Value of Electron-Beam Computed Tomography for the Noninvasive Detection of High-Grade Coronary-Artery Stenoses and Occlusions

Stephan Achenbach, M.D., Werner Moshage, M.D., Dieter Ropers, M.D., Jörg Nossen, M.D., and Werner G. Daniel, M.D.

N Engl J Med 1998; 339:1964-1971December 31, 1998

Abstract

Background

Reliable noninvasive assessment of coronary-artery stenoses and occlusions would constitute an advantage in the care of patients with known or suspected coronary artery disease. We investigated the accuracy of contrast-enhanced electron-beam computed tomography (CT) for the detection of high-grade coronary-artery stenoses and occlusions.

Methods

Electron-beam CT was performed in 125 patients. After intravenous injection of a contrast agent, 40 cross-sectional images of the heart were acquired during inspiration, triggered by the electrocardiogram in diastole. Three-dimensional reconstructions of the heart and coronary arteries were rendered to facilitate evaluation of the images. The proximal and middle segments of the major coronary arteries were evaluated for the presence or absence of high-grade stenoses and occlusions. The results were compared with those of invasive coronary angiography in a blinded fashion.

Results

Because of technical problems that impaired the quality of the images, 124 (25 percent) of the 500 coronary arteries studied (left main, left anterior descending, left circumflex, and right coronary) in a total of 125 patients were excluded from evaluation. No vessels could be evaluated in 19 patients (15 percent), and another 28 patients (22 percent) had one, two, or three vessels that could not be evaluated. In the remaining coronary arteries with adequate image quality, electron-beam CT permitted visualization of 69 of 75 high-grade stenoses and occlusions (sensitivity, 92 percent), whereas in 282 of 301 arteries, the absence of high-grade stenoses and occlusions was correctly detected (specificity, 94 percent).

Conclusions

When image quality is adequate, electron-beam CT may be useful to detect or rule out high-grade coronary-artery stenoses and occlusions.

Media in This Article

Figure 1Electron-Beam CT Images of the Heart.
Figure 2Frequently Observed Reasons for Inability to Evaluate Electron-Beam CT Images of the Coronary Arteries.
Article

Annually, more than 1 million coronary angiographic procedures are performed in the United States.1 Replacing even a fraction of these diagnostic procedures with a noninvasive imaging technique would constitute an important advance in the care of patients with known or suspected coronary artery disease. Several noninvasive imaging techniques have the potential for visualization of the coronary arteries, including transthoracic and transesophageal echocardiography,2-6 synchrotron dichromography,7,8 and most important, magnetic resonance imaging (MRI).9-11 However, because of limited spatial, temporal, or contrast resolution, none of these techniques have permitted reliable detection of coronary-artery stenoses and occlusions in a clinical setting.

Electron-beam computed tomography (CT) is a cross-sectional imaging technique with high spatial and temporal resolution; it is possible for image acquisition to be triggered by the patient's electrocardiogram. The technique is therefore well suited to cardiac imaging. It is widely used to assess coronary calcifications,12-18 and there have been preliminary reports of the use of electron-beam CT with intravenous injection of contrast agent to visualize the coronary-artery lumen.19-25 We prospectively evaluated a group of 125 patients by both electron-beam CT and conventional coronary angiography to determine the value of the method for the noninvasive detection of high-grade stenoses and occlusions of the coronary arteries.

Methods

Subjects

One hundred twenty-five patients (104 men and 21 women) were studied. Their mean age was 56 years (range, 23 to 82). Their average weight was 81 kg, but nine patients weighed more than 100 kg. All patients who were admitted to our institution for inpatient diagnostic coronary angiography on days on which we had access to the electron-beam CT scanner (a total of 44 days) were asked to participate in the study. Patients with coronary stents or coronary-artery bypass grafts, patients whose clinical condition was unstable, patients with atrial fibrillation, and patients with contraindications to the administration of contrast agent were excluded. All patients gave written informed consent according to institutional guidelines. Conventional invasive coronary angiography was performed by independent investigators within one week after the electron-beam CT investigation. All selected patients were included in the analysis.

Electron-Beam CT

Acquisition of Data

Electron-beam CT has previously been described in detail.26,27 In brief, the unique feature of this CT method is its very high imaging speed (one image can be acquired in 50 to 100 msec), which can be achieved because there is no rotating x-ray tube. The heart can therefore be imaged with substantially less artifact due to motion than with conventional CT scanners. In addition, scanning can be triggered by the patient's electrocardiogram. In our study, data were acquired with a C-150 XP electron-beam CT scanner (Imatron, South San Francisco, Calif.). Scanning was performed with the patient in a supine position and holding his or her breath after inspiration. Our imaging protocol consisted of three steps, as described below.

To determine the position of the heart, eight cross-sectional images of the chest (slice thickness, 7 mm; 4-mm gap between slices; field of view, 15 cm; acquisition time, 50 msec) were acquired simultaneously with the scanner in the multislice mode.

To determine exactly the transit time of the contrast agent from injection into an antecubital vein to the appearance of contrast enhancement in the aortic root and coronary arteries, a bolus of 10 ml of contrast agent (Ultravist-370 [iopromide], Schering, Berlin, Germany) was injected at a rate of 4 ml per second. As injection of the contrast agent was begun, the patient was instructed to breathe in, breathe out, breathe in again, and then hold the breath in inspiration. This sequence of commands was timed to have a total duration of 10 seconds. With the onset of breath-holding, 10 seconds after the start of contrast injection, acquisition of 10 axial cross-sectional images at the level of the aortic root was begun (single-slice mode; slice thickness, 3 mm; acquisition time, 100 msec per image). Image acquisition was triggered by the electrocardiogram, with an image being acquired after every other heartbeat at 80 percent of the RR interval. The average CT density in the aortic root was measured in each of the 10 acquired images with the scanner software. The interval between the onset of contrast injection and peak enhancement was considered to represent the transit time of the contrast agent.

In a final step, 40 axial cross-sectional images of the heart were acquired in single-slice mode. The acquisition time was 100 msec per image. With acquisition triggered by the electrocardiogram, one image was acquired after every heartbeat in diastole (at 80 percent of the cardiac cycle). The slice thickness was 3 mm, and after the acquisition of each image, the table was advanced 2 mm to generate overlapping images. The field of view was 15 cm, with an image matrix of 512 by 512, which generated a pixel size of 0.29 by 0.29 mm to make full use of the in-plane spatial resolution of the scanner (seven line pairs per centimeter).

Images were acquired with the patient holding his or her breath after inspiration and with breathing commands identical to those used for the determination of the transit time of the contrast agent. The scanner permits the acquisition of one image per cardiac cycle for heart rates up to 120 beats per minute. With heart rates between 44 and 104 beats per minute (mean, 69) during acquisition of the images, the duration of breath-holding was 23 to 54 seconds (mean, 35). Between 120 and 160 ml of contrast agent was injected intravenously at a rate of 4 ml per second. A delay corresponding to the individually determined transit time of the contrast agent was maintained between the initiation of injection of the contrast agent and the acquisition of the first image. The radiation dose was estimated to be less than 11 mSv.12,28

Evaluation of Data

The images obtained by electron-beam CT were independently evaluated by two investigators without knowledge of the findings on the patients' coronary angiograms. The image data were transferred to an off-line workstation (MagicView, Siemens, Erlangen, Germany). The cross-sectional images were screened for artifacts caused by motion and for calcifications of the coronary arteries. Calcifications were rated as too severe to permit further evaluation if a calcified plaque occupied more than half the diameter of the artery in two consecutive images. Manual segmentation of the images was performed on an image-by-image basis to eliminate all structures that would obstruct the view of the coronary arteries, such as the chest wall, pulmonary vessels, and atrial appendages (Figure 1AFigure 1Electron-Beam CT Images of the Heart., Figure 1B, and Figure 1C). Subsequently, three-dimensional reconstructions of the heart and coronary arteries were rendered by a shaded-surface display technique. A lower threshold of 80 Hounsfield units was used for the reconstruction in all patients to visualize selectively the contrast-enhanced lumina of the coronary arteries, while ignoring the vessel wall and surrounding connective tissue (Figure 1A, Figure 1B, and Figure 1C).19 Three-dimensional reconstructions were rendered interactively from at least six angles (equivalent to posteroanterior, left anterior oblique, and right anterior oblique projections, each in at least two craniocaudal angulations). If necessary, reconstructions were rendered from additional angles to depict all coronary segments from two orthogonal directions.

On the basis of the original cross-sectional images and the three-dimensional reconstructions, both investigators independently rated the left main coronary artery (segment 5 according to the system of the American Heart Association29) and the proximal and middle segments of the left anterior descending coronary artery (segments 6 and 7 and the first half of segment 8), the left circumflex coronary artery (segments 11 and 13), and the right coronary artery (segments 1 and 2) as being occluded or having a high-grade stenosis (more than 75 percent reduction in diameter), showing neither high-grade stenosis nor occlusion, or being impossible to evaluate. The side branches and the distal segments of the coronary arteries were not included in the evaluation. Determination of the severity of stenosis was based on visual estimation of both the axial cross sections and the three-dimensional reconstructions. In the reconstructions, the severity of the lesion was determined in the projection that showed the highest degree of stenosis. Cohen's kappa30 was calculated to determine agreement between observers on which of three categories was applicable (status impossible to evaluate, high-grade stenosis or occlusion absent, and high-grade stenosis or occlusion present).

Coronary Angiography

In all patients, coronary angiography was performed by the transfemoral Judkins approach. Angiograms were documented on cine film and evaluated by two cardiologists who were independent of the electron-beam CT team. Reductions in diameter of more than 75 percent were considered to represent high-grade stenoses. In cases of disagreement between the two readers, agreement was achieved in a joint reading. Cohen's kappa was calculated to determine variability between observers, with regard to the same categories as for electron-beam CT (status impossible to evaluate, high-grade stenosis or occlusion absent, and high-grade stenosis or occlusion present).

Results

In all the patients, electron-beam CT was performed without complications. The mean time needed for the investigation was 18±3 minutes per patient, including all preparations for scanning. Of the 500 coronary arteries (left main, left anterior descending, left circumflex, and right coronary artery in 125 patients), 124 were considered impossible to evaluate because of technical problems. The main reasons for impaired image quality were artifacts of respiration (40 vessels) and severe calcifications (33 vessels) (Figure 2AFigure 2Frequently Observed Reasons for Inability to Evaluate Electron-Beam CT Images of the Coronary Arteries. and Figure 2B), as well as artifacts of movement of the left circumflex coronary artery (11 vessels) and the right coronary artery (14 vessels). Other reasons included reduced signal-to-noise ratio (12 vessels), superposition of veins (5 vessels), malpositioning of the imaging volume (5 vessels), and a very small arterial lumen (4 vessels). In 11 patients, one coronary artery could not be evaluated; in 14 patients, two arteries could not be evaluated; and in 3 patients, three vessels could not be evaluated. In 19 patients (15 percent), the quality of the image was too poor for any of the four arteries to be evaluated. The reduction in image quality in these patients was caused by artifacts of respiration in eight, severe calcifications in five, reduced signal-to-noise ratio in three, and more than one of these factors in three.

The left main coronary artery could be evaluated in 105 patients (84 percent), the left anterior descending coronary artery in 100 patients (80 percent), the left circumflex coronary artery in 83 patients (66 percent), and the right coronary artery in 88 patients (70 percent).

With the 124 coronary arteries that were graded as impossible to evaluate excluded from the analysis, 41 of 42 substantial lesions (high-grade stenoses and occlusions) of the left anterior descending coronary artery, 14 of 18 substantial lesions of the left circumflex coronary artery, and 14 of 15 substantial lesions of the right coronary artery were correctly detected by electron-beam CT (Figure 3AFigure 3Cardiac Images of a 56-Year-Old Patient with a Complete Occlusion of the Left Circumflex Coronary Artery (Arrow) According to Electron-Beam CT (Panel A) and Coronary Angiography (Panel B)., Figure 3B, Figure 4AFigure 4Cardiac Images of an 81-Year-Old Patient with a High-Grade Stenosis of the Right Coronary Artery (Arrow) According to Electron-Beam CT (Panels A and B) and Coronary Angiography (Panel C)., Figure 4B, and Figure 4C). In the arteries that could be evaluated, the absence of high-grade stenosis or complete occlusion was correctly determined in 104 of 105 cases for the left main coronary artery, 51 of 58 cases for the left anterior descending coronary artery, 57 of 65 cases for the left circumflex coronary artery, and 70 of 73 cases for the right coronary artery (Table 1Table 1Accuracy of Electron-Beam CT for the Detection of High-Grade Stenoses and Occlusions of the Coronary Arteries.). These numbers correspond to a sensitivity of 92 percent, a specificity of 94 percent, a positive predictive value of 78 percent, and a negative predictive value of 98 percent for the detection of substantial coronary lesions in the arteries that could be evaluated.

Whether a vessel could be evaluated by electron-beam CT was not associated with the angiographic results for that vessel (Table 2Table 2Relation between Angiographic Status of the Coronary Arteries and Results of Electron-Beam CT.). Of the 376 vessels that could be evaluated by electron-beam CT, 75 (20 percent) had a high-grade stenosis or occlusion on angiography; of the 124 arteries that could not be evaluated by electron-beam CT, 24 (19 percent) had a high-grade stenosis or occlusion on angiography.

Overall, when arteries that could not be evaluated were included, electron-beam CT permitted visualization of 69 of the 99 high-grade coronary-artery stenoses and occlusions that were detected by angiography in the group of patients we investigated. Of the 401 coronary arteries that were documented by angiography to be free of high-grade lesions, 282 were correctly identified by electron-beam CT.

Interobserver agreement was achieved for 453 of the 500 coronary arteries (91 percent). Cohen's kappa was 0.82, indicating close agreement between observers. Disagreement concerned mainly the quality of the images: in 31 of the 47 coronary vessels with discrepant readings, one of the two observers considered image quality to be too poor for evaluation. With conventional angiography, agreement between the observers was achieved in 96 percent of cases without joint reading, and Cohen's kappa was 0.90.

Discussion

Because of its high spatial and temporal resolution as well as the fact that image acquisition can be triggered by the electrocardiogram, electron-beam CT is well suited for cardiac imaging. In a clinical setting, this method, which has also been called “ultrafast CT” or “cine CT,” has been applied principally to visualize and quantify coronary calcifications12-17 and to evaluate cardiac function.31 However, preliminary reports have indicated the potential of the method, after intravenous injection of contrast agent, to visualize the coronary-artery lumen19,20 and to detect stenoses noninvasively.21-25 In our investigation, we found a high sensitivity and specificity, as well as a very high negative predictive value of 98 percent, for the detection of complete occlusions and high-grade stenoses of the coronary arteries, provided that sufficiently good image quality could be obtained. In the vessels that could be evaluated, the number of false negative results for stenoses was small (92 percent sensitivity). Most stenoses that were missed were located in the left circumflex and right coronary arteries. These vessels are perpendicular to the imaging plane and are therefore visualized with lower spatial resolution than those oriented parallel to the imaging plane. As in earlier, smaller studies, most false positive results were for stenoses in smaller vessels23 (again, the left circumflex coronary artery in most cases).

The main drawback of electron-beam CT currently seems to be that 25 percent of all coronary segments had to be excluded from evaluation because of inadequate image quality. Because of their position in the coronary groove, the left circumflex and right coronary arteries have more rapid diastolic motion than the left anterior descending coronary artery. The motion is caused mainly by atrial contraction during end-diastole. In addition, the interpretation of results for all coronary arteries may be compromised by artifacts of respiration if patients are unable to obey the breath-holding commands. Although mildly calcified vessels can be evaluated, extensive calcifications are a frequent reason for false negative23,25 and false positive22 results. We therefore excluded all arteries with more than mild calcifications from evaluation. Mistriggering of single cross sections (from premature beats, for example) was not observed as frequently as previously reported.22

Before the method is applied on a broad clinical scale, the number of studies that cannot be evaluated will have to be reduced. To do this, it will be necessary to ensure, for example, that through more careful instruction and preparation patients hold their breath for the full scanning period. Alternatively, the heart rate may be medically increased to shorten the overall image-acquisition time. Furthermore, choosing different imaging planes could increase spatial resolution in the vessel segments that are perpendicular to the axial imaging plane we used in the present study (the middle right coronary artery and the left circumflex coronary artery).23 Further reducing the image-acquisition time or optimizing the time interval in the cardiac cycle that is used for image acquisition would probably diminish artifacts of motion. Higher spatial resolution of the scanner might improve the detection of short lesions and of stenoses in small arteries. It might also permit the development of image-processing techniques to subtract calcified plaques, something that is currently not possible, because of the overlap of density values between calcifications and contrast-enhanced vessel lumen.

Our evaluation was limited to the proximal and middle segments of the major coronary arteries. Distal segments and side branches were excluded because their small diameters do not permit reliable imaging with the spatial resolution the scanner currently provides (seven line pairs per centimeter in the imaging plane), and partial-volume effects would lead to an unacceptably high number of false positive results. However, since lesions in distal segments and side branches are rarely the target of catheter intervention or bypass revascularization, the fact that electron-beam CT imaging is limited to segments of larger diameter does not mean that the technique may not be clinically useful. Similarly, no attempt was made to differentiate between high-grade stenoses and complete occlusions of coronary arteries. In electron-beam CT, the partial-volume effect can cause a very tight stenosis to appear to be a complete occlusion of the vessel lumen. Consequently, it may be impossible to decide whether such a lesion represents a high-grade stenosis or a complete occlusion with filling of the distal segment by collateral flow. For clinical applications, however, mistaking complete occlusions for high-grade stenoses would in most cases be irrelevant, since both conditions usually require further invasive testing.

The accuracy of electron-beam CT for the diagnosis of high-grade stenoses and occlusions of the coronary arteries seems to be superior to that of other noninvasive techniques. Echocardiography permits the assessment of only the most proximal segments of the coronary arteries, and the technique is unreliable for identifying coronary lesions. Dichromatic synchrotron angiography has not yet been evaluated on a broad scale. Magnetic resonance coronary-angiography techniques are evolving rapidly. As compared with electron-beam CT, they have the advantage of requiring no exposure to radiation and, in most cases, no injection of contrast agent. However, the spatial and temporal resolution of currently available MRI techniques32-37 is inferior to that of electron-beam CT, and except for respiratory-gated acquisitions,36,37 the necessary respiratory maneuvers require a high degree of cooperation from patients.32-35 The reported accuracy of MRI protocols that have been proposed for the detection of coronary-artery stenoses varies widely, from 38 to 90 percent,32-37 and (with the possible exception of visualizing the course of anomalous coronary arteries38-40) magnetic resonance angiography is currently considered an investigational technique.10,11,40,41

The high negative predictive value of electron-beam CT for the detection of high-grade stenoses in patients with adequate image quality may make this method a clinically useful test. Since we found that the results of electron-beam CT could be completely evaluated in 62 percent of all patients, and that whether or not they could be evaluated was not related to the angiographic status of the arteries, it can be calculated that up to two thirds of the coronary angiographic procedures that demonstrate normal coronary arteries could currently be averted by the use of this noninvasive test. Further research must be directed toward reducing the number of nondiagnostic studies. Once this goal has been achieved, electron-beam CT may become a useful tool to rule out high-grade coronary-artery stenoses or occlusions in, for example, patients with a low likelihood of obstructive coronary disease, and also in patients who have undergone balloon angioplasty or bypass grafting. In this way, a substantial number of the invasive coronary angiographic studies that have negative results, which constitute up to 20 percent of all studies performed,42 could be replaced by noninvasive outpatient examinations.

Source Information

From the Department of Internal Medicine II, University of Erlangen-Nürnberg, Erlangen, Germany.

Address reprint requests to Dr. Achenbach at the Medizinische Klinik II mit Poliklinik, Friedrich-Alexander-Universität Erlangen-Nürnberg, Östliche Stadtmauerstr. 29, D-91054 Erlangen, Germany.

References

References

  1. 1

    1993 Heart and stroke facts statistics. Dallas: American Heart Association, 1993:8.

  2. 2

    Hozumi T, Yoshida K, Ogata Y, et al. Noninvasive assessment of significant left anterior descending coronary artery stenosis by coronary flow velocity reserve with transthoracic color Doppler echocardiography. Circulation 1998;97:1557-1562
    Web of Science | Medline

  3. 3

    Feigenbaum H. Transthoracic ultrasonic visualization of coronary atherosclerosis. J Am Soc Echocardiogr 1989;2:253-258
    Medline

  4. 4

    Tardif JC, Vannan MA, Taylor K, Schwartz SL, Pandian NG. Delineation of extended lengths of coronary arteries by multiplane transesophageal echocardiography. J Am Coll Cardiol 1994;24:909-919
    CrossRef | Web of Science | Medline

  5. 5

    Caiati C, Aragona P, Iliceto S, Rizzon P. Improved detection of proximal left anterior descending coronary artery stenosis after intravenous injection of a lung-crossing contrast agent: a transesophageal Doppler echocardiographic study. J Am Coll Cardiol 1996;27:1413-1421
    CrossRef | Web of Science | Medline

  6. 6

    Mulvagh SL, Foley DA, Aeschbacher BC, Klarich KK, Seward JB. Second harmonic imaging of an intravenously administered echocardiographic contrast agent: visualization of coronary arteries and measurement of coronary blood flow. J Am Coll Cardiol 1996;27:1519-1525
    CrossRef | Web of Science | Medline

  7. 7

    Rubenstein E, Brown GS, Chapman D, et al. Synchrotron radiation coronary angiography in humans. In: Chance B, Deisenhofer J, Ebashi S, et al., eds. Synchrotron radiation in the biosciences. New York: Oxford University Press, 1994:639-50.

  8. 8

    Hamm CW, Meinertz T, Dix WR, et al. Intravenous coronary angiography with dichromography using synchrotron radiation. Herz 1996;21:127-131
    Web of Science | Medline

  9. 9

    Manning WJ, Edelman RR. Magnetic resonance coronary angiography. Magn Reson Q 1993;9:131-151
    Medline

  10. 10

    Taylor AM, Pennell DJ. Recent advances in cardiac magnetic resonance imaging. Curr Opin Cardiol 1996;11:635-642
    CrossRef | Web of Science | Medline

  11. 11

    van der Wall EE, van Rugge FP, Vliegen HW, Reiber JH, de Roos A, Bruschke AV. Ischemic heart disease: value of MR techniques. Int J Card Imaging 1997;13:79-89
    CrossRef | Web of Science | Medline

  12. 12

    Agatston AS, Janowitz WR, Hildner FJ, Zusmer NR, Viamonte M Jr, Detrano R. Quantification of coronary artery calcium using ultrafast computed tomography. J Am Coll Cardiol 1990;15:827-832
    CrossRef | Web of Science | Medline

  13. 13

    Ultrafast CT for coronary calcificationLancet 1991;337:1449-1450
    Medline

  14. 14

    Breen JF, Sheedy PF II, Schwartz RS, et al. Coronary artery calcification detected with ultrafast CT as an indication of coronary artery disease. Radiology 1992;185:435-439
    Web of Science | Medline

  15. 15

    Budoff MJ, Georgiou D, Brody A, et al. Ultrafast computed tomography as a diagnostic modality in the detection of coronary artery disease: a multicenter study. Circulation 1996;93:898-904
    Web of Science | Medline

  16. 16

    Arad Y, Spadaro LA, Goodman K, et al. Predictive value of electron beam computed tomography of the coronary arteries: 19-month follow-up of 1173 asymptomatic subjects. Circulation 1996;93:1951-1953
    Web of Science | Medline

  17. 17

    Detrano R, Hsiai T, Wang S, et al. Prognostic value of coronary calcification and angiographic stenoses in patients undergoing coronary angiography. J Am Coll Cardiol 1996;27:285-290
    CrossRef | Web of Science | Medline

  18. 18

    Rumberger JA, Sheedy PF II, Breen JF, Fitzpatrick LA, Schwartz RS. Electron beam computed tomography and coronary artery disease: scanning for coronary artery calcification. Mayo Clin Proc 1996;7:369-377
    CrossRef

  19. 19

    Moshage WE, Achenbach S, Seese B, Bachmann K, Kirchgeorg M. Coronary artery stenoses: three-dimensional imaging with electrocardiographically triggered, contrast agent-enhanced, electron-beam CT. Radiology 1995;196:707-714
    Web of Science | Medline

  20. 20

    Chernoff DM, Ritchie CJ, Higgins CB. Evaluation of electron beam CT coronary angiography in healthy subjects. AJR Am J Roentgenol 1997;169:93-99
    Web of Science | Medline

  21. 21

    Budoff MJ, Oudiz RJ, Zalace CP, et al. Intravenous three dimensional coronary angiography using contrast enhanced electron beam computed tomography. J Am Coll Cardiol 1997;29:Suppl A:393A-393A abstract.

  22. 22

    Nakanishi T, Ito K, Imazu M, Yamakido M. Evaluation of coronary artery stenoses using electron-beam CT and multiplanar reformation. J Comput Assist Tomogr 1997;21:121-127
    CrossRef | Web of Science | Medline

  23. 23

    Schmermund A, Rensing BJ, Sheedy PF, Bell MR, Rumberger JA. Intravenous electron-beam computed tomographic coronary angiography for segmental analysis of coronary artery stenoses. J Am Coll Cardiol 1998;31:1547-1554
    CrossRef | Web of Science | Medline

  24. 24

    Reddy GP, Chernoff DM, Siripornpitak S, Adams JR, Outtrim RA, Higgins CB. Assessment of coronary artery stenoses with contrast-enhanced electron-beam CT and axial reconstructions. Radiology 1997;205:Suppl:459-459 abstract.
    Web of Science | Medline

  25. 25

    Achenbach S, Moshage W, Bachmann K. Detection of high-grade restenosis after PTCA using contrast-enhanced electron beam CT. Circulation 1997;96:2785-2788
    Web of Science | Medline

  26. 26

    Boyd D, Gould RG, Quinn J, Sparks R, Stanley R, Hermannsfeldt W. A proposed cardiac 3-d densitometer for easy detection and evaluation of heart disase. IEEE Trans Nucl Sci 1979;26:2724-2727
    CrossRef | Web of Science

  27. 27

    Gould RG. Principles of ultrafast computed tomography: historical aspects, mechanism of action, and scanner characteristics. In: Stanford W, Rumberger JA, eds. Ultrafast computed tomography in cardiac imaging: principles and practice. Mount Kisco, N.Y.: Futura Publishing, 1992:1-16.

  28. 28

    Stanford W, Thompson BH, Weiss RM. Coronary artery calcification: clinical significance and current methods of detection. AJR Am J Roentgenol 1993;161:1139-1146
    Web of Science | Medline

  29. 29

    A reporting system on patients evaluated for coronary artery disease: report of the Ad Hoc Committee for Grading of Coronary Artery Disease, Council on Cardiovascular Surgery, American Heart Association. Circulation 1975;51:Suppl:5-40
    Medline

  30. 30

    Cohen J. A coefficient for agreement on nominal scales. Educ Psychol Meas 1960;20:37-46
    CrossRef | Web of Science

  31. 31

    Rumberger JA. Quantifying left ventricular regional and global systolic function using ultrafast computed tomography. Am J Card Imaging 1991;5:29-37
    Medline

  32. 32

    Manning WJ, Li W, Edelman RR. A preliminary report comparing magnetic resonance coronary angiography with conventional angiography. N Engl J Med 1993;328:828-832[Erratum, N Engl J Med 1994;330:152.]
    Full Text | Web of Science | Medline

  33. 33

    Post JC, van Rossum AC, Hofman MB, Valk J, Visser CA. Three-dimensional respiratory-gated MR angiography of coronary arteries: comparison with conventional coronary angiography. AJR Am J Roentgenol 1996;166:1399-1404
    Web of Science | Medline

  34. 34

    Pennell DJ, Bogren HG, Keegan J, Firmin DN, Underwood SR. Assessment of coronary artery stenosis by magnetic resonance imaging. Heart 1996;75:127-133
    CrossRef | Web of Science | Medline

  35. 35

    Mohiaddin RH, Bogren HG, Lazim F, et al. Magnetic resonance coronary angiography in heart transplant recipients. Coron Artery Dis 1996;7:591-597
    CrossRef | Web of Science | Medline

  36. 36

    Kessler W, Achenbach S, Moshage W, et al. Usefulness of respiratory gated magnetic resonance coronary angiography in assessing narrowings> or = 50% in diameter in native coronary arteries and in aortocoronary bypass conduits. Am J Cardiol 1997;80:989-993
    CrossRef | Web of Science | Medline

  37. 37

    Muller MF, Fleisch M, Kroeker R, Chatterjee T, Meier B, Vock P. Proximal coronary artery stenosis: three-dimensional MRI with fat saturation and navigator echo. J Magn Reson Imaging 1997;7:644-651
    CrossRef | Web of Science | Medline

  38. 38

    McConnell MV, Ganz P, Selwyn AP, Li W, Edelman RR, Manning WJ. Identification of anomalous coronary arteries and their anatomic course by magnetic resonance coronary angiography. Circulation 1995;92:3158-3162
    Web of Science | Medline

  39. 39

    Post JC, van Rossum AC, Bronzwaer JGF, et al. Magnetic resonance angiography of anomalous coronary arteries: a new gold standard for delineating the proximal course? Circulation 1995;92:3163-3171
    Web of Science | Medline

  40. 40

    Task Force of the European Society of Cardiology, Association of European Paediatric Cardiologists. The clinical role of magnetic resonance in cardiovascular disease. Eur Heart J 1998;19:19-39
    CrossRef | Web of Science | Medline

  41. 41

    Duerinckx AJ. MR angiography of the coronary arteries. Top Magn Reson Imaging 1995;7:267-285
    CrossRef | Medline

  42. 42

    Johnson LW, Lozner EC, Johnson S, et al. Coronary arteriography 1984-1987: a report of the Registry of the Society for Cardiac Angiography and Interventions. I. Results and complications. Cathet Cardiovasc Diagn 1989;17:5-10
    CrossRef | Medline

Citing Articles (113)

Citing Articles

  1. 1

    Gianluca Pontone, Daniele Andreini, Antonio L. Bartorelli, Erika Bertella, Saima Mushtaq, Andrea Annoni, Alberto Formenti, Luisa Chiappa, Sarah Cortinovis, Andrea Baggiano, Edoardo Conte, Francesca Bovis, Fabrizio Veglia, Claudia Foti, Giovanni Ballerini, Cesare Fiorentini, Mauro Pepi. (2011) Radiation dose and diagnostic accuracy of multidetector computed tomography for the detection of significant coronary artery stenoses. International Journal of Cardiology
    CrossRef

  2. 2

    Gary Y.H. Liew, Michael Feneley, Stephen G. Worthley. (2011) Noninvasive Coronary Artery Imaging: Current Clinical Applications. Heart, Lung and Circulation 20:7, 425-437
    CrossRef

  3. 3

    Zilai Pan, Lifang Pang, Jianying Li, Huan Zhang, Wenjie Yang, Bei Ding, Weimin Chai, Kemin Chen, Weiwu Yao. (2011) Achieving consistent image quality with dose optimization in 64-row multidetector computed tomography prospective ECG gated coronary calcium scoring. The International Journal of Cardiovascular Imaging 27:4, 611-617
    CrossRef

  4. 4

    Koen Nieman. (2011) Coronary computed tomography angiography versus stress testing in suspected coronary disease. Expert Review of Cardiovascular Therapy 9:1, 93-104
    CrossRef

  5. 5

    Jack A. Ziffer. (2010) President's Page: Prospective and retrospective preeminence: What is right about coronary CT. Journal of Cardiovascular Computed Tomography 4:4, 288-290
    CrossRef

  6. 6

    Nina Ghosh, Ronnen Maze, Benjamin Chow, Carole Dennie, Alexander Dick, Terrence Ruddy. (2010) Noninvasive cardiovascular imaging in coronary artery disease. Imaging in Medicine 2:3, 271-288
    CrossRef

  7. 7

    Jack Ziffer. (2009) President's page: The way ahead. Journal of Cardiovascular Computed Tomography 3:6, 421-423
    CrossRef

  8. 8

    Philipp A. Kaufmann. (2009) Cardiac hybrid imaging: state-of-the-art. Annals of Nuclear Medicine 23:4, 325-331
    CrossRef

  9. 9

    Gregory S. Hurlock, Hiroshi Higashino, Teruhito Mochizuki. (2009) History of cardiac computed tomography: single to 320-detector row multislice computed tomography. The International Journal of Cardiovascular Imaging 25:S1, 31-42
    CrossRef

  10. 10

    Simona Sitia, Fabiola Atzeni, Piercarlo Sarzi-Puttini, Vitantonio Di Bello, Livio Tomasoni, Luigi Delfino, Francesco Antonini-Canterin, Giovanni Di Salvo, Vito De Gennaro Colonna, Salvatore La Carrubba, Scipione Carerj, Maurizio Turiel. (2009) Cardiovascular involvement in systemic autoimmune diseases. Autoimmunity Reviews 8:4, 281-286
    CrossRef

  11. 11

    A. Schmermund, B. Nowak, T. Voigtlander. (2008) Non-invasive CT coronary angiography: can high diagnostic image quality be achieved with less radiation exposure?. European Heart Journal 29:24, 2955-2956
    CrossRef

  12. 12

    R. Dikkers. (2008) The Visualization of the Coronary Arteries and Coronary Calcifications. Imaging Decisions MRI 12:2, 1-6
    CrossRef

  13. 13

    Wajeh Qunibi, Moustafa Moustafa, Larry R. Muenz, David Y. He, Paul D. Kessler, Jose A. Diaz-Buxo, Mathew Budoff. (2008) A 1-Year Randomized Trial of Calcium Acetate Versus Sevelamer on Progression of Coronary Artery Calcification in Hemodialysis Patients With Comparable Lipid Control: The Calcium Acetate Renagel Evaluation-2 (CARE-2) Study. American Journal of Kidney Diseases 51:6, 952-965
    CrossRef

  14. 14

    John A Rumberger. (2008) Role of noninvasive imaging using computed tomography for detection and quantification of coronary atherosclerosis. Future Cardiology 4:3, 269-283
    CrossRef

  15. 15

    Nan-Hung Pan, Hung-Yu Yang, Ming-Hsiung Hsieh, Yi-Jen Chen. (2008) Coronary calcium score from multislice computed tomography correlates with QT dispersion and left ventricular wall thickness. Heart and Vessels 23:3, 155-160
    CrossRef

  16. 16

    João A.C. Lima. (2008) Another Step Forward in CT Angiography. JACC: Cardiovascular Imaging 1:2, 187-189
    CrossRef

  17. 17

    Bertrand Janne d’Othée, Uwe Siebert, Ricardo Cury, Hossein Jadvar, Edward J. Dunn, Udo Hoffmann. (2008) A systematic review on diagnostic accuracy of CT-based detection of significant coronary artery disease. European Journal of Radiology 65:3, 449-461
    CrossRef

  18. 18

    Teruhito Kido, Akira Kurata, Hiroshi Higashino, Yuma Inoue, Rene Epunza Kanza, Hideki Okayama, Jitsuo Higaki, Kenya Murase, Teruhito Mochizuki. (2008) Quantification of Regional Myocardial Blood Flow Using First-Pass Multidetector-Row Computed Tomography and Adenosine Triphosphate in Coronary Artery Disease. Circulation Journal 72:7, 1086-1091
    CrossRef

  19. 19

    Maurizio Turiel, Simona Sitia, Fabiola Atzeni, Piercarlo Sarzi-Puttini. (2007) Diagnosis of cardiac involvement in systemic autoimmune diseases. Expert Review of Clinical Immunology 3:6, 839-843
    CrossRef

  20. 20

    Julie J. Ramos, Marcus Williams, Andreas Synetos, Stamatios Lerakis. (2007) Clinical Utility of Cardiac Computed Tomography. The American Journal of the Medical Sciences 334:5, 350-355
    CrossRef

  21. 21

    R. Marano, L. Bonomo. (2007) Coronary artery calcium score: has anything changed?. La radiologia medica 112:7, 949-958
    CrossRef

  22. 22

    Elie Chammas, Chadi Dib, Maya Rahhal, Toni Helou, Georges Ghanem, Walid Tarcha. (2007) Noninvasive Assessment of Coronary Flow Reserve in the Left Anterior Descending Artery by Transthoracic Echocardiography before and after Stenting. Echocardiography 24:8, 789-794
    CrossRef

  23. 23

    Stephan Achenbach. (2007) Cardiac CT: State of the art for the detection of coronary arterial stenosis. Journal of Cardiovascular Computed Tomography 1:1, 3-20
    CrossRef

  24. 24

    Ayman A. El-Menyar, Jassim Al Suwaidi, David R. Holmes. (2007) Left Main Coronary Artery Stenosis: State-of-the-Art. Current Problems in Cardiology 32:3, 103-193
    CrossRef

  25. 25

    Judd E. Hollander, Harold I. Litt, Maureen Chase, Aaron M. Brown, Woojin Kim, William G. Baxt. (2007) Computed Tomography Coronary Angiography for Rapid Disposition of Low-risk Emergency Department Patients with Chest Pain Syndromes. Academic Emergency Medicine 14:2, 112-116
    CrossRef

  26. 26

    Hidemasa Tanaka, Kenei Shimada, Ken Yoshida, Satoshi Jissho, Junichi Yoshikawa, Minoru Yoshiyama. (2007) The Simultaneous Assessment of Aortic Valve Area and Coronary Artery Stenosis Using 16-Slice Multidetector-Row Computed Tomography in Patients With Aortic StenosisComparison With Echocardiography. Circulation Journal 71:10, 1593-1598
    CrossRef

  27. 27

    Yiannis S. Chatzizisis, George D. Giannoglou, Antonis Matakos, Chrysanthi Basdekidou, George Sianos, Alexandros Panagiotou, Christos Dimakis, George E. Parcharidis, George E. Louridas. (2006) In-vivo accuracy of geometrically correct three-dimensional reconstruction of human coronary arteries: is it influenced by certain parameters?. Coronary Artery Disease 17:6, 545-551
    CrossRef

  28. 28

    Mario J Garcia. (2006) Clinical experience and progress of computed tomographic coronary angiography. Coronary Artery Disease 17:2, 99-105
    CrossRef

  29. 29

    Mani Vembar, Matthew J. Walker, Peter C. Johnson. (2006) Cardiac imaging using multislice computed tomography scanners: technical considerations. Coronary Artery Disease 17:2, 115-123
    CrossRef

  30. 30

    Alexander Lembcke, Patrick A. Hein, Pascal M. Dohmen, Christian Klessen, Till H. Wiese, Udo Hoffmann, Bernd Hamm, Christian N.H. Enzweiler. (2006) Pictorial review: Electron beam computed tomography and multislice spiral computed tomography for cardiac imaging. European Journal of Radiology 57:3, 356-367
    CrossRef

  31. 31

    Christian N.H. Enzweiler, Sebastian Höhn, Matthias Taupitz, Alexander E. Lembcke, Till H. Wiese, Bernd Hamm, Dietmar E. Kivelitz. (2006) Contrast Enhancement in Electron Beam Tomography of the Heart: Comparison of a Monomeric and a Dimeric Iodinated Contrast Agent in 59 Patients1. Academic Radiology 13:1, 95-103
    CrossRef

  32. 32

    Katrina Barnett, James A. Feldman. (2005) Noninvasive Imaging Techniques to Aid in the Triage of Patients with Suspected Acute Coronary Syndrome: A Review. Emergency Medicine Clinics of North America 23:4, 977-998
    CrossRef

  33. 33

    Aseem Vashist, Brian G Abbott. (2005) Noninvasive cardiac imaging in the evaluation of suspected acute coronary syndromes. Expert Review of Cardiovascular Therapy 3:3, 473-486
    CrossRef

  34. 34

    Javier Sanz, Teresa Rius, Paola Kuschnir, Valentin Fuster, Judith Goldberg, Xiang Y. Ye, Paul Wisdom, Michael Poon. (2005) The Importance of End-Systole for Optimal Reconstruction Protocol of Coronary Angiography With 16-Slice Multidetector Computed Tomography. Investigative Radiology 40:3, 155-163
    CrossRef

  35. 35

    William Stanford. (2005) Advances in cardiovascular CT imaging: CT clinical imaging. The International Journal of Cardiovascular Imaging 21:1, 29-37
    CrossRef

  36. 36

    Bernd M. Ohnesorge, Lars K. Hofmann, Thomas G. Flohr, U. Joseph Schoepf. (2005) CT for imaging coronary artery disease: defining the paradigm for its application. The International Journal of Cardiovascular Imaging 21:1, 85-104
    CrossRef

  37. 37

    Paul Schoenhagen, Arthur E. Stillman, Sandy S. Halliburton, Stacie A. Kuzmiak, Tracy Painter, Richard D. White. (2005) Non-invasive coronary angiography with multi-detector computed tomography: comparison to conventional X-ray angiography. The International Journal of Cardiovascular Imaging 21:1, 63-72
    CrossRef

  38. 38

    Matthew J Budoff. (2005) Noninvasive coronary angiography using computed tomography. Expert Review of Cardiovascular Therapy 3:1, 123-132
    CrossRef

  39. 39

    Jae-Youn Moon, Namsik Chung, Byoung Wook Choi, Kyu Ok Choe, Hye Sun Seo, Young-Guk Ko, Seok-Min Kang, Jong-Won Ha, Se-Joong Rim, Yangsoo Jang, Won-Heum Shim, Seung-Yun Cho. (2005) The Utility of Multi-detector Row Spiral CT for Detection of Coronary Artery Stenoses. Yonsei Medical Journal 46:1, 86
    CrossRef

  40. 40

    Akira Kurata, Teruhito Mochizuki, Yasushi Koyama, Toyoaki Haraikawa, Jun Suzuki, Yuji Shigematsu, Jitsuo Higaki. (2005) Myocardial Perfusion Imaging Using Adenosine Triphosphate Stress Multi-Slice Spiral Computed Tomography. Circulation Journal 69:5, 550-557
    CrossRef

  41. 41

    MORDECHAI YIGLA, ZOHAR KEIDAR, ISRA SAFADI, NAVEH TOV, SHIMON A REISNER, FARID NAKHOUL. (2004) Pulmonary calcification in hemodialysis patients: Correlation with pulmonary artery pressure values. Kidney International 66:2, 806-810
    CrossRef

  42. 42

    Brad H. Thompson, William Stanford. (2004) Imaging of coronary calcification by computed tomography. Journal of Magnetic Resonance Imaging 19:6, 720-733
    CrossRef

  43. 43

    John A. Rumberger. (2004) Noninvasive Coronary Angiography Using Electron Beam Computed Tomography: Technique, Clinical Application, Future Prospective. The American Heart Hospital Journal 2:3, 153-164
    CrossRef

  44. 44

    Christoph R. Becker, Andreas Knez. (2004) Past, present, and future perspective of cardiac computed tomography. Journal of Magnetic Resonance Imaging 19:6, 676-685
    CrossRef

  45. 45

    Kyongtae T. Bae, Cheng Hong, Bruce R. Whiting. (2004) Radiation dose in multidetector row computed tomography cardiac imaging. Journal of Magnetic Resonance Imaging 19:6, 859-863
    CrossRef

  46. 46

    Stephan Achenbach, Dieter Ropers, Falk-Karsten Pohle, Udo Hoffmann, Maros Ferencik, R Cury, S Abbara, Tom Brady, Werner G. Daniel. (2004) The Use of CTA in the Chest Pain Center: A Perspective. Critical Pathways in Cardiology: A Journal of Evidence-Based Medicine 3:2, 87-93
    CrossRef

  47. 47

    Marc Dewey, Michael Laule, Lasse Krug, Dirk Schnapauff, Patrik Rogalla, Wolfgang Rutsch, Bernd Hamm, Alexander Lembcke. (2004) Multisegment and Halfscan Reconstruction of 16-Slice Computed Tomography for Detection of Coronary Artery Stenoses. Investigative Radiology 39:4, 223-229
    CrossRef

  48. 48

    Stephan Achenbach. (2004) Detection of coronary stenoses by multidetector computed tomography: it's all about resolution. Journal of the American College of Cardiology 43:5, 840-841
    CrossRef

  49. 49

    F Philippe. (2004) Quelle méthode d’angiographie pour prédire le risque coronaire en 2003 ?. Annales de Cardiologie et d'Angéiologie 53:2, 79-90
    CrossRef

  50. 50

    S.H. Lee, B.W. Choi, H.-J. Kim, H. Jung, H.-K. Son, W.-S. Kang, S.K. Yoo, K.O. Choe, H.S. Yoo. (2004) Analysis of the Heart Rate and Its Variation Affecting Image Quality and Optimized Reconstruction Window in Retrospective ECG-Gated Coronary Angiography Using Multidetector Row CT. IEEE Transactions on Nuclear Science 51:1, 225-231
    CrossRef

  51. 51

    Bin Lu, David M. Shavelle, Songshou Mao, Lynn Chen, Janis Child, Sivi Carson, Matthew J. Budoff. (2004) Improved Accuracy of Noninvasive Electron Beam Coronary Angiography. Investigative Radiology 39:2, 73-79
    CrossRef

  52. 52

    Stephen Schroeder, Axel Kuettner, Martin Leitritz, Jan Janzen, Andreas F Kopp, Christian Herdeg, Martin Heuschmid, Christof Burgstahler, Andreas Baumbach, Manfred Wehrmann, Claus D Claussen. (2004) Reliability of Differentiating Human Coronary Plaque Morphology Using Contrast-Enhanced Multislice Spiral Computed Tomography. Journal of Computer Assisted Tomography 28:4, 449-454
    CrossRef

  53. 53

    Nobusada Funabashi, Yoshiki Kobayashi, Masayuki Kudo, Miki Asano, Kiyomi Teramoto, Issei Komuro, Geoffrey D Rubin. (2004) New Method of Measuring Coronary Diameter by Electron-Beam Computed Tomographic Angiography Using Adjusted Thresholds Determined by Calibration With Aortic Opacity. Circulation Journal 68:8, 769-777
    CrossRef

  54. 54

    Matthew J Budoff, Stephan Achenbach, Andre Duerinckx. (2003) Clinical utility of computed tomography and magnetic resonance techniques for noninvasive coronary angiography. Journal of the American College of Cardiology 42:11, 1867-1878
    CrossRef

  55. 55

    Jacob Gurevitch, Tamar Gaspar, Boris Orlov, Ron Amar, Dan Dvir, Nathan Peled, Dan J Aravot. (2003) Noninvasive evaluation of arterial grafts with newly released multidetector computed tomography. The Annals of Thoracic Surgery 76:5, 1523-1527
    CrossRef

  56. 56

    Francisco Llach, Elvira Fernandez. (2003) Overview of renal bone disease: Causes of treatment failure, clinical observations, the changing pattern of bone lesions, and future therapeutic approach. Kidney International 64:s87, 113-119
    CrossRef

  57. 57

    Michael Bursztyn, Michael Motro, Ehud Grossman, Joseph Shemesh. (2003) Accelerated coronary artery calcification in mildly reduced renal function of high-risk hypertensives. Journal of Hypertension 21:10, 1953-1959
    CrossRef

  58. 58

    C. H. C. Janssen, P. M. van Ooijen, M. Oudkerk. (2003) Noninvasive Coronary Imaging Using Electron Beam Computed Tomography. Imaging Decisions MRI 7:2, 15-22
    CrossRef

  59. 59

    Junichiro Takasu, Songshou Mao, Matthew J Budoff. (2003) Aortic Atherosclerosis Detected with Electron-Beam CT as a Predictor of Obstructive Coronary Artery Disease. Academic Radiology 10:6, 631-637
    CrossRef

  60. 60

    A. H. Mahnken, J. E. Wildberger, A. M. Sinha, K. Dedden, S. Stanzel, R. Hoffmann, T. Schmitz-Rode, R. W. Gunther. (2003) Value of 3D-Volume Rendering in the Assessment of Coronary Arteries with Retrospectively Ecg-Gated Multislice Spiral Ct. Acta Radiologica 44:3, 302-309
    CrossRef

  61. 61

    KOEN NIEMAN, AAD VAN DER LUGT, PETER MT PATTYNAMA, PIM J DE FEYTER. (2003) Noninvasive Visualization of Atherosclerotic Plaque with Electron Beam and Multislice Spiral Computed Tomography. Journal of Interventional Cardiology 16:2, 123-128
    CrossRef

  62. 62

    STEPHAN ACHENBACH, TOM GIESLER, DIETER ROPERS, STEFAN ULZHEIMER, KATHARINA ANDERS, EVELYN WENKEL, KARSTEN POHLE, MARC KACHELRIESS, HANS DERLIEN, WILLI A. KALENDER, WERNER G. DANIEL, WERNER BAUTZ, ULRICH BAUM. (2003) Comparison of Image Quality in Contrast-enhanced Coronary-artery Visualization by Electron Beam Tomography and Retrospectively Electrocardiogram-gated Multislice Spiral Computed Tomography. Investigative Radiology 38:2, 119-128
    CrossRef

  63. 63

    Nobusada Funabashi, Kazuo Misumi, Hiroyuki Ohnishi, Mai Watanabe, Yutaka Suzuki, Noriyuki Imai, Katsuya Yoshida, Issei Komuro. (2003) Endoluminal Perspective Volume Rendering of Coronary Arteries Using Electron-Beam Computed Tomography. Circulation Journal 67:12, 1064-1067
    CrossRef

  64. 64

    Fouad Albaaj, Alastair J Hutchison. (2003) Hyperphosphataemia in Renal Failure. Drugs 63:6, 577-596
    CrossRef

  65. 65

    Thomas C. Gerber, Ronald S. Kuzo, Gary E. Lane, Peter C. O'Brien, Nolan Karstaedt, Richard L. Morin, Robert E. Safford, Joseph L. Blackshear, Jerald H. Pietan. (2003) Image Quality in a Standardized Algorithm for Minimally Invasive Coronary Angiography with Multislice Spiral Computed Tomography. Journal of Computer Assisted Tomography 27:1, 62-69
    CrossRef

  66. 66

    Yuichi Sato, Naoya Matsumoto, Masahiko Kato, Fumio Inoue, Toshiyuki Horie, Junji Kusama, Akihiro Yoshimura, Takako Imazeki, Takahiro Fukui, Satoru Furuhashi, Motoichiro Takahashi, Katsuo Kanmatsuse. (2003) Noninvasive Assessment of Coronary Artery Disease by Multislice Spiral Computed Tomography Using a New Retrospectively ECG-Gated Image Reconstruction Technique. Circulation Journal 67:5, 401-405
    CrossRef

  67. 67

    M. Vembar, M. J. Garcia, D. J. Heuscher, R. Haberl, D. Matthews, G. E. Böhme, N. L. Greenberg. (2003) A dynamic approach to identifying desired physiological phases for cardiac imaging using multislice spiral CT. Medical Physics 30:7, 1683
    CrossRef

  68. 68

    Shoji Yamakami, Junji Toyama, Mitsuhiro Okamoto, Toyoaki Matsushita, Yoshimasa Murakami, Masaki Ogata, Shigenori Ito, Tatsuya Fukutomi, Naotsuka Okayama, Makoto Itoh. (2003) Noninvasive Detection of Coronary Artery Bypass Graft Patency by Intravenous Electron Beam Computed Tomographic Angiography. Japanese Heart Journal 44:6, 811-822
    CrossRef

  69. 69

    Thomas Neunteufl, Gerald Maurer. (2003) Noninvasive Ultrasound Techniques for the Assessment of Atherosclerosis in Coronary Artery Disease. Circulation Journal 67:3, 177-186
    CrossRef

  70. 70

    Yuichi Sato, Katsuo Kanmatsuse, Fumio Inoue, Toshiyuki Horie, Masahiko Kato, Junji Kusama, Akihiro Yoshimura, Takako Imazeki, Satoru Furuhashi, Motoichiro Takahashi. (2003) Noninvasive Coronary Artery Imaging by Multislice Spiral Computed Tomography. Circulation Journal 67:2, 107-111
    CrossRef

  71. 71

    Hendrik Treede, Christoph Becker, Hermann Reichenspurner, Andreas Knez, Christian Detter, Maximilian Reiser, Bruno Reichart. (2002) Multidetector computed tomography (MDCT) in coronary surgery: first experiences with a new tool for diagnosis of coronary artery disease. The Annals of Thoracic Surgery 74:4, 1398-1402
    CrossRef

  72. 72

    G Morgan-Hughes. (2002) Multislice Computed Tomography Cardiac Imaging: Current Status. Clinical Radiology 57:10, 872-882
    CrossRef

  73. 73

    Christoph R. Becker, Andreas Knez, Alexander Leber, Hendrik Treede, B. Ohnesorge, U. Joseph Schoepf, Maximilian F. Reiser. (2002) Detection of Coronary Artery Stenoses With Multislice Helical CT Angiography. Journal of Computer Assisted Tomography 26:5, 750-755
    CrossRef

  74. 74

    Jonathan Valabhji, Robert S. Elkeles. (2002) Non-invasive measurement of coronary heart disease using electron beam computed tomography. Current Opinion in Lipidology 13:4, 409-414
    CrossRef

  75. 75

    DIETER ROPERS, MATTHIAS REGENFUS, NIKOLAOS STILIANAKIS, SUSANNE BIRKE, WINFRIED KESSLER, WERNER MOSHAGE, GERHARD LAUB, WERNER G. DANIEL, STEPHAN ACHENBACH. (2002) A Direct Comparison of Noninvasive Coronary Angiography by Electron Beam Tomography and Navigator-Echo-Based Magnetic Resonance Imaging for the Detection of Restenosis Following Coronary Angioplasty. Investigative Radiology 37:7, 386-392
    CrossRef

  76. 76

    Christoph R Becker. (2002) Assessment of coronary arteries with CT. Radiologic Clinics of North America 40:4, 773-782
    CrossRef

  77. 77

    Koen Nieman, Benno Rensing, Arie Munne, Robert-Jan van Geuns, Peter Pattynama, Pim de Feyter. (2002) Three-Dimensional Coronary Anatomy in Contrast-Enhanced Multislice Computed Tomography. Preventive Cardiology 5:2, 79-83
    CrossRef

  78. 78

    Bin Lu, Nan Zhuang, Song-Shou Mao, Hamid Bakhsheshi, Steve C. K. Liu, Matthew J. Budoff. (2002) Image Quality of Three-Dimensional Electron Beam Coronary Angiography. Journal of Computer Assisted Tomography 26:2, 202-209
    CrossRef

  79. 79

    S NEUBAUER. (2002) MRI and CT. Medicine 30:3, 40-44
    CrossRef

  80. 80

    Ernst Wellnhofer, Andreas Wahle, Eckart Fleck. (2002) Progression of coronary atherosclerosis quantified by analysis of 3-D reconstruction of left coronary arteries. Atherosclerosis 160:2, 483-493
    CrossRef

  81. 81

    S WORTHLEY, H OMARFAROUQUE, G HELFT, I MEREDITH. (2002) Coronary artery imaging in the new millennium. Heart, Lung and Circulation 11:1, 19-25
    CrossRef

  82. 82

    T. C. Gerber, R. S. Kuzo, N. Karstaedt, G. E. Lane, R. L. Morin, P. F Sheedy, R. E. Safford, J. L. Blackshear, J. H. Pietan. (2002) Current Results and New Developments of Coronary Angiography With Use of Contrast-Enhanced Computed Tomography of the Heart. Mayo Clinic Proceedings 77:1, 55-71
    CrossRef

  83. 83

    Kim, W. Yong, Danias, Peter G., Stuber, Matthias, Flamm, Scott D., Plein, Sven, Nagel, Eike, Langerak, Susan E., Weber, Oliver M., Pedersen, Erik M., Schmidt, Matthias, Botnar, René M., Manning, Warren J., . (2001) Coronary Magnetic Resonance Angiography for the Detection of Coronary Stenoses. New England Journal of Medicine 345:26, 1863-1869
    Full Text

  84. 84

    Achenbach, Stephan, Daniel, Werner G., . (2001) Noninvasive Coronary Angiography — An Acceptable Alternative?. New England Journal of Medicine 345:26, 1909-1910
    Full Text

  85. 85

    O Hergesell, E Ritz. (2001) Compounds in development to combat hyperphosphataemia. Expert Opinion on Investigational Drugs 10:12, 2185-2190
    CrossRef

  86. 86

    (2001) Noninvasive Tests in Patients with Stable Coronary Artery Disease. New England Journal of Medicine 345:18, 1351-1351
    Full Text

  87. 87

    Songshou Mao, Matthew J Budoff, Lu Bin, Steve C.K Liu. (2001) Optimal ECG Trigger Point in Electron-Beam CT Studies. Academic Radiology 8:11, 1107-1115
    CrossRef

  88. 88

    Harvey S. Hecht. (2001) New developments in atherosclerosis imaging: Electron beam tomography. Current Atherosclerosis Reports 3:5, 417-424
    CrossRef

  89. 89

    Bin Lu, Ru-ping Dai, Shi-liang Jiang, Hua Bai, Sha He, Nan Zhuang, Xiao-gang Sun, Matthew J. Budoff. (2001) Effects of Window and Threshold Levels on the Accuracy of Three-Dimensional Rendering Techniques in Coronary Artery Electron-Beam CT Angiography. Academic Radiology 8:8, 754-761
    CrossRef

  90. 90

    F. D. Knollmann, P. Muschick, W. Krause, H. Hausmann, R. Hetzer, R. Felix. (2001) DETECTION OF MYOCARDIAL ISCHEMIA BY ELECTRON BEAM CT. Experimental studies. Acta Radiologica 42:4, 386-392
    CrossRef

  91. 91

    Stephan Achenbach, Frank Meissner, Dieter Ropers, Karsten Pohle, Magda Kusus, Gerd Muschiol, Werner G. Daniel, Werner Moshage. (2001) Overlapping Cross-Sections Significantly Improve the Reproducibility of Coronary Calcium Measurements by Electron Beam Tomography: A Phantom Study. Journal of Computer Assisted Tomography 25:4, 569-573
    CrossRef

  92. 92

    Stephen Schroeder, Andreas F. Kopp, Andreas Baumbach, Axel Kuettner, Christian Herdeg, Albert Rosenberger, Hans-K. Selbmann, Claus D. Claussen, Martin Oberhoff, Karl R. Karsch. (2001) Noninvasive detection of coronary lesions by multislice computed tomography: Results of the new age pilot trial. Catheterization and Cardiovascular Interventions 53:3, 352-358
    CrossRef

  93. 93

    B. E. Cham. (2001) Plaque cholesterol and calcium: the value of EBCT in the detection of coronary atherosclerosis. European Journal of Clinical Investigation 31:6, 467-468
    CrossRef

  94. 94

    Christopher R deFilippi, Salvatore Rosanio, Monica Tocchi, Rohit J Parmar, Marjorie A Potter, Barry F Uretsky, Marschall S Runge. (2001) Randomized comparison of a strategy of predischarge coronary angiography versus exercise testing in low-risk patients in a chest pain unit: in-hospital and long-term outcomes. Journal of the American College of Cardiology 37:8, 2042-2049
    CrossRef

  95. 95

    S Schroeder. (2001) Accuracy and Reliability of Quantitative Measurements in Coronary Arteries by Multi-slice Computed Tomography: Experimental and Initial Clinical Results. Clinical Radiology 56:6, 466-474
    CrossRef

  96. 96

    Christian N. H. Enzweiler, Matthias Taupitz, Jan Petersein, Patrick F. Sheedy, Jerome F. Breen, Bernd Hamm. (2001) Rib Artifacts in Electron Beam Tomography: Incidence and Severity Without and With the Cone Beam Reconstruction Algorithm. Journal of Computer Assisted Tomography 25:3, 365-370
    CrossRef

  97. 97

    Ralph Haberl, Alexander Becker, Alexander Leber, Andreas Knez, Christoph Becker, Christine Lang, Roland Brüning, Maximilian Reiser, Gerhard Steinbeck. (2001) Correlation of coronary calcification and angiographically documented stenoses in patients with suspected coronary artery disease: results of 1,764 patients. Journal of the American College of Cardiology 37:2, 451-457
    CrossRef

  98. 98

    Koen Nieman, Matthijs Oudkerk, Benno J Rensing, Peter van Ooijen, Arie Munne, Robert-Jan van Geuns, Pim J de Feyter. (2001) Coronary angiography with multi-slice computed tomography. The Lancet 357:9256, 599-603
    CrossRef

  99. 99

    S. G. Worthley, G. Helft, A. G. Zaman, V. Fuster, J. J. Badimon. (2000) Atherosclerosis and the vulnerable plaque - imaging: Part II. Australian and New Zealand Journal of Medicine 30:6, 704-710
    CrossRef

  100. 100

    Tedd L. Mitchell, John J. Pippin, Susan M. Devers, Thomas E. Kimball, Larry W. Gibbons, Lori L. Cooper, Veronica Gonzalez-Dunn, Kenneth H. Cooper. (2000) Incidental Detection of Preclinical Renal Tumors with Electron Beam Computed Tomography: Report of 26 Consecutive Operated Patients. Journal of Computer Assisted Tomography 24:6, 843-845
    CrossRef

  101. 101

    Shaul Atar, Tomoo Nagai, Bojan Cercek, Tasneem Z Naqvi, Huai Luo, Robert J Siegel. (2000) Pacing stress echocardiography: an alternative to pharmacologic stress testing. Journal of the American College of Cardiology 36:6, 1935-1941
    CrossRef

  102. 102

    Ira Tabas, Petri T. Kovanen. (2000) In search of an endpoint. Current Opinion in Lipidology 11:5, 447-450
    CrossRef

  103. 103

    Veronica Villa-Colinayo, Weibin Shi, Jesus Araujo, Aldons J. Lusis. (2000) Genetics of atherosclerosis: The search for genes acting at the level of the vessel wall. Current Atherosclerosis Reports 2:5, 380-389
    CrossRef

  104. 104

    Michael V. McConnell. (2000) Imaging techniques to predict cardiovascular risk. Current Cardiology Reports 2:4, 300-307
    CrossRef

  105. 105

    Goodman, William G., Goldin, Jonathan, Kuizon, Beatriz D., Yoon, Chun, Gales, Barbara, Sider, Donna, Wang, Yan, Chung, Joanie, Emerick, AlethaGreaser, Lloyd, Elashoff, Robert M., Salusky, Isidro B., . (2000) Coronary-Artery Calcification in Young Adults with End-Stage Renal Disease Who Are Undergoing Dialysis. New England Journal of Medicine 342:20, 1478-1483
    Full Text

  106. 106

    Masami Nishino, Mary J Malloy, Josefina Naya-Vigne, Julie Russell, John P Kane, Rita F Redberg. (2000) Lack of association of lipoprotein(a) levels with coronary calcium deposits in asymptomatic postmenopausal women. Journal of the American College of Cardiology 35:2, 314-320
    CrossRef

  107. 107

    Jan Trautvetter, Stephan Achenbach, Dieter Ropers, Werner G. Daniel, Werner Moshage. (2000) EVALUATION VON 2- UND 3-DIMENSIONALEN BILDREKONSTRUKTIONSVERFAHREN ZUR KORONARARTERIENDARSTELLUNG MITTELS ELEKTRONENSTRAHLTOMOGRAPHIE. Biomedizinische Technik/Biomedical Engineering 45:s1, 519-520
    CrossRef

  108. 108

    Marc Kachelrieß, Stefan Ulzheimer, Willi A. Kalender. (2000) ECG-correlated image reconstruction from subsecond multi-slice spiral CT scans of the heart. Medical Physics 27:8, 1881
    CrossRef

  109. 109

    MANFRED HOFMANN. (1999) Will Noninvasive Techniques Replace Coronary Angiography?. Journal of Interventional Cardiology 12:6, 405-410
    CrossRef

  110. 110

    Andrew M. Kates, Giridhar Vedala, Pamela K. Woodard, Victor G. Davila-Roman, Robert J. Gropler. (1999) Noninvasive coronary artery imaging in the diagnosis and management of patients with ischemic heart disease. Current Opinion in Cardiology 14:4, 314-320
    CrossRef

  111. 111

    (1999) Noninvasive Detection of Coronary Artery Disease. New England Journal of Medicine 340:20, 1595-1596
    Full Text

  112. 112

    Celermajer, David S., . (1998) Noninvasive Detection of Atherosclerosis. New England Journal of Medicine 339:27, 2014-2015
    Full Text

  113. 113

    Taylor, Allen J., , O'Malley, Patrick G., . (1998) Self-Referral of Patients for Electron-Beam Computed Tomography to Screen for Coronary Artery Disease. New England Journal of Medicine 339:27, 2018-2020
    Full Text

Letters