. 36

    RAFAEL BEYAR. (2004) Novel Approaches to Reduce Restenosis. Annals of the New York Academy of Sciences 1015:1, 367-378
    CrossRef

  37. 37

    Caroline A Kavanagh, Yuri A Rochev, William M Gallagher, Kenneth A Dawson, Alan K Keenan. (2004) Local drug delivery in restenosis injury: thermoresponsive co-polymers as potential drug delivery systems. Pharmacology & Therapeutics 102:1, 1-15
    CrossRef

  38. 38

    Eric Ducasse, Collette Creusy, Jocelyne Mazurier, Jean-Marc Cosset, François Eschwege, Paul Puppinck, Francesco Speziale, Enrico Sbarigia, Paolo Fiorani, Eric Lartigau. (2004) Morphometric and histological changes in the vascular wall after external radiation for the prevention of intimal hyperplasia. Journal of Surgical Research 117:2, 316-322
    CrossRef

  39. 39

    Dominick J. Angiolillo, Manel Sabatá, Fernando Alfonso, Carlos Macaya. (2004) “Candy wrapper” effect after drug-eluting stent implantation: Déjà vu or stumbling over the same stone again?. Catheterization and Cardiovascular Interventions 61:3, 387-391
    CrossRef

  40. 40

    Balram Bhargava, Ganesan Karthikeyan, Prabhakar Tripuraneni. (2004) Intravascular Brachytherapy. American Journal of Cardiovascular Drugs 4:6, 385-394
    CrossRef

  41. 41

    Kwang-Il Kim, Jangwhan Bae, Hyun-Jae Kang, Bon-Kwon Koo, Tae-Jin Youn, Sang-Hyun Kim, In-Ho Chae, Hyo-Soo Kim, Dae-Won Sohn, Byung-Hee Oh, Myoung-Mook Lee, Young-Bae Park, Yun-Shik Choi, Dong-Soo Lee. (2004) Three-Year Clinical Follow-up Results of Intracoronary Radiation Therapy Using a Rhenium-188-Diethylene-Triamine-Penta-Acetic-Acid-Filled Balloon System. Circulation Journal 68:6, 532-537
    CrossRef

  42. 42

    Eric Ducasse, Jean-Marc Cosset, François Eschwege, Colette Creusy, Jacques Chevalier, Paul Puppinck, Eric Lartigau. (2004) External Beam Ionizing Radiation for Inhibition of Myointimal Hyperplasia after Dilatation and Anastomoses: Experimental Models and Results. Annals of Vascular Surgery 18:1, 108-114
    CrossRef

  43. 43

    E Nikolsky, L Gruberg, E Rosenblatt, E Grenadier, M Boulos, Z Bernstein, A Huber, R Gitman, R Bar‐Deroma, W Markiewicz, R Beyar. (2004) Chronic total occlusion due to diffuse in‐stent restenosis: is brachytherapy the solution?. Acute Cardiac Care 6:1, 33-38
    CrossRef

  44. 44

    Ariel Finkelstein, Joerg Hausleiter, Terence Doherty, Kaname Takizawa, Jonathan Bergman, Ming Liu, Vladimir Rukshin, Michael Fishbein, Neal Eigler, Prediman Shah, Tripathi Rajavashisth, Raj Makkar. (2004) Intracoronary β‐irradiation enhances balloon‐injury‐induced tissue factor expression in the porcine injury model. Acute Cardiac Care 6:1, 20-27
    CrossRef

  45. 45

    Michael Maeng, Martin Busk, Kari Tanderup, Henrik Mertz, Henning Rud Andersen, Leif Thuesen. (2003) Catheter-based32P ?-radiation after stent implantation in porcine coronary arteries: Role of source-centering and geographical miss. Catheterization and Cardiovascular Interventions 60:2, 247-257
    CrossRef

  46. 46

    E. Regar, A. Thury, W.J. van der Giessen, G. Sianos, J. Vos, P.C. Smits, S.G. Carlier, P. de Feyter, D.P. Foley, Patrick W. Serruys. (2003) Sonotherapy, antirestenotic therapeutic ultrasound in coronary arteries: The first clinical experience. Catheterization and Cardiovascular Interventions 60:1, 9-17
    CrossRef

  47. 47

    Francis Q. Almeda, David Y. Chua, Sandeep Nathan, Susie Kim, Peter M. Meyer, Cam Nguyen, James C.H. Chu, Clifford J. Kavinsky, R. Jeffrey Snell, Gary L. Schaer. (2003) Correlates of failure following treatment with Sr-90 beta irradiation for in-stent restenosis. Catheterization and Cardiovascular Interventions 59:2, 176-183
    CrossRef

  48. 48

    E Charpentier, N Danchin, C Edlinger, J.P Perrin, S Baffert, F Livinec, N Jakobi-Rodrigues, D Couturier, E Fery-Lemonnier. (2003) Curiethérapie endocoronaire : une méta-analyse. Annales de Cardiologie et d'Angéiologie 52:3, 162-168
    CrossRef

  49. 49

    Tapas Das, Sudipta Chakraborty, Sharmila Banerjee, Grace Samuel, H. D. Sarma, Meera Venkatesh, M. R. A. Pillai. (2003) Preparation and animal biodistribution of166Ho labeled DOTA for possible use in intravascular radiation therapy (IVRT). Journal of Labelled Compounds and Radiopharmaceuticals 46:3, 197-209
    CrossRef

  50. 50

    Robert Bonvini, Iris Baumgartner, Do Dai Do, Mario Alerci, Jeanne-Marie Segatto, Paolo Tutta, Kurt Jäger, Markus Aschwanden, Ernst Schneider, Beatrice Amann-Vesti, Richard Greiner, Felix Mahler, Augusto Gallino. (2003) Late acute thrombotic occlusion after endovascular brachytherapy and stenting of femoropopliteal arteries. Journal of the American College of Cardiology 41:3, 409-412
    CrossRef

  51. 51

    Ron Waksman. (2003) Vascular Brachytherapy: Is It Still a Viable Option?. Journal of Vascular and Interventional Radiology 14:2, P79-P83
    CrossRef

  52. 52

    Jiro Aoki, Yuji Ikari, Tokuichiro Sugimoto, Sachito Fukuda, Kazuhiro Hara. (2003) Clinical Outcome of Percutaneous Transluminal Coronary Rotational Atherectomy in Patients With End-Stage Renal Disease. Circulation Journal 67:7, 617-621
    CrossRef

  53. 53

    Georgios Sianos, William Wijns, Pim J de Feyter, Patrick W Serruys. (2003) Geographical miss during centered intracoronary beta‐radiation with 90Yttrium: incidence and implications for recurrence rates after vascular brachytherapy for de novo lesions. Acute Cardiac Care 5:4, 181-189
    CrossRef

  54. 54

    Weon Kim, Myung Ho Jeong, Ok Young Park, Jay Young Rhew, Hee Seung Bom, Sun Joo Choi, Kyung Bae Park, Eun Hee Kim, Ju Han Kim, Young Keun Ahn, Jong Tae Park, Jeong Gwan Cho, Jong Chun Park, Jung Chaee Kang. (2003) Effects of β-Radiation Using a Holmium-166 Coated Balloon on Neointimal Hyperplasia in a Porcine Coronary Stent Restenosis Model. Circulation Journal 67:7, 625-629
    CrossRef

  55. 55

    David Faxon. (2002) Restenosis: do we need to understand it to treat it?. Journal of the American College of Cardiology 40:12, 2090-2091
    CrossRef

  56. 56

    V RODRIGUEZ, J GROVE, S YELICH, D PEARSON, M STEIN, W PEVEC. (2002) Effects of Brachytherapy on Intimal Hyperplasia in Arteriovenous Fistulas in a Porcine Model. Journal of Vascular and Interventional Radiology 13:12, 1239-1246
    CrossRef

  57. 57

    E.C. Sims, M.T. Rothman, T.D. Warner, M.E.B. Powell. (2002) Coronary Artery Brachytherapy. Clinical Oncology 14:4, 313-326
    CrossRef

  58. 58

    Akiko Maehara, Neil S Patel, Louis B Harrison, Neil J Weissman, Anh B Bui, Han-Soo Kim, Andrew E Ajani, Marco T Castagna, Taya L McMillan, Nathan Yang, Rosanna Chan, Julliana Pisch, Harry Quan, Sou-Tung Chiu-Tsao, Ron Waksman, Gary S Mintz. (2002) Dose heterogeneity may not affect the neointimal proliferation after gamma radiation for in-stent restenosis. Journal of the American College of Cardiology 39:12, 1937-1942
    CrossRef

  59. 59

    William H. Frishman. (2002) Coronary In-Stent Restenosis: Is It Preventable?. Heart Disease 4:3, 139-140
    CrossRef

  60. 60

    Eugenia Nikolsky, Edward Rosenblatt, Ehud Grenadier, Monther Boulos, Ariel Roguin, Chanderashekhar Patil, Zvi Bernstein, Akiva Huber, Margalit Ben-Zvi, Raquel Bar-Deroma, Walter Markiewicz, Rafael Beyar. (2002) A prospective single-center registry for the use of intracoronary gamma radiation in patients with diffuse in-stent restenosis. Catheterization and Cardiovascular Interventions 56:1, 46-52
    CrossRef

  61. 61

    Monica Hansrani, Klaus Overbeck, Jonathan J Smout, Gerard P Stansby, Gerard P Stansby. 2002. Intravascular brachytherapy for peripheral vascular disease. .
    CrossRef

  62. 62

    G Patrick Clagett. (2002) What’s new in vascular surgery. Journal of the American College of Surgeons 194:2, 165-201
    CrossRef

  63. 63

    Christoph Hehrlein. (2002) Intracoronary radiation: Are the frequent flyers still around?. Catheterization and Cardiovascular Interventions 55:2, 237-238
    CrossRef

  64. 64

    Ron Waksman, Albert E Raizner, Alan C Yeung, Alexandra J Lansky, Lynn Vandertie. (2002) Use of localised intracoronary β radiation in treatment of in-stent restenosis: the INHIBIT randomised controlled trial. The Lancet 359:9306, 551-557
    CrossRef

  65. 65

    Edoardo De Benedetti, Donald Latchem, Christian Roguelov, Philippe Coucke, Charles Seydoux, Jean-Jacques Goy, Philip Urban, Eric Eeckhout. (2002) Repeated intracoronary beta radiation for recurrent in-stent restenosis. Catheterization and Cardiovascular Interventions 55:2, 233-236
    CrossRef

  66. 66

    Balram Bhargava, Prabhakar Tripuraneni. (2002) Role of intracoronary brachytherapy for in-stent restenosis?. The Lancet 359:9306, 543-544
    CrossRef

  67. 67

    Sekhara Rao Basavaraju, Clay E. Easterly. (2002) Pathophysiological effects of radiation on atherosclerosis development and progression, and the incidence of cardiovascular complications. Medical Physics 29:10, 2391
    CrossRef

  68. 68

    Leif Thuesen, Henning Rud Andersen, Henning Bagger, Hans Erik Bøtker, Dorthe Dalsgaard, Steen Dalby Kristensen, Lars Romer Krusell, Henrik Munkholm. (2002) Long Genuine Coronary Artery Lesions Treated with Stiff Tubular or Flexible Coiled Stents. A Randomized Angiographic Follow-up Study. Scandinavian Cardiovascular Journal 36:2, 91-94
    CrossRef

  69. 69

    Xin Qu, Judah Weinberger. (2002) Deposition of90YPO4 and144CePO4 radioisotopes on polymer surfaces for radiation delivery devices. Journal of Biomedical Materials Research 63:2, 98-105
    CrossRef

  70. 70

    Patrick K. Coussement, Pieter Stella, Hubert Vanbilloen, Alfons Verbruggen, Peter van Rijk, A. Hoekstra, Erik Van Limbergen, Peter de Jaegere, Ivan De Scheerder. (2002) Intracoronary beta-radiation of de novo coronary lesions using a186Re liquid-filled balloon system: Six-month results from a clinical feasibility study. Catheterization and Cardiovascular Interventions 55:1, 28-36
    CrossRef

  71. 71

    N. Kayhan, R. Krempien, W. Harms, M. Wannenmacher, S. Hagl, C. F. Vahl. (2002) RECHNERÜNTERSTÜTZTE REKONSTRUKTION VON HERZKRANZGEFÄßEN. Biomedizinische Technik/Biomedical Engineering 47:s1b, 930-932
    CrossRef

  72. 72

    Marleen Piessens, Nick Reynaert, Jaroslaw Potempa, Hubert Thierens, William Wijns, Luc Verbeke. (2002) Dose distributions for [sup 90]Y intravascular brachytherapy sources used with balloon catheters. Medical Physics 29:7, 1562
    CrossRef

  73. 73

    Harry C Lowe, Stephen N Oesterle, Levon M Khachigian. (2002) Coronary in-stent restenosis: Current status and future strategies. Journal of the American College of Cardiology 39:2, 183-193
    CrossRef

  74. 74

    Xin Qu, Judah Weinberger. (2001) Encapsulation of isotope on novel ?-emitting poly(ethylene terephthalate) surfaces. Journal of Biomedical Materials Research 57:4, 619-623
    CrossRef

  75. 75

    David J A Goldsmith, Adrian Covic. (2001) Coronary artery disease in uremia: Etiology, diagnosis, and therapy. Kidney International 60:6, 2059-2078
    CrossRef

  76. 76

    Dougal R. McClean, Martyn R. Thomas. (2001) Non-flow limiting dissection leading to late coronary restenosis following intracoronary brachytherapy. Catheterization and Cardiovascular Interventions 54:3, 355-357
    CrossRef

  77. 77

    Nestor Mercado, Eric Boersma, William Wijns, Bernard J Gersh, Carlos A Morillo, Vincent de Valk, Gerrit-Anne van Es, Diederick E Grobbee, Patrick W Serruys. (2001) Clinical and quantitative coronary angiographic predictors of coronary restenosis. Journal of the American College of Cardiology 38:3, 645-652
    CrossRef

  78. 78

    Martin R. Bennett, Michael O'Sullivan. (2001) Mechanisms of angioplasty and stent restenosis: implications for design of rational therapy. Pharmacology & Therapeutics 91:2, 149-166
    CrossRef

  79. 79

    R WAKSMAN, J LAIRD, C JURKOVITZ, A LANSKY, F GERRITS, A KOSINSKI, N MURRAH, W WEINTRAUB. (2001) Intravascular Radiation Therapy after Balloon Angioplasty of Narrowed Femoropopliteal Arteries to Prevent Restenosis: Results of the PARIS Feasibility Clinical Trial. Journal of Vascular and Interventional Radiology 12:8, 915-921
    CrossRef

  80. 80

    Luis Gruberg, Ron Waksman. (2001) Intravascular radiation for the prevention of recurrence of restenosis in coronary arteries. Expert Opinion on Investigational Drugs 10:5, 891-907
    CrossRef

  81. 81

    Sheppard, Richard, Eisenberg, Mark J., . (2001) Intracoronary Radiotherapy for Restenosis. New England Journal of Medicine 344:4, 295-297
    Full Text

  82. 82

    Sapirstein, Wolf, Zuckerman, Bram, Dillard, James, . (2001) FDA Approval of Coronary-Artery Brachytherapy. New England Journal of Medicine 344:4, 297-299
    Full Text

  83. 83

    Mark Kester, Peter Waybill, Mark Kozak. (2001) New Strategies to Prevent Restenosis. American Journal of Cardiovascular Drugs 1:2, 77-83
    CrossRef

  84. 84

    G Jan-Willem Bech, Bernard de Bruyne, T Akasaka, Francesco Liïstro, Hans JRM Bonnier, Jacques J Koolen, Nico HJ Pijls. (2001) Coronary pressure and FFR predict long-term outcome after PTCA. Acute Cardiac Care 4:2, 67-76
    CrossRef