Clinical Consequences of Electrocardiographic Artifact Mimicking Ventricular Tachycardia

Bradley P. Knight, M.D., Frank Pelosi, M.D., Gregory F. Michaud, M.D., S. Adam Strickberger, M.D., and Fred Morady, M.D.

N Engl J Med 1999; 341:1270-1274October 21, 1999

Article

Electrocardiographic artifact can simulate ventricular tachycardia. The literature regarding electrocardiographic artifact is limited to case reports,1-7 proposed classifications,8,9 and diagnostic criteria.10 There is little information regarding the clinical implications of the misdiagnosis of artifact as ventricular tachycardia. We describe 12 patients who underwent unnecessary diagnostic or therapeutic interventions as a result of such a misdiagnosis.

Methods

We included in this series patients who were seen in consultation by a cardiac electrophysiologist at our institution between 1995 and 1999 and who underwent a diagnostic procedure or received treatment unnecessarily, solely as a result of the misdiagnosis of artifact as ventricular tachycardia. We obtained information on the characteristics of the patients, features of the artifact, and the interventions from the medical records. For patients who had more than one documented recording of artifact that simulated ventricular tachycardia, we selected for analysis the longest recording. Continuous variables are presented as means ±SD.

Results

Characteristics of the Patients

Twelve patients received at least one unnecessary intervention solely because of the misdiagnosis of artifact as ventricular tachycardia (Table 1Table 1Characteristics of the Patients, Features of the Electrocardiographic Artifact, and Consequences to the Patient of the Misdiagnosis of Artifact as Ventricular Tachycardia.). In seven cases, artifact that mimicked ventricular tachycardia was recorded during telemetric monitoring of inpatients. Five of these patients had been admitted to the hospital after presenting with cardiac symptoms (syncope or presyncope in three patients, palpitations in one, and chest pain in one); one patient had been admitted after cardiac arrest, and the remaining patient was recovering from cardiac surgery. Three patients were undergoing electrocardiographic monitoring in the emergency department when the artifact was recorded, after presenting with chest pain (one patient), acute respiratory failure in the context of chronic obstructive pulmonary disease (one patient), or a constellation of neurologic symptoms (one patient). Artifact was also recorded in a patient who was undergoing outpatient 24-hour Holter monitoring for the evaluation of palpitations and in another patient who was undergoing an exercise treadmill test for the evaluation of atypical chest pain.

The mean age of the patients was 57±14 years, and 6 of the 12 patients were men. Five patients had structural heart disease. The mean left ventricular ejection fraction was 49±17 percent. When the artifact was recorded, nine patients were asymptomatic, one patient was experiencing chest pain, one patient was unresponsive because of acute respiratory failure, and one patient had paresthesias of the left arm.

Characteristics of the Artifact

Artifact simulated monomorphic ventricular tachycardia in five patients (Figure 1Figure 1Rhythm Strip of Electrocardiographic Artifact That Mimicked Monomorphic Ventricular Tachycardia and Led to the Patient's Being Treated with Lidocaine.) and polymorphic ventricular tachycardia in seven patients (Figure 2Figure 2Rhythm Strip of Electrocardiographic Artifact That Mimicked Polymorphic Ventricular Tachycardia and Led to the Patient's Being Treated with Lidocaine.). The mean rate of simulated ventricular tachycardia was 226±35 beats per minute (range, 180 to 280), and the mean number of apparent tachycardia complexes was 22±7 (range, 11 to 33). Six patients had more than one episode of artifact that simulated ventricular tachycardia. The onset of the artifact was recorded in 75 percent of the patients, and the termination was recorded in 92 percent. Each recording could be recognized as artifact by the presence of native QRS complexes at the cycle length of the base-line rhythm within the artifact.

Clinical Consequences of Misdiagnosis

Diagnostic cardiac catheterization was performed in three patients solely because of the incorrect diagnosis of ventricular tachycardia. Unnecessary medical therapies included intravenous lidocaine in seven patients, intravenous nitroglycerin in one patient, and sublingual nitroglycerin in one patient. Two patients were given a precordial thump that was interpreted as a successful cardioversion. One patient in whom torsade de pointes was incorrectly diagnosed underwent implantation of a permanent pacemaker to prevent a recurrence by keeping bradycardia, which typically precedes polymorphic ventricular tachycardia, from occurring. One patient underwent placement of an implantable cardioverter–defibrillator after an episode of artifact was misdiagnosed as polymorphic ventricular tachycardia. Another patient was given a blood transfusion because it was thought that his ventricular tachycardia had been caused by anemia.

The misdiagnosis of artifact increased the use of health care resources. Two patients were transferred to a tertiary care institution for electrophysiologic testing, one patient was transferred to an intensive care unit from a telemetry ward, and one patient was admitted to the hospital from home after a Holter-monitor recording was misinterpreted as showing ventricular tachycardia.

Characteristics of the Physicians

An initial misdiagnosis of ventricular tachycardia was made by a cardiologist in four cases, a medical house officer in four, an emergency-medicine physician in three, and an electrophysiologist in one. A board-certified cardiologist agreed with the diagnosis of ventricular tachycardia in 10 cases before an electrophysiologist was consulted. A board-certified cardiologist referred an additional patient for a second opinion several months after the patient underwent placement of an implantable cardioverter–defibrillator by a practicing electrophysiologist.

Discussion

These cases demonstrate that when artifact is misdiagnosed as ventricular tachycardia, patients may be subjected to a broad range of unnecessary diagnostic and therapeutic procedures, including cardiac catheterization and the implantation of cardiac devices.

Arrhythmias are unique among transient pathologic states because, even in the absence of symptoms, they often lead to intensive investigations and treatments that have long-term repercussions. Because unsustained ventricular tachycardia can be a sign of structural heart disease, acute ischemia, electrolyte abnormalities, or drug toxicity, its identification often prompts a thorough cardiac evaluation that includes blood analysis, stress testing, echocardiography, and angiography.

Several therapies may be prescribed for patients in whom transient ventricular tachyarrhythmias are diagnosed. Patients with impaired ventricular function after myocardial infarction who are found to have unsustained ventricular tachycardia are often referred for electrophysiologic testing for risk stratification. If sustained ventricular tachycardia is induced at the time of electrophysiologic testing, a prophylactic implantable cardioverter–defibrillator is usually implanted.11,12 However, the Multicenter Unsustained Tachycardia Trial found a 4 to 5 percent risk of cardiac arrest or death from arrhythmia per year among patients with coronary artery disease, a left ventricular ejection fraction of 40 percent or less, and asymptomatic unsustained ventricular tachycardia who did not have inducible sustained ventricular tachycardia (Buxton A: personal communication). Therefore, unsustained ventricular tachycardia alone, regardless of the results of electrophysiologic testing, may become an indication for an implantable cardioverter–defibrillator. If so, an erroneous diagnosis of unsustained ventricular tachycardia in a patient with ischemic cardiomyopathy may lead to the placement of an implantable cardioverter–defibrillator — a consequence that underscores the importance of correctly identifying artifact that simulates unsustained ventricular tachycardia.

Pacing therapy may also be instituted in some patients who have ventricular tachycardia in the context of bradycardia. For example, a pacemaker is often implanted in patients with torsade de pointes to prevent the bradycardia that typically precedes the polymorphic ventricular tachycardia.13

Episodes of ventricular tachycardia may lead to increased use of health care resources. Patients are often transferred to an intensive care unit for closer monitoring after an episode of ventricular tachycardia. A diagnosis of ventricular tachycardia may also result in transfer to a tertiary care medical center, where electrophysiologic testing and defibrillator therapy are available.

Electrocardiographic artifact with sufficient amplitude and duration can closely simulate ventricular tachycardia. Therefore, artifact can readily lead to unnecessary testing and therapy. The results of this study demonstrate that patients in whom ventricular tachycardia is incorrectly diagnosed on the basis of electrocardiographic artifact may be subjected to much the same testing and treatment as patients with actual ventricular tachycardia.

Characteristics that differentiate artifact from ventricular tachycardia include the absence of hemodynamic deterioration during the event; normal QRS complexes within the artifact; an unstable base line on the electrocardiogram before the event, after the event, or both; and an association with body movement.10 In this study, the cases of artifact that mimicked ventricular tachycardia and resulted in unnecessary interventions were usually not associated with symptoms and could be distinguished from ventricular tachycardia by the presence of QRS complexes that were visible in the electrocardiographic artifact at intervals that coincided with the cycle length of the base-line rhythm.

The phenomenon of electrocardiographic artifact was recognized and categorized shortly after electrocardiographic monitoring became available.1-10 Artifact has been categorized as pseudoarrhythmic and nonarrhythmic.8 Artifact can simulate ventricular tachycardia, supraventricular tachycardia, Mobitz type II atrioventricular block, and sinus arrest. The potential clinical effect of each type of artifact was categorized by one group of researchers as possible, minimal, or serious.8 However, the outcomes of patients in whom arrhythmia is incorrectly diagnosed because of artifact have been unclear.

Previous studies have concluded that the most likely causes of electrocardiographic artifact that mimics ventricular tachycardia are body movement and intermittent skin–electrode contact.8 The rapid manipulation of a recording electrode on the skin has been shown to simulate ventricular tachycardia.7 Although body movement or a poor skin–electrode contact may have caused the electrocardiographic artifacts described in this series, the actual causes of the artifacts could not be determined.

Although the magnitude of the problem remains unclear, the cases in this report demonstrate that physicians with a broad spectrum of expertise in the diagnosis and treatment of arrhythmia, including board-certified cardiologists and practicing electrophysiologists, at times may mistake artifact for ventricular tachycardia. Furthermore, these cases show that the misdiagnosis of electrocardiographic artifact as ventricular tachycardia may lead to unnecessary interventions as drastic as the implantation of a permanent pacemaker or an implantable cardioverter–defibrillator. These findings indicate the importance of improved training in the recognition of artifact and the need for a heightened index of suspicion among physicians who treat patients with arrhythmias.

Source Information

From the Department of Internal Medicine, Division of Cardiology, University of Michigan Medical Center, Ann Arbor.

Address reprint requests to Dr. Knight at the University of Michigan Health System, 1500 E. Medical Center Dr., B1F245, Ann Arbor, MI 48109-0022, or at .

References

References

  1. 1

    Dyke DBS, Rich PB, Morady F. Wide complex tachycardia in a critically ill patient: what is the rhythm? J Cardiovasc Electrophysiol 1997;8:1327-1328
    CrossRef | Web of Science | Medline

  2. 2

    Kleinman B, Shah K, Belusko R, Blakeman B. Electrocardiographic artifact caused by extracorporeal roller pump. J Clin Monit 1990;6:258-259
    CrossRef | Medline

  3. 3

    Schiller EC, Heerdt P, Roberts J. Life-threatening ECG artifact dur-ing extracorporeal shock wave lithotripsy. Anesthesiology 1988;68:477-478
    CrossRef | Web of Science | Medline

  4. 4

    Paulsen AW, Pritchard DG. ECG artifact produced by crystalloid administration through blood/fluid warming sets. Anesthesiology 1988;69:803-804
    CrossRef | Web of Science | Medline

  5. 5

    Kantor HI, Bowman A, Abbott PD Jr. Misdiagnosis of fetal life from an artifact in the electrocardiogram. Am J Obstet Gynecol 1966;94:287-289
    Web of Science | Medline

  6. 6

    Lampert BA, Sundstrom FD. ECG artifact simulating supraventricular tachycardia during automated percutaneous lumbar discectomy. Anesth Analg 1988;67:1096-1098
    Web of Science | Medline

  7. 7

    Falk RH, Knowlton AA. Atypical ventricular tachycardia or motion artifact? Am J Cardiol 1987;59:924-924
    CrossRef | Web of Science | Medline

  8. 8

    Krasnow AZ, Bloomfield DK. Artifacts in portable electrocardiographic monitoring. Am Heart J 1976;91:349-357
    CrossRef | Web of Science | Medline

  9. 9

    Murdock DK, Moran JF, Stafford M, King L, Lowb HS, Scanlon PJ. Pacemaker malfunction: fact or artifact? Heart Lung 1986;15:150-154
    Web of Science | Medline

  10. 10

    Lin SL, Wang SP, Kong CW, Chang MS. Artifact simulating ventricular and atrial arrhythmia. Jpn Heart J 1991;32:847-851
    CrossRef | Medline

  11. 11

    Moss AJ, Hall WJ, Cannom DS, et al. Improved survival with an implanted defibrillator in patients with coronary disease at high risk for ventricular arrhythmia. N Engl J Med 1996;335:1933-1940
    Full Text | Web of Science | Medline

  12. 12

    Gregoratos G, Cheitlin MD, Conill A, et al. ACC/AHA guidelines for implantation of cardiac pacemakers and antiarrhythmia devices: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (Committee on Pacemaker Implantation). J Am Coll Cardiol 1998;31:1175-1209
    CrossRef | Web of Science | Medline

  13. 13

    Moss AJ, Liu JE, Gottlieb S, Locati EH, Schwartz PJ, Robinson JL. Efficacy of permanent pacing in the management of high-risk patients with long QT syndrome. Circulation 1991;84:1524-1529
    Web of Science | Medline

Citing Articles (27)

Citing Articles

  1. 1

    2011. Artifacts. , 180-182.
    CrossRef

  2. 2

    Payam Yazdan-Ashoori, Hoshiar Abdollah, Rodrigo Miranda, Christopher Simpson, Damian Redfearn, Kevin Michael, Adrian Baranchuk. (2011) Inappropriate external defibrillation. International Journal of Cardiology
    CrossRef

  3. 3

    Emre Aslanger. (2010) An unusual electrocardiogram artifact in a patient with near syncope. Journal of Electrocardiology 43:6, 686-688
    CrossRef

  4. 4

    Jason M. Tarkin, Nearchos Hadjiloizou, Sam Kaddoura, Julian Collinson. (2010) Variable presentation of ventricular tachycardia-like electrocardiographic artifacts. Journal of Electrocardiology 43:6, 691-693
    CrossRef

  5. 5

    Gökmen Gemici, Ali Kemal Kalkan, Muzaffer Degertekin, Ertan Demirtas. (2010) Artifact mimicking non-sustained polymorphic ventricular tachycardia in a patient with recent myocardial infarction. Central European Journal of Medicine 5:2, 154-155
    CrossRef

  6. 6

    Phil Jevon. (2010) An introduction to electrocardiogram monitoring. Nursing in Critical Care 15:1, 34-38
    CrossRef

  7. 7

    Javier García-Niebla, Pablo Llontop-García, Juan Ignacio Valle-Racero, Guillem Serra-Autonell, Velislav N. Batchvarov, Antonio Bayés de Luna. (2009) Technical Mistakes during the Acquisition of the Electrocardiogram. Annals of Noninvasive Electrocardiology 14:4, 389-403
    CrossRef

  8. 8

    Adrian Baranchuk, Jaskaran Kang, Cathy Shaw, Debra Campbell, Sebastian Ribas, Wilma M. Hopman, Haitham Alanazi, Damian P. Redfearn, Christopher S. Simpson. (2009) Electromagnetic Interference of Communication Devices on ECG Machines. Clinical Cardiology 32:10, 588-592
    CrossRef

  9. 9

    C. C. M. Poon, M. G. Irwin. (2009) Anaesthesia for deep brain stimulation and in patients with implanted neurostimulator devices. British Journal of Anaesthesia 103:2, 152-165
    CrossRef

  10. 10

    Richard Kobza, Markus Roos, Stefan Toggweiler, Michel Zuber, Paul Erne. (2008) Recorded heart sounds for identification of ventricular tachycardia. Resuscitation 79:2, 265-272
    CrossRef

  11. 11

    L. Domínguez-Pérez, M.A. Arias, L. Rodríguez-Padial. (2008) Seudotaquicardia ventricular. Revista Clínica Española 208:11, 577-578
    CrossRef

  12. 12

    R. Mahamdia, C.L. Peng, F. Héritier, K. Djédaïni, M. Roger. (2008) Pseudotorsade de pointe chez une patiente âgée de 88 ans. La Revue de Médecine Interne 29:5, 415-417
    CrossRef

  13. 13

    Santosh I. Patel, Michael J. Souter. (2008) Equipment-related Electrocardiographic Artifacts. Anesthesiology 108:1, 138-148
    CrossRef

  14. 14

    2008. Misplacement of Leads and Electrocardiographic Artifacts. , 586-597.
    CrossRef

  15. 15

    Carmen Nogueras, Ramón Miralles, Alba Roig, Immaculada Llorach, Antoni M. Cervera. (2007) POLYPHARMACY AS PART OF COMPREHENSIVE GERIATRIC ASSESSMENT: DISCLOSURE OF FALSE DIAGNOSIS OF ATRIAL FIBRILLATION BY DRUG REVISION. Journal of the American Geriatrics Society 55:9, 1476-1478
    CrossRef

  16. 16

    Norman C. Wang. (2007) Now You See It; Now You Don’t. The American Journal of Medicine 120:6, 512-514
    CrossRef

  17. 17

    William D. Brearley, Lee Taylor, Michael W. Haley, Laszlo Littmann. (2007) Pneumomediastinum mimicking acute ST-segment elevation myocardial infarction. International Journal of Cardiology 117:2, e73-e75
    CrossRef

  18. 18

    Alexandre Trigano, Olivier Blandeau, Christian Dale, Man-Faï Wong, Joe Wiart. (2007) Risk of cellular phone interference with an implantable loop recorder. International Journal of Cardiology 116:1, 126-130
    CrossRef

  19. 19

    Patric Blomstedt, Mazen Jabre, Boulos-Paul Bejjani, Lars-Owe D. Koskinen. (2006) Electromagnetic Environmental Influences on Implanted Deep Brain Stimulators. Neuromodulation: Technology at the Neural Interface 9:4, 262-269
    CrossRef

  20. 20

    Mori J. Krantz, Joel A. Garcia. (2005) Pseudo-torsade de pointes with underlying wandering atrial pacemaker. Clinical Cardiology 28:7, 355-355
    CrossRef

  21. 21

    Julián Ortega-Carnicer. (2005) Tremor-related artefact mimicking ventricular tachycardia. Resuscitation 65:3, 243-244
    CrossRef

  22. 22

    Jens Lönnblad, Göran Nilsson, Mia Folke, Bertil Hök, Maria Lindén, Ylva Bäcklund. (2005) Artefacts in continuous ECG recordings – Provoking and preventing manoeuvres. Scandinavian Cardiovascular Journal 39:3, 167-171
    CrossRef

  23. 23

    Alan H. Kadish, Alfred E. Buxton, Harold L. Kennedy, Bradley P. Knight, Jay W. Mason, Claudio D. Schuger, Cynthia M. Tracy, William L. Winters, Alan W. Boone, Michael Elnicki, John W. Hirshfeld, Beverly H. Lorell, George P. Rodgers, Cynthia M. Tracy, Howard H. Weitz. (2001) ACC/AHA clinical competence statement on electrocardiography and ambulatory electrocardiography. Journal of the American College of Cardiology 38:7, 2091-2100
    CrossRef

  24. 24

    Amos Katz, Yoram Snir, Alan B. Wagshal, Gisela Cohn, Reuben Ilia. (2001) Pseudoventricular fibrillation. Critical Care Medicine 29:9, 1835-1836
    CrossRef

  25. 25

    A A Obwegeser, R J Uitti, M F Turk, U M Wszolek, T R Flipse, R C Smallridge, R J Witte, R E Wharen. (2001) Simultaneous thalamic deep brain stimulation and implantable cardioverter-defibrillator.. Mayo Clinic Proceedings 76:1, 87-89
    CrossRef

  26. 26

    Hirokatsu Toyoyama, MD Koh Mizutani, MD Humiko Kuki, DDS Yoshiro Toyoda. (2000) Misdiagnosis of ECG Artifacts May Lead to Drastic and Unnecessary Interventions. Anesthesia & Analgesia 91:2, 498-499
    CrossRef

  27. 27

    (2000) Electrocardiographic Artifact. New England Journal of Medicine 342:8, 590-592
    Full Text

Letters