Interference with an Implantable Defibrillator by an Electronic Antitheft-Surveillance Device

Peter A. Santucci, M.D., Janet Haw, R.N., Richard G. Trohman, M.D., and Sergio L. Pinski, M.D.

N Engl J Med 1998; 339:1371-1374November 5, 1998

Article

Implantable cardioverter–defibrillators have an important role in the treatment of patients with ventricular arrhythmias. Electromagnetic interference with permanent pacemakers and implantable defibrillators may have deleterious effects.1-11 Electronic antitheft-surveillance devices, which are widely used in stores, libraries, and other places to prevent theft, are a potential source of electromagnetic interference. Approximately 400,000 of these devices are in use worldwide.

Electromagnetic interference with implantable defibrillators can generally be divided into four types. The most common type involves an overcounting of the ventricular rate. Misinterpretation of rapid rates may lead to inappropriate antitachycardia pacing or the delivery of shocks. Since the current generation of implantable defibrillators provides bradycardia pacing, overcounting may also lead to inappropriate inhibition of pacing. With a second type of electromagnetic interference, a noise-reversion mode is triggered, which results in asynchronous pacing and inhibits the detection of true tachycardias and defibrillation for them. Much less common types of interference are inadvertent reprogramming of the functions of the defibrillator and permanent damage to the circuitry. Clinical problems caused by electromagnetic interference with implantable defibrillators have been infrequent and without severe consequences. We report a life-threatening interaction between an electronic antitheft-surveillance system and an implantable defibrillator.

Case Report

While standing in a bookstore, a 72-year-old man with an implantable defibrillator (placed because of cardiac arrest due to ventricular tachycardia) had complete atrioventricular block in the absence of antecedent symptoms. He received a shock from his defibrillator. Presyncope developed, and a second shock was delivered. Shortly thereafter, third and fourth shocks were delivered. A quick-thinking bystander, who was a registered nurse, noticed that he was standing next to the store's electronic antitheft-surveillance equipment and pulled him away. No further shocks were delivered, and he soon recovered. The antitheft system was a Sensormatic Ultra-Max (Sensormatic, Boca Raton, Fla.).

Four months earlier, in September 1997, the patient's defibrillator system had been upgraded to allow dual-chamber pacing. The new system included a CPI-Ventak atrioventricular generator (model 1815, Guidant, St. Paul, Minn.) coupled to an endocardial right ventricular defibrillator lead (Transvene, model 6966, Medtronic, Minneapolis) and an active-fixation right atrial lead (model 4058, Medtronic). The device was programmed to respond with 33-J shocks if the heart rate exceeded 165 bpm.

Stored intracardiac electrograms that had been recorded during the episode revealed electrical “noise” (nonphysiologic electrical activity) on both atrial and ventricular channels. The defibrillator responded to the noise by overcounting the ventricular rate, which was interpreted as a tachycardia above the programmed detection threshold, with the result that pacing was inhibited and multiple shocks were delivered (Figure 1A, 1B, and 1CFigure 1Stored Intracardiac Recordings from an Implantable Defibrillator, Showing the Atrial-Rate, Ventricular-Rate, and Shock Electrograms (Top, Middle, and Bottom Tracings, Respectively) during an Episode of Electromagnetic Interference.). The presyncope that occurred before the second shock was attributable to ventricular asystole during the inhibition of pacing.

The simultaneous presence of electrical noise on the atrial and ventricular channels strongly suggested an external source of interference rather than a lead abnormality. Lead impedances and pacing thresholds were unremarkable. Manipulation of the defibrillator pocket (generator site) did not reproduce noise on either channel.

The patient confirmed that the episode had started while he was at a magazine rack located approximately 0.3 m (1 ft) from the electronic surveillance equipment. This fact, coupled with the resolution of the symptoms (and shocks) when he was pulled away, strongly suggested that the episode was caused by electromagnetic interference from the surveillance equipment.

To confirm that the surveillance equipment was the cause of the electromagnetic interference (and to assess the effect of changing the detection threshold to the least sensitive setting), a “rechallenge” was performed under controlled circumstances. Another antitheft device of the same model was set up, and a Sensormatic representative confirmed that it was functioning properly. At a distance of about 0.3 m, electromagnetic interference from the antitheft device affected the defibrillator on three separate occasions, resulting in an abrupt inhibition of ventricular pacing and asystole (the mechanism for delivering shocks was disabled). Presyncope occurred, and the antitheft device had to be disabled rapidly to prevent syncope. The fact that the interference was reproduced with a surveillance unit exactly like the one in the bookstore makes it highly unlikely that the episode in the bookstore was caused by a malfunction of the particular unit. Intracardiac electrograms confirmed the presence of oversensing in both chambers (which was interpreted as sustained ventricular fibrillation), and the defibrillator would have delivered shocks if the mechanism for doing so had not been deactivated. Reducing the ventricular detection threshold to the least sensitive setting did not eliminate the interference, although it occurred only when the patient was in closer proximity to the electronic surveillance device.

Discussion

We describe a case of life-threatening interference with the function of an implantable defibrillator by electronic antitheft-surveillance equipment. Previously reported interactions with sources of electromagnetic interference (i.e., slot machines, remote-control devices for toys, and electronic antitheft-surveillance equipment)3-6 have caused only transient oversensing, resulting in either the inhibition of pacing for bradycardia or tachycardia or the delivery of a single inappropriate shock without clinically significant sequelae. With the increasing use of both implantable defibrillators and antitheft-surveillance devices, potentially serious interactions may be more common in the future unless preventive measures are undertaken.

Interference from an electronic-surveillance device poses several risks for patients with defibrillators, including prolonged bradycardia or asystole due to the inhibition of ventricular pacing, the induction of ventricular arrhythmias as a result of inappropriate pauses or shocks, injury from falling, and the physical discomfort and psychological effects of multiple shocks. Induction of ventricular arrhythmias in this setting could be disastrous. The defibrillator could exhaust its programmed therapies and be unable to convert a true tachyarrhythmia to normal rhythm. Another concern is the possible induction of cardiac ischemia in susceptible patients.

It is not unusual for patients with defibrillators to present with aborted shocks due to transient electrical noise (demonstrable on the stored electrograms). The source of noise may be uncertain and is often thought to be related to an intermittent abnormality in the lead conductor. Correlating the episode with exposure to surveillance equipment will clarify the cause in some patients and avoid unnecessary replacement of the lead. Replicating the exposure under controlled conditions may be helpful in uncertain cases.

Electromagnetic interference with the function of pacemakers has been well documented.7-11 Since implantable defibrillators are designed to be more sensitive to intracardiac electrical activity than pacemakers (permitting optimal detection of ventricular fibrillation), their sensitivity to electromagnetic interference may also be increased. There are ways to decrease the risk of interference. The sensitivity of some defibrillators can be reduced. This must be done with extreme caution, however, because a low level of sensitivity carries the risk that true ventricular fibrillation will not be detected and that potentially lifesaving shocks will be withheld or delayed. The noise-reversion mode offers some protection against electromagnetic interference. In the noise-reversion mode, the defibrillation function is inhibited, and asynchronous pacing is automatically initiated. However, the noise-reversion mode may fail, as in this case. Even if it operates properly, there is a slight possibility that asynchronous pacing will induce tachyarrhythmia. In addition, induced (or spontaneous) true arrhythmias will not be sensed or treated in the noise-reversion mode. Thus, it is at best a partial solution.

These points underscore the limitations of protective mechanisms for defibrillators. The most important protection is the patient's awareness of the possibility of electromagnetic interference and avoidance of prolonged exposure to electronic surveillance equipment. With the surveillance devices typically used in stores and libraries, the customer passes through a detection system on exiting. Only brief exposure to electromagnetic interference is involved in walking through the equipment at a normal pace. Transient exposure does not result in serious problems of interference, and in most cases, the interference is not manifested clinically. Thus, under usual circumstances, the risk of inappropriate shocks should be minimal. At most, an aborted shock may be triggered, without untoward clinical consequences.

Determining the susceptibility of various defibrillator sensing systems to this type of interference will require further investigation. It is noteworthy, however, that the sensing system in this patient was a true bipolar configuration (involving a separate tip and ring electrode rather than a right ventricular coil). This configuration is generally considered the least susceptible to external signals.

The Sensormatic equipment that caused the interference in this patient relies on the emission of a pulsed electromagnetic signal, which can be a source of interference. Although this type of equipment is quite common, not all antitheft devices rely on this mechanism. The problem we describe may not occur with other surveillance systems involving different mechanisms. Interference with pacemaker function has occasionally been documented with other types of surveillance systems.11

Several measures can be undertaken to reduce the risk of electromagnetic interference. The most important is to make physicians and patients aware of the potential for interference. Patients should be instructed not to linger near electronic surveillance equipment. Although the equipment is usually easily visible, it may be concealed near an exit. All customers should be made aware of the presence of such units. Ideally, merchandise should not be displayed next to electronic surveillance equipment. Unless there is a strong reason not to do so, defibrillators should be programmed, after charging, to confirm the persistence of arrhythmia (the current practice with implantable defibrillators) in order to prevent the delivery of shocks in response to transient electromagnetic interference. In this regard, it may be problematic that confirmation of persistent arrhythmia generally occurs only with the first shock of any episode. Finally, for patients dependent on pacing support from their defibrillators, cautious programming to reduce the sensitivity to electromagnetic interference may be considered.

Source Information

From the Section of Cardiology, Rush–Presbyterian–St. Luke's Medical Center, 1753 W. Congress Pky., Chicago, IL 60612, where reprint requests should be addressed to Dr. Santucci.

References

References

  1. 1

    Eisenberg E, Waisbrod H. Spinal cord stimulator activation by an antitheft device: case report. J Neurosurg 1997;87:961-962[Erratum, J Neurosurg 1998;88:791.]
    CrossRef | Web of Science | Medline

  2. 2

    Marco D, Eisinger G, Hayes DL. Testing of work environments for electromagnetic interference. Pacing Clin Electrophysiol 1992;15:2016-2022
    CrossRef | Web of Science | Medline

  3. 3

    Schmitt C, Brachmann J, Waldecker B, et al. Implantable cardioverter defibrillator: possible hazards of electromagnetic interference. Pacing Clin Electrophysiol 1991;14:982-984
    CrossRef | Web of Science | Medline

  4. 4

    Madrid A, Sanchez A, Bosch E, Fernandez E, Moro Serrano C. Dysfunction of implantable defibrillators caused by slot machines. Pacing Clin Electrophysiol 1997;20:212-214
    CrossRef | Web of Science | Medline

  5. 5

    Mathew P, Lewis C, Neglia J, Krol RB, Saksena S. Interaction between electronic article surveillance systems and implantable defibrillators: insights from a fourth generation ICD. Pacing Clin Electrophysiol 1997;20:2857-2859
    CrossRef | Web of Science | Medline

  6. 6

    Sticherling C, Klingenheben T, Hohnloser SH. An unusual cause of ICD shock. Pacing Clin Electrophysiol 1997;20:2265-2267
    CrossRef | Web of Science | Medline

  7. 7

    Hayes DL, Wang PJ, Reynolds DW, et al. Interference with cardiac pacemakers by cellular telephones. N Engl J Med 1997;336:1473-1479
    Full Text | Web of Science | Medline

  8. 8

    Dodinot B, Godenir JP, Costa AB. Electronic article surveillance: a possible danger for pacemaker patients. Pacing Clin Electrophysiol 1993;16:46-53
    CrossRef | Web of Science | Medline

  9. 9

    Beaugeard D, Kacet S, Bricout M, Camblin J. Interférences entre les stimulateurs cardiaques et les détecteurs de vol dans les magasins. Arch Mal Coeur Vaiss 1992;85:1457-1461
    Medline

  10. 10

    Moraes JCTB. Effects of electronic autodefense devices on cardiac pacemakers. Artif Organs 1995;19:238-240
    CrossRef | Web of Science | Medline

  11. 11

    Lucas EH, Johnson D, McElroy BP. The effects of electronic article surveillance systems on permanent cardiac pacemakers: an in vitro study. Pacing Clin Electrophysiol 1994;17:2021-2026
    CrossRef | Web of Science | Medline

Citing Articles (23)

Citing Articles

  1. 1

    Joo Won Chung, Hye Jin Hwang, Moon Jae Chung, Jeong Youp Park, Hui-Nam Pak, Si Young Song. (2012) Safety of Capsule Endoscopy Using Human Body Communication in Patients with Cardiac Devices. Digestive Diseases and Sciences
    CrossRef

  2. 2

    Karoly Kaszala, Saman Nazarian, Henry Halperin. 2011. Electromagnetic Interference and CIEDs. , 1004-1027.
    CrossRef

  3. 3

    MAHAZARIN GINWALLA, SMITHA BATTULA, JONATHAN DUNN, WILLIAM R. LEWIS. (2010) Termination of Electrocution-Induced Ventricular Fibrillation by an Implantable Cardioverter Defibrillator. Pacing and Clinical Electrophysiology 33:4, 510-512
    CrossRef

  4. 4

    Richard G. Trohman. (2010) Trust in technology: Straddling the line between faith and reason*. Critical Care Medicine 38:2, 712-713
    CrossRef

  5. 5

    Giuseppe Lippi, Norbert Blanckaert, Pierangelo Bonini, Sol Green, Steve Kitchen, Vladimir Palicka, Anne J. Vassault, Camilla Mattiuzzi, Mario Plebani. (2009) Causes, consequences, detection, and prevention of identification errors in laboratory diagnostics. Clinical Chemistry and Laboratory Medicine 47:2, 143-153
    CrossRef

  6. 6

    Chunliu Zhan, William B. Baine, Artyom Sedrakyan, Claudia Steiner. (2008) Cardiac Device Implantation in the United States from 1997 through 2004: A Population-based Analysis. Journal of General Internal Medicine 23:S1, 13-19
    CrossRef

  7. 7

    Lakshminarayan Yerra, Pratap C. Reddy. (2007) Effects of Electromagnetic Interference on Implanted Cardiac Devices and Their Management. Cardiology in Review 15:6, 304-309
    CrossRef

  8. 8

    J. SCOTT ALLISON, ANDREW E. EPSTEIN, VANCE J. PLUMB. (2007) Electrical Noise Artifact: From Without or Within?. Journal of Cardiovascular Electrophysiology 18:3, 332-333
    CrossRef

  9. 9

    J. Rod Gimbel, James W. Cox. (2007) Electronic Article Surveillance Systems and Interactions With Implantable Cardiac Devices: Risk of Adverse Interactions in Public and Commercial Spaces. Mayo Clinic Proceedings 82:3, 318-322
    CrossRef

  10. 10

    J. R. Gimbel, J. W. Cox. (2007) Electronic Article Surveillance Systems and Interactions With Implantable Cardiac Devices: Risk of Adverse Interactions in Public and Commercial Spaces. Mayo Clinic Proceedings 82:3, 318-322
    CrossRef

  11. 11

    Beste Ozben, Ahmet Kaya Bilge, Ercument Yilmaz, Kamil Adalet. (2006) Implantation of a Permanent Pacemaker in a Patient With Severe Parkinson's Disease and a Preexisting Bilateral Deep Brain Stimulator. International Heart Journal 47:5, 803-810
    CrossRef

  12. 12

    Satoshi Yamaji, Shinobu Imai, Fumio Saito, Hiroshi Yagi, Toshio Kushiro, Takahisa Uchiyama. (2006) Does High-Power Computed Tomography Scanning Equipment Affect the Operation of Pacemakers?. Circulation Journal 70:2, 190-197
    CrossRef

  13. 13

    M. Windirsch, H. J. Hennes, B. Wolcke. (2004) Beeinflussung des Elektrokardiogramms durch elektromagnetische Felder. Notfall & Rettungsmedizin 7:6, 407-410
    CrossRef

  14. 14

    Joshua M. Rosenow, Howard Tarkin, Elias Zias, Carmine Sorbera, Alon Mogilner. (2003) Simultaneous use of bilateral subthalamic nucleus stimulators and an implantable cardiac defibrillator. Journal of Neurosurgery 99:1, 167-169
    CrossRef

  15. 15

    Christof Kolb, Sebastian Schmieder, Günter Lehmann, Bernhard Zrenner, Martin R Karch, Andreas Plewan, Claus Schmitt. (2003) Do airport metal detectors interfere with implantable pacemakers or cardioverter-defibrillators?. Journal of the American College of Cardiology 41:11, 2054-2059
    CrossRef

  16. 16

    A. A. Klein, G. N. Djaiani. (2003) Mobile phones in the hospital - past, present and future. Anaesthesia 58:4, 353-357
    CrossRef

  17. 17

    Michael Glikson, Paul A Friedman. (2001) The implantable cardioverter defibrillator. The Lancet 357:9262, 1107-1117
    CrossRef

  18. 18

    Michael Glikson, David L. Hayes. (2001) Cardiac Pacing. Medical Clinics of North America 85:2, 369-421
    CrossRef

  19. 19

    A A Obwegeser, R J Uitti, M F Turk, U M Wszolek, T R Flipse, R C Smallridge, R J Witte, R E Wharen. (2001) Simultaneous thalamic deep brain stimulation and implantable cardioverter-defibrillator.. Mayo Clinic Proceedings 76:1, 87-89
    CrossRef

  20. 20

    H. Leon Greene. (2000) The implantable cardioverter-defibrillator. Clinical Cardiology 23:5, 315-326
    CrossRef

  21. 21

    JACQUES MUGICA, LAURE HENRY, HERVE PODEUR. (2000) Study of Interactions Between Permanent Pacemakers and Electronic Antitheft Surveillance Systems. Pacing and Clinical Electrophysiology 23:3, 333-337
    CrossRef

  22. 22

    SANJEEV SAKSENA. (1999) Antiarrhythmic Device Advisories and Recalls: Managing an Increasingly Vulnerable Process. Pacing and Clinical Electrophysiology 22:6, 950-952
    CrossRef

  23. 23

    (1999) Implantable Defibrillators, Pacemakers, and Electronic Antitheft Devices. New England Journal of Medicine 340:14, 1117-1119
    Full Text

Letters